INSTRUCTION MANUAL CS1 15 Barometric Pressure Sensor 3/03 Copyright (c) 2002-2003 Campbell Scientific, Inc..
Warranty and Assistance The CS115 BAROMETR IC PRESS URE SE NSOR is w arranted by CAMPBELL SCIENTIFIC, INC. to be f ree from def ects in m aterials and workmanship und er no rmal use a nd ser vice fo r twelve ( 12 ) months fr om date of shipment unle ss spec ified othe rwise.
i CS115 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. Introducti on ......................................................
This is a blank page..
1 CS115 Barometric Pressure Sensor The CS115 Barometric Pressure Sensor us es the resonant silicon technology pressure sensor developed by Druck. The sensor outputs a variable frequency which can be measured using the datalogger' s period averaging instruction or pulse count instruction.
CS115 Barometric Pr essure Sensor 2 2. Specifications 2.1 Performance Measurem ent R ange: 600 mb t o 1100 mb (hPa) Operating Tem perature Range: -40°C to +60°C Hum idity Ran ge: non- condens ing Media Compatibility: non-corrosive gas Total Accuracy 1 : at 20°C ±0.
CS115 Barometric Pr essure Sensor 3 3. Installation 3.1 Venting and Condensati on To p revent condensation, install the se nsor in an en vironm entally protected enclosure, com plete w ith desiccant w hich should be chan ged at regular intervals.
CS115 Barometric Pr essure Sensor 4 As y ou m ount the sen sor onto the back plate of the enclosu re, place the fork lu g under the m ountin g screw to accomplish the proper groun ding of the s ensor as show n i n Fig ure 3-1. 3.3 Jumper Settings The CS115 ha s tw o operating m odes, ‘ contin uous’ and ‘tri ggered’ .
CS115 Barometric Pr essure Sensor 5 4. Wiring The CS115 w iring instru ctions f or the exam ple program s are sh own in Table 4-1. For datal oggers C R500, CR510, C R10(X), CR 23X, and CR5000, analog ch annel is us ed for period averagin g meas uremen t.
CS115 Barometric Pr essure Sensor 6 Once th e CS115 is tu rned on one m inu te before every hour, an d the correct measu rement is m ade, it is then copied into the current variable called "Pressure" . The sensor is then turned off im m ediately follow ing the meas uremen t.
CS115 Barometric Pr essure Sensor 7 5.1 Period Averaging Measurement Examples 5.1.1 Period Averaging Measurement Example fo r CR10(X) EXAMPLE 1. Sam ple Program for CR 10(X) ;{CR 10X} ; *Table 1 Prog .
CS115 Barometric Pr essure Sensor 8 8: Sam ple (P70) 1: 1 Reps 2: 1* Loc [ P_m b ] *Table 2 Prog ram 02: 0.0000 Execution In terval (secon ds) *Table 3 S ubroutin es End Program -Input Loca tio ns- 1 P_m b * Proper entries w ill vary w ith program an d datalogger channel, and input location assign ments.
CS115 Barometric Pr essure Sensor 9 ;T urn off CS1 15 ; 4: Do (P86) 1: 58* Set Port 8* Low 5: If tim e is (P92) 1: 0 Minutes (Secon ds --) in to a 2: 60 Interval (sam e un its as above) 3: 1 0 Set Out.
CS115 Barometric Pr essure Sensor 10 3: Period A verage (SE) (P27) 1: 1 Reps 2: 14 200 kHz Max Freq @ 500 m V Peak to Peak, F req Output 3: 24* SE Chan nel 4: 10 No. of Cy cles 5: 5 Timeou t (unit s = 0.01 seconds) 6: 1* Loc [ P_m b ] 7: 1000 Mult 8: 0.
CS115 Barometric Pr essure Sensor 11 5.1.4 Period Averaging Measurement Example fo r CR5000 in a Fast Executing Program at 10Hz EXAMPLE 4. Sam pl e Program for C R5000 ‘CR5000 ' 'Sample Pr.
CS115 Barometric Pr essure Sensor 12 5.2 Pulse Count Measure ment Exampl e 5.2.1 Pulse Count Measu rement Example for 21X EXAMPLE 5. Sam ple Program for 21X ;{21X} ; *Table 1 Prog ram 01: 1 Execution .
CS115 Barometric Pr essure Sensor 13 *Table 3 S ubroutin es End Program -Input Loca tio ns- 1 P_mb 1 1 1 * Proper entries w ill vary w ith program an d datalogger channel, and input location assign ments.
CS115 Barometric Pr essure Sensor 14 The corrections in volved can be signi ficant: e.g ., at 1000 mb an d 20°C, barometric pressure w ill decrease by 1.1 mb for ev ery 10 meter in crease in altitude. 7. Maintenance and Calibration Since the sen sor is sem i-sealed, m inim um m aintenan ce is required.
Thi s is a b lank p age ..
Campbell Scie ntific Com panies Campbell Scientific, Inc. (CS I) 815 West 1 800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com info@campb ellsci.com Campbell Scientif ic Africa Pty. Ltd. (CSAf) PO B ox 24 50 Somers et West 7129 SOUTH AFRICA www.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Campbell Hausfeld CS115 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Campbell Hausfeld CS115 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Campbell Hausfeld CS115, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Campbell Hausfeld CS115 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Campbell Hausfeld CS115, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Campbell Hausfeld CS115.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Campbell Hausfeld CS115. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Campbell Hausfeld CS115 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.