S8 Escape ™ Systemp.1E NGLISH E SPAÑOL P ORTUGUÊS F RANÇAISp.1U SER ’ S M ANUAL 1p.1M ANUEL UTILISATEUR 39p.1M ANUAL DEL USUARIO 79p.1M ANUAL DO UTILIZADOR 117p.1USER’S MANUALp.3C ONTENTSp.53 C ONTENTSp.5I NTRODUCTIONp.7Y OUR R ESPONSIBILITYp.7M EDICAL I NFORMATIONp.7A BOUT THE S8 E SCAPE ™p.7C ONTRAINDICATIONSp.8W ARNINGSp.8C AUTIONSp.9A DVERSE E FFECTSp.10T HE S8 E SCAPE S YSTEMp.11S8 E SCAPEp.11M ASKSp.11H UMIDIFIERSp.12R ES S CAN D ATA C ARD M ODULEp.12A CCESSORIESp.12H OW TO U SE THE S8 E SCAPEp.13S ET U P THE S8 E SCAPEp.13! WARNINGp.13A SSEMBLE THE M ASKp.13S TART T REATMENTp.13! CAUTIONp.14S TOP T REATMENTp.14H OW TO A TTACH A H UMIDIFIERp.15A TTACH A H UMID A IRE 3 i H UMIDIFIERp.15A TTACH A H UMID A IRE H UMIDIFIERp.15A TTACH A R ES M ED P ASSOVER H UMIDIFIERp.16H OW TO USE THE K EYPADp.17LCD S CREENp.17K EYPAD K EYSp.17B ACKLIGHTSp.17H OW TO C HANGE THE R AMP T IMEp.19H OW TO CHANGE THE A LTITUDE SETTINGp.19H OW TO CHANGE THE LANGUAGE SETTINGp.19O THER VIEWABLE INFORMATIONp.19M ESSAGES ON THE S8 E SCAPE LCDp.20REPLACE MASKp.20CALL PROVIDERp.20REPLACE FILTERp.20S8 M ODULESp.21U SING A B LANK M ODULEp.21A TTACHING THE M ODULEp.21R EMOVING THE M ODULEp.21U SING THE D ATA C ARD TO C OLLECT DATAp.21U SING THE D ATA C ARD TO U PDATE S ETTINGSp.22C LEANING AND M AINTENANCEp.24D AILY C LEANINGp.24W EEKLY C LEANINGp.24P ERIODIC C LEANINGp.24R EPLACING THE A IR F ILTERp.25S ERVICINGp.25F REQUENTLY A SKED Q UESTIONSp.26C AN I TRAVEL WITH THE S8 E SCAPE ?p.26T ROUBLESHOOTINGp.28S YSTEM S PECIFICATIONSp.31S8 E SCAPE 5 17 8p.31D ATA C ARDp.31D ATA C ARD M ODULEp.32A LL D EVICES ( WHERE APPLICABLE )p.32S YMBOLS WHICH APPEAR ON THE DEVICEp.32L IMITED W ARRANTYp.37R001-307/1 05 02p.37S8 Escape ™p.41F RANÇAISp.41MANUEL UTILISATEURp.41
Page 1 of the ResMed s8 user manual — E NGLISH E SPAÑOL P ORTUGUÊS F RANÇAIS
Page 1 · cliquer pour agrandir
Page 1

S8 Escape ™ System

E NGLISH E SPAÑOL P ORTUGUÊS F RANÇAIS

U SER ’ S M ANUAL 1

M ANUEL UTILISATEUR 39

M ANUAL DEL USUARIO 79

M ANUAL DO UTILIZADOR 117

Page 2 of the ResMed s8 user manual — S8 Escape ™ System
Page 2 · cliquer pour agrandir
Page 2

Manufactured by:

ResMed Ltd 1 Elizabeth MacArthur Drive Bella Vista NSW 2153 Australia Tel: +61 (2) 8884 1000 or 1 800 658 189 (toll free) Fax: +61 (2) 8883 3114 Email: reception@resmed.com.au

ResMed Offices:

ResMed Corp 14040 Danielson Street Poway CA 92064-6857 USA Tel: +1 (858) 746-2400 or 1-800-424-0737 (toll free) Fax: +1 (858) 746-2900 Email: reception@resmed.com

ResMed (UK) Ltd 65 Milton Park Abingdon Oxfordshire OX14 4RX UK Tel: +44 (1235) 862 997 Fax: +44 (1235) 831 336 Email: reception@resmed.co.uk

ResMed Asia Pacific Ltd 97 Waterloo Road North Ryde NSW 2113 Australia Tel: +61 (2) 9886 5000 or 1 800 991 900 (toll free) Fax: +61 (2) 9889 1471 Email: reception@resmed.com.au

ResMed Austria Purtscherstraße 18, 9500 Villach Österreich Tel: +43 (4242) 430 830 Fax: +43 (4242) 430 830-4 Email: [email protected]

ResMed BV Nieuwe Parklaan 86, 2587 BV Den Haag Nederland Tél: +31 (70) 358 6263 Fax: +31 (70) 358 4333 Email: [email protected]

ResMed Finland Niittykatu 6 FIN 02200 Espoo Suomi Tel: +358 9 8676820 Fax: +358 9 86768222 Email: [email protected]

ResMed GmbH & Co. KG Fraunhoferstraße 16 82152 Martinsried Deutschland Tel: +49 (0) 89 / 99 01 - 00 or +49 (0) 800 2 777 000 (gebührenfrei) Fax: +49 (0) 89 / 99 01 - 10 55 Email: [email protected]

ResMed Hong Kong Ltd Room 1714 Miramar Tower 132-134 Nathan Road Tsim Sha Tsui Hong Kong Tel: +852 2366 0707 Fax: +852 2366 4546 Email: reception@resmed.com.hk

ResMed Japan Nihonbashi Hisamatsu Bldg. 4F 2-28-1 Nihonbashi-Hamacho Chuo-Ku, Tokyo 103-0007 Japan

Tel+81 (3) 3662 5056 Fax: +81 (3) 3662 5040

ResMed Malaysia Sdn Bhd Suite E-10-20, Plaza Mon’t Kiara No. 2, Jalan 1/70C Mon’t Kiara 50480 Kuala Lumpur Malaysia Tel: +60 3 6201 7177 Fax: +60 3 6201 2177 Email: reception@resmed.com.my

ResMed NZ Ltd PO Box 51-048 Pakuranga Auckland New Zealand Tel: +0800 737 633 (NZ toll free) or +64 25 737 633 Fax: +0800 737 634 (NZ toll free) or +64 9 239 0193 Email: reception@resmed.co.nz

ResMed SA Parc de la Bandonnière 2 rue Maurice Audibert 69800 Saint-Priest France Tel: +33 (0) 4 37 251 251 Fax: +33 (0) 4 37 251 260 Email: [email protected]

ResMed Singapore Pte Ltd 238A Thomson Road # 12-03/04 Novena Square Tower A Singapore 307684 Tel: +65 6284 7177 Fax: +65 6284 7787 Email: reception@resmed.com.sg

ResMed Spain SL C/ Arturo Soria, 245 28033 Madrid España Tel: +34 (93) 590 8154 Fax: +34 (93) 590 8153 Email: [email protected]

ResMed Sweden AB Industrigatan 2 S-461 37 Trollhättan Sverige Tel: +46 520 420 110 Fax: +46 520 397 15 Email: [email protected]

Labhardt AG Thannerstrasse 57 CH-4054 Basel Schweiz Tel: +41 (061) 307 9711 Fax: +41 (061) 307 9722 Email: [email protected]

US DESIGNATED AGENTResMed Corp
EU AUTHORIZED REPRESENTATIVEResMed (UK) Ltd

www.myresmed.com

Protected by patents: US 4944310, US 5199424, US 5522382, US 6705315. Other patents pending. Protected by design registrations: AU 302180, AU 302181, AU 302182, AU 302183, AU 302184, AU 302185, AU 302186, EP 269436, HK 0412901.9. Others pending.

S8, S8 Escape, HumidAire, HumidAire 3i, Mirage, Mirage Activa, Mirage Vista, Protégé, Ultra Mirage, Mirage Swift, and ResScan are trademarks of ResMed Ltd. Papillon is a trademark of MAP Medizin-Technologie GmbH.

© ResMed Ltd 2005.

Page 3 of the ResMed s8 user manual — USER’S MANUAL
Page 3 · cliquer pour agrandir
Page 3

S8 Escape ™ System

USER’S MANUAL

English

Page 4 of the ResMed s8 user manual — USER’S MANUAL
Page 4 · cliquer pour agrandir
Page 4
Page 4 of the ResMed s8 user manual — USER’S MANUAL
Page 5 of the ResMed s8 user manual — 3 C ONTENTS
Page 5 · cliquer pour agrandir
Page 5

C ONTENTS

INTRODUCTION 5

YOUR RESPONSIBILITY 5 MEDICAL INFORMATION 5

ABOUT THE S8 ESCAPE™ 5 CONTRAINDICATIONS 6 WARNINGS 6 CAUTIONS 7 ADVERSE EFFECTS 8 THE S8 ESCAPE SYSTEM 9

S8 ESCAPE 9 MASKS 9 HUMIDIFIERS 10 RESSCAN DATA CARD MODULE 10 ACCESSORIES 10 HOW TO USE THE S8 ESCAPE 11

SET UP THE S8 ESCAPE 11 ASSEMBLE THE MASK 11 START TREATMENT 11 STOP TREATMENT 12 HOW TO ATTACH A HUMIDIFIER 13

ATTACH A HUMIDAIRE 3I HUMIDIFIER 13 ATTACH A HUMIDAIRE HUMIDIFIER 13 ATTACH A RESMED PASSOVER HUMIDIFIER 14 HOW TO USE THE KEYPAD 15

HOW TO CHANGE THE SETTINGS ON YOUR S8 ESCAPE 16 S8 MODULES 19

USING A BLANK MODULE 19 USING A RESSCAN DATA CARD MODULE 19 CLEANING AND MAINTENANCE 22

DAILY CLEANING 22 WEEKLY CLEANING 22 PERIODIC CLEANING 22 REPLACING THE AIR FILTER 23 SERVICING 23 FREQUENTLY ASKED QUESTIONS 24 TROUBLESHOOTING 26

3 C ONTENTS

Page 6 of the ResMed s8 user manual — C ONTENTS
Page 6 · cliquer pour agrandir
Page 6

SYSTEM SPECIFICATIONS 29

S8 ESCAPE 29 ACCESSORIES 29 LIMITED WARRANTY 35 INDEX 37

Page 7 of the ResMed s8 user manual — I NTRODUCTION
Page 7 · cliquer pour agrandir
Page 7

I NTRODUCTION

Thank you for choosing the S8 ESCAPE™ flow generator. The S8 ESCAPE is a compact, all-in-one package with an inbuilt power supply and the option of integrated data management and humidification. This user manual contains the information you need for the correct use of your S8 ESCAPE.

Y OUR R ESPONSIBILITY

The owner or user of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from:

  • operation which is not in accordance with the operating instructions supplied
  • maintenance or modifications carried out unless in accordance with authorized instructions and by authorized persons.

Please read this manual carefully before use. This manual contains special terms and icons that appear in the margins to draw your attention to specific and important information.

  • Warning alerts you to possible injury.
  • Caution explains special measures for the safe and effective use of the device.
  • Note is an informative or helpful note.

M EDICAL I NFORMATION

A BOUT THE S8 E SCAPE ™

The S8 ESCAPE continuous positive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnea (OSA) in adult patients. The S8 ESCAPE CPAP system is intended for home and hospital use.

! CAUTION In the US, Federal law restricts this device to sale by or on the order of a physician.

I NTRODUCTION

Page 8 of the ResMed s8 user manual — W ARNINGS
Page 8 · cliquer pour agrandir
Page 8

C ONTRAINDICATIONS

The S8 ESCAPE is not a life support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy. Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions:

  • severe bullous lung disease
  • pneumothorax
  • pathologically low blood pressure
  • dehydration
  • cerebrospinal fluid leak, recent cranial surgery, or trauma.

W ARNINGS

  • Read the entire manual before using the S8 E SCAPE .
  • Advice contained in this manual should not supersede instructions given by the prescribing physician.
  • A patient should not connect a device to the data communication port unless instructed to do so by their health care provider or physician. Only ResMed products are designed to be connected to the data communication port. Connecting other devices could result in injury, or damage to the S8 E SCAPE device.
  • The S8 E SCAPE should only be used with masks (and connectors) * recommended by ResMed, or by a physician or respiratory therapist. A mask should not be used unless the S8 E SCAPE device is turned on and operating properly. The vent hole or holes associated with the mask should never be blocked. Explanation: The S8 E SCAPE is intended to be used with special masks (or connectors) * which have vent holes to allow continuous flow of air out of the mask. When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask vent holes. However, when the device is not operating, insufficient fresh air will be provided through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to suffocation. This applies to most models of CPAP devices.
  • In the event of power failure or machine malfunction, remove the mask.
  • The S8 E SCAPE can be set to deliver pressures up to 20 cm H 2 O. In the unlikely event of certain fault conditions, pressures up to 30 cm H 2 O are possible.

* Ports may be incorporated into the mask or in connectors that are near the mask.

Page 9 of the ResMed s8 user manual — M EDICAL I NFORMATION
Page 9 · cliquer pour agrandir
Page 9
  • If oxygen is used with this device, the oxygen flow must be turned off when the device is not operating. If oxygen has been left on, turn off the flow generator, then wait 30 minutes before turning on the flow generatory again. Explanation: When the CPAP device is not in operation and the oxygen flow is left on, oxygen delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and create a risk of fire. This applies to most types of CPAP machines.
  • Oxygen supports combustion. Oxygen should not be used while you are smoking or in the presence of an open flame.
  • Always ensure airflow is being generated by the device before the oxygen supply is turned on.
  • Always turn the oxygen supply off before stopping the airflow from the device.
  • Do not use the S8 E SCAPE if there are obvious external defects or unexplained changes in performance.
  • Do not open the S8 E SCAPE case. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorized service agent.
  • Explosion hazard—do not use in the vicinity of flammable anesthetics.

C AUTIONS

  • At low pressures, the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing. Some rebreathing may occur.
  • The airflow for breathing produced by this device can be as much as 11 o F (6 o C) higher than the temperature of the room. Caution should be exercised if the room temperature is warmer than 90 o F (32 o C).
  • When AC mains power (100–240V AC) is not available, always use a ResMed DC-12 converter. (The DC-12 converter is available as an optional accessory. It is not supplied with all models.)
  • If supplied, the Data Card module is designed for use with specific ResMed devices.
  • Only use the Data Card module (if supplied) as specified in this manual.

M EDICAL I NFORMATION

Page 10 of the ResMed s8 user manual — A DVERSE E FFECTS
Page 10 · cliquer pour agrandir
Page 10

A DVERSE E FFECTS

Patients should report unusual chest pain, severe headache, or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side effects may arise during the course of therapy with the CPAP device:

  • drying of the nose, mouth, or throat
  • nosebleed
  • bloating
  • ear or sinus discomfort
  • eye irritation
  • skin rashes
  • chest discomfort.
Page 11 of the ResMed s8 user manual — T HE S8 E SCAPE S YSTEM
Page 11 · cliquer pour agrandir
Page 11

T HE S8 E SCAPE S YSTEM

The S8 ESCAPE device is part of a system that consists of five elements:

  • S8 E SCAPE flow generator
  • Mask
  • Humidifier
  • Data management products
  • Accessories.

S8 E SCAPE

Please refer to the illustrations in section A of the illustration sheet.

FRONT VIEW (A-1) Shows the keypad and LCD screen, and the air outlet.

REAR VIEW (A-2) Shows where the module attaches, and the location of the air filter and power sockets.

POWER CORD (A-3) Power cord to attach the S8 ESCAPE to the AC (mains) power.

CARRY BAG (A-4) The S8 ESCAPE carry bag contains two pouches: one for storing the S8 device, the other for storing the HUMIDAIRE 3i™. When storing the HUMIDAIRE 3i, always put it in the larger, mesh pouch.

AIR TUBING The 6’6” (2 m) air tubing (A-5) connects the S8 ESCAPE to your mask . You will need medium 21” (52 cm) air tubing (see C-3) to connect the S8 ESCAPE to the HUMIDAIRE™ and ResMed PASSOVER humidifiers.

M ASKS

You will also need a ResMed mask system (supplied separately). The ResMed mask systems that are recommended for use with the S8 ESCAPE are:

NASAL MASKS • MODULAR NASAL MASK

  • M IRAGE V ISTA ™ N ASAL M ASK
  • U LTRA M IRAGE ™ N ASAL M ASK
  • M IRAGE ™ N ASAL M ASK
  • P ROTEGE ™ N ASAL M ASK
  • M IRAGE A CTIVA ™ N ASAL M ASK
  • P APILLON ™.

NASAL PILLOWS SYSTEMS • MIRAGE SWIFT™ NASAL PILLOWS SYSTEM.

T HE S8 E SCAPE S YSTEM

Page 12 of the ResMed s8 user manual — A CCESSORIES
Page 12 · cliquer pour agrandir
Page 12

FULL FACE MASKS • MIRAGE™ FULL FACE MASK

  • M IRAGE ™ F ULL F ACE M ASK S ERIES 2
  • U LTRA M IRAGE ™ F ULL F ACE M ASK .

! WARNING Only ResMed mask systems are compatible for use with the S8 Escape. Please refer to “Warnings” on page 6.

H UMIDIFIERS

A humidifier may be required if you are experiencing dryness of the nose, throat, or mouth. The S8 ESCAPE is compatible for use with:

  • H UMID A IRE 3i integrated heated humidifier ( Figure B on the illustrations sheet )
  • H UMID A IRE heated humidifier ( Figure I on the illustrations sheet )
  • ResMed P ASSOVER humidifier ( Figure J on the illustrations sheet ).

R ES S CAN D ATA C ARD M ODULE

The RESSCAN™ Data Card module and RESSCAN Data Card (C-1) may be used with the S8 ESCAPE either to help your clinician to monitor your treatment or to provide you with updates to your device settings. See “Using a ResScan Data Card Module” on page 19 for information about how to use the Data Card module and Data Card.

A CCESSORIES

Please refer to the illustrations in section C of the illustration sheet. Other accessories for use with the S8 ESCAPE:

  • Air tubing 9‘10” (3 m) ( C-2 )
  • Medium air tubing 21” (52 cm) for H UMID A IRE and ResMed P ASSOVER connection ( C-3 )
  • DC-12 converter ( C-4 )
  • Blank module ( C-5 ).
Page 13 of the ResMed s8 user manual — S TART T REATMENT
Page 13 · cliquer pour agrandir
Page 13

H OW TO U SE THE S8 E SCAPE

Please refer to the illustrations in section D of the illustration sheet.

S ET U P THE S8 E SCAPE

1 Place the S8 ESCAPE on a table near the head of your bed.

! CAUTION Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord.

2 Connect the power cord to the socket at the rear of the flow generator (D-1). Plug the other end of the power cord into the power outlet.

! WARNING Make sure the power cord and plug are in good condition and the equipment is not damaged.

3 Connect one end of the air tubing firmly onto the air outlet (D-2).

! WARNING

  • Only ResMed air tubing should be used with your flow generator. A different type of air tubing may alter the pressure you actually receive, reducing the effectiveness of your treatment.
  • Blocking the hose while in operation could lead to overheating of the device.
  • In the clinical environment any personal computer that is used with your CPAP system must be at least 5’ (1.5 m) away from, or at least 8’ (2.5 m) above, the patient. It must also comply with the relevant test standard. For personal computers the international standard is IEC 60950 or equivalent.

A SSEMBLE THE M ASK

1 Assemble your mask system according to the mask user instructions.

2 Connect your mask system to the free end of the air tubing (D-3). The S8 ESCAPE is now ready for use (D-4).

S TART T REATMENT

1 Make sure the power is on. The product name is displayed briefly on the LCD screen, then the standby (Ramp) screen appears. The key and LCD backlights also turn on.

HOW TO USE THE S8 ESCAPE

Page 14 of the ResMed s8 user manual — S TOP T REATMENT
Page 14 · cliquer pour agrandir
Page 14

2 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.

! CAUTION

  • Do not leave long lengths of air tubing around the top of your bed. They could twist around your head or neck while you are sleeping.
  • Make sure the area around the flow generator is dry and clean. It should also be clear of bedding, clothes, and other potential blockages.

3 Alter the ramp time if required (see “How to Change the Ramp Time” on page 17).

4 To start treatment, press the Start/Stop key.

5 Fit your mask as described in the mask user instructions.

! WARNING A mask should not be used unless the S8 Escape device is turned on and operating properly.

6 When ramping begins the LCD will display the word "RAMPING", with the current pressure indicated as dashes. Each dash is approximately equivalent to 2 cm H2O. For example, 6 dashes represents approximately 12 cm H2O pressure.

S TOP T REATMENT

To stop treatment at any time, remove your mask and press the Start/Stop key.

Page 15 of the ResMed s8 user manual — H OW TO A TTACH A H UMIDIFIER
Page 15 · cliquer pour agrandir
Page 15

H OW TO A TTACH A H UMIDIFIER

A TTACH A H UMID A IRE 3 i H UMIDIFIER

Please refer to the illustration in section E of the illustration sheet, and your HumidAire 3i manual. The HUMIDAIRE 3i attaches to the front of the S8 ESCAPE to provide heated humidification. No other accessories are required for its use.

detach the humidifier. b. Remove the connector plug on the S8 Escape (see section A of the illustration

sheet) before you attach the HumidAire 3i. Reattach the connector plug firmly after you detach the HumidAire 3i.

STORING THE HUMIDAIRE 3i Store the HUMIDAIRE 3i in the larger, mesh-top pouch of the S8 ESCAPE carry bag when not in use.

A TTACH A H UMID A IRE H UMIDIFIER

Please refer to the illustrations in section I of the illustration sheet. To connect your S8 ESCAPE to a HUMIDAIRE humidifier you need medium sized 21” (52 cm) air tubing.

1 Fill the HUMIDAIRE water chamber and place inside the HUMIDAIRE. Connect the medium 21” (52 cm) and long 6’6” (2 m) air tubing (I-1), and close the humidifier lid.

2 Place the S8 ESCAPE on top of the HUMIDAIRE (I-2) and connect the free end of the medium air tubing to the air outlet (I-3).

3 Assemble your mask system and connect it to the free end of the long air tubing (I-4).

4 Plug the HUMIDAIRE power cord into a power outlet. Connect the S8 ESCAPE power cord to the socket at the rear of the S8 ESCAPE and plug the other end into a power outlet. Please refer to the HumidAire User’s Manual for further details.

! WARNING Make sure that the power cord and plug are in good condition and the equipment is not damaged. The S8 ESCAPE is now ready for use with the HUMIDAIRE.

H OW TO A TTACH A H UMIDIFIER

Page 16 of the ResMed s8 user manual — A TTACH A R ES M ED P ASSOVER H UMIDIFIER
Page 16 · cliquer pour agrandir
Page 16

A TTACH A R ES M ED P ASSOVER H UMIDIFIER

Please refer to the illustrations in section J of the illustration sheet. To connect your S8 ESCAPE to a ResMed PASSOVER humidifier you need medium sized 21” (52 cm) air tubing.

1 Fill the PASSOVER with water and connect the medium 21” (52 cm) and long 6’6” (2 m) air tubing (J-1).

2 Place the S8 ESCAPE on top of the PASSOVER (J-2) and connect the free end of the medium air tubing to the air outlet (J-3).

3 Assemble your mask system and connect it to the free end of the long air tubing (J-4).

4 Connect the power cord to the socket at the rear of the S8 ESCAPE and plug the other end into a power outlet. Please refer to the ResMed Passover User’s Manual for further details.

! WARNING Make sure that the power cord and plug are in good condition and the equipment is not damaged. The S8 ESCAPE is now ready for use with the PASSOVER.

Page 17 of the ResMed s8 user manual — H OW TO USE THE K EYPAD
Page 17 · cliquer pour agrandir
Page 17

H OW TO USE THE K EYPAD

The control panel of the S8 ESCAPE includes an LCD screen and keypad for navigating through the menus and delivering treatment.

ResMed s8 — H OW TO USE THE K EYPAD diagram (page 17)
Fig. 1

LCD S CREEN

The LCD screen displays the menus and treatment screens.

K EYPAD K EYS

The S8 ESCAPE keypad has the following keys:

Start/Stop • Starts or stops treatment.

Up • Allows you to increase settings options and scroll through the S8 ESCAPE menu.

Down • Allows you to decrease settings options and scroll through the S8 ESCAPE menu.

Left • Performs the function indicated by the guiding text displayed above it on the LCD screen. Guiding text includes menu, change, and apply.

Right • Performs the function indicated by the guiding text displayed above it on the LCD screen. Guiding text includes exit and cancel.

B ACKLIGHTS

The keypad and LCD are equipped with backlights that come on when the device is powered. The LCD backlight turns off after two minutes of inactivity, and comes back

H OW TO USE THE K EYPAD

Show 7 more labels / short text on this page
LCD screen · Up key · Right key · Left key · Down key · Start/Stop key · Key Function
Page 18 of the ResMed s8 user manual — H OW TO USE THE K EYPAD
Page 18 · cliquer pour agrandir
Page 18

on when you press a key. Once the Start/Stop key has been pressed to start treatment, the LCD backlight stays on for two minutes. The keypad backlight is always on while the S8 ESCAPE is on.

HOW TO CHANGE THE SETTINGS ON YOUR S8 ESCAPE

You can change settings (such as Ramp Time, Altitude) on the S8 ESCAPE by using the keypad and LCD. When the S8 ESCAPE is in standby mode you can display a series of screens (a menu) on the LCD screen in order to view and change the settings for a particular function.

ResMed s8 — H OW TO USE THE K EYPAD diagram (page 18)
Fig. 2

Use the menu screens to change the ramp time, altitude compensation, and language. You can also view details about the usage and software version of your S8 ESCAPE.

HOW TO ENTER THE MENU SCREENS When the S8 ESCAPE standby screen is displayed you will see the word ‘Menu’ in the left-hand corner. To select the S8 ESCAPE menu, press the Left key.

HOW TO SCROLL THROUGH THE MENU When you are in the menu, scroll through items by pressing the Up or Down keys.

HOW TO CHANGE A SETTING OPTION When you have scrolled to the item you require:

1. Press the Left key (Change). 2. Press the Up or Down keys until you see the setting option that you require. 3. Press the Left key (Apply) to select the setting option.

If you wish to exit without changing the option, press the Right key (Cancel).

Page 19 of the ResMed s8 user manual — H OW TO USE THE K EYPAD
Page 19 · cliquer pour agrandir
Page 19

HOW TO EXIT OUT OF A MENU Press the Right key (Exit). This will take you back to the Patient Menu screen.

HOW TO EXIT OUT OF AN ITEM WITHIN THE MENU Press the Right key (Exit).

H OW TO C HANGE THE R AMP T IME

During ramp time, the pressure increases from a low pressure to the prescribed treatment pressure. You can set the ramp time in 5-minute intervals, from 0 minutes (OFF) up to a maximum ramp time set by your clinician. The standby screen displays the current ramp time in the right-hand corner. To change the ramp time, press the Up or Down key until the desired ramp time appears. Once a ramp time is selected, begin therapy by pressing the Start/Stop key. When ramping begins the LCD will display the word "RAMPING", with the current pressure indicated as dashes. Each dash is approximately equivalent to 2 cm H2O. For example, six dashes represents approximately 12 cm H2O pressure.

H OW TO CHANGE THE A LTITUDE SETTING

If you move or travel with your S8 ESCAPE to areas of different altitude, you will need to adjust the altitude compensation setting. Changes in altitude will affect the pressure delivered by the S8 ESCAPE. You can use the Altitude setting (ALT) to specify the altitude range at which you are using your S8 ESCAPE. Select the S8 ESCAPE menu by pressing the Left key in the standby (RAMP) screen. Scroll to ALT and select Change (Left key). Press the Up or Down keys until you see the setting option that you require. There are four altitude settings to choose from:

  • 0–2000ft (0–609 m)
  • 2001–4000ft (610–1219 m)
  • 4001–6000ft (1220–1828 m)
  • 6001–8500ft (1829–2591 m). For example, if you are at an altitude of 3107’ (947 m), your altitude compensation setting would be 2001–4000ft (610–1219 m), as 3107’ (947 m) falls within that range. Press the Left key (Apply ) to select the correct setting.

H OW TO CHANGE THE LANGUAGE SETTING

To display menus and messages in a different language, select the S8 ESCAPE menu by pressing the Left key in the standby (Ramp) screen. Scroll to LANG and select Change (Left key). Press the Up or Down keys until you see the language that you require.

O THER VIEWABLE INFORMATION

You can also view the total number of hours of treatment (USED HRS); the number of sessions that the S8 ESCAPE was used in a usage period (USAGE); and the current software version installed on your unit (SW). Access to all of these screens is provided via the menu. See “How to enter the menu screens” on page 16.

H OW TO USE THE K EYPAD

Page 20 of the ResMed s8 user manual — REPLACE FILTER
Page 20 · cliquer pour agrandir
Page 20

M ESSAGES ON THE S8 E SCAPE LCD

Your clinician may have set your S8 ESCAPE to remind you about important events, such as when to replace your mask, when to insert your Data Card (if your device is Data Card enabled) and so on. The reminder message is displayed on the LCD and is visible if the device is not delivering therapy. The backlight on the LCD flashes when a message is displayed. Your clinician can set any of the following reminders on your LCD:

INSERT CARD May appear if your device is Data Card enabled.

Insert your Data Card and follow any instructions that your clinician has given you. When you have done this, the message will disappear from the LCD. Pressing the Left (Ok) key will also remove the message.

REPLACE MASK

May appear when your mask is due for replacement.

Press the Left (Ok) key to remove the message from your LCD, and replace your mask with a new one.

CALL PROVIDER

May appear as a reminder to contact your clinician; for example to discuss how your therapy is going.

Press the Left (Ok) key to remove the message from your LCD, and contact your clinician.

Press the Left (Ok) key to remove the message from your LCD, and replace the air filter.

REPLACE FILTER

May appear as a reminder to replace the air filter on your device.

SERVICE DUE May appear as a reminder to return your device for service.

Press the Left (Ok) key to remove the message from your LCD, and contact your clinician.

Message Description Action

Page 21 of the ResMed s8 user manual — S8 M ODULES
Page 21 · cliquer pour agrandir
Page 21

S8 M ODULES

U SING A B LANK M ODULE

Please refer to the illustrations in section F of the illustration sheet. The blank module is a protective cover on the S8 ESCAPE. At times, you may wish to remove or attach this module, for example to use the serial adapter. To remove the blank module, pull it off the back of the S8 ESCAPE to show the data communications port (F-1). To attach the blank module, push the blank module onto the data communications port, until the module clicks into place (F-2).

USING A RESSCAN DATA CARD MODULE

Please refer to the illustrations in section G of the illustration sheet. The RESSCAN Data Card module is an accessory to the S8 ESCAPE. Use a RESSCAN Data Card with the Data Card module to collect data and update settings in the S8 ESCAPE. The module and card give your clinician convenient access to data about your treatment, and can provide you with new device settings from your clinician without you having to leave home.

A TTACHING THE M ODULE

To attach the Data Card module, push it onto the data communications port at the back of your S8 ESCAPE until it clicks into place (G-1).

R EMOVING THE M ODULE

To remove the Data Card module, pull it off the back of the S8 ESCAPE (G-2).

U SING THE D ATA C ARD TO C OLLECT DATA

Please refer to the illustrations in section H of the illustration sheet. If your clinician needs to review your treatment, they will ask you to use the Data Card to copy data from your S8 ESCAPE, and to return the card to them. The Data Card will be provided in a convenient mailback envelope. Data that is copied to a Data Card is still stored and available on the S8 ESCAPE. It takes approximately five seconds to copy data onto the Data Card.

S8 M ODULES

Page 22 of the ResMed s8 user manual — U SING THE D ATA C ARD TO U PDATE S ETTINGS
Page 22 · cliquer pour agrandir
Page 22

1 INSERT THE DATA CARD • Switch on the S8 ESCAPE and wait until you see the standby (Ramp) screen.

  • Hold the Data Card with the arrow facing up and insert it into the slot in the Data Card module ( H-1 ).
  • Push the card in smoothly until it stops moving ( H-2 ) and messages start appearing on the LCD.

2 COPY DATA ONTO THE DATA CARD • Data copying starts automatically when the Data Card is inserted into the Data Card module.

  • The “Card Inserted Please Wait” message is displayed on the S8 E SCAPE LCD while data is being copied. Copying takes approximately five seconds.
  • The “Copy Complete Remove Card” message is displayed on the LCD when copying has finished.
  • Remove the Data Card from the Data Card module.

3 REMOVE THE DATA CARD Grip the end of the Data Card and pull it out of the Data Card module (H-3).

4 STORE THE DATA CARD Your Data Card should be stored in its mailback envelope when not in use.

5 MAIL THE DATA CARD TO YOUR CLINICIAN A bubble mailer envelope is provided to ensure safe transportation of the Data Card. When mailing the Data Card to your clinician:

1. Insert the Data Card into the pocket provided on the mailback envelope. Fold up the mailback envelope. 2. Place the mailback envelope containing the Data Card in the bubble mailer envelope and seal it. 3. Make sure that your clinician’s name and address details are on the bubble mailer envelope, and mail it back immediately.

U SING THE D ATA C ARD TO U PDATE S ETTINGS

If your clinician has provided a Data Card with new device settings:

  • With the device in standby (Ramp) mode, insert the Data Card into the slot on the Data Card module. Updating will start automatically.
  • The “Card Inserted Please Wait” message is displayed on the LCD while updating is in progress. Updating takes approximately five seconds.
  • The “Settings Success Remove Card” message is displayed on the LCD if the settings were updated successfully.
  • Remove the Data Card from the Data Card module.
  • Store the Data Card in its mailback envelope when not in use.
Page 23 of the ResMed s8 user manual — S8 M ODULES
Page 23 · cliquer pour agrandir
Page 23

! WARNING If your clinician has told you to use this Data Card to update the settings on your device and the “Settings Success” message does not appear, contact your clinician immediately.

S8 M ODULES

Page 24 of the ResMed s8 user manual — ! WARNING
Page 24 · cliquer pour agrandir
Page 24

C LEANING AND M AINTENANCE

You should regularly carry out the cleaning and maintenance described in this section. Refer to your mask manual for detailed instructions.

D AILY C LEANING

Mask Clean the mask according to the instructions supplied with the mask.

Air tubing Disconnect the air tubing from the S8 ESCAPE (and humidifier, if used) and hang the tubing and mask in a clean, dry place until next use.

Humidifier If you are using a humidifier, clean it according to the instructions in the humidifier user’s manual.

W EEKLY C LEANING

1. Remove the air tubing from the S8 E SCAPE and the mask. 2. Wash the air tubing in warm water using mild detergent. Rinse thoroughly, hang, and allow to dry. 3. Before next use, assemble the mask and headgear according to the user instructions. 4. Reconnect the air tubing to the air outlet and mask.

P ERIODIC C LEANING

1. Clean the exterior of the S8 E SCAPE with a damp cloth and mild detergent. 2. Check the air filter for blockage or holes (see “Replacing the Air Filter” on page 23).

! WARNING

  • Beware of electric shock. Do not immerse the unit or power cord in water. Always unplug the unit before cleaning and be sure it is dry before plugging back in.
  • The mask system and air tubing are subject to normal wear and tear. Inspect them regularly for damage.
Page 25 of the ResMed s8 user manual — C LEANING AND M AINTENANCE
Page 25 · cliquer pour agrandir
Page 25

R EPLACING THE A IR F ILTER

To replace the air filter:

1. Remove the air filter cover at the back of the S8 E SCAPE .
2. Remove and discard the old air filter. 3. Insert a new filter with the blue tinted side facing out from the device. 4. Replace the air filter cover.

Inspect the air filter every month to check if it is blocked by dirt or contains holes. With normal use of an S8 ESCAPE, the air filter needs to be replaced every six months (or more often if your device is in a dusty environment).

! WARNING

  • Do not wash the air filter. The air filter is not washable or reusable.
  • The air filter cover protects the device in the event of accidental liquid spillage onto the device. Ensure the air filter and air filter cover are fitted at all times.

S ERVICING

! CAUTION

  • Do not attempt to open the S8 Escape case. There are no user serviceable parts inside.
  • Inspection and repair should only be performed by an authorized agent. Under no circumstances should you attempt to service or repair the flow generator yourself.

This product (S8 ESCAPE) should be inspected by an authorized ResMed Service Center five years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an Authorized ResMed Service Center. If you feel that your flow generator is not performing properly, see “Troubleshooting” on page 26.

C LEANING AND M AINTENANCE

Air filter cover

Page 26 of the ResMed s8 user manual — C AN I TRAVEL WITH THE S8 E SCAPE ?
Page 26 · cliquer pour agrandir
Page 26

F REQUENTLY A SKED Q UESTIONS

HOW WILL IT FEEL WHEN I AM BREATHING WITH THE S8 ESCAPE? When you first use your S8 ESCAPE, you may feel uncomfortable breathing against the airflow. This is normal. You will continue to breathe normally while you sleep. Take time to adjust to this new sensation.

WHAT DO I DO WHEN I WANT TO GET UP IN THE NIGHT? If you need to get up during the night, remove your mask and stop therapy. Always remember to restart therapy and put your mask on when you return to bed.

WHAT DO I DO IF I FEEL AIR LEAKING OUT OF THE MASK? The flow generator delivers the most effective treatment when the mask is well fitted and comfortable. Treatment can be affected by leaks, so it is important to eliminate any leaks that may arise. For example, try repositioning the mask and adjusting the straps. If you are using a nasal mask, try to keep your mouth closed during treatment. Air leaks from your mouth can decrease the effectiveness of your treatment. If mouth leaks are a problem, a full face mask or chin strap may help. Contact your clinician or equipment supplier for further details. If you are using a full face mask and you have problems trying to get a comfortable mask fit, contact your sleep clinic or equipment supplier. You may benefit from a different size or style of mask.

ANYTHING SPECIAL I SHOULD DO BEFORE PUTTING THE MASK ON? Before wearing your mask, wash your face to remove excess facial oils. This will allow a better fit and prolong the life of the mask cushion. Do not use moisturizer when wearing a mask.

MY MOUTH FEELS DRY DURING TREATMENT—WHAT CAN I DO? You may experience dryness of the nose, mouth, and/or throat during the course of treatment, especially during winter. In many cases, a humidifier may resolve this discomfort. Contact your clinician for advice.

MY NOSE IS IRRITATED DURING TREATMENT—WHAT CAN I DO? You may experience sneezing and/or a runny or blocked nose during the first few weeks of treatment. In many cases, nasal irritation can be resolved with a humidifier. Consult your clinician or equipment supplier for advice.

C AN I TRAVEL WITH THE S8 E SCAPE ?

International use Your S8 ESCAPE flow generator has an internal power supply that enables it to operate in other countries. It will operate on power supplies of 100–240 V and 50/60 Hz. No special adjustment is necessary, but you may need a plug adapter for the power outlet.

Page 27 of the ResMed s8 user manual — F REQUENTLY A SKED Q UESTIONS
Page 27 · cliquer pour agrandir
Page 27

Altitude changes If you are moving or traveling to a higher or lower altitude, see “How to change the Altitude setting” on page 17.

On an aircraft Please consult the medical services department of your carrier if you intend to use your S8 ESCAPE on an aircraft. The S8 ESCAPE complies with US FCC Part 15, Class B requirements.

CAN I USE MY S8 ESCAPE IF MAINS (AC) POWER IS NOT AVAILABLE? A DC-12 converter is available to connect the S8 ESCAPE to a 12V or 24V DC power source. Contact your clinician for details.

! CAUTION The device should not be connected to both AC and DC power sources simultaneously.

F REQUENTLY A SKED Q UESTIONS

Page 28 of the ResMed s8 user manual — T ROUBLESHOOTING
Page 28 · cliquer pour agrandir
Page 28

T ROUBLESHOOTING

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem Possible Cause Solution

No display. Power not connected. Ensure the appropriate power cord is connected to your S8 Escape and the power switch (if available) is on.

Insufficient air delivered from S8 Escape.

Ramp Time is in use. Wait for air pressure to build up or change ramp time.

Air filter is dirty. Replace air filter.

Air tubing is kinked or punctured.

Straighten or replace tubing.

Air tubing not connected properly.

Check air tubing.

Mask and headgear not positioned correctly.

Adjust position of mask and headgear.

Cap(s) missing from access port(s) on mask.

Replace cap(s).

Altitude compensation setting is incorrect.

Adjust altitude compensation setting to the correct altitude range.

Pressure required for treatment may have changed.

See your clinician to adjust the pressure.

Excessive air delivered from the S8 Escape.

Altitude compensation setting is incorrect.

Adjust altitude compensation setting to the correct altitude range.

Page 29 of the ResMed s8 user manual — T ROUBLESHOOTING
Page 29 · cliquer pour agrandir
Page 29
ResMed s8 — T ROUBLESHOOTING diagram (page 29)
Fig. 3

Display error message: Exxxx Call Service (where xxxx defines an error) or MOTOR FAULT Call Service or UPLOAD LANGUAGES Call Service or DATA LOST Call Service

The following message is displayed on the LCD after you try to update settings using the Data Card: Card Error Remove Card

The Data Card is not inserted correctly.

Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device. Ensure that the device is switched on and in standby mode before you insert the Data Card.

The Data Card is only partly inserted.

Ensure that the Data Card is inserted into the slot on the Data Card module as far as it will go.

You may have removed the Data Card before settings were copied to the S8 Escape.

Reinsert the Data Card and wait for the Settings Success Remove Card message to appear on the LCD.

The card is not a ResScan Data Card.

Only use a ResScan Data Card.

The following message is displayed on the LCD after you try to copy data to a Data Card: Card Error Remove Card

The Data Card is not inserted correctly.

Ensure that the Data Card is inserted with the arrow facing up and pointing toward the device.

The card is not a ResScan Data Card.

Use a ResScan Data Card, as specified in this manual.

The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Invalid Remove Card

The identification details on the Data Card do not match the details on your device.

Contact your clinician immediately.

T ROUBLESHOOTING

Page 30 of the ResMed s8 user manual — T ROUBLESHOOTING
Page 30 · cliquer pour agrandir
Page 30

Problem Possible Cause Solution

The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Error Remove Card

There is a data error on the Data Card.

Contact your clinician immediately.

The following message is NOT displayed on the LCD after you try to update the settings using the Data Card: Settings Success Remove Card

The settings were not updated. Contact your clinician immediately.

Page 31 of the ResMed s8 user manual — S YSTEM S PECIFICATIONS
Page 31 · cliquer pour agrandir
Page 31

S YSTEM S PECIFICATIONS

S8 E SCAPE

Performance: Operating pressure range: 4 to 20 cm H2O Dimensions (H x W x D): 4.6” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm) Weight: 2.9lb (1.3 kg)

Power Supply • Input range for S8 ESCAPE with HUMIDAIRE 3i:100–240V, 50/60Hz; 110V, 400Hz; 2.5A < 140VA (110W) (maximum power consumption). Instantaneous peak power consumption

  • Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed DC-12 converter for DC input. Actual power consumption will vary, depending on factors such as the use of accessories, height above sea level, and ambient temperature. Power consumption values for typical treatment conditions (using a breathing machine set to 0.5 L and 15 breaths/min with 2 m (6’6”) air tubing and an U LTRA M IRAGE mask) are given in the following table.

Device Pressure (cm H2O)

S8 E SCAPE 5 17 8

Power outlet for HumidAire 3i Outlet range: 100–240V, 50/60Hz; 110V, 400Hz; 2.0A Air Filter: Two-layered, powder-bonded, polyester non-woven fiber Air Tubing: Flexible plastic, 6’6” (2 m) Air Outlet: The 22 mm conical outlet complies with EN 1281-1 IEC 60601-1 Classifications: Class II (double insulation), Type CF Noise Level: Radiated sound pressure is measured at

A CCESSORIES

D ATA C ARD

Dimensions (L x W x D): 2.56” x 0.87” x 0.03” (65.04 mm x 22.15 mm x 0.85 mm)

S YSTEM S PECIFICATIONS

Show 5 more labels / short text on this page
Power (VA) · Power (W) · 10 23 12 · 15 29 16 · 20 34 20
Page 32 of the ResMed s8 user manual — S YMBOLS WHICH APPEAR ON THE DEVICE
Page 32 · cliquer pour agrandir
Page 32

D ATA C ARD M ODULE

Dimensions (H x W x D): 1.95” x 1.66” x 0.47” (49.7 mm x 42.2 mm x 12 mm)

A LL D EVICES ( WHERE APPLICABLE )

Housing ConstructionFlame retardant engineering thermoplastic

Environmental Conditions Operating Temperature: +41°F to +104°F (+5°C to +40°C) Operating Humidity: 10%–95% non-condensing Storage and Transport Temperature: -4°F to +140°F (-20°C to +60°C) Storage and Transport Humidity: 10%–95% non-condensing Operating Altitude: sea level to 8500’ (2591 m) Electromagnetic Compatibility Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments. For further details, see “Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity” on page 31.

S YMBOLS WHICH APPEAR ON THE DEVICE

Show 4 more labels / short text on this page
Type CF equipment · Class II equipment · Drip proof · Start/Stop
Page 33 of the ResMed s8 user manual — S YSTEM S PECIFICATIONS
Page 33 · cliquer pour agrandir
Page 33

GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS AND IMMUNITY

Guidance and manufacturer’s declaration—electromagnetic emissions

The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.

Emissions test Compliance Electromagnetic environment— guidance

RF emissions CISPR11 Group 1 The S8 device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11 Class B* The S8 device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage network that supplies buildings used for domestic purposes.

Harmonic Emissions IEC 61000-3-2

Voltage Fluctuations/Flicker Emissions IEC 61000-3-3

* Class B for all system configurations except when a PC is attached to the S8 device via a serial adapter, in which case the system is Class A compliant. Class A is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document.

The use of accessories (eg humidifiers) other than those specified in this manual is not recommended. They may result in increased emissions or decreased immunity of the S8 device.

S YSTEM S PECIFICATIONS

Class A · Complies

Page 34 of the ResMed s8 user manual — D ATA C ARD M ODULE
Page 34 · cliquer pour agrandir
Page 34

Guidance and manufacturer’s declaration—electromagnetic immunity

The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.

Immunity test IEC60601-1-2 test level Compliance level Electromagnetic environment—guidance

Electrostatic discharge (ESD) IEC 61000-4-2

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst IEC 61000-4-4

±2 kV for power supply lines

Mains power quality should be that of a typical commercial or hospital environment.

±1 kV for input/ output lines

±1 kV differential mode ±2 kV common mode

±1 kV differential mode

Mains power quality should be that of a typical commercial or hospital environment.

±2 kV common mode

Voltage dips, short interruptions and voltage variations on power supply input lines. IEC 61000-4-11

95% dip in Ut) for 0.5 cycle

< 12V (>95% dip in 240V) for 0.5 cycle

Mains power quality should be that of a typical commercial or hospital environment. If the user of the S8 device requires continued operation during power mains interruptions, it is recommended that the S8 device be powered from an uninterruptible power source.

40% Ut (60% dip in Ut) for 5 cycles

96V (60% dip in 240V) for 5 cycles

70% Ut (30% dip in Ut) for 25 cycles

168V (30% dip in 240V) for 25 cycles

95%dip in Ut) for 5 sec

95%dip in 240V) for 5 sec

Power frequency (50/60 Hz) magnetic field IEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Show 7 more labels / short text on this page
±6 kV contact · ±6 kV contact · ±8 kV air · ±8 kV air · ±2 kV · Not Applicable · Surge IEC 61000-4-5
Page 35 of the ResMed s8 user manual — S YSTEM S PECIFICATIONS
Page 35 · cliquer pour agrandir
Page 35

Guidance and manufacturer’s declaration—electromagnetic immunity

The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.

Immunity test IEC60601-1-2 test level

Compliance level Electromagnetic environment—guidance

Portable and mobile RF communications equipment should be used no closer to any part of the S8 device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Conducted RF IEC 61000-4-6

3 Vrms 150 kHz to 80 MHz 3 Vrms d = 1.17 √P

Radiated RF IEC 61000-4-3

10 V/m 80 MHz to 2.5 GHz 10 V/m d = 0.35 √P 80 MHz to 800 MHz

d = 0.70 √P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the S8 device is used exceeds the applicable RF compliance level above, the S8 device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the S8 device.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

S YSTEM S PECIFICATIONS

Recommended separation distance

Page 36 of the ResMed s8 user manual — D ATA C ARD M ODULE
Page 36 · cliquer pour agrandir
Page 36

Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices

The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the S8 device as recommended below, according to the maximum output power of the communications equipment.

ResMed s8 — D ATA C ARD M ODULE diagram (page 36)
Fig. 4

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Page 37 of the ResMed s8 user manual — R001-307/1 05 02
Page 37 · cliquer pour agrandir
Page 37

L IMITED W ARRANTY

ResMed s8 — L IMITED W ARRANTY diagram (page 37)
Fig. 5

ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable.

Accessories, mask systems (including mask frame, cushion, headgear and tubing). Excludes single-use devices.

If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; c) any damage or contamination due to cigarette, pipe, cigar or other smoke; d) any damage caused by water being spilled on or into a flow generator. To make a warranty claim, the initial consumer must return the defective product to the point of purchase, freight prepaid.

This warranty is in lieu of all other express or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.

ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office.

R001-307/1 05 02

L IMITED W ARRANTY

Page 38 of the ResMed s8 user manual — L IMITED W ARRANTY
Page 38 · cliquer pour agrandir
Page 38
Page 38 of the ResMed s8 user manual — L IMITED W ARRANTY
Page 39 of the ResMed s8 user manual — L IMITED W ARRANTY
Page 39 · cliquer pour agrandir
Page 39
Page 39 of the ResMed s8 user manual — L IMITED W ARRANTY
Page 40 of the ResMed s8 user manual — L IMITED W ARRANTY
Page 40 · cliquer pour agrandir
Page 40
Page 40 of the ResMed s8 user manual — L IMITED W ARRANTY
Page 41 of the ResMed s8 user manual — F RANÇAIS
Page 41 · cliquer pour agrandir
Page 41

S8 Escape ™

F RANÇAIS

MANUEL UTILISATEUR

Français

Page 42 of the ResMed s8 user manual — S8 Escape ™
Page 42 · cliquer pour agrandir
Page 42
Page 42 of the ResMed s8 user manual — S8 Escape ™
Page 44 of the ResMed s8 user manual — S8 Escape ™
Page 44 · cliquer pour agrandir
Page 44

CARACTÉRISTIQUES 68

S8 ESCAPE 68 ACCESSOIRES 69 TOUS LES APPAREILS (LE CAS ÉCHÉANT) 69 GARANTIE LIMITÉE 75 INDEX 77

Sources

Références et liens

Pages mentionnées dans ce manuel — fabricant, assistance et téléchargements.

Poser une question sur : ResMed s8

Les utilisateurs et notre équipe répondent généralement sous 24 h — et votre question servira aussi aux suivants.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Aucune question pour l'instant sur le ResMed s8

Soyez le premier à poser une question — d'autres utilisateurs ou notre équipe y répondront.

Document d'origine

ResMed s8 manuel PDF

158 pages·3.7 MB·Humidificateur·PDF
Cliquez sur une page ci-dessus pour l'ouvrir ici
Page 1 / …
100%
Chargement des pages…