ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorized Representative) 65 Milton Park Abingdon Oxfordshire OX14 4RX UK ResMed Offices Australia, Austria, Finland, France, Germany, Hong Kong, Japan, Malaysia, Netherlands, New Zealand, Singapore, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
Patents and designs pending.
HumidAire, HumidAire 3i, S8 and SmartStart are trademarks of ResMed Ltd and HumidAire is registered in U.S. Patent and Trademark Office.
© 2006 ResMed Ltd.

English

WELCOME 3 WHAT IS THE HUMIDAIRE 3i? 3
YOUR RESPONSIBILITY 3 MEDICAL INFORMATION 4
INTENDED USE 4 CONTRAINDICATIONS 4 WARNINGS 4 COMPONENTS OF YOUR HUMIDAIRE 3i 5 SETTING UP YOUR HUMIDAIRE 3i FOR USE 5
REMOVING THE FRONT CAP (IF FITTED) 5 REMOVING AND STORING THE CONNECTOR PLUG 6 ATTACHING YOUR HUMIDAIRE 3i 6 FILLING THE WATER CHAMBER 6 STARTING AND STOPPING TREATMENT 7
STARTING YOUR TREATMENT 7 STOPPING YOUR TREATMENT 8 DETACHING YOUR HUMIDAIRE 3i 8 FREQUENTLY ASKED QUESTIONS 9 CLEANING AND MAINTENANCE 10
DISASSEMBLING YOUR HUMIDAIRE 3i 10 CLEANING YOUR HUMIDAIRE 3i 10 REASSEMBLING YOUR HUMIDAIRE 3i 12 SERVICING 12 TROUBLESHOOTING 13 TECHNICAL SPECIFICATIONS 15 LIMITED WARRANTY 17 INDEX 19

Thank you for choosing the ResMed HUMIDAIRE 3i™ humidifier. The HUMIDAIRE 3i is designed to humidify (moisten) the air delivered to your airway during CPAP (continuous positive airway pressure) therapy. The HUMIDAIRE 3i attaches to your ResMed S8™ Series CPAP flow generator with a simple ‘click-in’ mechanism. This user manual contains the information you need for the correct use of your HUMIDAIRE 3i.
The HUMIDAIRE 3i is designed to form a single unit with a ResMed CPAP flow generator. The combined unit provides humidification during CPAP therapy, and can make therapy more comfortable. You may have experienced nasal congestion or dryness of the nose, mouth, or throat during CPAP therapy. These symptoms are caused by dry climates, cold air temperatures, the airflow delivered from your flow generator, or mouth leak. Moist air may help relieve these symptoms of dryness. Refer to your flow generator manual to see whether it is compatible with the HUMIDAIRE 3i. Contact your equipment supplier or ResMed office if you require further information.
You or the user of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from:
Please read this manual carefully before use. This manual contains special terms and icons that appear in the margins to draw your attention to specific and important information.

The HUMIDAIRE 3i is indicated for humidification of the air delivered from the ResMed S8 Series CPAP devices. The HUMIDAIRE 3i is for use only as recommended by a physician.
The HUMIDAIRE 3i is contraindicated for use with patients whose upper (supraglottic) airway has been bypassed. Please also refer to the relevant system manual for contraindications associated with CPAP therapy.
The following are general warnings. Specific warnings and cautions appear next to the relevant instructions in the manual.

Please refer to the illustrations in section A of the illustration sheet.
SETTING UP YOUR HUMIDAIRE 3i FOR USE
Make sure your flow generator is switched off at the power outlet before connecting the HumidAire 3i.
Please refer to the illustrations in section B of the illustration sheet. Your flow generator may be fitted with a front cap (B-1). Before you can attach the HUMIDAIRE 3i, press the button underneath the front cap (B-2) and pull it away from the body of your flow generator (B-3 & B-4).
5 COMPONENTS OF YOUR HUMIDAIRE 3i

Please refer to the illustrations in section C of the illustration sheet. Before attaching the HUMIDAIRE 3i, remove the connector plug from your flow generator (C-1) and attach it to the back of the docking station (C-2).
Please refer to the illustrations in section D of the illustration sheet.
1 Align your HUMIDAIRE 3i with your flow generator (D-1) and push them together until they click into place (D-2).
2 Insert the power cord into the socket at the rear of your flow generator. Plug the other end of the power cord into a power outlet (D-3).
3 Place the combined unit on a firm, level surface near your bed at a level lower than your sleeping position. Ensure the surface is not sensitive to heat and/or moisture, or use a protective mat.
Be careful not to place the unit where it can be bumped into and knocked over.
4 Connect one end of the air tubing firmly onto the air outlet of your HUMIDAIRE 3i and the other end to your mask (D-4).
Please refer to the illustrations in section E of the illustration sheet.
Take care when handling your HumidAire 3i. The water and the water chamber base may both be hot. Allow ten minutes for the heater plate and any excess water to cool.
1 Release the cover latch (E-1) and lift the cover until it rests in the upright position (E-2).

2 Remove the water chamber (E-3) and carefully fill it with fresh, clean water (E-4). Make sure you keep the level of the water under the maximum water level mark (E-4a). Always ensure that the water chamber is clean. See “Cleaning and Maintenance” on page 10.
Do not overfill the water chamber, as water may enter your flow generator and air tubing.
Do not use any additives (eg scented oils and perfumes). They may reduce the humidification output of the HumidAire 3i and/or cause deterioration of the water chamber materials.
3 Carefully place the filled water chamber into the humidifier docking station (E-5) and latch the cover (E-6). The HUMIDAIRE 3i is now ready for use.
Please refer to the illustration in section F of the illustration sheet.
1 When your HUMIDAIRE 3i is attached and the flow generator is powered, the indicator light lights up on the HUMIDAIRE 3i. A prompt for the Warm-Up feature (not available on all models) may appear.
When activated, this feature allows the water to warm up before you start treatment.
2 If the Warm-Up feature is available, use it to pre-heat the water before starting your treatment. Skip this step if you wish to start treatment immediately.
WARM-UP FEATURE (NOT AVAILABLE IN ALL MODELS) Note: Check your flow generator user’s manual to see if this feature is available.
To start the Warm-Up feature:
i. Select Warmup by pressing the Right key on the control panel. The following message is displayed for 2 seconds:
WARMING
ii. After 2 seconds the WARMING screen is displayed.
WARMING menu exit

iii. While your HUMIDAIRE 3i is in Warm-Up mode you can:
The Warm-Up feature will automatically turn off after 30 minutes if treatment has not started.
3 Turn the humidity control dial to the desired setting. The setting switches heated humidification off and delivers minimum humidification. The indicator light will remain lit to indicate standby mode. A setting of 6 ( ) delivers maximum humidification. We recommend that you initially set the control dial to 3 ( ). If necessary, adjust the control dial up or down to find the setting that is most comfortable and effective for you. You can adjust the humidity setting at any time. The aim is to have moist air delivered to your mask but to avoid condensation forming in the air tubing. The temperature and humidity of the surrounding air will affect the setting you need. See “Frequently Asked Questions” for further tips on the correct humidity setting.
4 Fit your mask as described in your mask user instructions.
5 Press the Start/Stop key on the control panel. Alternatively, if your flow generator has the SmartStart™ function available, simply breathe into your mask and air should begin to flow.
6 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep.
Do not leave an excessive length of air tubing around the top of your bed. It could twist around your head or neck while you are sleeping.
To stop your treatment at any time, remove your mask and press the Start/Stop key. Alternatively, if your flow generator has the SmartStart function available, simply remove your mask and treatment will end. Some flow generators continue to blow air gently after you have stopped your treatment. This can continue for up to one hour to assist the heater plate in cooling. However, you can unplug the device from the power outlet at any time and allow the heater plate to cool without airflow.
Please refer to the illustrations in section G of the illustration sheet.

1 Unplug the power cord from the power outlet (G-1).
Take care when handling your HumidAire 3i. The water and the water chamber base may both be hot. Allow ten minutes for the heater plate and any excess water to cool (G-2).
2 Remove the water chamber from the docking station (G-3).
3 Press the large silver button located midway underneath the docking station (G-4) and pull the HUMIDAIRE 3i and flow generator apart (G-5).
WHY IS THERE CONDENSATION IN MY MASK AND/OR AIR TUBING? You may experience excessive condensation in your mask or air tubing. This may occur seasonally, particularly when the room temperature is low or cool air is circulating around the tube. Turn the control dial down to a lower setting until the condensation decreases. You may have to adjust your humidity setting from season to season.
DO I HAVE TO USE DISTILLED WATER IN THE WATER CHAMBER? Using distilled water will maximise the life of the water chamber and reduce mineral deposits. See “Cleaning your HumidAire 3i” on page 10 for further details.
WHAT IF I STILL HAVE SYMPTOMS OF DRYNESS? If you still experience symptoms of dryness, then your humidity setting may be set too low to be effective. Gradually turn the humidity control dial up over a few nights until the symptoms disappear. If you are still experiencing symptoms of dryness, even after increasing the humidity setting, you may be breathing through your mouth. Mouth breathing or mouth leaks cause drying and irritation of the upper airway. If this is a problem, a full face mask or chin strap may help. Consult your clinician or equipment supplier for details.
Large silver button

You should regularly carry out the cleaning and maintenance of the water chamber, cover and cover seal as described in this section.
Please refer to the illustrations in section H of the illustration sheet.
Beware of electrocution. Do not immerse your flow generator, docking station or power cord in water. Always unplug the device before cleaning and be sure that it is dry before plugging it back in.
1 Unplug the power cord from the power outlet (H-1).
Take care when handling your HumidAire 3i. The water and the water chamber base may both be hot. Allow ten minutes for the heater plate and any excess water to cool (H-2).
2 Remove the water chamber from the docking station (H-3) and discard any excess water (H-4).
3 Detach the air tubing from the air outlet (H-5).
4 Gently flex one cover arm away and up from the docking station until the cover detaches completely (H-6 to H-8). Peel the cover seal away from the cover (H-9).
Do not remove the tub plate from the water chamber.
! CAUTION

DAILY Wash the water chamber and tubing in warm water, using a mild detergent (see previous caution). Use the yellow cleaning plug to assist in the cleaning process (see steps 1 to 5 below).

Rinse them thoroughly with clean water and allow them to dry away from direct sunlight.
WEEKLY Inspect the water chamber for wear and deterioration. Replace the water chamber if any component has cracked, become cloudy or pitted. Replace immediately if there is any leaking from the water chamber. If you notice white powder or deposits in the water chamber:
i. Fill the water chamber to the maximum water level mark with a solution of one part vinegar to ten parts water. ii. Let the solution soak for ten minutes. iii. Discard the solution and rinse the water chamber with clean water.
COVER AND COVER SEAL Inspect the cover and cover seal weekly for wear and deterioration. Replace any damaged parts. When required, wash the cover and cover seal in warm water, using a mild detergent (see previous caution). Rinse them thoroughly with clean water and allow them to dry away from direct sunlight.
DOCKING STATION Periodically wipe the outside of the docking station (including the connector seal) with a damp cloth.

Please refer to the illustrations in section I of the illustration sheet. Reassemble the parts when they are clean and dry.
1 Align the cover seal to the guides on the cover (I-1) and press along the edges to ensure it is fitted properly (I-2).
2 Put one cover arm into its pivot point on the docking station (I-3) and gently flex the other arm into its corresponding point (I-4 & I-5). The cover should be able to open and close freely while securely attached to the docking station.
3 Replace the water chamber in the humidifier docking station (I-6).
With normal use of a HUMIDAIRE 3i, the water chamber needs to be replaced every two years. This product (HUMIDAIRE 3i) should be inspected by an authorised ResMed Service Centre 5 years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorised ResMed Service Centre.


If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the unit.
Service and repair should only be performed by an authorised service agent.
Device not working (no display on LCD screen and/or indicator light not lighting up)
Power cord is not plugged in properly.
Check that the power cord is plugged firmly into the power outlet and the switch (if available) is on at the outlet.
Faulty power outlet.
Try another power outlet.
Flow generator and docking station not connected correctly.
Check that the docking station is securely attached to the flow generator.
Indicator light periodically becomes brighter to indicate that the heating element is active.
Indicator light is permanently dull and/or air does not feel warm
HumidAire 3i reflector (see illustration C-2a) is dirty.
Clean the reflector with a damp cloth.
The water chamber may be damaged or cracked.
Contact your supplier for a replacement.
Cover is not attached correctly.
Make sure the cover is fitted correctly.
Indicator light flickering · Leaking water chamber

Problem Possible Cause Solution
No/low air flow The air tubing is not connected properly.
Connect the air tubing as described in “Setting Up your HumidAire 3i For Use” on page 5.
The air tubing is blocked, pinched or punctured.
Unblock or free the air tubing. Check the air tubing for punctures.
The flow generator is not working correctly.
See the Troubleshooting section in the flow generator manual.
The water chamber is overfilled.
Make sure that the water level is not above the maximum water level mark.
The cover is not latched correctly.
Make sure the cover is latched onto the docking station correctly.
The cover seal is not fitted correctly.
Make sure the cover seal is facing the right way up and fitted securely.
Condensation is forming in the air tube.
Water splashing on your face
The water chamber is overfilled.
Empty some water from the water chamber.
Condensation is forming in the air tube and mask.
Air does not feel moist
The control dial is set too low.
Turn the control dial up.
HumidAire 3i heating element is not working.
Contact your equipment supplier.
Air too moist Control dial is set too high.
Turn the control dial down.
High relative humidity in your room.
Adjust your humidity setting.
White powder or deposits in the water chamber
The water leaves scaly mineral deposits.

PERFORMANCE Maximum heater plate temperature: 85°C Temperature cut-out: 93°C Maximum output humidity: 95% relative humidity (control dial set to maximum setting, flow rate 50 L/min, ambient temperature of 20°C, ambient relative humidity of 40%). Maximum gas temperature: 41°C
DIMENSIONS (H X W X D) Docking station and water chamber: 118 mm x 170 mm x 212 mm
WEIGHT Docking station and unfilled water chamber: 670 g Water capacity to maximum fill line: 415 mL
POWER SUPPLY Input range for RESMED S8 flow generator with HUMIDAIRE 3i: 100–240V, 50/60Hz; 110V, 400Hz; 2.5A
HOUSING CONSTRUCTION Docking station: Flame retardant engineering thermoplastic, plated cast aluminium. Water chamber: Injection molded plastic, stainless steel.
ENVIRONMENTAL CONDITIONS Operating temperature: +5°C to +40°C Operating humidity: 10–95% non-condensing Storage and transport temperature: -20°C to +60°C Storage and transport humidity: 10–95% non-condensing
IEC 60601-1 CLASSIFICATION Class II (double insulation) Type CF
ELECTROMAGNETIC COMPATIBILITY Please refer to your flow generator user's manual for details.

SYMBOLS WHICH APPEAR ON THE DEVICE
Environmental information WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and electronic equipment. This device should be disposed of separately, not as unsorted municipal waste. To dispose of your device, you should use appropriate collection, reuse and recycling systems available in your region. The use of these collection, reuse and recycling systems is designed to reduce pressure on natural resources and prevent hazardous substances from damaging the environment. If you need information on these disposal systems, please contact your local waste administration. The crossed-bin symbol invites you to use these disposal systems. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to www.resmed.com/environment


ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable.
Accessories, mask systems (including mask frame, cushion, headgear and tubing). Excludes single-use devices.
If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; c) any damage or contamination due to cigarette, pipe, cigar or other smoke; d) any damage caused by water being spilled on or into a flow generator. Warranty is void on product sold, or resold, outside the region of original purchase. Warranty claims on defective product must be made by the initial consumer at the point of purchase.
This warranty is in lieu of all other express or implied warranties, including any implied warranty of merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office.
17 L IMITED W ARRANTY







BEDIENUNGSANLEITUNG
Deutsch
Pages mentionnées dans ce manuel — fabricant, assistance et téléchargements.
Les utilisateurs et notre équipe répondent généralement sous 24 h — et votre question servira aussi aux suivants.
Soyez le premier à poser une question — d'autres utilisateurs ou notre équipe y répondront.