EN Instructions for use | DE Gebrauchsanweisung FR Mode d’emploi | IT Istruzioni per l’uso NL Gebruiksaanwijzing | ID Petunjuk penggunaan


Bisphenol-A · free

Thank you for choosing the Swing Breastpump. Breastmilk is the natural choice for your baby, but there are reasons and situations where breastfeeding is not possible. For over 50 years, it has been Medela’s conviction that these mothers and babies can be supported. In close cooperation with breastfeeding experts, Medela, one of the market leaders, has researched and developed breastpumps that perfectly meet breastfeeding mothers needs. For mothers who insist on the best for their babies and themselves. Because you care.
Herzlichen Dank, dass Sie sich für die Swing Milchpumpe entschieden haben. Muttermilch ist an sich das Beste für Ihr Baby, doch es gibt viele Gründe, weshalb Stillen nicht immer möglich ist. Als einer der Marktführer setzt sich Medela seit über 50 Jahren mit Überzeugung für das Wohl von Babys und Müttern ein. In Zusammenarbeit mit Stillexperten und mit führenden Forschern hat Medela Milchpumpen entwickelt, die optimal auf die Bedürfnisse stillender Frauen abgestimmt sind. Für Mütter, die das Beste für ihr Baby und sich wollen. «Because you care».
Medela vous remercie d’avoir acheté un tire-lait Swing. Le lait maternel est le meilleur aliment pour votre enfant, mais de nombreuses raisons et circonstances empêchent l’allaitement au sein. Depuis plus de 50 ans, l’objectif de Medela, l’un des leaders sur le marché des tire-laits et accessoires d’allaitement, est de soutenir le bien-être des mamans et de leurs bébés. Medela a développé des tire-laits qui répondent de façon optimale aux besoins des mamans allaitantes, en collaboration avec des experts de l’allaitement et des chercheurs avant-gardistes. Pour les mamans qui ne veulent que le meilleur pour elles et leurs bébés. „Because you care.“
Grazie per aver scelto il tiralatte Swing. Il latte materno è la scelta naturale per il tuo bambino, ma in alcuni casi non è possibile allattare al seno. Da oltre 50 anni Medela è impegnata ad aiutare madri e bambini che si trovano in questa situazione. In stretta cooperazione con gli esperti dell’allattamento al seno, Medela, uno dei leader del mercato, ha condotto ricerche volte a sviluppare tiralatte che soddisfino pienamente le esigenze delle mamme. Per madri che vogliono il meglio per i loro bambini e per loro stesse. „Because you care“.
Hartelijk dank dat u voor de Swing borstkolf hebt gekozen. Moedermelk is de natuurlijke keuze voor uw baby, maar er zijn redenen en omstandigheden waarin het niet mogelijk is om borstvoeding te geven. Medela is er al meer dan 50 jaar van overtuigd dat deze moeders en baby’s ondersteund kunnen worden. In nauwe samenwerking met borstvoedingsdeskundigen heeft Medela, een van de marktleiders, borstkolven onderzocht en ontwikkeld die perfect tegemoetkomen aan de behoeften van moeders die borstvoeding geven. Voor moeders die het beste willen voor hun baby en voor zichzelf. Uit liefde.
Terima kasih telah memilih Pompa ASI Swing. ASI adalah pilihan alami bagi bayi Anda, tetapi situasi dan kondisi kadang membuat menyusui tidak mungkin. Selama lebih dari 50 tahun, Medela yakin ada yang dapat dilakukan untuk membantu ibu-ibu dan bayi-bayi ini. Medela, salah satu pemimpin pasar, bekerja sama dengan pakar per-ASI-an, berhasil meneliti dan mengembangkan pompa ASI yang benar-benar memenuhi kebutuhan ibu menyusui. Untuk para ibu yang menginginkan yang terbaik bagi bayi mereka juga mereka sendiri. Karena Anda peduli.

DE 30–55
FR 56–81
IT 82–107
NL 108–133
ID 134–159




The intended use and indications of breastpumps Breastpumps are intended to be used by lactating women in the hospital or home environment for expressing breastmilk. Breastpumps are indicated to alleviate the symptoms of engorgement by expressing milk. In case of mastitis breastpumps support the healing process by removing breastmilk from the affected breast. Breastpumps are indicated to relieve sore and cracked nipples and to bring out flat or inverted nipples. Also breastpumps enable mothers to provide breastmilk for babies who cannot feed directly at the breast for a variety of reasons such as latch-on problems, cleft palate or premature babies.
Intended patient population/user Adolescent paediatric or adult lactating women. Many women find it convenient, or even necessary, to use a breastpump to express and store their breastmilk once they have returned to work, are travelling or are otherwise separated from their baby. Women are often delighted to learn that a breastpump can be used as a supplement to breastfeeding and that some pumps are designed to mimic the suckling of a nursing baby.
Contraindications There are no known contraindications for the Swing breastpump.

The warning symbol identifies all instructions that are important for safety. Failure to observe these instructions can lead to injury or damage to the breastpump! When used in conjunction with the following words, the warning symbol stands for:
Info Useful or important information that is not related to safety.
Symbols on the packaging
This symbol indicates that the material is part of a recovery/recycling process.
This symbol indicates a carton package.
This symbol indicates to keep the device away from sunlight.
This symbol indicates to handle the fragile device with care.
°C This symbol indicates the temperature limitation for operation, transport and storage.
This symbol indicates the humidity limitation for operation, transport and storage.
This symbol indicates the atmospheric pressure limitation for operation, transport and storage.
This symbol indicates to keep the device dry.
This symbol indicates unique GSI Global Trade Item Numbers (GTIN)
This symbol indicates do not dispose the device together with unsorted municipal waste (for EU only).
This symbol indicates to follow instructions for use.
Symbols on the device
This symbol indicates to follow instructions for use.
0123 This symbol indicates the compliance with the essential requirements of the Council Directive 93/42/EEC of 14 June 1993 concerning medical devices.

This symbol indicates the manufacturer.
This symbol indicates do not dispose the device together with unsorted municipal waste (for EU only).
This symbol indicates the compliance with additional USA and Canada safety requirements for medical electrical equipment.
This symbol indicates a type BF applied part.
SN This symbol indicates manufacturer’s serial number of the device.
+ This symbol indicates the positioning of battery cell.
IP22 This symbol indicates the protection against ingress of solid foreign objects and against harmful effects due to the ingress of water.
This symbol indicates the date of manufacture (four digits for the year and two digits for the month).
Symbols on the mains adapter

This symbol indicates that the mains adapter is a class II device.
This symbol indicates the compliance with China Compulsory Certification.
This symbol indicates that the mains adapter is safety tested.
This symbol indicates that the mains adapter is for indoor use only.
This symbol indicates the compliance with USA and Canada safety requirements.
This symbol indicates the compliance with Federal Communications Commission requirements.
This symbol indicates polarity of d.c. power connector.
This symbol indicates the compliance with AUS/NZ regulatory requirements (Regulatory compliance mark).
This CE-mark indicates the compliance with the low voltage and electromagnetic compatibility directive.
This symbol indicates to not dispose the device together with unsorted municipal waste (for EU only). The solid bar symbol indicates that mains adapter is put on the market after 13 August 2005.
This symbol indicates the compliance with Japanese safety requirements.
This symbol indicates the compliance with energy efficiency requirements.
This symbol indicates alternating current.
This symbol indicates direct current.

Failure to follow these instructions/safety information can lead to danger from the device. Subject to technical changes.

The Swing breastpump is not heat-resistant: keep away from radiators and open flames.
Do not expose the motor unit to direct sunlight.
Repairs must be performed only by an authorised service agency. Do not repair yourself! No modifications to the device are permitted.
Never use a damaged device. Replace damaged or worn parts.
MAX. 275 h The operating life of the Swing breastpump is 275 hours. The service life is two years.

Separation from the mains is only assured through the disconnection of the mains adapter from the wall socket outlet.
Keep the mains adapter away from heated surfaces.
Do not reach for any electrical device if it has fallen into water. Unplug immediately.
The breastpump should never be left unattended when plugged into a power source.
Device · Electrical use

Use the Swing breastpump only for its intended use as described in this manual.
Never operate an electrical device if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water.
Do not use the Swing breastpump while bathing or showering.
Do not use the Swing breastpump while sleeping or overly drowsy.
This is a single user product. Use by more than one person may present a health risk.
Do not drive whilst hands-free pumping.
Supervision is necessary when the Swing breastpump is used in the vicinity of children.
Contact your lactation consultant or physician if you experience problems or pain.
Portable and mobile radio frequency communications equipment can affect the Swing breastpump.
l Plastic bottles and component parts become brittle when frozen and
may break when dropped. l Also, bottles and component parts may become damaged if mishandled,
e.g. dropped, over-tightened, or knocked over. l Take appropriate care in handling bottles and components. l Do not use the breastmilk if bottles or components become damaged.
Use

Spare parts and accessories available for order can be found on page 28

1x PersonalFit™ Breastshield M (24mm) (for other breastshield sizes see Chapter 12, page 28) 810.7084
1x 2-component connector 800.7196
1x valve head 810.0475
2x valve membrane 820.7013

1x bottle 811.0021
1x bottle stand 810.0462

1x sling 800.0734
1x Swing storage bag 800.0838

1x Swing tubing 800.0836
1x motor unit 600.0975



1x Calma Solitaire 200.3386
1x Calma multi lid 200.2714
1x mains adapter 200.4729 Euro 4.8V 200.4730 UK and US/JP 4.8V 200.4731 AUS/NZ 4.8V 200.4732 Plugs CN and KO 4.8V

5.1 Before using for the first time and after every use
l Only use drinking-quality water
for cleaning. l Take apart and wash all parts that
come in contact with the breast and breastmilk immediately after use to avoid dry up of milk residues and to prevent growth of bacteria.
Disassemble the pump set into individual parts.

l The pump set requires no main-
tenance. l Take care not to damage parts
of the pump set during cleaning. l If the individual components of the
pump set are cleaned in a dish washer parts may be discoloured by food pigments. This has no impact on the function.
Rinse all parts with cold, clear water (approx. 20 °C) for 10 – 15 seconds.



Rinse all parts with cold, clear water (approx. 20 °C).
Clean all parts with plenty of warm soapy water (approx. 30 °C).


to steps 2, 3 and 4.
Dry with a clean cloth or leave to dry on a clean cloth.
If you clean the individual parts of the pump set in the dish- washer, place them on the top rack or in the cutlery section.
or · As an alternative

5.2 Before using for the first time and once a day
l Only use drinking-quality water
for cleaning. l Take apart and wash all parts that
come in contact with the breast and breastmilk immediately after use to avoid dry up of milk residues and to prevent growth of bacteria.
page 12, chapter 5.1
Disassemble the pump set into individual parts.
l When boiling, a tea spoon of citric
acid may be added to avoid a build- up of limescale. l Store the pump set in a clean bag/

container until next use. Or store it in clean paper or a cloth.
Use QuickClean* bags in the microwave in accordance with the instructions.
More information on · steps 2–4



Cover all parts with water and boil for five minutes.
Use B-Well Steam Steriliser* in accordance with the instructions.
5.3 Cleaning the motor unit


Wipe off with a clean, damp cloth.
Dry with a clean cloth or leave to dry on a clean cloth.
*More information at www.medela.com
or · or

6.1 Mains operation
l re 6.1 Mains operation:
Follow the exact order of the steps.

l Only use the mains adapter supplied
with the Swing. l Make sure the voltage of the mains
adapter is compatible with the power source.
Plug the mains adapter cord into the motor unit.
6.2 Battery operation
l re 6.2 Battery operation:
The batteries do not need to be removed when the main adapter is used for pumping.

l Regularly check the condition of
the batteries in the motor unit. l If Swing will remain unused
for a longer period of time, remove the batteries from the battery compartment. l Batteries last for two to three pump-
Open the battery compart- ment on the back in the direction of the arrow.
ing sessions (approx. 1 ½ hours pumping time).
Info


Plug the mains adapter into the mains socket.


Insert four AA Mignon LR6 batteries.
Close the battery compartment.

6.3 Assembling the pump set
l Wash hands thoroughly (at least one

minute) with soap and water before touching the breast and pump parts.
l Use Medela original accessories
only. l Check pump set components for
Carefully push the breastshield onto the connector.
wear or damage before use and replace if necessary. l Hold the tubing by the connecting
piece while attaching and detaching it. l All components must be completely
dry before use.

l Carry out all the steps with care and
assemble the pump set correctly. Otherwise a good vacuum may not be achieved.
Screw bottle into connector.
Info



Carefully push the valve membrane onto the valve head until it engages.
Push the valve head and mem- brane onto the connector. Position valve head laterally.


Push the yellow connecting piece onto the connector with a slight twist.
Insert the other end of the tubing into the motor unit.

7.1 Preparing for single pumping
l re step 2 (chapter 7.1):
Answer the following questions to find out whether the available breastshield size will fit:

l Is your nipple moving freely in
the tunnel? l Is minimal or no areolar tissue
being pulled into the tunnel of the breastshield? l Do you see a gentle, rhythmical
motion in the breast with each cycle of the pump? l Do you feel the breast emptying
Place the breastshield on your breast so that your nipple is properly centred in the tunnel.
all over? l Is your nipple pain-free?
If the answer to any of these ques- tions is “No” or if you experience problems or pain during expression, consult your breastfeeding and lactation consultant or midwife.
7.2 Pumping

l re step 2 (chapter 7.1):
Do not hold the pump set by the bottle. This can lead to blockage of the milk ducts and engorgement.
Switch on the breastpump with
l Wipe the breast with a warm
. The stimulation phase lasts
washcloth (do not use alcohol).
a maximum of two minutes. The stimulation phase will then automatically change to the expression phase.


Hold the breastshield onto your breast with your thumb and index finger. Support your breast with the palm of your hand.


If your milk starts to flow earlier, press the button to change to the expression phase. Make sure that the milk
The vacuum can be adjusted in every phase. Comfort vacuum: Use to increase the vacuum until it is slightly uncomfortable. Then press to dial it back minimally.
is flowing into the bottle properly.

l Contact your midwife or lactation
consultant if you can express only minimal or no milk or if expression is painful.
Switch off the breastpump with
. l Always disconnect the Swing from the
power source after expression. l Only fill the bottle to the 150 ml mark.
l Swing switches off automatically if
it has been running for 30 minutes with no manipulations. l Do not kink the tubing during
expression.
does not light up Off
even, regular flashing Stimulation phase
light constantly on Expression phase
rapid flashing Min./Max. vacuum
Info · Overview of LEDs · lights up On



page 12.
Use the stand to prevent the bottle from tipping over.
Close the bottle with a lid. Follow the instructions in
Chapter 8 “Storing and thawing breastmilk”.
Clean in accordance · with Chapter 5,

8.1 Storage For healthy, term babies:
12 months at approx. -18 °C
Do not store at room temperature
3–5 days at 4 °C
6 months at approx. -16 °C
Thawed breast- milk (previously frozen)
Do not store 10 hours Do not refreeze after thawing!
l Do not store breastmilk in the door of the refrigerator. Instead choose the coolest
point in the refrigerator (above the vegetable compartment).
8.2 Freezing
l Never fill bottles or “Pump & Save”* bags more than ¾ full to allow space for
possible expansion during freezing. l Label bottles or “Pump & Save” bags with the date of expression.
8.3 Thawing
Do not thaw frozen breastmilk in a microwave or in a pan of boiling water (danger of burning).
l In order to preserve the components of the breastmilk, thaw in the refrigerator
overnight or by holding the bottle or “Pump & Save” bag under running warm water (max. 37 °C). l Gently shake the bottle or “Pump & Save” bag to blend any fat that has separated. l With the B-Well bottle warmer*, breastmilk and baby food in glass jars can be gently
warmed to avoid any loss of vitamins and nutrients.



With Calma l your baby can feed, pause and
breathe regularly. l milk will only flow when a vacuum is
Calma – unique feeding solution for breastmilk.
created. l the milk flow stops if the baby wants
to breathe or pause while feeding. l your baby’s natural feeding behaviour
will be maintained supporting easy transition back to the breast.
Detailed information about Calma can be found in the enclosed instructions.
As a result of its comprehensive research work, Medela is one of the first companies in the world to incorporate babies’ breastfeeding behaviour into its 2-Phase Expression and Calma innovations. 2-Phase Expression is based on the baby’s natural suckling rhythm so that more milk can be expressed in less time. Short, stimulating patterns are followed by longer ones, providing optimal milk flow. The unique Calma breastmilk teat allows the baby to pause and breathe during feeding whilst maintaining the suckling behaviour learned at the breast. Developed for mother and baby for a happy breastfeeding time.


If the motor is not running
Check presence of mains current. Check the position of the batteries in the battery compartment.
Verify the valve head and membrane are clean and not damaged. Ensure the membrane is attached correctly to the valve head. Check the lateral position of the valve. Check all connections. Check the ends of the tubing fit securely into the back of the connector and into the port on the breastpump. Do not kink the tubing during expression. Ensure the breastshield forms a complete seal around the breast.
If there is low or no suction
If condensation appears in the tubing
Rinse out the tubing. Shake out any water droplets. Hang up the tubing to air-dry.
If milk gets into the tubing
Turn off the pump and unplug from power source. Remove and clean the tubing. Shake out water droplets and hang to air-dry.
Turn off the pump and unplug from power source. 1. Open the battery cover, remove the expanded-plastic liner. Wash the expanded-plastic liner with clear, lukewarm water and squeeze out well. Allow to dry thoroughly. 2. Pour about 200 ml of warm soapy water into a container. 3. Connect the tubing and the power cable onto the pump. 4. Hold the device with the control button upwards over a basin and switch on. Press and ensure that the lowest vacuum level is set by pressing twice. 5. Hold the end of the tubing that is not connected to the pump in the container with the soapy water. 6. The water is pumped through the tubing into the device. Allow it to pump until all the water has been flushed through. 7. Repeat procedure with 100 ml of clear, warm water. 8. Then leave the device switched on for 3 minutes, until the pump is completely dry. 9. As soon as the expanded-plastic liner is dry, put it in place again and close the battery cover. 10. If necessary, clean the pump set according to cleaning instruc- tions „Prior to first use and once daily“.
If milk gets into the pump
Problem Solution

The “International Warranty” brochure contains detailed warranty information.
The unit is made of various metal and plastics. Before disposal, the device is to be rendered unusable and it must not be disposed of as unsorted municipal waste in accordance with local regulations. Use your local return and collection system for waste electrical and electronic equipment (incl. batteries). Improper disposal may have harmful effects on the environment and on public health.
Warranty · Disposal

The following accessories for Swing are available from Medela distribution partners by specifying the product number.
099.0016 099.0017 099.0019 099.0018
Swing Adapter (230V; 4.8V) Euro Swing Adapter (240V; 4.8V) UK Swing Adapter (120V; 4.8V) US / JP Swing Adapter (240V; 4.8V) AUS / NZ
099.0015 Swing Motor Unit
800.0843 Swing Tubing
008.0033* 008.0035* 008.0036* 008.0037*
PersonalFit Breastshield S (21 mm) PersonalFit Breastshield L (27 mm) PersonalFit Breastshield XL (30 mm) PersonalFit Breastshield XXL (36 mm)
800.0796 200.1659 150 ml Breastmilk Bottles (3 pcs) 250 ml Breastmilk Bottles (2 pcs)
008.0139 Calma Solitaire
200.1756 200.0625 Cooling Bag CityStyle Breastpump Bag
Other Medela products are available on the www.medela.com web site.
* If the result of expression is unsatisfactory or expression is painful, contact your midwife or lactation consultant. A different size of PersonalFitTM Breastshield may help make expression more comfortable and successful.



The Swing breastpump needs special precautions regarding EMC (Electromagnetic compatibility) and needs to be installed and put into service according to the EMC information provided in these instructions for use. Wireless communications equip- ment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect the electric breastpump Swing and should be kept at least a distance 1.0 m away from the equipment. Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 1)
The electric breastpump Swing is intended for use in the electromagnetic environment speci- fied below. The customer or the user of the electric breastpump Swing should assure that it is used in such an environment.
Emission tests Compliance Electromagnetic environment – guidance
The electric breastpump Swing uses RF en- ergy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B
The electric breastpump Swing is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic pur- poses.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3
Electromagnetic emissions · Complies

Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)
The electric breastpump Swing is intended for use in the electromagnetic environment specified below. The customer or the user of the electric breastpump Swing should assure that it is used in such an environment.
Electromagnetic environment – guidance
Immunity test IEC 60601 test level
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrostatic discharge ( ESD ) IEC 61000-4-2
± 6 kV contact ± 8 kV air
± 6 kV contact ± 8 kV air
Electrical fast transient / burst IEC 61000-4-4
± 2 kV for power supply lines ± 1 kV for input / output lines
± 2 kV for power supply lines ± 1 kV for input/ output lines
Mains power quality should be that of a typical commercial or hospital environment.
± 1 kV line( s ) to line(s) ± 2 kV line( s ) to earth
± 1 kV differential mode ± 2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Electromagnetic immunity · Compliance level · Surge IEC 61000-4-5

Mains power qual- ity should be that of a typical commercial or hospital environment. If the user of the elec- tric breastpump Swing requires continued operation during power mains interruptions, it is recommended that the electric breastpump Swing is powered from an uninterruptible power supply or a battery.
95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles 95 % dip in UT) for 5 s
95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles 95 % dip in UT) for 5 s
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Power frequency mag- netic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
3 A/m 30 A/m

Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
The electric breastpump Swing is intended for use in the electromagnetic environment specified below. The customer or the user of the electric breastpump Swing should assure that it is used in such an environment.
Immunity test IEC 60601 test level
Electromagnetic environment – guidance
Portable and mobile RF communi- cations equipment should be used no closer to any part of the electric breastpump Swing, including cables, than the recommended separation distance calculated from the equation applicable to the fre- quency of the transmitter.
Recommended separation distance d = 0,35√P
3 Vrms 150 kHz to 80 MHz
Conducted RF IEC 61000-4-6
d = 0,35√P 80 MHz to 800 MHz d = 0,7√P 800 MHz to 2,5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF trans- mitters, as determined by an electro- magnetic site survey,a should be less than the compliance level in each fre- quency range.b
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2,5 GHz
Interference may occur in the vicinity of equipment marked with the follow- ing symbol:

affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele- phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the meas- ured field strength in the location in which the electric breastpump Swing is used exceeds the applicable RF compliance level above, the electric breastpump Swing should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the electric breastpump Swing.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 6)
Recommended separation distances between portable and mobile RF communications equipment and the electric breastpump Swing
The electric breastpump Swing is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the electric breastpump Swing can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the electric breastpump Swing as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter M
Rated maximum output power of transmitter W
80 MHz to 800 MHz d = 0,35√P
150 kHz to 80 MHz d = 0,35√P
800 MHz to 2,5 GHz d = 0,7√P
0,01 0,04 0,04 0,07
0,1 0,11 0,11 0,22
1 0,35 0,35 0,7
10 1,11 1,11 2,21
100 3,50 3,50 7,00
For transmitters rated at a maximum output power not listed above, the recommen- ded separation distance d in metres (m) can be estimated using the equation ap- plicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Environment Locations General characteristics
Partly controlled, covered by the general requirements of this collateral standard
Typical health care Hospital, large clinic, doctor’s office
Residential Doctor’s office, small clinic Not controlled, health care professional present
Not controlled, health care professional not normally present
Not controlled, wide vari- ations, critical receivers nearby, harsh environments for ESD, RF, electric and magnetic fields
Transport, mobile Car, aircraft (fixed-wing and helicopter), ambulance
Special Operating theatre, emergency room
Case-by-case examination of environment
Electromagnetic environments · Residential Home

Operation Betrieb En Service Utilizzo Gebruik Pengoperasian
vacuum (approx.) -15 -295 mmHg -2 -39 kPa 45 … 120 cpm
Transport / Storage Transport / Lagerung Transport / Stockage Transport / Stoccaggio Transport / Opslag Pengangkutan / Penyimpanan
100-240V~ 50/60Hz 230-240V~ 50Hz AUS/NZ only
4 x 1.5V, Alkaline AA, Mignon, LR6
121 x 121 x 57
Operation Betrieb En Service Utilizzo Gebruik Pengoperasian
Transport / Storage Transport / Lagerung Transport / Stockage Transport / Stoccaggio Transport / Opslag Pengangkutan / Penyimpanan

Medela AG Lättichstrasse 4b 6341 Baar, Switzerland www.medela.com
Netherlands & Belgium Medela Benelux BV Uilenwaard 31 5236 WB ’s-Hertogenbosch Netherlands Phone +31 73 690 40 40 Fax +31 73 690 40 44 [email protected] [email protected] www.medela.nl www.medela.be
International Sales Medela AG Lättichstrasse 4b 6341 Baar Switzerland Phone +41 (0)41 769 51 51 Fax +41 (0)41 769 51 00 [email protected] www.medela.com
United Kingdom Medela UK Ltd. Huntsman Drive Northbank Industrial Park Irlam, Manchester M44 5EG / UK Phone +44 (0)161 776 0400 Fax +44 (0)161 776 0444 info@medela.co.uk www.medela.co.uk
Germany & Austria Medela Medizintechnik GmbH & Co. Handels KG Postfach 1148 85378 Eching Germany Phone +49 (0)89 31 97 59-0 Fax +49 (0)89 31 97 59 99 [email protected] www.medela.de
Australia Medela Pty Ltd, Medical Technology 3 Arco Lane Heatherton, Vic 3202 Australia Phone +61 (0) 3 9552 8600 Fax +61 (0) 3 9552 8699 contact@medela.com.au www.medela.com.au
France Medela France Sarl 14, rue de la Butte Cordière 91154 Etampes cedex France Phone +33 (0)1 69 16 10 30 Fax +33 (0)1 69 16 10 32 [email protected] www.medela.fr
Switzerland Medela AG Lättichstrasse 4b 6341 Baar Switzerland Phone +41 (0)848 633 352 Fax +41 (0)41 769 51 00 [email protected] www.medela.ch
©Medela AG/200.4102/2013-04/F
Italy Medela Italia Srl Via Turrini, 19 – Loc. Bargellino 40012 Calderara di Reno (BO) Italy Phone + 39 051 72 76 88 Fax + 39 051 72 76 89 [email protected] www.medela.it
Pages mentionnées dans ce manuel — fabricant, assistance et téléchargements.
Problèmes fréquents, causes probables et solutions.
| Problème | Cause possible | Solution | |
|---|---|---|---|
| If the motor is not running | Check presence of mains current. Check the position of the batteries in the battery compartment. | p.26 | |
| If there is low or no suction | Verify the valve head and membrane are clean and not damaged. Ensure the membrane is attached correctly to the valve head. Check the lateral position of the valve. Check all connections. Check the ends of the tubing fit securely into the back of the connector and into the port on the breastpump. Do not kink the tubing during expression. Ensure the breastshield forms a complete seal around the breast. | p.26 | |
| If condensation appears in the tubing | Rinse out the tubing. Shake out any water droplets. Hang up the tubing to air-dry. | p.26 | |
| If milk gets into the tubing | Turn off the pump and unplug from power source. Remove and clean the tubing. Shake out water droplets and hang to air-dry. | p.26 | |
| If milk gets into the pump | Turn off the pump and unplug from power source. 1. Open the battery cover, remove the expanded-plastic liner. Wash the expanded-plastic liner with clear, lukewarm water and squeeze out well. Allow to dry thoroughly. 2. Pour about 200 ml of warm soapy water into a container. 3. Connect the tubing and the power cable onto the pump. 4. Hold the device with the control button upwards over a basin and switch on. Press and ensure that the lowest vacuum level is set by pressing twice. 5. Hold the end of the tubing that is not connected to the pump in the container with the soapy water. 6. The water is pumped through the tubing into the device. Allow it to pump until all the water has been flushed through. 7. Repeat procedure with 100 ml of clear, warm water. 8. Then leave the device switched on for 3 minutes, until the pump is completely dry. 9. As soon as the expanded-plastic liner is dry, put it in place again and close the battery cover. 10. If necessary, clean the pump set according to cleaning instruc- tions „Prior to first use and once daily“. | p.26 |