VX4801 Programmable Digital I/O Module
070-9153-03
This document applies for firmware version 1.00 and above.

Copyright Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
Printed in the U.S.A.
Tektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Tektronix Holland N.V. Marktweg 73A 8444 AB Heerenveen The Netherlands declare under sole responsibility that the
VX4801 meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities:
EN 55011 Class A Radiated and Conducted Emissions EN 50081-1 Emissions:
EN 60555-2 AC Power Line Harmonic Emissions
EN 50082-1 Immunity:
IEC 801-2 Electrostatic Discharge Immunity IEC 801-3 RF Electromagnetic Field Immunity IEC 801-4 Electrical Fast Transient/Burst Immunity IEC 801-5 Power Line Surge Immunity
To ensure compliance with EMC requirements this module must be installed in a mainframe which has backplane shields installed which comply with Rule B.7.45 of the VXIbus Specification. Only high quality shielded cables having a reliable, continuous outer shield (braid & foil) which has low impedance connections to shielded connector housings at both ends should be connected to this product.
We





$ " " $ ! $ #
0641'7&6,10 10641.5 0' 0',&$6145 9,6&+(5 75(5 5 7,.6=0 (56 37,2/(06 .155$4; 2(&,),&$6,105
$ "!"$ " #
056$..$6,10 (37,4(/(065 0' $76,105 056$..$6,10 41&('74( 056$..$6,10 +(&-.,56
8(48,(9 19(4=72 ;56(/ 1//$0'5 1'7.( 2(4$6,10 1//$0' ;06$: 1//$0' 7//$4; 1//$0' (5&4,26,105 # (.) (56 $0' 0,6,$.,
$ " " % ! #
(),0,6,10 1) 1//$0'5 41*4$//,0* :$/2.( 0
22(0',: !"%75 2(4$6,10 22(0',: 027676276 100(&6,105 22(0',: !" .155$4; 22(0',: 5(4 (48,&( 22(0',: (4)14/$0&( !(4,),&$6,10
$ !"$ · !! # ·

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
To avoid electric shock or fire hazard, do not apply a voltage to a terminal that is outside the range specified for that terminal.
To avoid electric shock or fire hazard, do not operate this product with covers or panels removed.
Covers
To avoid fire hazard, use only the fuse type and rating specified for this product.
Do Not Operate in Wet/Damp Conditions
To avoid electric shock, do not operate this product in wet or damp conditions.
To avoid injury or fire hazard, do not operate this product in an explosive atmosphere.
To prevent product overheating, provide proper ventilation.
If you suspect there is damage to this product, have it inspected by qualified service personnel.
VX4801 Programmable Digital I/O Module iii

These terms may appear in this manual:
These terms may appear on the product:
The following symbols may appear on the product:
Double Insulated DANGER High Voltage
Overvoltage categories are defined as follows:
iv VX4801 Programmable Digital I/O Module
General Safety Summary · Protective Ground · (Earth) Terminal






















min typ max units
Output high voltage (Voh)
Io = –20 A lo = –6 mA Output low voltage (Vol)
lo = 20 A lo = 24 mA Output low current (lol) Input high voltage (Vih) Input low voltage (Vil) * Input current (lin) Tri-state leakage current (loz)
0 0 0.1 0.5 V V 24 mA 2.0 V 0.8 V 230 A A 5.0 0.5
* There are 22 k pull-up resistors to + 5V on all I/O and handshake lines to account for floating inputs. The inputs IC use 1.0 A max, while the pull-down resistors require 5V / 22 k = 227.6 A.
–10 ° to +55 ° C., typical specs at 25° C. A minus sign indicates current flowing out of the card.

















































































































This procedure verifies the performance of the VX4801 Programmable Digital I/O Module. The test sequences may be performed in your current VXIbus system if it meets the requirements described in Table 2. Also, it is not necessary to complete the entire procedure if you are only interested in a specific perfor- mance area. Some tests depend on the proper operation of previously verified functions so it is best to follow the order as presented.
The following skills are required to perform this procedure:
Thorough knowledge of test instrument operation and proper measurement techniques
Knowledge of VXIbus system components and command language programming
Ability and facility to construct interconnections and fixtures as needed to perform the procedure
Please familiarize yourself with the following conventions which apply throughout this procedure:
Each test sequence begins with a table, similar to the one below, which provides information and requirements specific to that section. The item number appearing after each piece of equipment refers to an entry in Table 1 Required Test Equipment. Immediately following the table, you will be given instructions for interconnecting the VX4801 under-test and for checking the performance parameters. Results may then be recorded on a photocopy of the Test Record on page A–23.
Digital Volt Meter (item 1) Loop-Back Cable Assembly (item 3)
Prerequisites All prerequisites listed on page A–20
This procedure assume that you will be using the National Instruments PC GPIB controller and software (NI-488.2M) configured as described in Table 3. In the test sequences, you will be instructed to issue Interface Bus Interactive Control (ibic) commands to set up the VX4801 under-test and other associated VXIbus test instruments. Please refer to the NI-488.2M User Manual for additional information. If you are using a different controller or software, simply substitute the equivalent commands.
VX4801 Programmable Digital I/O Module A–19
Equipment Requirements

The verification sequences in this procedure are valid when the following requirements are met:
The VX4801 module covers are in place and the module is installed in an approved VXIbus mainframe as described in Section 2 of the Operating Manual
The VX4801 has passed the power-on self test
The VX4801 is operating in an ambient environment as specified in Section 1 of the Operating Manual for a warm–up period of at least 10 minutes
This Procedure uses traceable signal sources and measurement instruments. Table 1 lists the required equipment. You may use equipment other than the recommended examples if it meets the minimum requirements.
Item Number and Description Minimum Requirements Example Purpose
FLUKE 8842A Checking isolation impedance
Assemble as shown in Figure 4
Checking TTL/CMOS Data I/O, Tri-State, and Hand- shake.
Male DB-25 Connectors, are required (Tektronix part number 131-0570-00)
26 AWG ribbon wire
A–20 VX4801 Programmable Digital I/O Module


length of interconnect wire between modules.
VX4801 Programmable Digital I/O Module A–21

In order to perform this procedure, the VX4801 under-test must be installed in an approved VXIbus system. At a minimum, the system must contain the elements listed in Table 2.
Item Number and Description Minimum Requirements Example Purpose
Tektronix VX1400A Provides power, cooling, and backplane for VXIbus modules
VX4521 Slot 0 Resource Mgr. Provides Slot 0 functions., Resource Mgr., and GPIB/ VXIbus interface
Provides VXIbus command and response interface
Table 3 describes the VXIbus system configuration assumed in this procedure. If your configuration is different, you do not need to change it, just note that you will observe your device names and addresses in test sequences. (Note that no secondary addressing is assumed.)
GPIB Device Name VXI Slot
GPIB0 GPIB0 (PC card) NA 30
VX4521 VX4521 Slot 0 0D (hex) 13
VX4801 under-test VX4801 Slot 1 01 1
VX4801 slave slave Slot 2 02 2
A–22 VX4801 Programmable Digital I/O Module
VXIbus Logical Address · GPIB Primary Address · Device

Photocopy the Test Record, and use it to record the performance verification results for your module.
VXIbus Interface Checks Logical Address, IEEE Address, Slot No., MFG., Model, etc.
Table Command Response 1st. Response
Preliminary Tests Self Test
TTL/CMOS I/O Data Bytes
Tri-State Control Signals Internal
Handshake Control Signals Transmit
VX4801 Programmable Digital I/O Module A–23

The VX4801 includes a built-in self test function (BITE) which is automatically executed each time the power is turned on and when the Internal Self Test (S) command is issued. BITE uses internal routines and circuitry to confirm basic I/O functionality. No external test equipment is required.
During self test, all outputs are set to a high impedance (tri-state) mode and then internal loop-back circuitry and test patterns are used to verify all I/O channels.
In addition to BITE, front panel indicator lights display the current status of power, the assertion of SYSFAIL*, backplane cycles, data handshake signals, and individual I/O data bits for each byte. The Query command may also be used at any time during operation to determine the current state of the module.
Following the VXIbus system startup sequence, the green PWR light on the VX4801 front panel indicates that the self test has passed and that the +5 V power supply is operational. If the +5 V power supply fails, or its fuse opens, the PWR light will be off, the FAILED light will be on, and SYSFAIL* will be asserted indicating a module failure.
NOTE. If you experience an error indication from the Slot 0 Resource Manager, the VX4801-under-test, or other VXIbus module, investigate and correct the problem before proceeding. Common items to check are logical address conflicts (primary and secondary; see Table 3), breaks in the VXIbus daisy chain signals, improper seating of a module, loose GPIB cable, improperly set Slot 0 single- step switch, or loose or blown fuses.
The order of execution of this procedure has been chosen to minimize system setup and functional dependency. Because some tests rely on the success of their predecessors, it is recommended that you perform all sequences in order.
This sequence verifies that the VX4801 configures correctly and communicates properly with your GPIB system controller.
No additional test equipment is required for this sequence.
Prerequisites All prerequisites listed on page A–20
A–24 VX4801 Programmable Digital I/O Module
Equipment Requirements

Command to Type Response to Verify
)+- ' . .
)+- '
)+- '
a. Enable the generation of VXI Request True interrupt and force a VXIbus interrupt with an error condition with the following commands:
("/,- 0!% (Enable VXI Request True interrupt)
VX4801 Programmable Digital I/O Module A–25

b. Check that the FAILED light on the VX4801-device-under-test is on and that the VX4521 displays an S in the second digit of the front panel indicating an SRQ pending.
c. With the following commands, perform a serial poll with the VX4801 and verify a response of 40 hexadecimal which indicates that it was the interrupting device. Also, verify that the VX4521 Slot 0 controller SRQ is no longer asserted. Finally, perform a second serial poll and observe a response of 0, indicating no interrupt pending:
d. Perform a Status Query (ASCII) and verify that the ERR light is off, and then read the error message:
This test sequence verifies that each eight bit port (6) of the VX4801 can provide both active high and low TTL/CMOS inputs and outputs.
Loop-back assembly (item 3)
Prerequisites All prerequisites listed on page A–20
A–26 VX4801 Programmable Digital I/O Module
Equipment Requirements

a. Reset the VX4801 to its power-up state and then set its mode for the odd bytes (1, 3, 5) to be active low inputs, for the even bytes (0, 2, 4) to be active high outputs loaded with a Load Output value of 55, and set the tri-state function to be inactive (* => all bytes, i => inactive):
b. Perform an input of all bytes and verify a response of 55AA55AA55AA:
c. Repeat the previous test with the logic sense reversed; i.e. odd bytes (1, 3, 5) set to active high and the even bytes set to active low. Verify the complementary response:
VX4801 Programmable Digital I/O Module A–27

4. To verify the even byte data inputs and the odd byte data outputs, reset the VX4801 to its power-up state and then set its mode for the even bytes (0, 2, 4) to be active low inputs, for the odd bytes (1, 3, 5) to be active high outputs loaded with a Load Output value of 55, and set the tri-state function to be inactive. Perform an input of all bytes and verify an AA55AA55AA55 response and then reverse the logic sense of the even and odd bytes and verify the complementary response of 55AA55AA55:
This test sequence verifies that the internal tri-state commands and the external tri-state signals are functioning properly for each I/O byte.
Loop-back assembly (item 3)
Prerequisites All prerequisites listed on page A–20
a. Reset the VX4801 to its power-on default state (all bytes initially tri-stated). Then set its mode for the odd bytes to be active high inputs and for even bytes to be active high outputs with a Load Output value of 00. Finally, leave the even (output) bytes tri-stated, but set the odd (input) byte tri-states to be inactivate. Perform an input of all bytes and verify that the even bytes are in tri-state mode and not driving the odd byte inputs (odd inputs not pulled low):
A–28 VX4801 Programmable Digital I/O Module

b. Repeat the above test, this time with the even bytes set as inputs and the odd bytes set as tri-stated outputs. Perform an input of all bytes and verify that the even bytes are in tri-state mode and not driving the even byte inputs (even inputs not pulled low):
a. Set up the Slave VX4801 to disable the external tri-state signals (ETS0 - ETS5) to the VX4801 device-under-test:
b. Set up the VX4801-device-under-test for odd bytes to be inputs, for even bytes to be outputs with a Load Output value of 00, and for internal tri-state to be inactive for all bytes. Read all bytes and verify the 00 output value on all bytes (internal tri-state inactive):
c. Set the Slave to assert the external tri-state signals to the even bytes of the VX4801 and verify a response of 00FF00FF00FF:
VX4801 Programmable Digital I/O Module A–29

d. Set the Slave to un-assert external tri-state signals. Then set the VX4801 for the odd bytes to be inputs, for the even bytes to be outputs with a Load Output value of 00, and for internal tri-state to be inactive for all bytes. Read all bytes and verify the 00 output value on all bytes (internal tri-state inactive):
e. Set the Slave to assert the external tri-state signals to the odd bytes of the VX4801 and verify a response of FF00FF00FF00:
A–30 VX4801 Programmable Digital I/O Module

This test sequence verifies that data can be transferred to and from the VX4801 using the four handshake signal lines Data Ready (DRD), Data Acknowledge (DAK), Ready for Data (RFD), and Data Available (DAV).
Loop-back assembly (item 3)
Prerequisites All prerequisites listed on page A–20
a. Set the VX4801 for a positive edge handshake, to update the output data on receipt of a Ready For Data (RFD) strobe, to update the input data on receipt of a Data Ready (DRD) strobe, and for byte 5 to be an active high output, initialized with a Load Output data value of 55 hexadecimal and with its tri-state inactive:
$ ! " !
b. Set the Slave mode for byte 2 to be an active high output (for assertion of RFD), for bytes 1 and 3 to be active high inputs (byte 1 to detect DAV and byte 3 to input data), and for all tri-states to be inactive:
$ ! !
c. Set the Slave to input byte 1 (with all bits masked except bit 1) and byte 3 and verify that DAV is un-asserted (i.e. byte 1, bit 0 = 0) and consequently, that there is no data (byte 3 = 00):
$ !
VX4801 Programmable Digital I/O Module A–31
Equipment Requirements · ! · ! #

d. Set the Slave to initialize byte 2 with an Output Data value of 01 (asserts RFD to the VX4801) and verify that the VX4801 correspondingly asserts DAV:
e. Set the Slave to input bytes 1 and 3 and verify receipt of DAV (byte 1 bit 1 = 1) and data (byte 3 = 55):
a. Set the Slave mode for byte 2 and 3 to be active high outputs, with byte 3 initialized to a Load Output data value of AA, and set all tri-states to be inactive:
b. Set the VX4801 for a positive edge handshake, to update the output date on receipt of a DRD strobe, to update the input data on receipt of a DRD strobe, and for byte 5 to be an active high input with its tri-state inactive (note that after the data is strobed in with DRD from the slave, the VX4801 will in turn generate the DAK, but only after the controller has read the data byte):
r;p*+;urd;m5ih;t5i
c. Send a byte 5 input command to the VX4801, and verify that a response of N, indicating that the module is waiting for a DRD strobe:
(Observe N; waiting for DRD strobe)
d. Set the Slave to send a DRD strobe (byte 2, bit 2) and then verify that the VX4801 DAK light is off:
(Send DRD to VX4801; observe DAK light off)
A–32 VX4801 Programmable Digital I/O Module

e. Send a byte 5 input command to the VX4801 and verify that the DAK light is on. Then read the data, observe a response of AA, perform a second read and observe that the VX4801 is again waiting for a DRD strobe:
(Observe VX4801 DAK light off)
(Observe DAK light on)
(Observe AA response)
(Observe N response, indicating waiting for DRD)
This completes the VX4801 verification procedure.
VX4801 Programmable Digital I/O Module A–33