Velleman DVM601

50 pages·666 KB·Multimètre·PDF·EN
Velleman DVM601
Table des matières 10 sections User manual · 1. Introduction · 2. Safety Instructions · 3. General Guidelines · … AfficherMasquer
Page 1 of the Velleman DVM601 user manual — 6 IN 1 DIGITAL MULTIMETER
Page 1 · cliquer pour agrandir
Page 1

6 IN 1 DIGITAL MULTIMETER

6-IN-1 DIGITALE MULTIMETER

MULTIMÈTRE NUMÉRIQUE 6-EN-1

MULTÍMETRO DIGITAL 6 EN 1

6-IN-1 DIGITALMULTIMETER

Velleman DVM601 — 6 IN 1 DIGITAL MULTIMETER diagram (page 1)
Fig. 1

USER MANUAL 3 GEBRUIKERSHANDLEIDING 11 NOTICE D’EMPLOI 19 MANUAL DEL USUARIO 27 BEDIENUNGSANLEITUNG 36

Page 2 of the Velleman DVM601 user manual — re 1
Page 2 · cliquer pour agrandir
Page 2
Velleman DVM601 diagram (page 2)
Fig. 2

re 1

Page 3 of the Velleman DVM601 user manual — User manual
Page 3 · cliquer pour agrandir
Page 3

User manual 1. Introduction To all residents of the European Union Important environmental information about this product

This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the environment. Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized company for recycling. This device should be returned to your distributor or to a local recycling service. Respect the local environmental rules. If in doubt, contact your local waste disposal authorities.

For indoor use only. Keep this device away from rain, moisture, splashing and dripping liquids. Keep the device away from children and unauthorised users. Risk of electric shock during operation. Be very careful when measuring live circuits. There are no user-serviceable parts inside the device. Refer to an authorized dealer for service and/or spare parts.
Page 4 of the Velleman DVM601 user manual — 4. Features
Page 4 · cliquer pour agrandir
Page 4
  • unauthorised way will void the warranty.
  • Caution : risk of electroshock when measuring voltages > 36VDC, 25VAC, currents > 10mA, AC power lines with inductance load and AC power lines with fluctuating power.
  • Only use the included test leads. When damaged, replace them with test leads of the same type and with the same specifications.
  • Always verify that all connections are reliable and safe.
  • Before measuring, always check the selected range.
  • Avoid body contact with ground potential (e.g. metallic terminals, output sockets, lead clamp…) while measuring. Make sure to be electrically insulated from ground during measurement.
  • Always use the device within its specified range.
  • Calibration and repair must be performed by a qualified technician. Refer to your local dealer.
  • Never attempt to measure voltages when connected to the current terminal. 4. Features
  • CAT. III - 1000V, CAT. IV - 600V
  • measure: sound-level, light, humidity, temperature, DC & AC voltage, DC & AC current, frequency, capacitance, resistance, diode & continuity test, live wire test ...
  • low-battery indication
  • auto ranging (except ranges Vac 400mV, Iac 10A and Idc 10A)
  • built-in non-contact AC voltage detector (50-1000VAC) 5. Overview and use Refer to the illustrations on page 2 of this manual.
1humidity & temperature sensor measure room temperature and relative humidity 2LCD 3 4/5 digits LCD display 3function rotary switch select function 4V/Hz%/Ω/Cap/°C input input jack for V, Hz%, Ω, Cap or °C measurement (red lead) 5COM input common input jack (black lead) 6µA/mA input input jack for µA or mA range measurement (red lead) 710A input input jack for 10A range measurement (red lead) 8microphone condenser microphone for noise measurement 22/08/2013 4 ©2008 Velleman® Components nv
Page 5 of the Velleman DVM601 user manual — 6. Use
Page 5 · cliquer pour agrandir
Page 5
DVM601 9light sensor long life silicon light sensor diode 10Hz/% button switch to Hz/% measurement – available when measuring AC/DC voltage, AC/DC current and Hz/% 11hold/backlight button Freeze/unfreeze all values on the display. Press and hold for ±3s to switch backlight on/off. 12°C/°F mode button Switch room temperature reading between °C and °F 13mode button Switch between AC and DC current measurement; switch between Ω, diode and continuity test. 14range button Select displayed measurement range manually. 15REL button Press to store the current measuring result and start measuring against this value. Press again to exit relative measurement mode. 16Non-Contact Voltage (NCV) indicator turns on when the built-in non-contact AC voltage detector detects AC voltages
  • 6. Use General
  • When the range of the measured signal is unknown, start measurement in auto-range mode, and then select range according to indicated value.
  • The display shows “OL” when the present range limit is exceeded. Select a higher range to continue measurement.
  • Do not measure voltages > 600 V or current higher than 500mA (mA/µA jack [6] ) or 10 A (10A jack [7] ). Note that the meter can only handle 10A for less than 30 seconds (with 15 minutes cool-down time).
  • The device will go in sleep mode when no activity is detected for ±30 minutes. Press any button to re-activate. Sound level measurement
  • Set the rotary switch [3] to the dBC -position.
  • Point the microphone [8] towards the sound source for more accurate measurement. The sound level will be displayed. Note: strong wind (over 10m/s) will have a negative impact on the
  • measurement. Humidity measurement
  • Set the rotary switch [3] to any position (except OFF).
  • The relative humidity at the current location is measured ( [1] ) and the value is shown on the top line of the LCD [2] . For more accurate results, let the multimeter acclimatize for ±2h before measuring.
Page 6 of the Velleman DVM601 user manual — 6. Use
Page 6 · cliquer pour agrandir
Page 6
  • Light measurement
  • Set the rotary switch [3] to the Lux - or Lux 10x -position.
  • Point the light sensor [9] towards the light source for more accurate measurement. The light level will be displayed [7] .
Velleman DVM601 — 6. Use diagram (page 6)
Fig. 3 · 6. Use
  • Room temperature
  • Set the rotary switch [3] to any position (except OFF).
  • The room temperature at the current location is measured ( [1] ) and the value is shown on the top line of the LCD [2] .
  • Thermocouple temperature
  • Set the rotary switch [3] in the °C or °F position. Do not apply input voltage at this setting!
  • Insert the thermocouple adaptor into the temp input [4] and COM input [5] .
  • Plug the thermocouple itself into the adaptor (if not done yet). There is only one way possible, do not force .
  • Use the other end of the thermocouple to measure temperatures. Temperature is shown on the LCD [5] . Notes:
  • do not expose the multi-meter itself to temperatures lower than 0°C (32°F) or higher than 40°C (104°F).
  • Do not change the probe, as this will influence measuring accuracy.
  • AC/DC voltage measurement
  • Connect the red test lead to the V input jack [4] and the black lead to the COM input jack [5] .
  • Set the rotary switch [3] in the desired position ( V , V , mV ). If the voltage to be measured is unknown beforehand, set the range switch in the highest range position and then reduce gradually until the ideal resolution is obtained.
  • Use the mode button [13] to choose between AC and DC voltage.
  • Connect the test leads to the source being measured.
  • Read the voltage value on the LCD display [2] along with the polarity of
Page 7 of the Velleman DVM601 user manual — 6. Use
Page 7 · cliquer pour agrandir
Page 7

the red lead connection. • Press the Hz/% button to switch to frequency; press again to see the duty cycle (%).

RangeResolutionRangeResolutionRangeResolution 400.0mV*0.1mV40.00V10mV600V1V 4.000V1.0mV400.0V100mV
  • AC/DC current measurement
  • Connect the red test lead to the µA/mA [6] input jack (max. 400mA) or 10A [7] input jack and the black test lead to the COM jack [5] .
  • Set the rotary switch [3] in the desired current range (<4mA: µA , <400mA: mA , <10A: 10A ) and use the mode button [3] to choose between AC and DC current.
  • Open the circuit in which the current is to be measured and connect the test leads to the circuit IN SERIES .
  • Read the current value and the polarity of the red lead connection on the LCD display.
RangeResolutionRangeResolution 400.0µA0.1µA400.0mA100µA 4000µA1µA10A10mA
RangeResolutionRangeResolutionRangeResolution 400.0Ω0.1Ω40.00KΩ10Ω4.000MΩ10kΩ 4.000kΩ1Ω400.0kΩ100Ω40.00MΩ1MΩ
Page 8 of the Velleman DVM601 user manual — 7. Battery
Page 8 · cliquer pour agrandir
Page 8
  • For continuity tests, connect the test leads to two points of the circuit to be tested. If continuity exists (resistance < 50Ω), the built-in buzzer will sound. If the circuit is open, “ OL ” will be indicated.
  • Non-contact AC voltage test
  • Set the rotary switch [3] to any position (except OFF).
  • The AC-sensor is located at the top of the meter, near the light sensor [9] . Point the AC-sensor towards the suspected AC-source. When AC voltage is present, the indicator [16] will light up. The stronger the ACsource, the more the indicator [16] will be lit. 7. Battery
  • When the low battery indication ( ) appears, replace the internal batteries. Turn the rotary switch [3] to the OFF-position.
  • Always disconnect test leads when replacing the battery. Do not us the device without batteries installed.
  • The battery cover is located on the back of the device and is closed with two screws. Remove the screws and open the battery compartment.
  • Remove the battery and insert a new 9V (E-block) battery following the polarity as indicated in the battery holder.
  • Reinstall the cover and secure it with the two screws.
  • Remove the battery when the device is not in use for a longer period of time to avoid leakage.
  • Do not attempt to recharge non-rechargeable batteries, do not puncture and do not throw in fire as they may explode.
  • WARNING: handle batteries with care, observe warnings on battery casing. Dispose of batteries in accordance with local regulations. Keep batteries away from children. 8. Fuses
  • To replace an internal fuse, turn the rotary switch [3] to the OFF-position.
  • Always disconnect test leads when replacing fuses. Do not use the device without the proper fuses installed.
  • The access the fuses, the whole back cover must be removed. Loosen the 6 outer screws (2 are located under the foldable stand). Do not remove the battery cover screws (see §7). Lift the whole back cover. Do not touch the electronic circuitry inside.
  • Replace fuses only by new ones with following specifications:
  • Fuse 1 (large): F 10A/600V fast blow
  • Fuse 2 (small): F 500mA/660V fast blow Using wrong fuses or short-circuiting fuse holders can lead to potentially life-threatening situations.
  • Close the back cover and secure the 6 screws.
Page 9 of the Velleman DVM601 user manual — 9. Technical specifications
Page 9 · cliquer pour agrandir
Page 9
Velleman DVM601 — 9. Technical specifications diagram (page 9)
Fig. 4 · 9. Technical specifications
capacitanceoverload protection - 600 V AC or DC frequency sensitivity< 1MHz: >0.5V RMS / > 1MHz: >3V RMS sound level measurements35-100dB resolution0.1dB typ. instrument freq.30Hz to 10kHz frequency weightingC accuracy± 5dB at 94dB sound level (1kHz sine wave) microphoneelectric condenser light measurementsup to 40 000lux accuracy± 5% of rdg + 10 digits temp. characteristics±0.1%/°C photodiode1 silicon photodiode with filter humidity measurements33%~99% RH resolution1% RH accuracy±3% RH internal temperature measurements0°C to +50°C resolution0.1°C accuracy3% of rdg ± 5 digits K-type thermocouple measuring range-20°C ~ +1300°C resolution1°C accuracy3% of rdg ± 5 digits DC voltage400mV/4V/40V/400V/600V basic accuracy± 1.0% of rdg ± 4 digits for 0.4V~40V range / ± 1.5% of rdg ± 4 digits for 400V~600V range input impedance10Mohm maximum input600V AC voltage400mV/4V/400V/600V basic accuracy± 1.5% of rdg ± 15 digits for 400mV range / ± 1.0% of rdg ± 4 digits for 4V~40V range / ± 1.5% of rdg ± 4 digits for 400V~600V range
Page 10 of the Velleman DVM601 user manual — 9. Technical specifications
Page 10 · cliquer pour agrandir
Page 10
DVM601 input impedance10Mohm frequency range50 to 400Hz maximum input600Vac DC current400µ/4m/400m/10A basic accuracy± 1.0% of rdg ± 2 digits for 400µA~4mA range/ ± 1.2% of rdg ± 2 digits for 400mA range / ± 2.0% of rdg ± 5 digits for 10A range overload protection500mA/600V fuse & 10A/600V fuse AC current400µ/4m/400m/10A basic accuracy± 1.2% of rdg ± 2 digits for 400µA~4mA range/ ± 1.5% of rdg ± 2 digits for 400mA range / ± 2.0% of rdg ± 5 digits for 10A range overload protection500mA/600V fuse & 10A/600V fuse resistance400/4k/40k/400k/4M/40M basic accuracy± 1.5% of rdg ± 4 digits for 400 ohms range / ± 1.5~2.0% of rdg ± 2 digits for 4k~40M range overload protection15 seconds max. 250V on all ranges capacitance50.00nF/500.0nF/5.000uF/50.00uF/100.0uF accuracy±3.0~5.0% of rdg ± 5~7 digits frequency5Hz/50Hz/500Hz/5KHz/50KHz/500KHz/10MHz basic accuracy±1.2% of rdg ± 3digits overload protection250V AC or DC maximum display3999 power supply9V battery (E-block) dimensions170 x 78 x 48mm weight (with battery)approx. 330g
Velleman DVM601 — 9. Technical specifications diagram (page 10)
Fig. 5 · 9. Technical specifications

Use this device with original accessories only. Velleman nv cannot be held responsible in the event of damage or injury resulted from (incorrect) use of this device. For more info concerning this product, please visit our website www.velleman.eu The information in this manual is subject to change without prior notice.

© COPYRIGHT NOTICE This manual is copyrighted. The copyright to this manual is owned by Velleman Components nv. All worldwide rights reserved. No part of this manual may be copied, reproduced, translated or reduced to any electronic medium or otherwise without the prior written consent of the copyright holder.

Sources

Références et liens

Pages mentionnées dans ce manuel — fabricant, assistance et téléchargements.

Poser une question sur : Velleman DVM601

Les utilisateurs et notre équipe répondent généralement sous 24 h — et votre question servira aussi aux suivants.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Aucune question pour l'instant sur le Velleman DVM601

Soyez le premier à poser une question — d'autres utilisateurs ou notre équipe y répondront.

Document d'origine

Velleman DVM601 manuel PDF

50 pages·666 KB·Multimètre·PDF
Cliquez sur une page ci-dessus pour l'ouvrir ici
Page 1 / …
100%
Chargement des pages…