U SER ’ S M ANUAL 1
M ANUEL UTILISATEUR 39
M ANUAL DEL USUARIO 79
M ANUAL DO UTILIZADOR 117

Manufactured by:
ResMed Ltd 1 Elizabeth MacArthur Drive Bella Vista NSW 2153 Australia Tel: +61 (2) 8884 1000 or 1 800 658 189 (toll free) Fax: +61 (2) 8883 3114 Email: reception@resmed.com.au
ResMed Offices:
ResMed Corp 14040 Danielson Street Poway CA 92064-6857 USA Tel: +1 (858) 746-2400 or 1-800-424-0737 (toll free) Fax: +1 (858) 746-2900 Email: reception@resmed.com
ResMed (UK) Ltd 65 Milton Park Abingdon Oxfordshire OX14 4RX UK Tel: +44 (1235) 862 997 Fax: +44 (1235) 831 336 Email: reception@resmed.co.uk
ResMed Asia Pacific Ltd 97 Waterloo Road North Ryde NSW 2113 Australia Tel: +61 (2) 9886 5000 or 1 800 991 900 (toll free) Fax: +61 (2) 9889 1471 Email: reception@resmed.com.au
ResMed Austria Purtscherstraße 18, 9500 Villach Österreich Tel: +43 (4242) 430 830 Fax: +43 (4242) 430 830-4 Email: [email protected]
ResMed BV Nieuwe Parklaan 86, 2587 BV Den Haag Nederland Tél: +31 (70) 358 6263 Fax: +31 (70) 358 4333 Email: [email protected]
ResMed Finland Niittykatu 6 FIN 02200 Espoo Suomi Tel: +358 9 8676820 Fax: +358 9 86768222 Email: [email protected]
ResMed GmbH & Co. KG Fraunhoferstraße 16 82152 Martinsried Deutschland Tel: +49 (0) 89 / 99 01 - 00 or +49 (0) 800 2 777 000 (gebührenfrei) Fax: +49 (0) 89 / 99 01 - 10 55 Email: [email protected]
ResMed Hong Kong Ltd Room 1714 Miramar Tower 132-134 Nathan Road Tsim Sha Tsui Hong Kong Tel: +852 2366 0707 Fax: +852 2366 4546 Email: reception@resmed.com.hk
ResMed Japan Nihonbashi Hisamatsu Bldg. 4F 2-28-1 Nihonbashi-Hamacho Chuo-Ku, Tokyo 103-0007 Japan
ResMed Malaysia Sdn Bhd Suite E-10-20, Plaza Mon’t Kiara No. 2, Jalan 1/70C Mon’t Kiara 50480 Kuala Lumpur Malaysia Tel: +60 3 6201 7177 Fax: +60 3 6201 2177 Email: reception@resmed.com.my
ResMed NZ Ltd PO Box 51-048 Pakuranga Auckland New Zealand Tel: +0800 737 633 (NZ toll free) or +64 25 737 633 Fax: +0800 737 634 (NZ toll free) or +64 9 239 0193 Email: reception@resmed.co.nz
ResMed SA Parc de la Bandonnière 2 rue Maurice Audibert 69800 Saint-Priest France Tel: +33 (0) 4 37 251 251 Fax: +33 (0) 4 37 251 260 Email: [email protected]
ResMed Singapore Pte Ltd 238A Thomson Road # 12-03/04 Novena Square Tower A Singapore 307684 Tel: +65 6284 7177 Fax: +65 6284 7787 Email: reception@resmed.com.sg
ResMed Spain SL C/ Arturo Soria, 245 28033 Madrid España Tel: +34 (93) 590 8154 Fax: +34 (93) 590 8153 Email: [email protected]
ResMed Sweden AB Industrigatan 2 S-461 37 Trollhättan Sverige Tel: +46 520 420 110 Fax: +46 520 397 15 Email: [email protected]
Labhardt AG Thannerstrasse 57 CH-4054 Basel Schweiz Tel: +41 (061) 307 9711 Fax: +41 (061) 307 9722 Email: [email protected]
Protected by patents: US 4944310, US 5199424, US 5522382, US 6705315. Other patents pending. Protected by design registrations: AU 302180, AU 302181, AU 302182, AU 302183, AU 302184, AU 302185, AU 302186, EP 269436, HK 0412901.9. Others pending.
S8, S8 Escape, HumidAire, HumidAire 3i, Mirage, Mirage Activa, Mirage Vista, Protégé, Ultra Mirage, Mirage Swift, and ResScan are trademarks of ResMed Ltd. Papillon is a trademark of MAP Medizin-Technologie GmbH.
© ResMed Ltd 2005.

English



INTRODUCTION 5
YOUR RESPONSIBILITY 5 MEDICAL INFORMATION 5
ABOUT THE S8 ESCAPE™ 5 CONTRAINDICATIONS 6 WARNINGS 6 CAUTIONS 7 ADVERSE EFFECTS 8 THE S8 ESCAPE SYSTEM 9
S8 ESCAPE 9 MASKS 9 HUMIDIFIERS 10 RESSCAN DATA CARD MODULE 10 ACCESSORIES 10 HOW TO USE THE S8 ESCAPE 11
SET UP THE S8 ESCAPE 11 ASSEMBLE THE MASK 11 START TREATMENT 11 STOP TREATMENT 12 HOW TO ATTACH A HUMIDIFIER 13
ATTACH A HUMIDAIRE 3I HUMIDIFIER 13 ATTACH A HUMIDAIRE HUMIDIFIER 13 ATTACH A RESMED PASSOVER HUMIDIFIER 14 HOW TO USE THE KEYPAD 15
HOW TO CHANGE THE SETTINGS ON YOUR S8 ESCAPE 16 S8 MODULES 19
USING A BLANK MODULE 19 USING A RESSCAN DATA CARD MODULE 19 CLEANING AND MAINTENANCE 22
DAILY CLEANING 22 WEEKLY CLEANING 22 PERIODIC CLEANING 22 REPLACING THE AIR FILTER 23 SERVICING 23 FREQUENTLY ASKED QUESTIONS 24 TROUBLESHOOTING 26

SYSTEM SPECIFICATIONS 29
S8 ESCAPE 29 ACCESSORIES 29 LIMITED WARRANTY 35 INDEX 37

Thank you for choosing the S8 ESCAPE™ flow generator. The S8 ESCAPE is a compact, all-in-one package with an inbuilt power supply and the option of integrated data management and humidification. This user manual contains the information you need for the correct use of your S8 ESCAPE.
The owner or user of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from:
Please read this manual carefully before use. This manual contains special terms and icons that appear in the margins to draw your attention to specific and important information.
The S8 ESCAPE continuous positive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnea (OSA) in adult patients. The S8 ESCAPE CPAP system is intended for home and hospital use.
! CAUTION In the US, Federal law restricts this device to sale by or on the order of a physician.

The S8 ESCAPE is not a life support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy. Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions:
* Ports may be incorporated into the mask or in connectors that are near the mask.


Patients should report unusual chest pain, severe headache, or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side effects may arise during the course of therapy with the CPAP device:

The S8 ESCAPE device is part of a system that consists of five elements:
Please refer to the illustrations in section A of the illustration sheet.
FRONT VIEW (A-1) Shows the keypad and LCD screen, and the air outlet.
REAR VIEW (A-2) Shows where the module attaches, and the location of the air filter and power sockets.
POWER CORD (A-3) Power cord to attach the S8 ESCAPE to the AC (mains) power.
CARRY BAG (A-4) The S8 ESCAPE carry bag contains two pouches: one for storing the S8 device, the other for storing the HUMIDAIRE 3i™. When storing the HUMIDAIRE 3i, always put it in the larger, mesh pouch.
AIR TUBING The 6’6” (2 m) air tubing (A-5) connects the S8 ESCAPE to your mask . You will need medium 21” (52 cm) air tubing (see C-3) to connect the S8 ESCAPE to the HUMIDAIRE™ and ResMed PASSOVER humidifiers.
You will also need a ResMed mask system (supplied separately). The ResMed mask systems that are recommended for use with the S8 ESCAPE are:
NASAL MASKS • MODULAR NASAL MASK
NASAL PILLOWS SYSTEMS • MIRAGE SWIFT™ NASAL PILLOWS SYSTEM.

FULL FACE MASKS • MIRAGE™ FULL FACE MASK
! WARNING Only ResMed mask systems are compatible for use with the S8 Escape. Please refer to “Warnings” on page 6.
A humidifier may be required if you are experiencing dryness of the nose, throat, or mouth. The S8 ESCAPE is compatible for use with:
The RESSCAN™ Data Card module and RESSCAN Data Card (C-1) may be used with the S8 ESCAPE either to help your clinician to monitor your treatment or to provide you with updates to your device settings. See “Using a ResScan Data Card Module” on page 19 for information about how to use the Data Card module and Data Card.
Please refer to the illustrations in section C of the illustration sheet. Other accessories for use with the S8 ESCAPE:

Please refer to the illustrations in section D of the illustration sheet.
1 Place the S8 ESCAPE on a table near the head of your bed.
! CAUTION Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord.
2 Connect the power cord to the socket at the rear of the flow generator (D-1). Plug the other end of the power cord into the power outlet.
! WARNING Make sure the power cord and plug are in good condition and the equipment is not damaged.
3 Connect one end of the air tubing firmly onto the air outlet (D-2).
1 Assemble your mask system according to the mask user instructions.
2 Connect your mask system to the free end of the air tubing (D-3). The S8 ESCAPE is now ready for use (D-4).
1 Make sure the power is on. The product name is displayed briefly on the LCD screen, then the standby (Ramp) screen appears. The key and LCD backlights also turn on.
HOW TO USE THE S8 ESCAPE

2 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.
3 Alter the ramp time if required (see “How to Change the Ramp Time” on page 17).
4 To start treatment, press the Start/Stop key.
5 Fit your mask as described in the mask user instructions.
! WARNING A mask should not be used unless the S8 Escape device is turned on and operating properly.
6 When ramping begins the LCD will display the word "RAMPING", with the current pressure indicated as dashes. Each dash is approximately equivalent to 2 cm H2O. For example, 6 dashes represents approximately 12 cm H2O pressure.
To stop treatment at any time, remove your mask and press the Start/Stop key.

Please refer to the illustration in section E of the illustration sheet, and your HumidAire 3i manual. The HUMIDAIRE 3i attaches to the front of the S8 ESCAPE to provide heated humidification. No other accessories are required for its use.
detach the humidifier. b. Remove the connector plug on the S8 Escape (see section A of the illustration
sheet) before you attach the HumidAire 3i. Reattach the connector plug firmly after you detach the HumidAire 3i.
STORING THE HUMIDAIRE 3i Store the HUMIDAIRE 3i in the larger, mesh-top pouch of the S8 ESCAPE carry bag when not in use.
Please refer to the illustrations in section I of the illustration sheet. To connect your S8 ESCAPE to a HUMIDAIRE humidifier you need medium sized 21” (52 cm) air tubing.
1 Fill the HUMIDAIRE water chamber and place inside the HUMIDAIRE. Connect the medium 21” (52 cm) and long 6’6” (2 m) air tubing (I-1), and close the humidifier lid.
2 Place the S8 ESCAPE on top of the HUMIDAIRE (I-2) and connect the free end of the medium air tubing to the air outlet (I-3).
3 Assemble your mask system and connect it to the free end of the long air tubing (I-4).
4 Plug the HUMIDAIRE power cord into a power outlet. Connect the S8 ESCAPE power cord to the socket at the rear of the S8 ESCAPE and plug the other end into a power outlet. Please refer to the HumidAire User’s Manual for further details.
! WARNING Make sure that the power cord and plug are in good condition and the equipment is not damaged. The S8 ESCAPE is now ready for use with the HUMIDAIRE.

Please refer to the illustrations in section J of the illustration sheet. To connect your S8 ESCAPE to a ResMed PASSOVER humidifier you need medium sized 21” (52 cm) air tubing.
1 Fill the PASSOVER with water and connect the medium 21” (52 cm) and long 6’6” (2 m) air tubing (J-1).
2 Place the S8 ESCAPE on top of the PASSOVER (J-2) and connect the free end of the medium air tubing to the air outlet (J-3).
3 Assemble your mask system and connect it to the free end of the long air tubing (J-4).
4 Connect the power cord to the socket at the rear of the S8 ESCAPE and plug the other end into a power outlet. Please refer to the ResMed Passover User’s Manual for further details.
! WARNING Make sure that the power cord and plug are in good condition and the equipment is not damaged. The S8 ESCAPE is now ready for use with the PASSOVER.

The control panel of the S8 ESCAPE includes an LCD screen and keypad for navigating through the menus and delivering treatment.

The LCD screen displays the menus and treatment screens.
The S8 ESCAPE keypad has the following keys:
Start/Stop • Starts or stops treatment.
Up • Allows you to increase settings options and scroll through the S8 ESCAPE menu.
Down • Allows you to decrease settings options and scroll through the S8 ESCAPE menu.
Left • Performs the function indicated by the guiding text displayed above it on the LCD screen. Guiding text includes menu, change, and apply.
Right • Performs the function indicated by the guiding text displayed above it on the LCD screen. Guiding text includes exit and cancel.
The keypad and LCD are equipped with backlights that come on when the device is powered. The LCD backlight turns off after two minutes of inactivity, and comes back

on when you press a key. Once the Start/Stop key has been pressed to start treatment, the LCD backlight stays on for two minutes. The keypad backlight is always on while the S8 ESCAPE is on.
HOW TO CHANGE THE SETTINGS ON YOUR S8 ESCAPE
You can change settings (such as Ramp Time, Altitude) on the S8 ESCAPE by using the keypad and LCD. When the S8 ESCAPE is in standby mode you can display a series of screens (a menu) on the LCD screen in order to view and change the settings for a particular function.

Use the menu screens to change the ramp time, altitude compensation, and language. You can also view details about the usage and software version of your S8 ESCAPE.
HOW TO ENTER THE MENU SCREENS When the S8 ESCAPE standby screen is displayed you will see the word ‘Menu’ in the left-hand corner. To select the S8 ESCAPE menu, press the Left key.
HOW TO SCROLL THROUGH THE MENU When you are in the menu, scroll through items by pressing the Up or Down keys.
HOW TO CHANGE A SETTING OPTION When you have scrolled to the item you require:
If you wish to exit without changing the option, press the Right key (Cancel).

HOW TO EXIT OUT OF A MENU Press the Right key (Exit). This will take you back to the Patient Menu screen.
HOW TO EXIT OUT OF AN ITEM WITHIN THE MENU Press the Right key (Exit).
During ramp time, the pressure increases from a low pressure to the prescribed treatment pressure. You can set the ramp time in 5-minute intervals, from 0 minutes (OFF) up to a maximum ramp time set by your clinician. The standby screen displays the current ramp time in the right-hand corner. To change the ramp time, press the Up or Down key until the desired ramp time appears. Once a ramp time is selected, begin therapy by pressing the Start/Stop key. When ramping begins the LCD will display the word "RAMPING", with the current pressure indicated as dashes. Each dash is approximately equivalent to 2 cm H2O. For example, six dashes represents approximately 12 cm H2O pressure.
If you move or travel with your S8 ESCAPE to areas of different altitude, you will need to adjust the altitude compensation setting. Changes in altitude will affect the pressure delivered by the S8 ESCAPE. You can use the Altitude setting (ALT) to specify the altitude range at which you are using your S8 ESCAPE. Select the S8 ESCAPE menu by pressing the Left key in the standby (RAMP) screen. Scroll to ALT and select Change (Left key). Press the Up or Down keys until you see the setting option that you require. There are four altitude settings to choose from:
To display menus and messages in a different language, select the S8 ESCAPE menu by pressing the Left key in the standby (Ramp) screen. Scroll to LANG and select Change (Left key). Press the Up or Down keys until you see the language that you require.
You can also view the total number of hours of treatment (USED HRS); the number of sessions that the S8 ESCAPE was used in a usage period (USAGE); and the current software version installed on your unit (SW). Access to all of these screens is provided via the menu. See “How to enter the menu screens” on page 16.

Your clinician may have set your S8 ESCAPE to remind you about important events, such as when to replace your mask, when to insert your Data Card (if your device is Data Card enabled) and so on. The reminder message is displayed on the LCD and is visible if the device is not delivering therapy. The backlight on the LCD flashes when a message is displayed. Your clinician can set any of the following reminders on your LCD:
INSERT CARD May appear if your device is Data Card enabled.
Insert your Data Card and follow any instructions that your clinician has given you. When you have done this, the message will disappear from the LCD. Pressing the Left (Ok) key will also remove the message.
May appear when your mask is due for replacement.
Press the Left (Ok) key to remove the message from your LCD, and replace your mask with a new one.
May appear as a reminder to contact your clinician; for example to discuss how your therapy is going.
Press the Left (Ok) key to remove the message from your LCD, and contact your clinician.
Press the Left (Ok) key to remove the message from your LCD, and replace the air filter.
May appear as a reminder to replace the air filter on your device.
SERVICE DUE May appear as a reminder to return your device for service.
Press the Left (Ok) key to remove the message from your LCD, and contact your clinician.
Message Description Action

Please refer to the illustrations in section F of the illustration sheet. The blank module is a protective cover on the S8 ESCAPE. At times, you may wish to remove or attach this module, for example to use the serial adapter. To remove the blank module, pull it off the back of the S8 ESCAPE to show the data communications port (F-1). To attach the blank module, push the blank module onto the data communications port, until the module clicks into place (F-2).
USING A RESSCAN DATA CARD MODULE
Please refer to the illustrations in section G of the illustration sheet. The RESSCAN Data Card module is an accessory to the S8 ESCAPE. Use a RESSCAN Data Card with the Data Card module to collect data and update settings in the S8 ESCAPE. The module and card give your clinician convenient access to data about your treatment, and can provide you with new device settings from your clinician without you having to leave home.
To attach the Data Card module, push it onto the data communications port at the back of your S8 ESCAPE until it clicks into place (G-1).
To remove the Data Card module, pull it off the back of the S8 ESCAPE (G-2).
Please refer to the illustrations in section H of the illustration sheet. If your clinician needs to review your treatment, they will ask you to use the Data Card to copy data from your S8 ESCAPE, and to return the card to them. The Data Card will be provided in a convenient mailback envelope. Data that is copied to a Data Card is still stored and available on the S8 ESCAPE. It takes approximately five seconds to copy data onto the Data Card.

1 INSERT THE DATA CARD • Switch on the S8 ESCAPE and wait until you see the standby (Ramp) screen.
2 COPY DATA ONTO THE DATA CARD • Data copying starts automatically when the Data Card is inserted into the Data Card module.
3 REMOVE THE DATA CARD Grip the end of the Data Card and pull it out of the Data Card module (H-3).
4 STORE THE DATA CARD Your Data Card should be stored in its mailback envelope when not in use.
5 MAIL THE DATA CARD TO YOUR CLINICIAN A bubble mailer envelope is provided to ensure safe transportation of the Data Card. When mailing the Data Card to your clinician:
If your clinician has provided a Data Card with new device settings:

! WARNING If your clinician has told you to use this Data Card to update the settings on your device and the “Settings Success” message does not appear, contact your clinician immediately.

You should regularly carry out the cleaning and maintenance described in this section. Refer to your mask manual for detailed instructions.
Mask Clean the mask according to the instructions supplied with the mask.
Air tubing Disconnect the air tubing from the S8 ESCAPE (and humidifier, if used) and hang the tubing and mask in a clean, dry place until next use.
Humidifier If you are using a humidifier, clean it according to the instructions in the humidifier user’s manual.

To replace the air filter:
Inspect the air filter every month to check if it is blocked by dirt or contains holes. With normal use of an S8 ESCAPE, the air filter needs to be replaced every six months (or more often if your device is in a dusty environment).
! CAUTION
This product (S8 ESCAPE) should be inspected by an authorized ResMed Service Center five years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an Authorized ResMed Service Center. If you feel that your flow generator is not performing properly, see “Troubleshooting” on page 26.
Air filter cover

HOW WILL IT FEEL WHEN I AM BREATHING WITH THE S8 ESCAPE? When you first use your S8 ESCAPE, you may feel uncomfortable breathing against the airflow. This is normal. You will continue to breathe normally while you sleep. Take time to adjust to this new sensation.
WHAT DO I DO WHEN I WANT TO GET UP IN THE NIGHT? If you need to get up during the night, remove your mask and stop therapy. Always remember to restart therapy and put your mask on when you return to bed.
WHAT DO I DO IF I FEEL AIR LEAKING OUT OF THE MASK? The flow generator delivers the most effective treatment when the mask is well fitted and comfortable. Treatment can be affected by leaks, so it is important to eliminate any leaks that may arise. For example, try repositioning the mask and adjusting the straps. If you are using a nasal mask, try to keep your mouth closed during treatment. Air leaks from your mouth can decrease the effectiveness of your treatment. If mouth leaks are a problem, a full face mask or chin strap may help. Contact your clinician or equipment supplier for further details. If you are using a full face mask and you have problems trying to get a comfortable mask fit, contact your sleep clinic or equipment supplier. You may benefit from a different size or style of mask.
ANYTHING SPECIAL I SHOULD DO BEFORE PUTTING THE MASK ON? Before wearing your mask, wash your face to remove excess facial oils. This will allow a better fit and prolong the life of the mask cushion. Do not use moisturizer when wearing a mask.
MY MOUTH FEELS DRY DURING TREATMENT—WHAT CAN I DO? You may experience dryness of the nose, mouth, and/or throat during the course of treatment, especially during winter. In many cases, a humidifier may resolve this discomfort. Contact your clinician for advice.
MY NOSE IS IRRITATED DURING TREATMENT—WHAT CAN I DO? You may experience sneezing and/or a runny or blocked nose during the first few weeks of treatment. In many cases, nasal irritation can be resolved with a humidifier. Consult your clinician or equipment supplier for advice.
International use Your S8 ESCAPE flow generator has an internal power supply that enables it to operate in other countries. It will operate on power supplies of 100–240 V and 50/60 Hz. No special adjustment is necessary, but you may need a plug adapter for the power outlet.

Altitude changes If you are moving or traveling to a higher or lower altitude, see “How to change the Altitude setting” on page 17.
On an aircraft Please consult the medical services department of your carrier if you intend to use your S8 ESCAPE on an aircraft. The S8 ESCAPE complies with US FCC Part 15, Class B requirements.
CAN I USE MY S8 ESCAPE IF MAINS (AC) POWER IS NOT AVAILABLE? A DC-12 converter is available to connect the S8 ESCAPE to a 12V or 24V DC power source. Contact your clinician for details.
! CAUTION The device should not be connected to both AC and DC power sources simultaneously.

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.
Problem Possible Cause Solution
No display. Power not connected. Ensure the appropriate power cord is connected to your S8 Escape and the power switch (if available) is on.
Insufficient air delivered from S8 Escape.
Ramp Time is in use. Wait for air pressure to build up or change ramp time.
Air filter is dirty. Replace air filter.
Air tubing is kinked or punctured.
Straighten or replace tubing.
Air tubing not connected properly.
Check air tubing.
Mask and headgear not positioned correctly.
Adjust position of mask and headgear.
Cap(s) missing from access port(s) on mask.
Replace cap(s).
Altitude compensation setting is incorrect.
Adjust altitude compensation setting to the correct altitude range.
Pressure required for treatment may have changed.
See your clinician to adjust the pressure.
Excessive air delivered from the S8 Escape.
Altitude compensation setting is incorrect.
Adjust altitude compensation setting to the correct altitude range.


Display error message: Exxxx Call Service (where xxxx defines an error) or MOTOR FAULT Call Service or UPLOAD LANGUAGES Call Service or DATA LOST Call Service
The following message is displayed on the LCD after you try to update settings using the Data Card: Card Error Remove Card
The Data Card is not inserted correctly.
Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device. Ensure that the device is switched on and in standby mode before you insert the Data Card.
The Data Card is only partly inserted.
Ensure that the Data Card is inserted into the slot on the Data Card module as far as it will go.
You may have removed the Data Card before settings were copied to the S8 Escape.
Reinsert the Data Card and wait for the Settings Success Remove Card message to appear on the LCD.
The card is not a ResScan Data Card.
Only use a ResScan Data Card.
The following message is displayed on the LCD after you try to copy data to a Data Card: Card Error Remove Card
The Data Card is not inserted correctly.
Ensure that the Data Card is inserted with the arrow facing up and pointing toward the device.
The card is not a ResScan Data Card.
Use a ResScan Data Card, as specified in this manual.
The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Invalid Remove Card
The identification details on the Data Card do not match the details on your device.
Contact your clinician immediately.

Problem Possible Cause Solution
The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Error Remove Card
There is a data error on the Data Card.
Contact your clinician immediately.
The following message is NOT displayed on the LCD after you try to update the settings using the Data Card: Settings Success Remove Card
The settings were not updated. Contact your clinician immediately.

Performance: Operating pressure range: 4 to 20 cm H2O Dimensions (H x W x D): 4.6” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm) Weight: 2.9lb (1.3 kg)
Power Supply • Input range for S8 ESCAPE with HUMIDAIRE 3i:100–240V, 50/60Hz; 110V, 400Hz; 2.5A < 140VA (110W) (maximum power consumption). Instantaneous peak power consumption
Device Pressure (cm H2O)
Power outlet for HumidAire 3i Outlet range: 100–240V, 50/60Hz; 110V, 400Hz; 2.0A Air Filter: Two-layered, powder-bonded, polyester non-woven fiber Air Tubing: Flexible plastic, 6’6” (2 m) Air Outlet: The 22 mm conical outlet complies with EN 1281-1 IEC 60601-1 Classifications: Class II (double insulation), Type CF Noise Level: Radiated sound pressure is measured at
D ATA C ARD
Dimensions (L x W x D): 2.56” x 0.87” x 0.03” (65.04 mm x 22.15 mm x 0.85 mm)

Dimensions (H x W x D): 1.95” x 1.66” x 0.47” (49.7 mm x 42.2 mm x 12 mm)
Environmental Conditions Operating Temperature: +41°F to +104°F (+5°C to +40°C) Operating Humidity: 10%–95% non-condensing Storage and Transport Temperature: -4°F to +140°F (-20°C to +60°C) Storage and Transport Humidity: 10%–95% non-condensing Operating Altitude: sea level to 8500’ (2591 m) Electromagnetic Compatibility Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments. For further details, see “Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity” on page 31.

GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS AND IMMUNITY
Guidance and manufacturer’s declaration—electromagnetic emissions
The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.
Emissions test Compliance Electromagnetic environment— guidance
RF emissions CISPR11 Group 1 The S8 device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B* The S8 device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage network that supplies buildings used for domestic purposes.
Harmonic Emissions IEC 61000-3-2
Voltage Fluctuations/Flicker Emissions IEC 61000-3-3
* Class B for all system configurations except when a PC is attached to the S8 device via a serial adapter, in which case the system is Class A compliant. Class A is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document.
The use of accessories (eg humidifiers) other than those specified in this manual is not recommended. They may result in increased emissions or decreased immunity of the S8 device.
Class A · Complies

Guidance and manufacturer’s declaration—electromagnetic immunity
The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.
Immunity test IEC60601-1-2 test level Compliance level Electromagnetic environment—guidance
Electrostatic discharge (ESD) IEC 61000-4-2
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4
±2 kV for power supply lines
Mains power quality should be that of a typical commercial or hospital environment.
±1 kV for input/ output lines
±1 kV differential mode ±2 kV common mode
±1 kV differential mode
Mains power quality should be that of a typical commercial or hospital environment.
±2 kV common mode
Voltage dips, short interruptions and voltage variations on power supply input lines. IEC 61000-4-11
95% dip in Ut) for 0.5 cycle
< 12V (>95% dip in 240V) for 0.5 cycle
Mains power quality should be that of a typical commercial or hospital environment. If the user of the S8 device requires continued operation during power mains interruptions, it is recommended that the S8 device be powered from an uninterruptible power source.
40% Ut (60% dip in Ut) for 5 cycles
96V (60% dip in 240V) for 5 cycles
70% Ut (30% dip in Ut) for 25 cycles
168V (30% dip in 240V) for 25 cycles
95%dip in Ut) for 5 sec
95%dip in 240V) for 5 sec
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Guidance and manufacturer’s declaration—electromagnetic immunity
The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.
Immunity test IEC60601-1-2 test level
Compliance level Electromagnetic environment—guidance
Portable and mobile RF communications equipment should be used no closer to any part of the S8 device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz 3 Vrms d = 1.17 √P
Radiated RF IEC 61000-4-3
10 V/m 80 MHz to 2.5 GHz 10 V/m d = 0.35 √P 80 MHz to 800 MHz
d = 0.70 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the S8 device is used exceeds the applicable RF compliance level above, the S8 device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the S8 device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Recommended separation distance

Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices
The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the S8 device as recommended below, according to the maximum output power of the communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.


ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable.
Accessories, mask systems (including mask frame, cushion, headgear and tubing). Excludes single-use devices.
If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; c) any damage or contamination due to cigarette, pipe, cigar or other smoke; d) any damage caused by water being spilled on or into a flow generator. To make a warranty claim, the initial consumer must return the defective product to the point of purchase, freight prepaid.
This warranty is in lieu of all other express or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office.
L IMITED W ARRANTY







MANUEL UTILISATEUR
Français



CARACTÉRISTIQUES 68
S8 ESCAPE 68 ACCESSOIRES 69 TOUS LES APPAREILS (LE CAS ÉCHÉANT) 69 GARANTIE LIMITÉE 75 INDEX 77
Pages mentionnées dans ce manuel — fabricant, assistance et téléchargements.
Les utilisateurs et notre équipe répondent généralement sous 24 h — et votre question servira aussi aux suivants.
Soyez le premier à poser une question — d'autres utilisateurs ou notre équipe y répondront.