This instrument may sometimes be used with hazardous materials, opera- tions, and equipment. It is beyond the scope of this manual to address all of the potential safety risks associated with its use in such applications. It is the responsibility of the user of this instrument to consult and establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Please read the following carefully! !

The instrument is designed to assist the filling and dispensing of gradu- ated and volumetric pipettes of glass or plastic in the volume range of 0.1 ml to 200 ml with a suction tube outer diameter < 9.2 mm for measuring liquids. If the instrument is used correctly, the pipetted liquid will only contact the pipette.
The instrument is designed for pipetting liquids, observing the following physical limits:
– from +10 °C to +40 °C (from 50 °F to 104 °F)
– Vapor pressure up to max. 500 mbar.
Aspirate slowly above 300 mbar, in order to prevent the liquid from boiling.
– Density up to 2.4 g/cm3
Never use the instrument with liquids whose vapors attack the materials silicone or PP.
The instrument is not designed for Pasteur pipettes.
Store the instrument and accessories only in cleaned condition in a cool and dry place. Storage temperature: -20 °C to +50 °C (-4 °F to +122 °F).


valve system (PP, SI, PTFE)
blow out bellows (SI)

1. Compress the suction bellows

Before attaching the pipette, squeeze the suction bellows.
Hold the pipette as near to its upper end as possible, and carefully insert it into the adapter until it fits tightly.
Once the pipette has been securely attached, always hold the instrument in a vertical posi- tion, tip down.

Be sure that the pipette fits tightly into the adapter. Never use force. Thin pipets are particularly liable to break. Avoid the risk of injury!
Immerse the pipette tip into the liquid. Press the pipetting lever slowly upwards. Fill the pipette so that the level of liquid is slightly above the required mark.

Please take care that no liquid enters into the instrument. This would impair the filtering function and reduce the suction capacity. If liquid does enter the instrument, exchange the filter (see pages 26 - 27).

The further up the lever is pressed, the stronger the suction becomes and the quicker the meniscus rises. In the case of large pipettes > 50 ml the vacuum contained in the suction bellows is not sufficient to draw in all the liquid at once. Therefore, squeeze the suction bellows again and continue drawing up liquid.
Use suitable lint-free tissue to wipe the pipette tip, if necessary. Press the pipetting lever down slowly, until precise adjustment of the meniscus has been achieved.

Hold the collecting vessel in an inclined posi- tion. Place the pipette tip against the inner vessel wall and press the pipetting lever down to dispense.
5.1 In the case of pipettes with a waiting time (Imprint e.g., 'Ex + 5 s'):
– As soon as the meniscus in the pipette
comes to a standstill, start waiting time as is indicated on the pipette (here: 5 s).
– Wipe the pipette tip a few millimeters
upward along the wall of the vessel.

5.2 In the case of blow-out pipettes (Imprint 'ausblasen – blow out'):

– As soon as the meniscus in the pipette tip
comes to a standstill, press the small blow- out bellows once.
– Wipe the pipette tip a few millimeters
upward along the wall of the vessel.

Hold the pipette as its extreme upper end, and gently twist and pull it out of the adapter.

When properly used, the instrument is maintenance-free. Before auto- claving, the instrument must be cleaned carefully. Wipe the housing with a damp cloth. The instrument is autoclavable at 121 °C (250 °F), 2 bar absolute (30 psi) with a holding time of at least 15 minutes according to DIN EN 285.
Replacing the Filter and Cleaning the Adapter
Only necessary if a pipette has been overfilled or the suction capacity is impaired.

Wear protective gloves and eye protection. Avoid the risk of contamination!




For autoclaving remove suction bellow.
Not every membrane filter is suitable. Only use genuine manufacturer’s recommended accessories. Filters of a pore size of 3 µm are a standard equipment of the instrument. The membrane filter is autoclavable up to 5 times at 121 °C. It is the user's responsibility to ensure effective autoclaving.

Replace the Valve System




When the meniscus has been set, liquid should not drip out of the pipette until the valves are activated. Should the pipette drip, see chapter 'Troubleshooting' (see page 31).
Leak Test

macro pipette controller, with spare membrane filter 3 µm and operating manual.
Color* Cat. No.
* Colors refer to accent colors on instrument adapter support and locking ring.
Adapter support, PP, length 49 mm
Spare membrane filter 3 µm, non-sterile, 1 pc. in blister pack
Cat. No. 260 52
Color Cat. No.
Spare membrane filter 3 µm, non-sterile, 10 pcs., in PE-bag
Cat. No. 260 56
Suction bellows with locking ring,
Silicone adapter, length 44 mm
Color Cat. No.
Cat. No. 261 46
Cat. No. 261 28

Trouble Possible Cause Action to be taken
Filter dirty or valve system damaged
Replace filter, see pages 26-27
Pipette drips Filter not properly posi- tioned or valve system damaged
Insert filter properly or replace valve system, see page 28
Pipette not held properly
Unscrew the adapter support of the instru- ment, replace silicone adapter
Compressed suction bellows take in air automatically
Suction bellows or valve system damaged
Replace the looking ring with the suction bellows or the valve system
Often visual control of the instrument is sufficient to discover a faulty device which can easily be replaced by a spare part. The advantages of do-it-yourself repair are obvious: It saves time and money.
Suction capacity impaired · Silicone adapter damaged

– Clean and decontaminate the instrument carefully.
– It is essential always to include an exact description of the type of
malfunction and the media used. If information regarding media used is missing, the instrument cannot be repaired.
– Shipment is at the risk and the cost of the sender.
Outside the U.S. and Canada:
– Complete the “Declaration on Absence of Health Hazards” and send the
instrument to the manufacturer or supplier. Ask your supplier or manufac- turer for the form.
In the U.S. and Canada:
– Return only cleaned and decontaminated instruments, with the Return
Authorization Number prominently displayed on the outside of the pak-kage to the address provided with the Return Authorization Number.
Return for repair


We shall not be liable for the consequences of improper handling, use, servicing, operating or unauthorized repairs of the instrument or the consequences of normal wear and tear especially of wearing parts such as pistons, seals, valves and the breakage of glass as well as the failure to follow the instructions of the operating manual.
Subject to technical modification without notice. Errors excepted.