Aller à la page of 16
USER MANUAL AIR-O XYGEN BLENDER (DIS S and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS CA UTION F ederal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA F ax: (+001) 610-262-6080 www .
1 Blender Air-Oxygen CONTENTS RECEIVING / INSPECTION ......................................................... 2 INTENDED USE .......................................................................... 2 READ ALL INSTRUCTIONS BEFORE USING .............
2 Blender Air-Oxygen RECEIVING / INSPECTION Remove the Precision Medical, Inc. Air -Oxygen Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider .
3 Blender Air-Oxygen SAFETY INFORMA TION - W ARNINGS AND CAUTIONS DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury . W ARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury .
4 Blender Air-Oxygen W ARNING An Oxygen Analyzer/Monitor must be used to verify oxygen concentration. Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender , and 3 lpm for the Low Flow Blender .
5 Blender Air-Oxygen SPECIFICA TIONS Model PM5200 High Flow PM5300 Low Flow Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 . 4 b a r ) w i t h N O B L E E D Auxiliary Outlet Flow Range 2 - 100 lpm 0 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 .
6 Blender Air-Oxygen SPECIFICA TIONS continued Dimensions: (without fittings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.4 cm) W eight: 2.
7 Blender Air-Oxygen DIAGRAMS CA UTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. Depending on model, your fittings may differ from these diagrams.
8 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A d ia l u se d f or se le ct ing ox yg en c on ce ntr at ion s b et we e n 21%-100%. The FiO 2 scale is used for reference only . This Dial does not rotate 360°.
9 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION F Air Inlet Fitting A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose . G Alarm An audible alarm that sounds due to an excessive pressure drop or deletion of either gas supply .
10 Blender Air-Oxygen ALARM TEST 1. The Air -Oxygen Blender is installed and the flowmeter is turned on. Disconnect or turn off the air supply line to the Air -Oxygen Blender . 2. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly .
11 Blender Air-Oxygen 3. T o activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover . 4. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor . 5. Confirm the flow of air and/or oxygen mixture to the patient.
12 Blender Air-Oxygen 1. Disconnect all gas connections and equipment before cleaning. 2. Clean exterior surfaces with a cloth dampened with mild detergent and water .
13 Blender Air-Oxygen TROUBLESHOOTING If the Air -Oxygen Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide refer to the Air -Oxygen Blender Service Manual (P/N 504827) available on the Internet; www .
14 Blender Air-Oxygen LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precision Medical, Inc. warrants that the Blender , (the Product), will be free of defects in workmanship and/or material for the .
DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo E.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Precision Medical PM5200 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Precision Medical PM5200 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Precision Medical PM5200, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Precision Medical PM5200 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Precision Medical PM5200, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Precision Medical PM5200.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Precision Medical PM5200. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Precision Medical PM5200 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.