Dealer: This manual MUST be given to the end user. User: BEFORE using this product, read this manual and save for future reference. Operator’s Manual For more information regarding Invacare products, parts, and services, please visit www.
WARNING Perfecto 2 ™ Series 2 Part No 114348 2 WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTAN DING THIS MANUAL. IF YOU ARE UN ABLE TO UNDERSTAND THE WARNINGS, CAUTIONS.
TABLE OF CONTENTS Part No 1143482 3 Perfecto 2 ™ Series Table of Contents ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 5 SECTION 1—GENERAL GUIDELINES ...............
Table of Contents TABLE OF CONTENTS Perfecto 2 ™ Series 4 Part No 114348 2 Indicators (If Your Unit Does Not Feature the O 2 Sensor) ... 24 SECTION 6—MAINTENANCE .......................... 25 Routine Maintenanc e ....................... ..........
SPECIAL NOTES Part No 1143482 5 Perfecto 2 ™ Series SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could re sult in personal injury or property damage. Refer to the following table for definitions of the signal words.
SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 6 Part No 114348 2 SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the P erfecto 2 concentrator these instructions MUST be followed.
SECTION 1—GENERAL GUIDELINES Part No 1143482 7 Perfecto 2 ™ Series DANGER The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air , and some that will no t, are easily ignited and burn rapidly in high concentrations of oxygen .
SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 8 Part No 114348 2 WARNING NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air op ening may be blocked. Keep the openings free from lint, hair and the like.
SECTION 1—GENERAL GUIDELINES Part No 1143482 9 Perfecto 2 ™ Series WARNING To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To Persons. A void using while bathing. If co ntinuous usage is required by the physician’ s prescription, the concentrator MUST be located in another room at least seven feet from the bath.
SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 10 Part No 1143482 WARNING Use of some administration acce ssories or certain humidifiers not specified for use with oxyge n concentrator may impair the performance. In certain circumstances oxygen therapy can be hazardous.
SECTION 2—FEATURES Part No 1143482 11 Perfecto 2 ™ Series SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault and Power Indicator Lights Power Switch Circuit Breake.
SECTION 3—HANDLING Perfecto 2 ™ Series 12 Part No 1143482 SECTION 3—HANDLING The concentrator should AL W A YS be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator .
SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1143482 13 Perfecto 2 ™ Series SECTION 4—TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/10 8 VAC), 60 Hz Rated Current Input: Perfecto 2 5: 3 A Sound Level: Perfecto 2 5: 44 ± 2 dBA Average Altitude: Perfecto 2 5 - Up to 6 ,000 ft.
SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 14 Part No 1143482 *NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes). Safety System: Current overload or li ne surge shutdown. High temperature compressor shutdown.
SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1143482 15 Perfecto 2 ™ Series IRC5PO2 INDICATO RS IRC5P INDICATORS LABEL SYMBOL O 2 PURITY INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A.
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 16 Part No 1143482 SECTION 5— OPERATING INSTRUCTIONS Introduction Y our oxygen concentrator is inte nded for individual use in the home. It is an electronically operated device that separates oxygen from room air .
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 17 Perfecto 2 ™ Series Recommended Guidelines for Optimum Performance NOTE: Ensure that your concentrator is at least thr ee inches away from walls, draperies or furnitur e to assure su fficient air flow .
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 18 Part No 1143482 Connect Humidifier (If So Prescribed) WARNING For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter pe riods of operation may reduce maximum product life.
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 19 Perfecto 2 ™ Series FIGURE 5.2 Humidifer Bottle Adapter 5. Attach the humidifier bottle adapter to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened.
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 20 Part No 1143482 FIGURE 5.3 Humidifier Compartmen t.
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 21 Perfecto 2 ™ Series Power Switch NOTE: For this procedur e, refer to FIGURE 5.4. 1. Press pow er switch to on po sition. All the panel lights and the audible alarm will come on for one second, indicating that the unit is functionin g properly .
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 22 Part No 1143482 2. If the flowrate on the flowme ter ev er falls below 0.5 L/min for P erfecto 2 5 for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm.
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 23 Perfecto 2 ™ Series Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min.) while waiting for the O 2 purity to reach maximum.
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 24 Part No 1143482 Indicators (If Your Unit Does Not Feature the O 2 Sensor) NOTE: For this procedur e, refer to FIGURE 5.7. IRC5P Model Indicator Light Explanation RED light ( ) - T otal Unit Shut-Down.
SECTION 6—MAINTENANCE Part No 1143482 25 Perfecto 2 ™ Series SECTION 6—MAINTENANCE NOTE: The Invacare concentrators ar e specifically designed to minimize routine pr even tive mainte nance. In places with high dust or soot levels, maintenance may need to be performed mor e often.
SECTION 6—MAINTENANCE Perfecto 2 ™ Series 26 Part No 1143482 FIGURE 6.1 Cleaning the Cabinet Filter Cabinet WARNING UNPLUG the conc entrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. 1. Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.
SECTION 6—MAINTENANCE Part No 1143482 27 Perfecto 2 ™ Series PREVENTIVE MAINTENANCE RECORD Model No. IRC _______________________________ Serial No.____________________________________ On Each Inspection *NOTE: Refer to Preventive Ma inte nance se ction of Ser vice Manual.
SECTION 7—TROUBLESHOOTING GUIDE Perfecto 2 ™ Series 28 Part No 1143482 SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord not plugged in. 1. Insert plug in to outlet. Concentrator not operating, power switch On.
SECTION 7—TROUBLESHOOTING GUIDE Part No 1143482 29 Perfecto 2 ™ Series YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to 5PO 2 models. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.
SECTION 7—TROUBLESHOOTING GUIDE Perfecto 2 ™ Series 30 Part No 1143482 Alarm: Rapid Beep....Beep... Beep...Beep LOW FLOW ALARM 1a. Kinked or blocked tubing, cannula or humidifier. 1b. Flowmeter set at 0.5 L/min for Perfecto 2 5. 1a. Inspect for kinks or blockages.
LIMITED WARRANTY Part No 1143482 31 Perfecto 2 ™ Series LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURE D AFTER JULY 4, 1975. This warranty is extended only to th e original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVI RONMENTS, OR OTH.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Invacare IRC5PO2 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Invacare IRC5PO2 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Invacare IRC5PO2, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Invacare IRC5PO2 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Invacare IRC5PO2, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Invacare IRC5PO2.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Invacare IRC5PO2. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Invacare IRC5PO2 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.