FR
Xantrex Technology logo
Notice d'utilisation officielle

Xantrex Technology XPH 75-2D

32 pages·799 KB·Alimentation électrique·PDF·ENNLJA
Xantrex Technology XPH 75-2D
Table des matières 70 sections About Xantrex · Trademarks · Notice of Copyright · Disclaimer · … AfficherMasquer
About Xantrexp.2Trademarksp.2Notice of Copyrightp.2Disclaimerp.2Date and Revisionp.2Part Numberp.2Contact Informationp.2Table of Contentsp.3Introductionp.4Specificationp.4OUTPUTSp.4OPERATING MODESp.5METER SPECIFICATIONSp.5GENERALp.5Safetyp.6EMCp.7Emissionsp.7Immunityp.7Cautionsp.7Installationp.8Mains Operating Voltagep.8Mains Leadp.8Connectionsp.8Operationp.8Setting Up the Outputp.8Constant Voltagep.8Constant Currentp.9Instantaneous Current Outputp.9Mode Selectionp.9p.9Caution. Turn both outputs off before changing mode. Possible damage to the circuit being powered, and to the unit itself, may occur if the mode is switched with either output still on.p.9Mode A: With the switch in this position the outputs operate independently and are isolated from one another. The voltage range on each output is 0-75V and the current range 0-2A.p.9Mode B: With the switch in this position only Output 2 is active; the voltage range is 0-75V and the current range is 0-4A. The controls of Output 1 are inoperative and the display is blanked. The terminals of Output 1 are isolated from those of Outpp.9Mode C: With the switch in this position only Output 2 is active; the voltage range is 0-150V and the current range is 0-2A. The controls of Output 1 are inoperative and the display is blanked. The red terminal of Output 1 is isolated but the black tp.9Protectionp.9Connection to the Loadp.10Series or Parallel Connection with Other Outputsp.10Ventilationp.10Maintenancep.10Fusep.10Cleaningp.10Calibrationp.11Access to Calibration Adjustmentsp.11Equipment Requiredp.11Mode A (75V/2A x 2)p.11Voltage Calibrationp.11Warranty Informationp.13Warrantyp.13How do you get service?p.13What does this warranty not cover?p.14Disclaimerp.14Productp.14Exclusionsp.14Warning: Limitations On Usep.14Return Material Authorization Policyp.15Return Procedurep.15Out of Warranty Servicep.152nd page of Calibration & Securite (EX752M).pdfp.1Current Calibrationp.12Voltage Regulationp.12Mode B (75V/4A)p.12Mode C (150V/2A)p.12Xantrex page 1.pdfp.1About Xantrexp.2Trademarksp.2Notice of Copyrightp.2Disclaimerp.2Date and Revisionp.2Part Numberp.2Contact Informationp.2
Page 1 of the Xantrex Technology XPH 75-2D user manual — Operating Manual
Page 1 · cliquer pour agrandir
Page 1

XPH 75-2D

Operating Manual

XPH Series Dual 75V 2A DC Power Supply

Page 2 of the Xantrex Technology XPH 75-2D user manual — Date and Revision
Page 2 · cliquer pour agrandir
Page 2

About Xantrex

Xantrex Technology Inc. is a world-leading supplier of advanced power electronics and controls with products from 50 watt mobile units to one MW utility-scale systems for wind, solar, batteries, fuel cells, microturbines, and backup power applications in both grid-connected and stand-alone systems. Xantrex products include inverters, battery chargers, programmable power supplies, and variable speed drives that convert, supply, control, clean, and distribute electrical power.

Trademarks

XPH series is a trademark of Xantrex International. Xantrex is a registered trademark of Xantrex International.

Other trademarks, registered trademarks, and product names are the property of their respective owners and are used herein for identification purposes only.

Disclaimer

UNLESS SPECIFICALLY AGREED TO IN WRITING, XANTREX TECHNOLOGY INC. (“XANTREX”)

(a) MAKES NO WARRANTY AS TO THE ACCURACY, SUFFICIENCY OR SUITABILITY OF ANY TECHNICAL OR OTHER INFORMATION PROVIDED IN ITS MANUALS OR OTHER DOCUMENTATION.

(b) ASSUMES NO RESPONSIBILITY OR LIABILITY FOR LOSS OR DAMAGE, WHETHER DIRECT, INDIRECT, CONSEQUENTIAL OR INCIDENTAL, WHICH MIGHT ARISE OUT OF THE USE OF SUCH INFORMATION. THE USE OF ANY SUCH INFORMATION WILL BE ENTIRELY AT THE USER’S RISK.

Date and Revision

July 2004 - Revision 2

Telephone1-800-670-0707 (toll free in North America)
Fax1-360-925-5143
EmailCustomerservice@ xantrex.com
Webwww.xantrex.com
Show 4 more labels / short text on this page
Part Number · 975-0102-01-02 · Contact Information · 1-360-925-5097 (direct)
Page 3 of the Xantrex Technology XPH 75-2D user manual — Table des matières
Page 3 · Table des matières
Table des matières

Table des matières

About Xantrexp.2Trademarksp.2Notice of Copyrightp.2Disclaimerp.2Date and Revisionp.2Part Numberp.2Contact Informationp.2Table of Contentsp.3Introductionp.4Specificationp.4OUTPUTSp.4OPERATING MODESp.5METER SPECIFICATIONSp.5GENERALp.5Safetyp.6EMCp.7Emissionsp.7Immunityp.7Cautionsp.7Installationp.8Mains Operating Voltagep.8Mains Leadp.8Connectionsp.8Operationp.8Setting Up the Outputp.8Constant Voltagep.8Constant Currentp.9Instantaneous Current Outputp.9Mode Selectionp.9p.9Caution. Turn both outputs off before changing mode. Possible damage to the circuit being powered, and to the unit itself, may occur if the mode is switched with either output still on.p.9Mode A: With the switch in this position the outputs operate independently and are isolated from one another. The voltage range on each output is 0-75V and the current range 0-2A.p.9Mode B: With the switch in this position only Output 2 is active; the voltage range is 0-75V and the current range is 0-4A. The controls of Output 1 are inoperative and the display is blanked. The terminals of Output 1 are isolated from those of Outpp.9Mode C: With the switch in this position only Output 2 is active; the voltage range is 0-150V and the current range is 0-2A. The controls of Output 1 are inoperative and the display is blanked. The red terminal of Output 1 is isolated but the black tp.9Protectionp.9Connection to the Loadp.10Series or Parallel Connection with Other Outputsp.10Ventilationp.10Maintenancep.10Fusep.10Cleaningp.10Calibrationp.11Access to Calibration Adjustmentsp.11Equipment Requiredp.11Mode A (75V/2A x 2)p.11Voltage Calibrationp.11Warranty Informationp.13Warrantyp.13How do you get service?p.13What does this warranty not cover?p.14Disclaimerp.14Productp.14Exclusionsp.14Warning: Limitations On Usep.14Return Material Authorization Policyp.15Return Procedurep.15Out of Warranty Servicep.152nd page of Calibration & Securite (EX752M).pdfp.1Current Calibrationp.12Voltage Regulationp.12Mode B (75V/4A)p.12Mode C (150V/2A)p.12Xantrex page 1.pdfp.1About Xantrexp.2Trademarksp.2Notice of Copyrightp.2Disclaimerp.2Date and Revisionp.2Part Numberp.2Contact Informationp.2
Page 4 of the Xantrex Technology XPH 75-2D user manual — Specification OUTPUTS
Page 4 · cliquer pour agrandir
Page 4

Introduction

The XPH 75-2D is a dual output 300 watt PSU with multi-mode capability in a compact and attractive case.

The multi-mode capability enables it to operate as a dual power supply with two independent and isolated outputs, or as a single power supply of double the power. As a dual, each output provides 0 to 75V at 0 to 2A (Mode A). As a single the output can be selected as either 0 to 75V at 0 to 4A (Mode B) or 0 to 150V at 0 to 2A (Mode C). In single modes, the unused half of the unit becomes completely inoperative and its displays are blanked.

Because of the higher voltages it can produce, the XPH 75-2D uses special safety binding posts which can accept spade terminals, 4mm plugs, or bare-end wires without exposing any metal parts.

The XPH series incorporates separate digital voltage and current meters on each output. The meters use bright 14mm (0.56”) LED displays and have an update rate of 4 per second providing near instantaneous response. Simultaneous metering of voltage and current provides accurate information “at a glance” and avoids any possibility of misinterpretation. When an output switch is set to “off”, the current limit setting is displayed enabling conditions to be set before the load is connected.

Excellent line and load regulation are matched by low noise and good transient response. High power efficiency ensures that the unit remains cool without any fan noise.

The XPH 75-2D has been designed to meet the stringent requirements of relevant IEC standards for safety and EMC, including harmonics emissions. All outputs are intrinsically short circuit proof, and are protected against external voltages and reverse currents.

Specification OUTPUTS

Voltage Range: 0V to 75V minimum, Modes A and B. 0V to 150V minimum, Mode C.

Current Range: 0A to 2A minimum, Modes A and C. 0A to 4A minimum, Mode B.

Output Voltage Setting: By coarse and fine controls.

Output Current Setting: By single logarithmic control.

Operating Mode: Constant voltage or constant current with automatic cross-over.

Output Switch: Electronic. Preset voltage and current displayed when off.

Output Terminals: 4mm safety terminals on 19mm (0.75”) pitch.

Output Impedance: Typically 50kΩ in constant current mode.

Output Protection: Output will withstand forward voltage of up to 85V (Modes A and B) or 170V (Mode C). Reverse protection by diode clamp for reverse currents up to 3A.

Load & Line Regulation:

Ripple & Noise (20MHz bandwidth):

Typically

Transient Response:

Temperature CoefficientTypically <100ppm/°C

Status Indication: Output on lamp. Constant current mode lamp.

Page 5 of the Xantrex Technology XPH 75-2D user manual — GENERAL
Page 5 · cliquer pour agrandir
Page 5

OPERATING MODES

Mode Selection: By front panel rotary switch.

Mode A: Independent outputs each capable of 0-75V at 0-2A.

Mode B: Output 2 only active, 0-75V at 0-4A.

Output 1 disabled.

Mode C: Output 2 only active, 0-150V at 0-2A.

Output 1 disabled.

METER SPECIFICATIONS

Meter Types: Dual 3 digit meters with 14mm (0.56") LEDs. Reading rate 4 Hz.

Meter Resolutions100mV, 10mA
Meter AccuraciesVoltage 0.3% of reading ± 1 digit, Current 0.6% of reading ± 1 digit

Operating Modes: Output 1 meters are blanked in Modes B and C.

GENERAL

AC Input: 110V-240V AC ±10%, 50/60Hz. Installation Category II.

Power Consumption: 500VA max.

Operating Range: +5ºC to +40ºC, 20% to 80% RH.

Storage Range: −40ºC to + 70ºC.

Environmental: Indoor use at altitudes up to 2000m, Pollution Degree 2.

Safety: Complies with EN61010-1.

EMC: Complies with EN61326.

Size: 260 x 160 x 320mm (WxHxD).

Weight4.3kg
Page 6 of the Xantrex Technology XPH 75-2D user manual — WARNING! THIS INSTRUMENT MUST BE EARTHED
Page 6 · cliquer pour agrandir
Page 6

Safety

This power supply is a Safety Class I instrument according to IEC classification and has been designed to meet the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for operation from a normal single phase supply.

This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe condition. This instruction manual contains some information and warnings which have to be followed by the user to ensure safe operation and to retain the instrument in a safe condition.

This instrument has been designed for indoor use in a Pollution Degree 2 environment in the temperature range 5°C to 40°C, 20% - 80% RH (non-condensing). It may occasionally be subjected to temperatures between +5°C and –10°C without degradation of its safety. Do not operate while condensation is present.

Use of this instrument in a manner not specified by these instructions may impair the safety protection provided. Do not operate the instrument outside its rated supply voltages or environmental range.

WARNING! THIS INSTRUMENT MUST BE EARTHED

Any interruption of the mains earth conductor inside or outside the instrument will make the instrument dangerous. Intentional interruption is prohibited. The protective action must not be negated by the use of an extension cord without a protective conductor.

When the instrument is connected to its supply, terminals may be live and opening the covers or removal of parts (except those to which access can be gained by hand) is likely to expose live parts. The apparatus shall be disconnected from all voltage sources before it is opened for any adjustment, replacement, maintenance or repair. Capacitors inside the power supply may still be charged even if the power supply has been disconnected from all voltage sources but will be safely discharged about 10 minutes after switching off power.

Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of the hazard involved.

If the instrument is clearly defective, has been subject to mechanical damage, excessive moisture or chemical corrosion the safety protection may be impaired and the apparatus should be withdrawn from use and returned for checking and repair.

Make sure that only fuses with the required rated current and of the specified type are used for replacement. The use of makeshift fuses and the short-circuiting of fuse holders is prohibited.

Do not wet the instrument when cleaning it.

The following symbols are used on the instrument and in this manual:-

Earth (ground) terminal.

mains supply OFF.

l mains supply ON.

alternating current (ac) · direct current (dc)

Page 7 of the Xantrex Technology XPH 75-2D user manual — Cautions
Page 7 · cliquer pour agrandir
Page 7

EMC

This instrument has been designed to meet the requirements of the EMC Directive 89/336/EEC.

Compliance was demonstrated by meeting the test limits of the following standards:

Emissions

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits used were:

a) RadiatedClass B
b) ConductedClass B

c) Harmonics: EN61000-3-2 (2000) Class A; the instrument is Class A by product category.

Immunity

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use.

Test methods, limits and performance achieved were:

a) EN61000-4-2 (1995) Electrostatic Discharge : 4kV air, 4kV contact, Performance A.

b) EN61000-4-3 (1997) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance B.

c) EN61000-4-11 (1994) Voltage Interrupt, 1 cycle, 100%, Performance B.

d) EN61000-4-4 (1995) Fast Transient, 1kV peak (AC line), 0.5kV peak (DC Outputs), Performance B.

e) EN61000-4-5 (1995) Surge, 0.5kV (line to line), 1kV (line to ground), Performance B.

f) EN61000-4-6 (1996) Conducted RF, 3V, 80% AM at 1kHz (AC line only; DC Output connections

According to EN61326 the definitions of performance criteria are:

Performance criterion A‘During test normal performance within the specification limits.’

Performance criterion B: ‘During test, temporary degradation, or loss of function or performance which is self-recovering’.

Performance criterion C: ‘During test, temporary degradation, or loss of function or performance which requires operator intervention or system reset occurs.’

Where Performance B is stated it is because DC Output regulation may deviate beyond Specification limits under the test conditions. However, the possible deviations are still small and unlikely to be a problem in practice.

Cautions

To ensure continued compliance with the EMC directive observe the following precautions:

a) after opening the case for any reason ensure that all signal and ground connections are

remade correctly and that case screws are correctly refitted and tightened.

b) In the event of part replacement becoming necessary, only use components of an identical

type, see the Service Manual.

Page 8 of the Xantrex Technology XPH 75-2D user manual — Constant Voltage
Page 8 · cliquer pour agrandir
Page 8

Installation

Mains Operating Voltage

This instrument has a universal input range and will operate from a nominal 115V or 230V mains supply without adjustment. Check that the local supply meets the AC Input requirement given in the Specification. Mains Lead

When a three core mains lead with bare ends is provided this should be connected as follows:

BROWN - MAINS LIVE

BLUE - MAINS NEUTRAL GREEN/YELLOW - EARTH Safety Earth Symbol When fitting a fused plug a 5 amp fuse should be fitted inside the plug. As the colours of the wires in the mains lead of this apparatus may not correspond with the coloured markings identifying the terminals in your plug proceed as follows: The wire which is coloured green-and-yellow must be connected to the terminal in the plug which is marked by the letter E or by the safety earth symbol shown above or coloured green or green- and-yellow. The wire which is coloured blue must be connected to the terminal which is marked with the letter N or coloured black. The wire which is coloured brown must be connected to the terminal which is marked with the letter L or coloured red.

Connections

All connections are made from the front panel. The load should be connected to the positive (red) and negative (black) terminals marked OUTPUT. Connect to one or both outputs as appropriate for the selected Mode, see Operation section.

The terminal marked is connected to the chassis and safety earth ground.

Operation

The operation of both outputs is identical in Mode A (independent mode); the following description applies to both. In Mode B and Mode C only Output 2 is active and Output 1 is disabled; the description therefore only applies to Output 2, see Mode Selection section.

Setting Up the Output

With the POWER switch on (l) and the output off the output voltage and current limit can be accurately preset using the VOLTAGE and CURRENT controls; the left-hand meter shows the set voltage and the right-hand meter shows the set maximum current.

When the output switch is switched on, the ON lamp lights; the left-hand meter now shows the actual voltage and the right-hand meter the actual load current.

Constant Voltage

The output voltage is adjusted using the coarse and fine VOLTAGE controls; the CURRENT control sets the maximum current that can be supplied.

Page 9 of the Xantrex Technology XPH 75-2D user manual — Protection
Page 9 · cliquer pour agrandir
Page 9

Constant Current

If the load resistance is low enough such that, at the output voltage set, a current greater than the current limit setting would flow, the power supply will automatically move into constant current operation. The current output is adjusted by the CURRENT control and the VOLTAGE controls set the maximum voltage that can be generated.

The CC lamp lights to show constant current mode.

Instantaneous Current Output

The current limit control can be set to limit the continuous output current to levels down to 10mA. However, in common with all precision bench power supplies, a capacitor is connected across the output to maintain stability and good transient response. This capacitor charges to the output voltage and short-circuiting of the output will produce a current pulse as the capacitor discharges which is independent of the current limit setting.

Mode Selection

The three operating modes are selected by the front panel rotary switch situated between the terminals.

Mode A: With the switch in this position the outputs operate independently and are isolated from one another. The voltage range on each output is 0-75V and the current range 0-2A.

Mode B: With the switch in this position only Output 2 is active; the voltage range is 0-75V and the current range is 0-4A. The controls of Output 1 are inoperative and the display is blanked. The terminals of Output 1 are isolated from those of Output 2.

Note also that the range and scaling of the current limit control doubles when changing from Mode A to Mode B, i.e. with the control set for a 1.5A limit in Mode A, the limit becomes 3A in Mode B. To avoid possible damage to the circuit being tested the outputs should be switched off before the mode is changed and the current limit set to the desired value before the output is switched on.

Mode C: With the switch in this position only Output 2 is active; the voltage range is 0-150V and the current range is 0-2A. The controls of Output 1 are inoperative and the display is blanked. The red terminal of Output 1 is isolated but the black terminal remains connected internally to the black terminal of Output 2.

Note also that the range and scaling of the voltage controls of Output 2 double when changing from Mode B to Mode C, i.e. with the controls set for 60V in Mode B the output voltage becomes 120V in Mode C. Care should therefore be taken to ensure that the outputs are switched off before the mode is changed and that the output is set to the correct voltage before it is switched on.

Protection

The output has intrinsic short-circuit protection and is protected from reverse voltages by a diode; the continuous reverse current must not exceed 3 Amps, although transients can be much higher.

The output is protected against externally applied forward voltages of up to 85V (Modes A and B) or 170V (Mode C).

Page 10 of the Xantrex Technology XPH 75-2D user manual — Maintenance
Page 10 · cliquer pour agrandir
Page 10

Connection to the Load

The load should be connected to the positive (red) and negative (black) output terminals. To provide protection from the high voltages that the unit can supply, the special terminals will accept 4mm plugs, bare-ended wires or spade terminals without exposing any metal parts. WARNING! Switch the output off before making connections to the terminals. Always check that there are no exposed connecting wires before switching the output on. Both terminals are fully floating and either can be connected to ground. Series or Parallel Connection with Other Outputs

The outputs of the power supply are fully floating and may be used in series with other power supply units to generate high DC voltages up to 300V DC.

The maximum permissible voltage between any terminal and earth ground ( ) is 300VDC; the maximum permissible voltage between either terminal of one output and any terminal of another output on the same supply is also 300VDC.

WARNING! All connections to the terminals must be made with the output switched off on all units. It should be noted that the unit can only source current and cannot sink it, thus units cannot be series connected in anti-phase. The unit can be connected in parallel with others to produce higher currents. Where several units are connected in parallel, the output voltage will be equal to that of the unit with the highest output voltage setting until the current drawn exceeds its current limit setting, upon which the output will fall to that of the next highest setting, and so on. In constant current mode, units can be connected in parallel to provide a current equal to the sum of the current limit settings. Ventilation

The power supply is very efficient but nevertheless can generate significant heat at full power. The supply relies on convection cooling only and it is therefore important that ventilation is never restricted if performance and safety are to be maintained.

Maintenance

The Manufacturers or their agents overseas will provide repair for any unit developing a fault. Where owner wish to undertake their own maintenance work, this should only be done by skilled personnel in conjunction with the service manual which may be purchased directly from the Manufacturers or their agents overseas. Fuse

The correct fuse type is:

10 Amp 250V HBC time-lag(T), 5 x 20mm. Note that the main function of the fuse is to make the instrument safe and limit damage in the event of failure of one of the switching devices. If a fuse fails it is therefore very likely that the replacement will also blow, because the supply has developed a fault; in such circumstances the instrument will need to be returned to the manufacturer for service. Make sure that only fuses of the required rated current and specified type are used for replacement. The use of makeshift fuses and the short-circuiting of fuse-holders is prohibited. To replace a fuse, first disconnect the instrument from the AC supply. Remove the 6 cover securing screws and lift off the cover. Replace the fuse with one of the correct type and refit the cover. Cleaning

If the PSU requires cleaning use a cloth that is only lightly dampened with water or a mild detergent. Polish the display window with a soft dry cloth. WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE PSU, NEVER ALLOW WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE OR DISPLAY WINDOW NEVER CLEAN WITH SOLVENTS.

Page 11 of the Xantrex Technology XPH 75-2D user manual — Voltage Calibration
Page 11 · cliquer pour agrandir
Page 11

Calibration Allow at least 5 minutes warm-up before commencing calibration.

Access to Calibration Adjustments

All adjustments are on the front panel control board; all the trimmers are adjusted from the reverse side of the board through holes in the board. To gain access to the board it is necessary to remove the top cover. WARNING! When the instrument is connected to its supply the removal of covers is likely to expose live parts. The instrument should be disconnected from all voltage sources before it is opened for any adjustment, replacement, maintenance or repair. Capacitors inside the power supply may still be charged even if the power supply has been disconnected from all voltage sources but will be safely discharged about 5 minutes after switching off power. Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of the hazard involved. Remove the 6 side screws and the front handle screw to release the top cover. Component adjustment references are given as Channel A/ Channel B, e.g. VR5/VR105; Channel A is the left-hand output.

Xantrex Technology XPH 75-2D — Access to Calibration Adjustments diagram (page 11)
Fig. 1 · Access to Calibration Adjustments

Control pcb calibration points (viewed from rear) Equipment Required

A 5½ digit multimeter with better than 0.05% accuracy on dc volts and better than 0.1% accuracy on dc current (to 4A); alternatively use a precision shunt for current measurement.

Rheostat or other high power load arrangement to provide up to 4A load at 75V. Mode A (75V/2A x 2)

Adjust Channel A and Channel B independently as follows.

Voltage Calibration

Connect the DMM (set to Volts) across the output. Set voltage and current controls to minimum. Switch output ON (Check LED is on) and check for a reading of 00.0V ± 0.1V on the Volts display and DMM; check the Amps display reads 0.00 ± 0.01A.

Set voltage and current controls to maximum. Adjust VR5/VR105 (maximum output volts) for a reading of 75.35V to 75.45V on the DMM. Adjust VR6/VR106 (measured output volts) until the Volts display matches the reading on the external DMM. Switch output OFF. Adjust VR10/110 (preset volts) until the Volts display shows 75.4.

Page 12 of the Xantrex Technology XPH 75-2D user manual — Mode C (150V/2A)
Page 12 · cliquer pour agrandir
Page 12

Current Calibration

Switch output OFF. Set output voltage to nominally 2V. Set current control to minimum. Connect the DMM (set to Amps) and load in series across the output. Switch output ON.

Adjust VR8/108 (offset compensation of current control error amp) for a reading of 0.003A ± .001A on the DMM. Check that the CC LED is ON.

Increase voltage controls and current control to maximum. Adjust load until the DMM reads 2.00A ± 0.02A. Adjust VR7/107 (measured output current) until the Amps display matches the DMM reading.

Reduce load until the CC LED is ON. Adjust VR4/104 (maximum output current) until the Amps display shows 2.05.

Voltage Regulation

Connect the DMM (set to Volts) across the output, no load, output ON.

Adjust voltage controls for a reading of 18.xxxx on the DMM; note exact reading.

Connect load, set current control to maximum and adjust load for 2A output current. Note that the DMM should be connected at the ‘back’ of the output terminals, closest to the internal sense connection, and the load plugged in to the 4mm sockets of the terminals. Adjust VR9/109 (differential voltage gain) until the external DMM matches the previous reading exactly.

Mode B (75V/4A)

With both outputs OFF select Mode B (75V/4A). Set the voltage and current controls to maximum and check that the current display (preset maximum current) reads ≥ 4.05 Amps. Connect the rheostat, switch the output ON and adjust the load until the CC LED is just ON; check that the current display (measured output current) matches the preset current ± 0.05 Amps.

Mode C (150V/2A)

With no loads connected and both outputs OFF select Mode C (150V/2A). Switch the output ON and check that the reading on the Volts display (measured output volts) is ≥ 150.1V. Switch the output OFF and adjust VR111 to give the same reading on the Volts display (preset output volts).

Page 13 of the Xantrex Technology XPH 75-2D user manual — Warranty
Page 13 · cliquer pour agrandir
Page 13

Warranty Information

Warranty

What does this warranty cover?

This Limited Warranty is provided by Xantrex Technology, Inc. ("Xantrex") and covers defects in workmanship and materials in your XPH series. This warranty lasts for a Warranty Period of three years (3 years) from the date of purchase at point of sale to you, the original end user customer.

What will Xantrex do?

Xantrex will, at its option, repair or replace the defective product free of charge, provided that you notify Xantrex of the product defect within the Warranty Period, and provided that Xantrex through inspection establishes the existence of such a defect and that it is covered by this Limited Warranty.

Xantrex will, at its option, use new and/or reconditioned parts in performing warranty repair and building replacement products. Xantrex reserves the right to use parts or products of original or improved design in the repair or replacement. If Xantrex repairs or replaces a product, its warranty continues for the remaining portion of the original Warranty Period or 90 days from the date of the return shipment to the customer, whichever is greater. All replaced products and all parts removed from repaired products become the property of Xantrex.

Xantrex covers both parts and labor necessary to repair the product, and return shipment to the customer via a Xantrex-selected non-expedited surface freight within the contiguous United States and Canada. Alaska and Hawaii are excluded. Contact Xantrex Customer Service for details on freight policy for return shipments outside of the contiguous United States and Canada.

How do you get service?

If your product requires troubleshooting or warranty service, contact your merchant. If you are unable to contact your merchant, or the merchant is unable to provide service, contact Xantrex directly at:

Phone1-800-670-0707 (toll free),
Fax1-360-925-5143
EmailCustomerservice@ xantrex.com

Direct returns may be performed according to the Xantrex Return Material Authorization Policy described in your product manual. For some products, Xantrex maintains a network of regional Authorized Service Centers. Call Xantrex or check our website to see if your product can be repaired at one of these facilities.

In any warranty claim, dated proof of purchase must accompany the product and the product must not have been disassembled or modified without prior written authorization by Xantrex.

Proof of purchase may be in any one of the following forms:

  • The dated purchase receipt from the original purchase of the product at point of sale to the end user, or
  • The dated dealer invoice or purchase receipt showing original equipment manufacturer (OEM) status, or
  • The dated invoice or purchase receipt showing the product exchanged under warranty

1-360-925-5097 (direct)

Page 14 of the Xantrex Technology XPH 75-2D user manual — Disclaimer Product
Page 14 · cliquer pour agrandir
Page 14

What does this warranty not cover?

This Limited Warranty does not cover normal wear and tear of the product or costs related to the removal, installation, or troubleshooting of the customer's electrical systems. This warranty does not apply to and Xantrex will not be responsible for any defect in or damage to:

a) the product if it has been misused, neglected, improperly installed, physically damaged or altered,

either internally or externally, or damaged from improper use or use in an unsuitable environment;

b) the product if it has been subjected to fire, water, generalized corrosion, biological infestations, or

input voltage that creates operating conditions beyond the maximum or minimum limits listed in the Xantrex product specifications including high input voltage from generators and lightning strikes;

c) the product if repairs have been done to it other than by Xantrex or its authorized service centers

d) the product if it is used as a component part of a product expressly warranted by another

e) the product if its original identification (trade-mark, serial number) markings have been defaced,

altered, or removed.

Disclaimer Product

THIS LIMITED WARRANTY IS THE SOLE AND EXCLUSIVE WARRANTY PROVIDED BY XANTREX IN CONNECTION WITH YOUR XANTREX PRODUCT AND IS, WHERE PERMITTED BY LAW, IN LIEU OF ALL OTHER WARRANTIES, CONDITIONS, GUARANTEES, REPRESENTATIONS, OBLIGATIONS AND LIABILITIES, EXPRESS OR IMPLIED, STATUTORY OR OTHERWISE IN CONNECTION WITH THE PRODUCT, HOWEVER ARISING (WHETHER BY CONTRACT, TORT, NEGLIGENCE, PRINCIPLES OF MANUFACTURER'S LIABILITY, OPERATION OF LAW, CONDUCT, STATEMENT OR OTHERWISE), INCLUDING WITHOUT RESTRICTION ANY IMPLIED WARRANTY OR CONDITION OF QUALITY, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE EXTENT REQUIRED UNDER APPLICABLE LAW TO APPLY TO THE PRODUCT SHALL BE LIMITED IN DURATION TO THE PERIOD STIPULATED UNDER THIS LIMITED WARRANTY.

IN NO EVENT WILL XANTREX BE LIABLE FOR ANY SPECIAL, DIRECT, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES, LOSSES, COSTS OR EXPENSES HOWEVER ARISING WHETHER IN CONTRACT OR TORT INCLUDING WITHOUT RESTRICTION ANY ECONOMIC LOSSES OF ANY KIND, ANY LOSS OR DAMAGE TO PROPERTY, ANY PERSONAL INJURY, ANY DAMAGE OR INJURY ARISING FROM OR AS A RESULT OF MISUSE OR ABUSE, OR THE INCORRECT INSTALLATION, INTEGRATION OR OPERATION OF THE PRODUCT.

Exclusions If this product is a consumer product, federal law does not allow an exclusion of implied warranties. To the extent you are entitled to implied warranties under federal law, to the extent permitted by applicable law they are limited to the duration of this Limited Warranty. Some states and provinces do not allow limitations or exclusions on implied warranties or on the duration of an implied warranty or on the limitation or exclusion of incidental or consequential damages, so the above limitation(s) or exclusion(s) may not apply to you. This Limited Warranty gives you specific legal rights. You may have other rights which may vary from state to state or province to province. Warning: Limitations On Use Please refer to your product manual for limitations on uses of the product. SPECIFICALLY, PLEASE NOTE THAT THE XPH SERIES SHOULD NOT BE USED IN CONNECTION WITH LIFE SUPPORT SYSTEMS OR OTHER MEDICAL EQUIPMENT OR DEVICES. WITHOUT LIMITING THE GENERALITY OF THE FOREGOING, XANTREX MAKES NO REPRESENTATIONS OR WARRANTIES REGARDING THE USE OF THE XANTREX XPH SERIES IN CONNECTION WITH LIFE SUPPORT SYSTEMS OR OTHER MEDICAL EQUIPMENT OR DEVICES. Please note that the XPH series is not intended for use as an uninterruptible power supply and Xantrex makes no warranty or representation in connection with any use of the product for such purposes.

(hereafter "ASCs"); · manufacturer;

Page 15 of the Xantrex Technology XPH 75-2D user manual — Out of Warranty Service
Page 15 · cliquer pour agrandir
Page 15

Return Material Authorization Policy

Before returning a product directly to Xantrex you must obtain a Return Material Authorization (RMA) number and the correct factory "Ship To" address. Products must also be shipped prepaid. Product shipments will be refused and returned at your expense if they are unauthorized, returned without an RMA number clearly marked on the outside of the shipping box, if they are shipped collect, or if they are shipped to the wrong location.

When you contact Xantrex to obtain service, please have your instruction manual ready for reference and be prepared to supply:

  • The serial number of your product
  • Information about the installation and use of the unit
  • Information about the failure and/or reason for the return
  • A copy of your dated proof of purchase

Return Procedure

1. Package the unit safely, preferably using the original box and packing materials. Please ensure that your product is shipped fully insured in the original packaging or equivalent. This warranty will not apply where the product is damaged due to improper packaging.
2. Include the following:
  • The RMA number supplied by Xantrex Technology, Inc. clearly marked on the outside of the box.
  • A return address where the unit can be shipped. Post office boxes are not acceptable.
  • A contact telephone number where you can be reached during work hours.
  • A brief description of the problem.
3. Ship the unit prepaid to the address provided by your Xantrex customer service representative.

If you are returning a product from outside of the USA or Canada

In addition to the above, you MUST include return freight funds and are fully responsible for all documents, duties, tariffs, and deposits.

If you are returning a product to a Xantrex Authorized Service Center (ASC)

A Xantrex return material authorization (RMA) number is not required. However, you must contact the ASC prior to returning the product or presenting the unit to verify any return procedures that may apply to that particular facility.

Out of Warranty Service

If the warranty period for your XPH series has expired, if the unit was damaged by misuse or incorrect installation, if other conditions of the warranty have not been met, or if no dated proof of purchase is available, your inverter may be serviced or replaced for a flat fee.

To return your XPH series for out of warranty service, contact Xantrex Customer Service for a Return Material Authorization (RMA) number and follow the other steps outlined in Return Procedure on page 28.

Payment options such as credit card or money order will be explained by the Customer Service Representative. In cases where the minimum flat fee does not apply, as with incomplete units or units with excessive damage, an additional fee will be charged. If applicable, you will be contacted by Customer Service once your unit has been received.

Page 16 of the Xantrex Technology XPH 75-2D user manual — Return Material Authorization Policy
Page 16 · cliquer pour agrandir
Page 16

Xantrex Technology Inc.

8999 Nelson Way Burnaby, British Columbia Canada V5A 4B5

604 422 8595 Tel 604 421 3056 Fax 800 667 8422 Toll Free North America

prg.info@xantrex.com www.xantrex.com

975-0102-01-02 48511-0720 Issue 2

Sources

Références et liens

Pages mentionnées dans ce manuel — fabricant, assistance et téléchargements.

Poser une question sur : Xantrex Technology XPH 75-2D

Les utilisateurs et notre équipe répondent généralement sous 24 h — et votre question servira aussi aux suivants.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Aucune question pour l'instant sur le Xantrex Technology XPH 75-2D

Soyez le premier à poser une question — d'autres utilisateurs ou notre équipe y répondront.

Document d'origine

Xantrex Technology XPH 75-2D manuel PDF

32 pages·799 KB·Alimentation électrique·PDF
Cliquez sur une page ci-dessus pour l'ouvrir ici
Page 1 / …
100%
Chargement des pages…