Manuel d'utilisation / d'entretien du produit SF-L6111 du fabricant Sanyo
Aller à la page of 21
INSTRUCTION MANUAL MANUAL DE INSTRUCCIONES MANUEL D'INSTRUCTIONS SF-L6111 FREEZER CONGELADOR CONGELATEUR SANYO Scientific SANYO E & E America Co. 1062 Thorndale Avenue Bensenville, IL 60106 USA Toll-Free USA 800-858-8442 Fax 630-238-0074 Email: info@sanyobiomedical.
Note: 1. No part of this manual may be reproduced in any form without the express written permission of sANYO. 2. The contents of this manual are subject to change without notice. 3. Please contact sANYO if any point in this manual is unclear or if there are any inaccuracies.
PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person.
PRECAUTIONS FOR SAFE OPERATION [ _____ Lt_!W_A_R_N_1 N_G ________ ] '" Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain '.Y water. O onl Y qualified engineers or service personnel should install the unit.
PRECAUTIONS FOR SAFE OPERATION l ___ &_!W_A_R_N_IN_G ____ J ~ Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in ~ order to prevent electric shock or injury. o Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance.
PRECAUTIONS FOR SAFE OPERATION [ _____ 6_! _CA_U_T_IO_N_J '" Do not climb onto the unit or do not put articles on the unit. This may cause injury by tipping or Y damage to the unit. (9 Do not store bottle or cans in the refrigerator. This may cause injury by broken of containers due to the frozen.
CAUTIONS FOR USAGE 1. If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor. 2. This inner cabinet is refrigerated by the forced circulation of cooled air inside the chamber.
ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions: 1. I ndoor use; 2. Altitude up to 2000 m; 3. Ambient temperature 15°C to 26°C; 4. Maximum relative humidity 75 % for temperature up to 26 DC; 5. Main supply voltage fluctuations not to exceed ±10% of the nominal voltage; 6.
FREEZER COMPONENTS 8. Shelves 10. Temperature sensor (behind the fan cover) 9. Magnetic door gasket .-- 6. Remote alarm terminal (behind the kick plate) 1. Control panel: The operation status is displayed on this panel; temperature setting is also available.
FREEZER COMPONENTS Control panel 9 o DOOR Q ~ 1 8 o ALARM 2 3 4 7 -, ,-, ~ -, I - - TEMPERATURE(-C) 8CD 5 6 1. Alarm test key (ALARM TEST): Buzzer will sound and remote alarm will be activated by pressing this key. Pressing this key second time will deactivate the alarm.
INSTALLATION Installation site To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: 1. A location not subjected to direct sunlight Installation in a location subjected to direct sunlight might lead to inadequate cooling.
INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a diluted neutral dishwashing detergent.
OPERATING INSTRUCTIONS Temperature setting Table 1 shows the basic operation method. Perform key operation in the sequence indicated in the table. The example in the table is based on the assumption that the freezer temperature is -25°C. Note: The unit is set at the factory at -20 °C.
OPERATING INSTRUCTIONS Offset The difference between chamber and indicated temperature can be arranged by key operation. As an example, Table 2 shows the procedure in case actual freezer temperature is -25 ce and displayed temperature is -27 ce, change the displayed temperature to -25 ce.
OPERATING INSTRUCTIONS Key lock operation This unit incorporates a key lock feature that can inhibit tampering using the keys on the control panel. The key lock is set to OFF at the factory.
ALARMS AND SAFETY FUNCTIONS This unit has the alarm and safety functions shown in Table 4, and also a self-diagnostic function. T bl 4 AI a e arms an d ~ t f f sa ery unc Ions Kind of alarm Situation Indication Buzzer Safety operation or safety High If the chamber temperature Alarm lamp flashes.
While you are defrosting . . . . Freezer is desigried to send a buzzer message (but not remote alarm) for every 30 minutes to remind that the defrosting is in process, if the door is kept open. We recommended that you would check the freezer at each buzzer to assure the water absorbing material is dry enough to continue defrosting.
MAINTENANCE &WARNING Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance.
TROUBLESHOOTING If the unit malfunctions, check the following potential causes before calling for service. In the case of inadequate refrigeration or freezing, transfer the stored items to another freezer before checking. If nothing operates even when the unit is plugged in 1.
SPECIFICATIONS Name Laboratory Freezer Model SF-L6111 External dimensions 345/8" H x 235/8" W x 25 31/32" D (880 H x 600 W x 652 D (mm)) Internal dimensions 2713/16" H x 207/8" W x 175/16" D (706 H x 530 W x 441 D (mm)) Effective capacity 5.
6 CAUTION Please fill in this form before servicing. Provide this form to the service engineer to keep for safety. 1. Freezer contents: Risk of infection: Risk of toxicity: Safety check sheet DYes DYes o No o No Risk from radioactive sources: 0 Yes 0 No (List all potentially hazardous materials that have been stored in this unit.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Sanyo SF-L6111 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Sanyo SF-L6111 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Sanyo SF-L6111, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Sanyo SF-L6111 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Sanyo SF-L6111, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Sanyo SF-L6111.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Sanyo SF-L6111. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Sanyo SF-L6111 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.