Manuel d'utilisation / d'entretien du produit 402 du fabricant Gilson
Aller à la page of 124
402 Dilutor Dispenser User’s Guide.
.
402 Dilutor Dispenser User’s Guide L T801310/C - ©2003 Gilson SAS All rights reserved November 2003.
.
Table of Contents-1 T able of Content s 0 Safety 1 Introduction Notational Conventions .................................................................. 1 - 2 Overview ..................................................................................
Table of Contents-2 Upgrading to Dual-Syringe Model ........................................... 3-10 Removing the Cover ............. ................................................ 3-11 Connecting the Flat Cable .....................................
Table of Contents-3 Changing the Piston Syringes and Piston Noses .................... 6- 7 Changing the 100 and 250 µL Piston Noses ...................... 6- 7 Changing the 500 µL Piston Noses ...................................... 6- 9 Changing the 1 mL Piston Noses .
.
Safety-1 Safety 0 Read this section carefully before installing and operating the instrument. For safe and correct use of the instrument, it is essential that both operating and service personnel follow generally accepted safety procedures as well as the safety instructions given in this document, the 402 Dilutor-Dispenser User’s Guide.
Safety-2 Safety This instrument 1 must not be sterilized, using an autoclave, or any other mechanical device. When you need to clean an instrument, use one of the three following methods: 1 - a clean dry cloth, 2 - a cloth dampened with water, 3 - a cloth dampened with soapy water.
1-1 Introduction 1 The following chapter describes the instrument, the service and warranty conditions..
1-2 Introduction 1 Notational Conventions Not ational Conventions The conventions in this User’s Guide are as follows. Used in the stand-alone mode, this instrument is referred to as ‘402 Dilutor-Dispenser’ or simply 402.
1-3 Introduction 1 Overview Overview The 402 Dilutor-Dispenser, is used to transfer liquids between vials according to user defined protocols (Methods). You program the 402 Dilutor-Dispenser via the keypad, which includes an LCD screen, for communication purposes.
1-4 Introduction 1 Good Laboratory Practice (GLP) is facilitated by the following: 씮 codes to protect the program files against unau- thorized modification, 씮 counters for the number of syringe an.
1-5 Introduction 1 Unpacking Unp acking Your 402 Dilutor-Dispenser is shipped in a box together with its various accessories. The original box and packing material should be kept in case the equipment has to be returned to your Gilson distributor.
1-6 Introduction 1 W arranty If the 402 Dilutor-Dispenser does not appear to function correctly, first verify the electrical connec- tions are correct and that the instrument is switched ON. Contact your Gilson distributor for technical advice or possibly a service visit.
1-7 Introduction 1 Customer Service Gilson and its worldwide network of authorized distributors provide you with four types of as si st - ance: sales, technical, applications and service.
1-8 Introduction 1.
2-1 Description 2 The following chapter describes the mechanics and hydraulics of the Dilutor-Dispenser. Two models of 402 Dilutor-Dispenser are available: Single-syringe and Dual-syringe. Each model comes with a keypad, a probe, and accessories. You order and fit the syringe(s), from those described on page 2-6, most suited to your application.
2-2 Description 2 Single-syringe Model Single-syringe Model The Single-syringe model has a Valve module at the left, and a blanking plate at the right.
2-3 Description 2 Dual-syringe Model Dual-syringe Model The Dual-syringe model is fitted with a Valve module at the left and a Tee-junction module at the right.
2-4 Description 2 Keypad and Probe Keyp ad and Probe Keypad The keypad includes an LCD display (two 24- character lines), membrane touch-keys, and a connecting cord. Numeric keys : the keypad has the numbers 0 to 9, a decimal point key, and the Enter key.
2-5 Description 2 Hydraulic Component s V alve The two-position valve, which is fitted to the left- hand side of the 402, has three ports. It switches the syringe port to either the reservoir or to the probe side (see opposite figure). The valve, which has a 0.
2-6 Description 2 Hydraulic Components There are seven syringe sizes available for the 402, all except the 25 mL are autoclavable. The lifetime of all pistons (see Chapter 6), when they are used with water at a pressure of up to 0.3 MPa is over 30000 cycles (each cycle contains 2 strokes: one up and one down).
3-1 Installation 3 This chapter explains how to install the fuses, syringes and hydraulics according to your system requirements. This chapter explains how to add the extra module to upgrade your 402 to a dual syringe pump.
3-2 Installation 3 Assembling the Probe For Standard Use In standard use the probe is used with the 2 mL tapered transfer-tubing (see opposite figure). 1 . Thread the tapered end of the transfer-tubing through the plastic ‘tunnels’ in each of the 5 pre- mounted cable supports (see figure below).
3-3 Installation 3 2 Special adapters for the probe 1 D200 tips, pack of 10 1 D1000 tips, pack of 10 1 Transfer-tubing (outlet) non-tapered, FEP (2 mL) 1 . Lay the tubing on the probe’s groove such that the end with the small washer is just at the orifice of the probe.
3-4 Installation 3 Connecting the T ubing Connecting the T ubing Tubing connectors should be hand-tightened, tools should not be used as their use will damage the connector’s thread.
3-5 Installation 3 Connecting the T ubing After assembling the probe, the other end of the tubing is installed by screwing the connector into the right-hand port of the valve. When you use Gilson Diamond Tips, the special non-tapered tubing, included in the kit, must be fitted.
3-6 Installation 3 The transfer tubing connects the outlet port (right- hand) of the Tee-junction to the probe. A 2 mL tapered transfer-tubing is supplied as standard ; 1 mL, 5 mL, and 10 mL are available as additional accessories.
3-7 Installation 3 Electrical Connections Electrical Connections The rear panel of the 402 Dilutor-Dispenser houses: - sector block containing the power socket, fuse drawer and ON/OFF switch, - keypad connector, - probe connector, - GSIOC connector. Sector Block The sector block contains the power socket, fuse holder and ON/OFF switch.
3-8 Installation 3 Electrical Connections Connecting the Power Cord Make sure that the ON/OFF switch is in the OFF (down) position. Connect the power cord to the power socket at the rear of the instrument and then to the power supply source.
3-9 Installation 3 Syringes Syringes The 402 Dilutor-Dispenser is supplied without syringes. Order a syringe (or syringes), appropriate to your Method, for your 402 Dilutor-Dispenser. Mount the syringe on the valve or Tee-junction. Assembling a Syringe The syringe and its piston are supplied assembled.
3-10 Installation 3 Upgrading to Dual-syringe Model Upgrading to Dual-syringe Model The single syringe 402 can be upgraded to improve the versatility of the system. To upgrade to the Dual-syringe 402, you must install a Tee-junction module (kit reference F4015010) at the right-hand-side of the 402.
3-11 Installation 3 Upgrading to Dual-syringe Model Removing the Cover 1 ) Undo the four screws (with two washers) at the base of the 402 that secure the top cover. Place the 402 in the upright position, pull the sides of the cover outwards, and then lift off the cover.
3-12 Installation 3 Inserting the Module Do not try to force the module into position as this may affect the alignment of the component parts. You should proceed carefully as follows. 1 ) Tilt the top of the module inwards and carefully insert the upper part of the module into the front aperture.
4-1 Operation - Standard Use 4 This chapter describes how to operate the 402 in standard use. Chapter 5 contains more detailed and advanced information, together with less frequently used options. You are advised to read the first section of Chapter 5, ‘Correct Use of the 402’, before proceeding.
4-2 Operation - Standard Use 4 Essential Steps Essential Steps Here is a summary of the things you must do, and the order in which you should do them. 1 ) Check that you have received all standard parts. 2 ) Connect the keypad. 3 ) Fit the fuses and connect the power supply cord.
4-3 Operation - Standard Use 4 Essential Steps If you have a Dual-syringe 402, install the tubings as described in Chapter 3, page 3-5. You must connect 4 pieces of tubing: inlet (reservoir), Valve to Tee-junction, transfer (probe), and waste.
4-4 Operation - Standard Use 4 Essential Steps Configuration (Conf) You only have to enter the Configuration details one-time; the 402 keeps this information in its memory, even after you switch off. The next time you switch on the 402, the Configuration and all Methods previously saved as files will be available to modify or run.
4-5 Operation - Standard Use 4 Essential Steps If you want to change the configured volume, press Chge (until you reach the correct value); when the volumes are the same press ENTER , press the ESC key to return to the Configuration Screen, or press Next , after which you see the Air Gap Volume Screen.
4-6 Operation - Standard Use 4 Essential Steps Next is used to go to the next screen (see Air Flow Screen ), Prev is used to go to the previous screen (see Resv flow Screen ). Air Flow is the rate in mL/min. that air is drawn into the transfer tubing when aspirating air.
4-7 Operation - Standard Use 4 Essential Steps Fit the Syringe(s) 1 ) First you must lower the piston rod(s), as follows: - press the Conf key to access the Configuration Screen : - press the Syr soft-key (ensure that the size is correct before continuing), - press the Exch soft-key.
4-8 Operation - Standard Use 4 In-run Prompts You can defer some parameters (volumes and flow rates) by entering a question mark (?) in place of the value. You enter the actual value after pressing the RUN key. This feature is particularly useful when optimizing parameters.
4-9 Operation - Standard Use 4 For the Dispense task, press Disp ; you see the first parameter screen, Dispense Screen (1). Press Chge until you see ‘Source: PROBE’, press ENTER to confirm, then press Next . Press Chge until you see ‘Fraction vol: SAME’, press ENTER to confirm, then press Next .
4-10 Operation - Standard Use 4 Prime the 402 (Man) Prime is used to thoroughly clean out all fluid pathways with fresh solvent and to clear bubbles from the tubings. Press the Man soft-key, the Manipulation Screen appears. Press Prime, Initialization Screen 2 appears.
4-11 Operation - Standard Use 4 ‘Homing ...’ appears on the screen until the syringe(s) have returned to their uppermost (home) positions, during which time any liquid contained in the transfer-tubing is expelled from the probe-end. After this the Action Screen reappears.
4-12 Operation - Standard Use 4 Essential Steps In Run Screen (1) 1 air gap 2 liqu id 3 frac tion1 4 frac ti on 2 5 frac ti on 3 6 fracti on 4 7 to wa ste Extra Vo lume In Run Screen (2a) In Run Screen (2b) Ensure that the probe-end of the transfer-tubing is in a suitable waste container, then press OK .
4-13 Operation - Standard Use 4 (ii) Press the command button to aspirate AIR. The command light is steady and you see alternately In Run Screens 3a and 3b. (iii) Put the tip of the transfer-tubing into the sample then press the command button to aspirate liquid.
4-14 Operation - Standard Use 4 Press ‘ Abort ’ to confirm that you wish to stop and you are returned to the Run Screen . Alternatively, you can bring the 402 to a halt by switching off the power, in which case you must ensure that the probe end of transfer-tubing is in a suitable waste container before switching on again.
5-1 Operation - Advanced Use 5 This chapter explains in details how liquids are aspirated and dispensed using the 402 Dilutor-Dispenser and gives in depth details of various software functions.
5-2 Operation - Advanced Use 5 Correct Use of the 402 Correct Use of a 402 Single-syringe Model • Samples and reagents are aspirated from the probe side.
5-3 Operation - Advanced Use 5 Correct Use of the 402 • If a drop of liquid is present at the end of the tubing after dispensing, you must include this drop in the result vial. • The larger capacity syringe must always be mounted on the left-hand side.
5-4 Operation - Advanced Use 5 Correct Use of the 402 Aspirating Large Volumes via the Probe 1 . The valve switches to connect the left- hand syringe to the Tee-junction and probe. 2 . The piston of the right-hand syringe remains in the top position to prevent liquid from entering the syringe.
5-5 Operation - Advanced Use 5 Software Sof tware You must first enter the configuration details for your 402 Dilutor-Dispenser. Then you may do one of three things: (i) Perform a direct manipulation (Prime, Measure, Titrate). (ii) Run a prefabricated Method (files 90 - 94).
5-6 Operation - Advanced Use 5 Software 4 Press the Next soft-key to go to the following screen. 5 Press the Chge soft-key to choose an option or to select from a list of possible values. 6 Press the ESC key to return to the level of the previous screen or menu.
5-7 Operation - Advanced Use 5 Software If you want to change the configured volume, press Chge (until you reach the correct value); when the volumes are the same press ENTER , press the ESC key to return to the Configuration Screen, or press Next , after which you see the Air Gap Volume Screen.
5-8 Operation - Advanced Use 5 When you specify left and right syringe capacities (left must be > right), you must enter compatible volumes. An error message is displayed if you try to enter an inc.
5-9 Operation - Advanced Use 5 Counters Under normal use you should expect to fit new (or refurbished) syringes after 30000 cycles with water. This feature is made easy by the 402, which automati- cally counts the number of operating cycles performed by the syringe(s) and valve.
5-10 Operation - Advanced Use 5 to access the required task and its parameters, (ii) if you have entered a syringe correction factor in the configuration process, the Syringe Correction Screen appears.
5-11 Operation - Advanced Use 5 When you select Yes , the file is deleted from the directory and you see the New-File Directory Screen . Press ESC to return to the New File Screen, or press Dir to access the file directory.
5-12 Operation - Advanced Use 5 To unlock this file you must key in the code that was used to lock the file. You may not edit a locked file, so the Edit soft-key is replaced by the View soft-key, which allows you to browse through the contents of that file, without making any changes.
5-13 Operation - Advanced Use 5 Press Abort , when you want to stop priming; Initialization Screen 2 reappears. Ensure that the probe-end of the transfer-tubing is in a suitable waste container, then press OK .
5-14 Operation - Advanced Use 5 After completely aspirating the liquid being measured, you see (for example) the Measurement Screen. Procedure 1 ) Aspirate all the liquid contained in the source vial, plus an air cushion, to ensure that you have aspirated all the liquid.
5-15 Operation - Advanced Use 5 This is realized using the 402 by pressing the probe’s command button with increasing rapidity as you approach the neutralization point. Care must be taken in selecting the dispense flow rate. • Press Titrate, Titrate Screen (1) appears.
5-16 Operation - Advanced Use 5 Keys The use of the soft-keys is as follows: - Prev is used to go to the previous task (or screen within a task), - Next is used to go to the next task (or screen withi.
5-17 Operation - Advanced Use 5 Chaining Tasks Because of the flexibility of the software many combinations of the tasks are possible in your Method. However, it is better to avoid certain combinations of task such as Dispense followed by Dilution.
5-18 Operation - Advanced Use 5 Dilute Three types of dilution task are available ‘Normal’, ‘Unknown’, and ‘Serial’. Three modes are available for ‘Normal’ and ‘Serial’: Total Volume, Diluent Volume, and Dilution Ratio.
5-19 Operation - Advanced Use 5 Press Next , then key in the sample volume, press ENTER. Depending on the mode, you see one of three screens after pressing Next. Key in the value, press ENTER. Optionally, for any mode, you can display the variables calculated by the software by pressing Data .
5-20 Operation - Advanced Use 5 For each dilution in the series: Vt = Vs x Rd, Vd = Vt - Vs For example, after selecting Dilution Ratio Mode, you enter Sample Volume = 10 and Dilution Ratio = 3; the software calculates the Total Volume as 30 and the required Diluent Volume as 20.
5-21 Operation - Advanced Use 5 Optionally, for any mode, you can display the variables calculated by the software by pressing Data . Optionally, you can modify the flow rates; press Flow , For any mode, key in a value for Asp flow, press ENTER , press Next; key in a value for Disp flow, press ENTER .
5-22 Operation - Advanced Use 5 Optionally, you can supply up to 12 reagents, press Reag , key in number of reagents, press ENTER . Press Next , then for each reagent, key in volume press ENTER. After the data are entered the Reag key changes to REAG.
5-23 Operation - Advanced Use 5 If for any reason you want to repeat the measure- ment (e.g. you need to select the LEFT syringe instead of the RIGHT), put tip of the transfer- tubing back into the original container, then press Redo to discharge the liquid.
5-24 Operation - Advanced Use 5 volume of liquid, is dispensed, at the end of the process, into the waste. When you configure disposable tips in air mode, ‘Extra volume’ is replaced by ‘Air push volume’ in this task. 1. Aspirate an air gap. 2.
5-25 Operation - Advanced Use 5 Press ESC (twice). Press ESC. Pipette You use this task to aspirate a series of fractions via the probe, then dispense the total volume aspirated into a recipient vessel. Each liquid aspirated is sepa- rated by an air gap (if air gap is configured as > 0).
5-26 Operation - Advanced Use 5 Optionally at the first of the volume screens, press Flow . Key in a value for Asp flow, press ENTER. Press Next. Key in a value for Disp flow, press ENTER. Optionally at the first of the volume screens, press Extra, Key in ‘Extra volume’, press ENTER.
5-27 Operation - Advanced Use 5 Press Mix , then select the mixing mode (air or asp/ disp). Press Chge to switch modes. Press ENTER , then Next . Key in volume, press ENTER. Optionally, press Flow . • For AIR mode, key in a value for Disp flow, press ENTER .
5-28 Operation - Advanced Use 5 Key in the file number of the second Method to link(8), press ENTER , press Next . Key in the file number of the third Method to link(9), press ENTER , press Next . Key in the file number of the fourth Method to link(8), press ENTER , press Next .
5-29 Operation - Advanced Use 5 V alue for use with File Number T ASK/Parameter Dual-syringe Model Single-syringe Model Method 92 Dispense Sample Source Reservoir (RESV) Reservoir (RESV) Fractions Sam.
5-30 Operation - Advanced Use 5 When you have finished, or wish to change programs, press STOP , to instantly halt the program. Alterna- tively, you can bring the 402 to a halt by switching off the power, in which case you must ensure that the probe end of transfer-tubing is in a suitable waste container before switching on again.
5-31 Operation - Advanced Use 5 Additional Information Flow Rates Flow rates depend on the nature of the liquid (viscosity), the capacity of the syringe, and the internal diameter of tubing. For liquids transferred via the probe, it is possible to adjust the aspirate and dispense flow rates according to the nature of the liquid (e.
5-32 Operation - Advanced Use 5 Additional Information Flow Rates - Dual-syringe Model Flow rates are the values entered in Flow menus (Asp Flow or Disp Flow). • It is not possible to enter a flow rate greater than the maximum without entering an air gap for the left-hand syringe.
5-33 Operation - Advanced Use 5 Syringe Correction Function A correction factor may be applied to all volumes of liquid to compensate for the effects of temperature (for example, when the 402 is used in a culture room) or for the effects of viscosity.
5-34 Operation - Advanced Use 5 Gilson D200 tips are used for volumes from 1 µL to 78 µL (with a 100 µL syringe, minimum recommended air-push-volume is 20 µL). Gilson D1000 tips are used for volumes from 50 µL to 900 µL (minimum recommended air-push-volume is 100 µL).
6-1 Maintenance 6 This chapter explains how to perform the routine maintenance procedures. The only maintenance operations necessary for the 402 DD are: • Cleaning the exterior of the instrument. • Cleaning of the valve, Tee, syringe assemblies and the tubings.
6-2 Maintenance 6 Cleaning Cleaning Cleaning the Exterior The instrument cover and front panel may be cleaned. Ensure that the instrument is switched OFF and the power cord is removed when you clean the instrument. When you need to clean the instrument, use one of the three following methods: 1 ) A clean dry cloth.
6-3 Maintenance 6 Cleaning 1 ) Pull the axle away from the valve main body. 2 ) Remove the ceramic stator from the valve axle. 3 ) Tap the valve inner body against a clean, solid surface to remove the spring and PTFE spring cover. 4 ) Do not remove the inner ceramic surface from the valve main body.
6-4 Maintenance 6 Cleaning Cleaning the Syringe Assembly You must dismount the syringe and piston assembly from the instrument and then separate the syringe and piston assembly before cleaning or autoclaving them. 1 ) Dismount the Syringe assembly from the instrument.
6-5 Maintenance 6 Autoclaving Autoclaving Certain parts of the 402 can be autoclaved (20 min. at 0.1 MPa, 121° C). They must be removed from the instrument for autoclaving. The autoclavable parts are: • The valve (disassembled) or the Tee-junction.
6-6 Maintenance 6 Chaning the Syringe Piston Assembly Changing the Syringe Piston Assembly For all syringe sizes, it is necessary to replace the piston assembly after approximately 30 000 cycles (using water).
6-7 Maintenance 6 Changing the Piston Syringes and Piston Noses Changing the Piston Syringes and Piston Noses The piston nose should be replaced after the pump has performed approximately 30 000 cycles.
6-8 Maintenance 6 1 . Remove the syringe from the pump as described in Chapter 4. 2 . Slide the piston out of the syringe barrel. 3 . Remove the piston nose from the piston rod by carefully pinching the piston nose with a pair of cutters and gently pulling the piston nose off the piston rod.
6-9 Maintenance 6 Changing the Syringe Piston and Piston Noses Changing the 500 µL Piston Noses 1 . Remove the syringe from the pump as described page 6-4.
6-10 Maintenance 6 Changing the Piston Syringes and Piston Noses Before Mounting a 500 µL Syringe: To optimize the fitting of the 500 µL syringe on the Dilutor valve, the 500 µL syringe is supplied with a cover seal. The orientation of this seal must be respected to ensure that the connection with the syringe is airtight.
6-11 Maintenance 6 Changing the Syringe Piston and Piston Noses Changing the 25 mL Piston Noses You don’t need a special tool for mounting the piston nose on the 25 mL piston. The nose screw- fits onto the piston shaft. 1 . Remove the syringe from the pump as described page 6-4.
6-12 Maintenance 6 Changing the Piston Syringes and Piston Noses Pisto n shaft Pis ton nose Piston shaft holder Piston no se holder O-ri n g 5 . Insert the new piston nose in the recess of the piston nose holder. Make sure that the aperture is facing outward and the O-ring is inside the seal.
7-1 T roubleshooting 7 This chapter explains how to diagnose possible problems and how to overcome them..
7-2 T roubleshooting 7 Mechanical Problems Mechanical Problems Problem Possible Cause Solution V alve does not fit on to the 402. V alve axle not properly aligned Ensure that the flat on the axle with the motor drive. is aligned with the drive slot on the motor .
7-3 T roubleshooting 7 Hydraulic Problems Hydraulic Problems Problem Possible Cause Solution Liquid leakage Connections incorrectly made. Tighten all tubing connections. Liquid leaving by the Tubing blocked. Check all tubing for blockages. waste tube.
7-4 T roubleshooting 7 Electrical Problems Electrical Problems Problem Possible Cause Solution Instrument does not Power cord not connected. Check power cord is properly connected. respond. 402 is not switched on. Check ON/OFF switch. Fuse blown. Check the fuses.
Appendix A-1 Parts & Accessories Lists Appendix A Parts lists for the 402 Dilutor Dispenser, consisting of Standard Equipment, Additional Accessories, and Spare parts.
Appendix A-2 Parts & Accessories Lists Appendix A 402 Single Syringe 402 Single Syringe Part Number Q t y Description F410514 1 402 Dilutor Dispenser (Single-syringe) F4015016 1 Probe (standard) F.
Appendix A-3 Parts & Accessories Lists Appendix A 402 Dual Syringe with T ee Junction Configuration 402 Dual Syringe with T ee Junction Configuration Part Number Qt y Description F410515 1 402 Dil.
Appendix A-4 Parts & Accessories Lists Appendix A Additional Accessories Additional Accessories Part Number Description F4015015 Adapter kit for disposable tips 25025341 100 µL syringe 25025342 2.
Appendix A-5 Parts & Accessories Lists Appendix A Spare Parts Sp are Part s Part Number Description F4015072 100 µL piston kit F4015073 250 µL piston kit F4015074 500 µL piston kit F4015075 1 m.
Appendix A-6 Parts & Accessories Lists Appendix A.
Appendix B-1 Chemical Comp atibility Appendix B This chapter contains information about the chemical compatibility of the syringes and the tubing..
Appendix B-2 Chemical Comp atibility Appendix B Chemical Compatibility Chemical Comp atibility Materials in Contact with Liquids In the 402 Dilutor Dispenser flow-path, several materials are in contac.
Appendix C-1 T echnical Dat a Appendix C This chapter contains the Technical Data concerning the 402 Dilutor Dispenser..
Appendix C-2 T echnical Data Appendix C Hydraulics Hydraulics Dilution Ratio When using the dual syringe model equipped with a 25 mL syringe on the left and a 100 µL syringe on the right, the dilution ratio is 1/25000, but in order to stay within the RSD range a maximum ratio of 1/500 is recommended.
Appendix C-3 T echnical Data Appendix C Hydraulics Syringes Performance Specifications The following values are obtained using a defined protocol. If this protocol is followed, the obtained values must be inferior or equal to the listed figures.
Appendix C-4 T echnical Data Appendix C Hydraulics T ypical values (all syringes) Syringe capacity Max. accuracy Max.error imprecision % + A % RSD % 100 0.50 0.10 50 0.80 0.20 10 1.00 0.50 The same specifications are obtained in the dispensing mode with liquid from the reservoir.
Appendix C-5 T echnical Data Appendix C Mechanics Mechanics V alve Type: Three port, two position. Angle between ports: 120°. Valve drive: Stepper motor. Rotation time: 0.5 Second. Liquid contact materials: PEEK and ceramic. Lifetime: 1 000 000 cycles.
Appendix C-6 T echnical Data Appendix C Environmental Environment al Conformity to Norms Refer to separate leaflet LT801354 ‘Declaration of Conformity’. Storage Indoor use only. Installation: Category II. Altitude: operating up to 2000 m. storage up to 4600 m.
Appendix D -1 Error Messages Appendix D This chapter contains the Error Messages concerning the 402 Dilutor Dispenser..
Appendix D-2 Error Messages Appendix D Error Messages Error Message Cause/Possible Action Adjust volume too small In DISPENSE, you have, you have entered a value for “Final vol.” that is too small. “Final vol.” > volume of biggest fraction dispensed.
Appendix D-3 Error Messages Appendix D Error Messages Error Message Cause/Possible Action Parameters incompatible! Refer to page 5-16. Ratio impossible Dilution ratio calculated by software is not valid; check parameters using formulae given in page 5-18 (DILUTE task) and correct one or both of the other values.
Appendix D-4 Error Messages Appendix D.
Index-1 Index 0 Abort 4-10, 4-11 Accessories list 4 - 5 Acetonitrile B- 2 Air bubble free filling 4-10 Air Gap 4-4, 4-12, 5-2, 5-6, 5-17 Alcohol B- 2 Alternative tubing 3 - 4 Autoclaving 6 - 5 parts 6.
Index-2 Index Installation 3 - 1 Keypad 2-2, 3-8 Large Volumes 5 - 4 Liquid contact materials B - 2 Link 5-27 Lock 5-11 Maintenance Autoclaving 6 - 5 Changing the piston assembly 6 - 7 Cleaning the in.
Index-3 Index Titrate 5-14 Troubleshooting electrical problems 7 - 4 hydraulic problems 7 - 3 mechanical problems 7 - 2 Tubing connecting 3 - 4 transfer 3 - 4 valve inlet 3 - 4 valve waste 3 - 5 Typic.
Index-4 Index.
.
W orld Wide Web : www .gilson.com E-mail : sales@gilson.com, service@gilson.com, training@gilson.com W orld Headquarters Gilson, Inc. 3000 W . Beltline Hwy .
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Gilson 402 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Gilson 402 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Gilson 402, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Gilson 402 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Gilson 402, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Gilson 402.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Gilson 402. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Gilson 402 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.