Read the operating instructions and the enclosed brochure „Guarantee and additional notices“ completely. Follow the instructions they contain. Safely keep these documents for future reference. !

Function/Application Multimeter for taking measurements in the range of the overvoltage category CAT III up to max. 1000 V/CAT IV up to max 600 V. The meter can be used to measure DC and AC voltages, direct and alternating currents, resistances, capacitances, frequencies and duty factors within the speciÖ ed ranges; it can also be used for continuity and diode testing. In addition, the meter is Ö tted with a non-contact voltage detector and a vibration alarm.

Symbols

Hazardous electrical voltage warning: Unprotected live components inside the device housing may pose a risk of electric shock.

Protection class II: The test device has reinforced or double insulation.

Overvoltage category III: Equipment in Ö xed installations and for applications where speciÖ c requirements with regard to the reliability and availability of equipment have to be met, e.g. circuit-breakers in Ö xed installations and devices used in industrial applications which are permanently connected to the Ö xed installation.

Overvoltage category IV: Devices such as electricity meters, overcurrent circuit breakers and ripple-control units, which are intended for use at or near the infeed into the electrical installation of buildings, and speciÖ cally from the main distribution to the supply system.

Page 2 of the Laserliner MultiMeter-Compact user manual — Symbols
Page 2 · cliquer pour agrandir
Page 2

Safety instructions • Make sure that you always select the correct connections, the correct rotary switch position and the correct range for the measurement to be taken.

  • Before measuring or checking the resistance, continuity, diodes or capacitance, disconnect the power supply to the electric circuit. Check that all high-voltage capacitors are discharged.
  • Isolate the device from all current sources before opening the battery compartment cover.
  • If possible, do not work alone.
  • If you have to take hold of the measuring spikes, do so by the grip sections only. Do not touch the measuring contacts whilst the measurement is being taken.
  • If the device comes into contact with moisture or other conductive residue, work must not be carried out under voltage. At and above voltages of 25 V AC/60 V DC, the presence of moisture creates the risk of life-threatening electric shocks. Clean and dry the device before use. When using the device outdoors, make sure that the weather conditions are appropriate and/or that suitable protection measures are taken.
  • If you are working with voltages higher than 25 V AC/60 V DC, exercise extreme caution. Touching the electrical conductors at such voltages poses a risk of life-threatening electric shocks.
  • Do not use the device in environments in which there are conductive particles or where the occurrence of moisture (in the form of condensation, for example) can create transient conductivity.
  • The device must only be used in accordance with its intended purpose and within the scope of the speci Ö cations.
  • If you are taking measurements in the hazardous vicinity of electrical installations, do not work alone and seek guidance from an electrically skilled person before starting work.
  • Before taking any measurements, make sure that both the area to be tested (e.g. a line), the test device and the accessories used (e.g. connection cable) are in proper working order. Test the device by connecting it to known voltage sources (e.g. a 230 V socket in the case of AC testing or a car battery in the case of DC testing). Stop using the device if one or a number of its functions fails.
Page 3 of the Laserliner MultiMeter-Compact user manual — Symbols
Page 3 · cliquer pour agrandir
Page 3
Laserliner MultiMeter-Compact — Symbols diagram (page 3)
Fig. 1 · Symbols

1 Rotary switch to set the measuring function

A Measured value display (4 segments, 4000 digits)

Laserliner MultiMeter-Compact — Symbols diagram (page 3)
Fig. 2 · Symbols

J Measurement units: mV, V, „A, mA, Ohm, kOhm, MOhm, nF, „F, Hz, kHz, MHz, %

DisplayO.L: open line/ over × ow: measuring circuit not closed or measuring range exceeded
Page 4 of the Laserliner MultiMeter-Compact user manual — Maximum input power
Page 4 · cliquer pour agrandir
Page 4

Maximum input power

Laserliner MultiMeter-Compact — Maximum input power diagram (page 4)
Fig. 3 · Maximum input power

V DC/V AC 1000 V DC, 1000 V AC

A DC/AC 10 A DC/AC (max. 30 seconds every 15 minutes)

Frequency, resistance, capacitance, duty factor, diode test, continuity test

AUTO OFF function In order to preserve the batteries, the meter switches off automatically if it is left idle for 15 minutes.

Laserliner MultiMeter-Compact — Maximum input power diagram (page 4)
Fig. 4 · Maximum input power

Attaching the test prods 2 In order to avoid the risk of injury, the test prods must always be kept in the holder on the rear of the meter when not in use and during transport.

Laserliner MultiMeter-Compact — Maximum input power diagram (page 4)
Fig. 5 · Maximum input power
Page 5 of the Laserliner MultiMeter-Compact user manual — Current measurement DC/AC 4
Page 5 · cliquer pour agrandir
Page 5
Laserliner MultiMeter-Compact — Current measurement DC/AC 4 diagram (page 5)
Fig. 6 · Current measurement DC/AC 4

The black test prod (–) must always be connected to the „COM socket“. When taking current measurements, the red test prod (+) must be connected to the „10 A socket“. For all other measuring functions, the red test prod must be connected to the „VąHZ socket“.

Please ensure that the test prods are connected correctly before each measurement. Taking a voltage measurement with 10 A current connections plugged in can cause the installed fuse to trip and result in damage to the measuring circuit. !

Current measurement DC/AC 4

To take a current measurement, turn the rotary switch to position „10 A“ and press the „Mode“ button to set the voltage mode (AC, DC). Disconnect the circuit before connecting the meter. Then connect the measuring contacts to the object to be tested. The measured value acquired and the polarity appear on the display. Disconnect the circuit again before disconnecting the meter.

Do not measure currents above 10 A for longer than 30 seconds. This could damage the device or the test prods. !

Page 6 of the Laserliner MultiMeter-Compact user manual — Resistance measurement 6
Page 6 · cliquer pour agrandir
Page 6

Voltage measurement DC/AC

To take a voltage measurement, turn the rotary switch to position „V“ and press the „Mode“ button to set the voltage mode (AC, DC). Then connect the measuring contacts to the object to be tested. The measured value acquired and the polarity appear on the display.

Resistance measurement 6

To measure the resistance, turn the rotary switch to position „ą“. Then connect the measuring contacts to the object to be tested. The measured value acquired appears on the display. If „O.L.“ appears on the display instead of a measured value, either the measuring range has been exceeded or the measuring circuit is not closed or has been interrupted. Resistances can only be measured correctly in isolation; therefore, the components might need to be disconnected from the remainder of the circuit.

Laserliner MultiMeter-Compact — Resistance measurement 6 diagram (page 6)
Fig. 7 · Resistance measurement 6

When measuring resistance, to avoid the risk of the results of a measurement being distorted, there must be no traces of dirt, oil, solder spray or other contamination on the test prods. !

Page 7 of the Laserliner MultiMeter-Compact user manual — Voltage measurement DC/AC
Page 7 · cliquer pour agrandir
Page 7
Laserliner MultiMeter-Compact — Voltage measurement DC/AC diagram (page 7)
Fig. 8 · Voltage measurement DC/AC

To test continuity, turn the rotary switch to position „ą“ and press the „Mode“ switch twice to activate the „Continuity test“ function. Then connect the measuring contacts to the object to be tested. A measured value of < 150 ohms is recognised as continuity; this is conÖ rmed by an audible signal. If „O.L.“ appears on the display instead of a measured value, either the measuring range has been exceeded or the measuring circuit is not closed or has been interrupted.

Laserliner MultiMeter-Compact — Voltage measurement DC/AC diagram (page 7)
Fig. 9 · Voltage measurement DC/AC

To test the diode, turn the rotary switch to position „ą“ and press the „Mode“ button once to activate the „Diode test“ function. Then connect the measuring contacts to the diode. The measured value acquired for the forward voltage appears on the display. If „O.L.“ appears on the display instead of a measured value, the diode has either been tested in the reverse direction or is faulty. If 0.0 V is measured, the diode is faulty or a short-circuit has occurred.

Reverse direction Forward direction

Forward direction

Page 8 of the Laserliner MultiMeter-Compact user manual — Voltage measurement DC/AC
Page 8 · cliquer pour agrandir
Page 8
Laserliner MultiMeter-Compact — Voltage measurement DC/AC diagram (page 8)
Fig. 10 · Voltage measurement DC/AC

To take a capacitance measurement, turn the rotary switch to position „CAP“ and press the „Mode“ button three times to activate the „Capacitance measurement“ function. Then connect the measuring contacts to the object to be tested. In the case of poled capacitors, connect the positive pole to the red test prod.

Laserliner MultiMeter-Compact — Voltage measurement DC/AC diagram (page 8)
Fig. 11 · Voltage measurement DC/AC

To take a frequency measurement, turn the rotary switch to position „Hz“. Then connect the measuring contacts to the object to be tested. Press the „Mode“ button to switch from Hz to % (duty factor).

Laserliner MultiMeter-Compact — Voltage measurement DC/AC diagram (page 8)
Fig. 12 · Voltage measurement DC/AC

The „Auto range“ function is activated automatically when the meter is switched on. Auto range searches for the best possible range for each measurement in the corresponding measuring functions. Press the „Range“ button to activate manual range selection, then press the „RANGE“ button several times until you reach the desired range. Pay attention to changes in decimal places or units. To return to the „Auto range“ range, press the „RANGE“ button and hold it down for 2 seconds. „AUTO“ reappears on the display. The „Range“ function can only be used for voltage, current and resistance measurements.

Page 9 of the Laserliner MultiMeter-Compact user manual — Hold function 13
Page 9 · cliquer pour agrandir
Page 9
Laserliner MultiMeter-Compact — Hold function 13 diagram (page 9)
Fig. 13 · Hold function 13

The comparative measurement function takes a measurement relative to a reference value that has been saved previously. This enables the difference between the current measured value and the saved reference value to be displayed. Press the „REL“ button whilst a reference measurement is being taken in the corresponding measuring function. The display now shows the difference between the current measurement and the set reference value. Press the „REL“ button again to deactivate the function. The „Rel“ function can only be used for voltage, current and capacitance measurements, as well as continuity tests.

Hold function 13

The „Hold“ function enables you to keep the current measured value on the display. Press the „HOLD“ button to activate and deactivate the function.

Laserliner MultiMeter-Compact — Hold function 13 diagram (page 9)
Fig. 14 · Hold function 13

The non-contact voltage detector integrated in the meter is able to detect AC voltages from 100 V to 600 V, thus enabling live cables or cable breaks, for example, to be identiÖ ed. Turn the rotary switch to position „V“ and run the voltage sensor along the object to be tested (5 - 10 mm). The display lights up and the device starts to vibrate if AC voltage is detected.

Laserliner MultiMeter-Compact — Hold function 13 diagram (page 9)
Fig. 15 · Hold function 13
Page 10 of the Laserliner MultiMeter-Compact user manual — Hold function 13
Page 10 · cliquer pour agrandir
Page 10
Laserliner MultiMeter-Compact — Hold function 13 diagram (page 10)
Fig. 16 · Hold function 13

Non-contact voltage detection is no substitute for conventional voltage testing. As the device detects an electrical Ö eld, it will react even to static charge. !

Voltage detection, single-pole phase test For reasons of safety, you must disconnect the black measuring lead from the COM socket on the device to take this measurement. Turn the rotary switch to position „V“. Connect the red test prod to the phase or neutral conductor. The red LED only lights up if the phase conductor is live. When the single-pole phase test is carried out on the outer conductor, the indicator function may be adversely affected under certain conditions (e.g. when insulating personnel protective equipment is used or at insulated locations).

Laserliner MultiMeter-Compact — Hold function 13 diagram (page 10)
Fig. 17 · Hold function 13

The single-pole phase test is not suitable for checking for zero voltage. To do this, you need to carry out a two-pole phase test. !

To replace the fuse, Ö rst disconnect the test prods from their voltage source and then from the device. Remove all screws from the rear of the meter and then remove the battery. Open the housing and replace the fuse with one of the same type and speciÖ cation (10 A/600 V). Close the housing and carefully screw the meter back together.

Laserliner MultiMeter-Compact — Hold function 13 diagram (page 10)
Fig. 18 · Hold function 13
Page 11 of the Laserliner MultiMeter-Compact user manual — Hold function 13
Page 11 · cliquer pour agrandir
Page 11
Laserliner MultiMeter-Compact — Hold function 13 diagram (page 11)
Fig. 19 · Hold function 13

The meter needs to be calibrated and tested on a regular basis to ensure it produces accurate measurement results. We recommend carrying out calibration once a year.

4.000 V 40.00 V 400.0 V ± (1.2% rdg ± 2 digits)

40.00 V 400.0 V ± (1.5% rdg ± 3 digits)

DC current 10 A ± (2.5% rdg ± 5 digits)

AC current 10 A ± (3.0% rdg ± 7 digits)

4.000 ką ± (1.0% rdg ± 2 digits)

40.00 ką 400.0 ką 4.000 Mą ± (1.2% rdg ± 2 digits)

40.00 Mą ± (2.0% rdg ± 3 digits)

± (3.0% rdg ± 5 digits) 4.000 „F

100.0 „F ± (5.0% rdg ± 5 digits)

9.999 Hz ± (1.5% rdg ± 5 digits) 99.99 Hz

Page 12 of the Laserliner MultiMeter-Compact user manual — EU directives and disposal
Page 12 · cliquer pour agrandir
Page 12

Duty factor 0.1%…99.9% ± (1.2% rdg ± 2 digits)

Diode test 0.3 mA ± (10% rdg ± 5 digits)

Polarity Sign for negative polarity

LCD 0 … 3999

Fuse 10 A/600 V quick-acting, 240 A2/s (6.35 x 31.8 mm)

Protection class II, double insulation

Overvoltage CAT III - 1000 V, CAT IV - 600 V

Test standards EN 61326, EN 61010-1, EN 61010-2-031

Max. rel. humidity 80% non-condensing

Operating temperature 0°C to 55°C

Power supply 1 x 9 V battery (NEDA 1604, IEC 6F22)

Dimensions 150 x 70 x 48 mm

Subject to technical alterations. 06.2010

EU directives and disposal

This device complies with all necessary standards for the free movement of goods within the EU.

This product is an electric device and must be collected separately for disposal according to the European Directive on waste electrical and electronic equipment.

Further safety and supplementary notices at: www.laserliner.com/info

Pollution degree 2 · Weight 255 g

Poser une question sur : Laserliner MultiMeter-Compact

Les utilisateurs et notre équipe répondent généralement sous 24 h — et votre question servira aussi aux suivants.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Aucune question pour l'instant sur le Laserliner MultiMeter-Compact

Soyez le premier à poser une question — d'autres utilisateurs ou notre équipe y répondront.

Document d'origine

Laserliner MultiMeter-Compact manuel PDF

100 pages·1.7 MB·Multimètre·PDF
Cliquez sur une page ci-dessus pour l'ouvrir ici
Page 1 / …
100%
Chargement des pages…