6000 COUNTS MULTI FUNCTION MULTIMETER MULTIFUNCTIONELE MULTIMETER - 6000 COUNTS MULTIMÈTRE MULTIFONCTIONS - 6000 POINTS
MULTIFUNKTIONSMULTIMETER - 6000 ZÄHLUNGEN

USER MANUAL 4 GEBRUIKERSHANDLEIDING 16 NOTICE D’EMPLOI 28 MANUAL DEL USUARIO 40 BEDIENUNGSANLEITUNG 52







To all residents of the European Union Important environmental information about this product
This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the environment. Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized company for recycling. This device should be returned to your distributor or to a local recycling service. Respect the local environmental rules. If in doubt, contact your local waste disposal authorities. Thank you for choosing Velleman! Please read the manual thoroughly before bringing this device into service. If the device was damaged in transit, don't install or use it and contact your dealer. Refer to the Velleman® Service and Quality Warranty on the last pages of this manual. 2. Used symbols

Follow the instructions below to guarantee a safe use of the meter and all its functionalities.
During use of the meter, respect all directives concerning protection against electroshocks and misuse. Never exceed the indicated limits.

Remark: refer to the warning on the back of the meter.
Avoid cold, heat and large temperature fluctuations. When the unit is moved from a cold to a warm location, leave it switched off until it has reached room temperature. This to avoid condensation and measuring errors.
Protect this device from shocks and abuse. Avoid brute force when operating.
Pollution degree 2-device. For indoor use only. Keep this device away from rain, moisture, splashing and dripping liquids. Not for industrial use. Refer to §5 Pollution degree.
Keep the device away from children and unauthorised users.
Risk of electric shock during operation. Be very careful when measuring live circuits. Use extreme caution when measuring voltages higher than 60Vdc or 30Vac rms.
Do not replace internal parts yourself. Replace damaged or lost accessories by identical ones with the same specifications. Order spare accessories e.g. test probes at your dealer.
This is an installation category CAT IV 600V / CAT III 1000V measuring instrument. Never use this equipment in a higher category than indicated. Refer to §4 Overvoltage /installation category.
Read this manual thoroughly. Familiarise yourself with the functions of the device before actually using it.
All modifications of the device are forbidden for safety reasons. Damage caused by user modifications to the device is not covered by the warranty.
Only use the device for its intended purpose. Using the device in an unauthorized way will void the warranty. Damage caused by disregard of certain guidelines in this manual is not covered by the warranty and the dealer will not accept responsibility for any ensuing defects or problems. Make sure the meter is in the appropriate measuring range before connecting it to a test circuit.
Do not measure circuits that may contain voltages > 1000V (DC or rms AC)
Do not measure current in circuits with voltages > 1000V
Do not conduct resistance, diode- or continuity measurements on live circuits.
Before each use, make sure the test probes are in good condition. Always place your fingers behind the protective edges of the test probes while measuring! Never touch free terminals when the meter is connected to a circuit.
Switch off the meter and remove test probes prior to replacing the battery or fuses.
When carrying out measurements on a TV set or switching power circuits, always be aware that high amplitude voltage pulses at the test points might damage the meter.

DMMs are categorized depending on the risk and severity of transient overvoltage that might occur at the point of test. Transients are short-lived bursts of energy induced in a system, e.g. caused by lightning strike on a power line.
The existing categories according EN 61010-1 are:
electronics circuits, control signals, circuits behind isolating transformer… circuits that are directly connected to mains power, but limited to:
o measurements on mono-phase appliances that are connected to the mains by
means of a plug o mono-phase appliances and circuits directly connected to the mains in a normal
domestic environment, provided that the circuit is at least 10m apart from a CAT IV-environment. E.g. household appliances, portable tools, light circuits at more than 10m from a distribution board … measurements in/on low-voltage distribution boards (distribution boards behind
meter box) measurements on (fixed) mono- or poly-phased appliances and circuits except in
CAT IV-environments (e.g. mains outlets, electric ovens, lighting circuits, bus bars, low-voltage distribution boards and circuit breakers). This device is suitable for measurements up to 600V: Measurements on distribution equipment and outdoor installations including meter
boxes and equipment/circuits outside or remote from the domestic environment e.g. circuits in sheds, garden houses and free-standing garages , or circuits using underground wiring e.g. garden lighting, pool-pump... This device is NOT suitable for: Voltages above 1000V Currents above 10A
This device is only suitable for measurements up to 600V in

CAT IV and up to 1000V in CAT III.
IEC 61010-1 specifies different types of pollution environments, for which different protective measures are necessary to ensure safety. Harsher environments require more protection, and the protection against the pollution which is to be found in a certain environment depends mainly on the insulation and the enclosure properties. The pollution degree rating of the DVM indicates in which environment the device may be used.
This device is only suitable for measurements in Pollution degree class 2 environments.
Refer to the illustrations on pages 2 and 3 of this manual. 6.1 Multi-meter



6.4 Mode description DATA HOLD mode In Data Hold mode, the display is not longer updated. Press the HOLD button to enter Data Hold mode. H is shown on the display [F]. When Data Hold mode is enabled in auto range mode, the meter switches to
manual ranging mode. To exit Data Hold mode, press the RANGE or HOLD button Manual / Auto range mode The meter has both manual and auto range options. In the auto range mode, the meter selects the best range for the input detected.
This allows you to switch test points without having to reset the range. When the meter is in the auto range mode, AUTO is displayed [D]. In the manual ranging mode, the range must be selected manually. This allows
the user to override auto range and lock the meter in a specific range. The meter defaults to auto range mode for measurement functions that have
more than one range. To enter and exit the manual range mode press the RANGE button. Press
successively to scroll through the available ranges. To exit the manual ranging mode, press and hold down RANGE key for two
seconds. The meter returns to the auto range mode and AUTO is displayed. Sleep mode The Meter enters the "sleep mode" and blanks the display if the meter is on but
not used for 30 minutes. Press the HOLD button or rotate the rotary switch to wake the meter up. To disable Sleep mode, hold down the SELECT key while turning the meter on. Relative measurement mode Relative measurement mode is available during all functions except frequency. To enter and exit the relative measurement mode, start the measurement. Press REL key to store the measured value and activate the relative
measurement mode. The difference between the reference value and subsequent reading is displayed. REL is shown on the display [G]. Press REL key for more than 2 seconds to return the meter to normal operation. 6.5 PC Link (model DVM4100 only) PC Link is only compatible with Windows XP, Vista and Windows 7 32 & 64 bit. The DVM4100 has a serial data output function. Before continuing the necessary USB driver and software must be installed on the
PC. Insert the included CD ROM into a suitable PC. Locate and open the file README.pdf for the installation procedure (English only). Follow the instructions on the screen Notes:
First install the PC-link software followed by the USB driver, than restart the PC. log in on the PC as administrator. For Vista/Windows 7 systems, disable the
UAC (User Account Control) (via Control Panel – System and Security – Action Center – User Account Control settings -> Never notify.) Connect the meter to a USB port of a PC using the included data cable. Press and hold the Hz % button while turn on the meter, the meter enter PC-Link
mode. T the symbol “PC-LINK” will appear on the LCD [E] and the serial data output function is active. Notes:
It is not possible to enter PC LINK during measurement; the meter must be
switched off first. To disable the auto power off function, simultaneously hold down the SELECT
and Hz% button while turning the meter on. Run the PC-LINK software, click the SET menu. Select the System Set. Then
select the proper COM port in the Serial Port Select. To determine the COM port, go to Device Manager:

Right-click the My Computer icon on the Windows desktop, and then click
Properties. Click the Hardware tab and then click Device Manager. Scroll through the list of installed devices and locate the Ports (Com and LPT)
entry. Click the plus (+) next to this entry to view the installed ports, If no errors occurred, the “Prolific USB-to-Serial COM Port (COM x)” will appear, COM x is the port, with x = a specific number. Select the default sampling rate or set a desired sampling rate. Now press the Start in the PC-LINK SOFT to measure and view the real-time data
or graph in the software interface. To disable the serial data output function, switch the meter OFF. More information about the PC-LINK SOFT can be found in the help-file of the
Risk of electric shock during operation. Be very careful when measuring live circuits.
Before measuring, always make sure the meter and/or test probes are not damaged and verify the connections, selected function and range.
Never exceed the limit value for protection. This limit value is listed separately in
the specifications for each range of measurement. Do not touch unused terminals when the meter is linked to a circuit which is being
tested. Only use the meter in the indicated overvoltage/installation category. Never
measure voltages that might exceed the indicated category values. When a measuring range is unknown, always select the highest possible range
and lower to the appropriate range. Disconnect the test leads from the tested circuit before rotating the range selector
in order to change functions. When carrying out measurements on a TV set or switching power circuits, always
remember that high amplitude voltage pulses at the test points might damage the meter. Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep
your fingers behind the probe barriers at all times during measurement. Do not measure current in circuits with voltages > 1000V. Never perform resistance, diode or continuity measurements on live circuits. Make
sure all capacitors in the circuit are depleted. 7.1 Voltage measurements
Do not measure circuits where voltages > 600V CAT IV or > 1000V CAT III may reside.
Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the probe barriers at all times during measurement. Do not touch unused terminals when the meter is linked to a circuit which is being tested.
Connect the red test lead to the "VΩ Hz " jack and the black lead to the "COM" jack. Set the rotary switch to V for DC measurements or to V for AC measurements. Press the RANGE button to manually select the range. Connect the test probes to the source being measured. The measured value appears on the LCD display. Notes:
For DC-measurements, when a negative polarity is present at the red test lead,
the indicated value is preceded by a “-” sign.

When measuring DC offset of an AC voltage, measure the AC voltage first and
select a range for the DC offset equal to or higher than the AC range to increase accuracy. 7.2 Resistance measurements
Do not perform resistance measurements on live circuits.
Connect the red test lead to the "VΩ Hz " jack and the black lead to the "COM" jack. Set the rotary switch to "”. When necessary, press the SELECT button until MΩ
is shown on the display. Press the RANGE button to manually select the range. Connect the test probes to the source being measured. The measured value appears on the LCD display. Notes:
o Never perform resistance measurements on a live circuit and make sure all
capacitors are completely depleted. o To increase accuracy when measuring low resistance values, first hold the tips
of the measuring probes together to determine the resistance value of the test leads. Subtract this value from the measured value of the circuit. o For resistance measurements above 10MΩ the meter needs a few seconds to
stabilize the read-out. o Should the measured resistance exceed the current range or in case of an open
circuit, the display will show “OL”. 7.3 Continuity measurements
Do not perform continuity measurements on live circuits.
Connect the red test lead to the "VΩ Hz " jack and the black lead to the "COM" jack. Set the rotary switch to . Press the SELECT button to select continuity test . Connect the test probes to the source being measured. When the measured resistance is less than 40Ω a continuous beep is produced
and the resistance value is showed on the display. Should the measured resistance exceed the current range or in case of an open circuit, the display will show “OL”. Nota: Never perform continuity measurements on a live circuit and make sure all
capacitors are completely depleted. 7.4 Diode measurements
Do not perform diode measurements on live circuits.
Connect the red test lead to the "VΩ Hz " jack and the black lead to the "COM" jack. Set the rotary switch to . Press the SELECT button to select diode test . Connect the red test probe to the anode of the diode and the black test probe to
the cathode. The meter shows the forward voltage drop of the diode. When the diode is
connected with a reversed polarity, the display shows “OL”. Notes:
o Never perform diode measurements on a live circuit and make sure all
capacitors are completely depleted. o Measuring diodes that are part of a circuit might produce faulty results.
Consider unmounting from the circuit.

Do not perform capacitance measurements on live circuits.
Connect the red test lead to the "VΩ Hz " jack and the black lead to the "COM" jack. Set the rotary switch to . Press the RANGE button to manually select the range. Connect the test probes to the capacitor and read its value from the display. Notes:
o It takes a few seconds before the meter shows the result. This is normal
behavior (e.g. 300.0µF range >30s). o To increase accuracy when measuring low capacitance values (
the tips of the measuring probes together to determine the capacitance value of the test leads. Subtract this value from the measured value of the circuit. o Never perform capacitance measurements on a live circuit and make sure all
capacitors are completely depleted. 7.6 Frequency measurements
Do not perform frequency measurements in circuits with voltages > 1000V.
Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the probe barriers at all times during measurement. Do not touch unused terminals when the meter is linked to a circuit which is being tested.
Connect the red test lead to the "VΩ Hz " jack and the black lead to the "COM" jack. Set the rotary switch to Hz. Press the RANGE button to manually select the range. Connect the test probes to the circuit and read the measured value from the LCD. Notes:
o The frequency can also be measured during AC voltage/current measurement.
Start the measurement and press the HZ % button to measure the frequency or duty cycle. o Use a shielded cable for measuring small signals in a noisy environment. 7.7 Temperature measurements
Do not touch any live parts with the temperature measuring probe.
Set the rotary switch to the °C range. When no temperature measuring probe is
connected, the current environmental temperature is displayed. Place the adaptor socket between the “VΩ Hz ” and the “COM” jack (align “+”
with “VΩ Hz ”). Plug the thermocouple (type K) into the adaptor socket (align “+” with “+”). Hold the tip of the temperature probe against the object under test. Read the temperature on the LCD. 7.8 Current Measurements
Do not measure current in circuits with voltages > 1000V
Current measuring jack µA/mA max. 600mA. For currents up to 10A use the 10A jack.
Always be careful when working with voltages above 60Vdc or 30Vac rms. Keep your fingers behind the probe barriers at all times during measurement.

and the black lead to the "COM" jack. For measurements up to 10A: connect the red test lead to the "10A" jack and
the black lead to the "COM" jack. Set the rotary switch to the µA range for measurements up to 600µA (only when
test probe is connected to the µAmA jack). Set the rotary switch to the mA range for measurements up to 600mA (only when
test probe is connected to the µAmA jack). Set the rotary switch to the A range for measurements up to 10A (only when test
probe is connected to the 10A jack). Press the SELECT button to select DC current (DCA) or AC current (ACA) mode. Connect the test probes in series with the circuit. Read the measured value from the display. Notes:
o for DC-current measurements, when a negative polarity is present at the red
test lead, the indicated value is preceded by a “-” sign. o The µAmA-range is protected against over current with a F630mA 1000V fuse,
the 10A-range is protected against over current with a F10A 1000V fuse. 8. Maintenance
Do not replace internal parts yourself. Replace damaged or lost accessories by identical ones with the same specifications. Order spare accessories e.g. test probes at your dealer. a. General maintenance Wipe the device regularly with a moist, lint-free cloth. Do not use alcohol or solvents. b. Fuse Replacement
Switch off the multi-meter, disconnect the test leads from the test points and remove the test leads from the measuring jacks before replacing the batteries or fuses. Refer to the illustration on pages 3 of this manual.
Remove the 4 screws at the back (indicated with in the illustration) and gently open the meter.
Remove the screw that holds the PCB in place (indicated with in the illustration) and gently lift the PCB. Remove the fuse from the fuse holder and replace it with a new fuse of the same
type and with the same specifications (F630mA/1000V - Ø10.3x38mm, F10A/1000V - Ø10.3x38mm). Close the meter meticoulsly. c. Battery Replacement
To avoid false readings, which could lead to possible electric shocks or personal injury, replace the battery as soon as the battery indicator appears.
Switch off the multi-meter, disconnect the test leads from the test points and remove the test leads from the measuring jacks before replacing the batteries or fuses. Refer to the illustration on pages 3 of this manual.
Remove the 2 screws at the back of the meter (indicated with in the illustration) and gently open the battery compartment. Replace the battery (9V 6LF22). Do not use rechargeable batteries. Insert
following the right polarity. Close the battery compartment and reseat the screws.

This device is not calibrated when purchased! Regulations concerning environment of use: Use this meter only for measurements in CATI, CATII, CAT III (
IV (
Temperature: 18 to 28°C (64°F to 82°F) relative humidity: max. 75% altitude: max. 2000m (6560ft) 9.1. General
Max. Voltage 600V CAT IV, 1000V CAT III Display 3 5/6-digit LCD, 2 to 3 samples/sec.
Fuse Protection µA/mA-range: 630mA/1000V (Ø10.3x38mm) 10A-range: 10A/1000V (Ø10.3x38mm) Temperature coefficient 0.1x(specified accuracy)/°C (28°C) Power Supply 9V NEDA 1604 6F22 006P (incl.) Ranging Method manual/automatic Polarity Indication " - " is displayed Overrange Indication "OL" is displayed automatically
Battery-Low Indication " " is displayed
Operating Temperature 0°C to 40°C (32~122°F) (
Range Resolution Accuracy 600mV 100µV (0.5% of rdg + 5 digits) 6V 1mV (0.8% of rdg + 5 digits) 60V 10mV 600V 100mV 1000V 1V (1.0% of rdg +2 digits) Input Impedance: >10M for all ranges 9.3. AC Voltage
Range Resolution Accuracy 600mV 100µV
(1.0% of rdg + 5 digits) 6V 1mV 60V 10mV 600V 100mV 1000V 1V (1.5% of rdg +2 digits) Input Impedance: >10M for all ranges Frequency Range: 40 to 400Hz Overload protection: 1000Vdc or ac rms. 9.4. DC Current
Range Resolution Accuracy 600µA 0.1µA
(1.5% of rdg + 3 digits) 6mA 1µA 60mA 10µA 600mA 0.1mA 6A 1mA (1.5% of rdg +5 digits)

When measuring currents > 6A, max. 15s continuous measurement followed by a 15 minutes break between 2 measurements. Overload Protection: 600mA range F0.63A/1000V– 10A range fuse F10A/1000V 9.5. Resistance

Range Resolution Accuracy 600 0.1 (0.5% of rdg + 3 digits) 6k 1
(0.5% of rdg + 2 digits) 60k 10 600k 100 6M 1k 60M 10k (1.5% of rdg + 5 digits) Note: in range 600Ω, short circuit test leads to determine lead resistance, than subtract this value from the measured value. 9.6. Diode and audible continuity test

Range Resolution Accuracy 60nF 10pF 10nF: ±(3.0% of rdg + 10 digits) 600nF 100pF
(5.0% of rdg +10 digits) 6µF 1nF 60µF 10nF 300µF 100nF 9.8. Temperature
Range Resolution Accuracy -55°C ~ 0°C 0.1°C ±(5.0% of rdg + 4°C) 1°C ~ 400°C ±(1.0% of rdg + 3°C) 401°C ~ 1,000°C 1°C ±(2.0% of rdg)

±(0.05% of rdg + 8 digits) 999.9Hz 0.1Hz 9.999kHz 1Hz 99.99kHz 10Hz N.A. 999.99kHz 100Hz
Use this device with original accessories only. Velleman nv cannot be held responsible in the event of damage or injury resulted from (incorrect) use of this device. For more info concerning this product and the latest version of this user manual, please visit our website www.velleman.eu The information in this manual is subject to change without prior notice.
© COPYRIGHT NOTICE This manual is copyrighted. The copyright to this manual is owned by Velleman nv. All worldwide rights reserved. No part of this manual may be copied, reproduced, translated or reduced to any electronic medium or otherwise without the prior written consent of the copyright holder.