Manuel d'utilisation / d'entretien du produit 265A du fabricant Teledyne
Aller à la page of 32
Teledyne API, Model 265A Chemiluminescence Ozone Analy zer Manual Addendum, 03389, Rev. B, DCN 5254 HIGHLIGHTS The purpose of this Highlights Page is to list the changes that were incorporated into this Manual per DCN 5254. Chapter / Page Number Description Title Page Changed Revision number and date.
THIS PAGE IS INTENTIONALLY LEFT BLANK.
TELEDYNE INSTRUMENTS Advanc ed Polluti on Instrumentati on A Tel e dyn e T ech nologies C ompany MANUAL ADDENDUM MODEL 265A CHEMILUMINESCENCE OZONE ANALYZER (ADDENDUM TO M200A INSTRUCTION MANUAL #0224.
THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 i PRINTED DOCUMENTS ARE UNCONTROLLED SAFETY MESSAGES Your safety and the safety of others is very im portant. We have provided m any important safety messages in this manual.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 ii PRINTED DOCUMENTS ARE UNCONTROLLED TABLE OF CONTENTS SAFETY MESSAGES ................................................................................................
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 iii PRINTED DOCUMENTS ARE UNCONTROLLED LIST OF FIGURES FIGURE 1-1: REMOVAL OF SHIPPING SCREWS & CHECK FOR CORRECT POWER ............ 3 FIGURE 1-2: M265A REAR PANEL .
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 iv PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 1 PRINTED DOCUMENTS ARE UNCONTROLLED 1 GETTING STARTED 1.0 Overview The Model 265A is a close derivative of the Model 200A Chemiluminescence NOx Analyzer.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 2 PRINTED DOCUMENTS ARE UNCONTROLLED 1.2 Electrical and Pneumatic Connections 1. See Figure 1-2 and Figure 1-3 to locate the rear panel electrical and pneumatic connections.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 3 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-1: Removal of Shipping Screws & Check for Correct Pow er.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 4 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-2: M265A Rear Panel.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 5 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-3: Inlet and Exhaust Venting Recommendations.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 Figure 1-4: V/F Board Switch Settings 6 PRINTED DOCUMENTS ARE UNCONTROLLED.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 7 PRINTED DOCUMENTS ARE UNCONTROLLED 1.3 Initial Operation 1. Turn on the instrument power. The display should imm ediately light, showing the computer’s mem ory configuration, then the instrument type - M265A.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 8 PRINTED DOCUMENTS ARE UNCONTROLLED Step 2 - Calibrate the instrument: Zero/Span Calibration Procedure Step Number Action Comment 1. Input zero gas Allow zero gas to enter the sample port on the rear of the instrument.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 9 PRINTED DOCUMENTS ARE UNCONTROLLED Step Number Action Comment 1. Scroll the TEST function menu until the O 3 SLOPE is displayed. The SLOPE value for O 3 should be 1.0 ± 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 10 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 11 PRINTED DOCUMENTS ARE UNCONTROLLED 2 SPECIFICATIONS, WARRANTY 2.1 Specifications Table 2.1: M265A Specifications Ranges In 1 ppb increments from 100 ppb to 20,000 ppb Measurement Mode Single range or AutoRange Measurement Units ppb, µg/m 3 Noise at zero 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 12 PRINTED DOCUMENTS ARE UNCONTROLLED 2.2 Warranty ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) (02024D) (DCN 4473) Prior to shipme nt, Teledyne API equipm ent is thoroughly inspected and tested.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 13 PRINTED DOCUMENTS ARE UNCONTROLLED 2.3 Test and Calibration Values Table 2-2: Final Test and Calibration Values TEST Values Observed Value Units Nominal Range Reference Section RANGE ppb 10-20000 NOISE ppb 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 14 PRINTED DOCUMENTS ARE UNCONTROLLED Table 2-2: Final Test and Calibration Values (Continued) Parameter Observed Value Units Nominal Range Reference Section O 3 Span Conc ppb 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 15 PRINTED DOCUMENTS ARE UNCONTROLLED 3 THE M265A OZONE ANALYZER 3.1 Principle of Operation The Teledyne API Model 265A analyzer is designed to measure the concentration of ozone using the chemiluminescence reaction below.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 16 PRINTED DOCUMENTS ARE UNCONTROLLED TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 17 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 3-2 : Pneumatic Diagram.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 18 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 19 PRINTED DOCUMENTS ARE UNCONTROLLED 4 SOFTWARE FEATURES The M265A software has two major operating modes. The SAMPLE m ode is the normal mode when the instrument is taking data.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 20 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 4-1: Sample Menu Tree.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 Figure 4-2: Setup Menu Tree 21 PRINTED DOCUMENTS ARE UNCONTROLLED.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 22 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 23 PRINTED DOCUMENTS ARE UNCONTROLLED 5 MAINTENANCE 5.1 Maintenance Schedule NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 24 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Teledyne 265A c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Teledyne 265A - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Teledyne 265A, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Teledyne 265A va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Teledyne 265A, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Teledyne 265A.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Teledyne 265A. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Teledyne 265A ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.