Anritsu Company 490 Jarvis Drive Morgan Hill, CA 95037-2809 USA

The Anritsu product(s) listed on the title page is (are) warranted against defects in materials and workmanship for three years from the date of shipment.
Anritsu’s obligation covers repairing or replacing products which prove to be defective during the warranty period. Buyers shall prepay transportation charges for equipment returned to Anritsu for warranty repairs. Obligation is limited to the original purchaser. Anritsu is not liable for consequential damages.
The foregoing warranty does not apply to defects resulting from improper or inadequate maintenance by the User, unauthorized modification or misuse, or operation outside of the specifications of the product. No other warranty is expressed or implied, and the remedies provided herein are the User’s sole and exclusive remedies.
DISCLAIMER OF WARRANTIES. TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, ANRITSU COMPANY AND ITS SUPPLIERS DISCLAIM ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, WITH REGARD TO THE SOFTWARE PRODUCT. THE USER ASSUMES THE ENTIRE RISK OF USING THE PROGRAM. ANY LIABILITY OF PROVIDER OR MANUFACTURER WILL BE LIMITED EXCLUSIVELY TO PRODUCT REPLACEMENT.
NO LIABILITY FOR CONSEQUENTIAL DAMAGES. TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, IN NO EVENT SHALL ANRITSU COMPANY OR ITS SUPPLIERS BE LIABLE FOR ANY SPECIAL, INCIDENTAL, INDIRECT, OR CONSEQUENTIAL DAMAGES WHATSOEVER (INCLUDING, WITHOUT LIMITATION, DAMAGES FOR LOSS OF BUSINESS PROFITS, BUSINESS INTERRUPTION, LOSS OF BUSINESS INFORMATION, OR ANY OTHER PECUNIARY LOSS) ARISING OUT OF THE USE OF OR INABILITY TO USE THE SOFTWARE PRODUCTS, EVEN IF ANRITSU COMPANY HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. BECAUSE SOME STATES AND JURISDICTIONS DO NOT ALLOW THE EXCLUSION OR LIMITATION OF LIABILITY FOR CONSEQUENTIAL OR INCIDENTAL DAMAGES, THE ABOVE LIMITATION MAY NOT APPLY TO YOU.
Microsoft, Windows, Windows Vista, Microsoft Office, Word, Access, Excel, PowerPoint, Visual Studio, Visual Basic, and .NET are all registered trademarks of the Microsoft Corporation.
Anritsu Company has prepared this manual for use by Anritsu Company personnel and customers as a guide for the proper installation, operation and maintenance of Anritsu Company equipment and computer programs. The drawings, specifications, and information contained herein are the property of Anritsu Company, and any unauthorized use or disclosure of these drawings, specifications, and information is prohibited; they shall not be reproduced, copied, or used in whole or in part as the basis for manufacture or sale of any equipment or software programs without the prior written consent of Anritsu Company.
Updates, if any, can be downloaded from the Documents area of the Anritsu web site at: http://www.us.anritsu.com

IMPORTANT-READ CAREFULLY BEFORE OPENING THE SOFTWARE PACKET. BY OPENING THE PACKET CONTAINING THE SOFTWARE, YOU ARE AGREEING TO BE BOUND BY THE TERMS OF THIS AGREEMENT. THIS IS A LEGAL AGREEMENT BETWEEN YOURSELF, AND YOUR EMPLOYER, IF APPLICABLE, AND ANRITSU COMPANY (“ACUS”). IF YOU DO NOT AGREE WITH ALL OF THE TERMS OF THIS AGREEMENT, PROMPTLY RETURN THE UNOPENED SOFTWARE PACKAGE, AND THE ACCOMPANYING DOCUMENTATION TO ANRITSU FOR FULL CREDIT.
2. COPYRIGHT. This SOFTWARE is owned by ACUS and is protected by United States copyright laws and international treaty provisions. Furthermore, it is considered by ACUS to be ACUS proprietary information. Therefore, you must treat the SOFTWARE like any other copyrighted or proprietary material except that you may either (a) make one copy of the SOFTWARE solely for backup or archival purposes, or (b) transfer the SOFTWARE to a single hard disk, provided you keep the original solely for backup or archival purposes. You may NOT copy any written documentation accompanying the SOFTWARE.
ACUS DOES NOT WARRANT THAT THE SOFTWARE WILL RUN PROPERLY ON ALL HARDWARE, THAT THE SOFTWARE WILL MEET LICENSEES REQUIREMENTS, OR THAT ALL SOFTWARE ERRORS WILL BE CORRECTED.
THE WARRANTIES ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
ACUS SHALL NOT BE LIABLE FOR DIRECT, INDIRECT, INCIDENTAL, SPECIAL, OR CONSEQUENTIAL DAMAGES, , OR DAMAGES FOR LOSS OF PROFITS, REVENUE, DATA, OR USE, INCURRED BY licensee, REGARDLESS OF WHETHER IN TORT OR CONTRACT, AND REGARDLESS OF WHETHER LICENSEE HAS ADVISED ACUS OF THE POSSIBILITY OF SUCH LOSSES.
6. Export Control. Licensee agrees that it shall not export or re-export the software covered hereunder, or any portion thereof, directly, or through others, to the proscribed countries listed in Section 779.4 and associated or successor sections of the U.S. Export Administration Regulations unless properly authorized by the U. S. Government. Licensee further agrees to fully comply with all of the rules and regulations regarding export of this software that may be issued from time to time by the Japanese Ministry of International Trade and Industry.





This manual provides general information, installation, and operating information for the Anritsu MA24106A USB power sensor. Throughout this manual, the terms MA24106A, USB power sensor, and power sensor will be used interchangeably to refer to the device. Manual organization is shown in the table of contents.
This chapter contains general information about the MA24106A power sensor. It includes a general description of the device and information on its identification number, information on initial inspection, and preparation for storage and shipment.
The MA24106A power sensor is a highly accurate, standalone instrument that communicates with a PC via USB. The power measurement capability of MA24106A is intended to mimic that of a traditional thermal (thermo-electric) power sensor. Therefore, it is ideal for measuring the average power of CW, modulated RF waveforms such as 3G, 4G, OFDM, and multi-tone signals. In other words, it measures true RMS power regardless of the type of input signal.
Inspect the shipping container for damage. If the shipping container is damaged, retain until the contents of the shipment have been checked against the packing list and the power sensor has been checked for mechanical and electrical operation. The following items are included with every MA24106A shipment:
If the shipment is incomplete or if the power sensor is damaged mechanically or electrically, notify your local sales representative or Anritsu Customer Service. If the shipping container is damaged or shows signs of stress, notify the carrier as well as Anritsu. Keep the shipping materials for the carrier's inspection.
All Anritsu power sensors are assigned a unique seven digit serial number, such as “0701012”. The serial number is imprinted on a label that is affixed to the unit. When ordering parts or corresponding with Anritsu Customer Service, please use the correct serial number with reference to the specific instrument's model number (for example, model MA24106A power sensor, serial number: 0701012).

Preparation for Storage/Shipment General Information
Preparing the power sensor for storage consists of cleaning the unit, packing the inside with moisture- absorbing desiccant crystals, and storing the unit in the recommended temperature environment. Please refer to the data sheet for storage temperature recommendations.
To provide maximum protection against damage in transit, the power sensor should be repackaged in the original shipping container. If this container is no longer available and the unit is being returned to Anritsu for repair, please advise Anritsu Customer Service; they will send a new shipping container free of charge. In the event neither of these two options is possible, instructions for packaging and shipment are given below:

This chapter provides information and instructions on operating the MA24106A power sensor. It contains the following:
Please make sure that the following minimum requirements are met for installing and using the software:
The driver must be installed before the MA24106A power sensor can be used. Follow the steps below as a guide for proper installation:

Figure 2-1. Anritsu Power Meter Installation Menu


Figure 2-2. Anritsu Power Meter Installation

Figure 2-3. Anritsu Power Meter Installation
Driver Installation Installation


Figure 2-4. License Agreement

Figure 2-5. Confirm Installation
Installation Driver Installation

The software will then install to the selected location.

Figure 2-6. Installing Anritsu Power Meter Application

Figure 2-7. Application Installation Complete
Driver Installation Installation


Figure 2-8. Found New Hardware Wizard

Figure 2-9. Found New Hardware Wizard
Installation Driver Installation


Figure 2-10. Found New Hardware Wizard

Figure 2-11. Found New Hardware Wizard
Driver Installation Installation


Figure 2-12. Found New Hardware Wizard
C:\Program Files\Anritsu\AnritsuPowerMeter\AnritsuMA24106A.inf

Figure 2-13. Install From Disk
Installation Driver Installation


Figure 2-14. Found New Hardware Wizard

Figure 2-15. Hardware Installation
Driver Installation Installation


Figure 2-16. Found New Hardware Wizard
Installation Driver Installation



This chapter provides information on using the Anritsu Power Meter application with the MA24106A USB power sensor. It provides a description of the Graphical User Interface, various settings of the application, basic procedures for Making Measurements, as well as information about Uncertainty of a Measurement.
The graphical user interface (Figure 3-1) layout is divided into five sections:

Data Entry Fields Status Bar
Figure 3-1. Graphical User Interface
The Anritsu Power Meter application always launches in the default state as described below:
Communication with the sensor does not take place until the Apply button is clicked or the Enter key on the keyboard is pressed.
Display Window · Menu Bar · Buttons

Buttons Using the Power Sensor
There are nine buttons available on the user interface to perform the most common tasks as described below:
The data entry fields become active when clicking a button to accept an appropriate entry (see the example in Figure 3-2). Communication with the power sensor does not take place until the Apply button is clicked or the Enter key on the keyboard is pressed. The following list summarizes the entry fields:

Figure 3-2. Example of an Active Entry Field

Using the Power Sensor Display Window
The display window contains the following information (see Figure 3-3 and Figure 3-4):

Figure 3-3. Display Window (Relative Mode with Fixed Offset On)

Figure 3-4. Display Window (Normal mode with Offset Table On)
The status bar displays the model number, serial number, cal factor frequency, averaging number, and operational status of the sensor (see Figure 3-5).
Figure 3-5. Status Bar

Menu Bar Using the Power Sensor
The Menu Bar contains the following menus:
The File menu contains the Exit command, which terminates the application.

Figure 3-6. File Menu
The Tools menu contains:

Figure 3-7. Tools Menu

Using the Power Sensor Menu Bar
The Data Logging menu provides choices between Data Logging On or Data Logging Off.

Figure 3-8. DataLogging Menu
xxxxxxx_yyyy_mm_dd_hhmmss.csv
where:
xxxxxxx: Serial number of the power sensor yyyy: Four-digit year mm: One- or two-digit month dd: One- or two-digit day hhmmss: Two digit hours (24-hour clock), minutes, and seconds
The filename and location can be selected or changed as desired.

Figure 3-9. DataLogging Save Dialog

Menu Bar Using the Power Sensor

Figure 3-10. Data Logging
The Power Graph feature provides the ability to plot measured power with respect to time. This feature can be used for drift testing, tuning circuits, and for monitoring circuit behaviors to external stimuli, etc. The graph is continuously updated in real time.

Figure 3-11. Power Graph Menu

Using the Power Sensor Menu Bar
Clicking Start in the Power Graph menu opens the power versus time graph (Figure 3-12).

Figure 3-12. Power Graph Screen
The default settings of the graph are as follows:
Clicking Setup opens the Graph Setup dialog (below) where the scales of time and power axes can be changed. Power is in dBm and time is in minutes.

Figure 3-13. PowerGraph Setup Dialog

Menu Bar Using the Power Sensor
The Offset Table feature provides the ability to apply corrections to measurements in cases of RF devices being used between the sensor and DUT. Offset Table is different from Fixed Offset as it provides the ability to enter different offset values at different frequencies for an RF device. The frequency response of that device needs to be known before it can be entered. Offset Table employs linear interpolation to determine offset values for intermediate frequencies. In cases where the cal factor frequency is higher than the highest frequency in the offset table, then the offset for the highest frequency in the table is used. Similarly, when the cal factor frequency is lower than the lowest frequency in the offset table, then the offset for the lowest frequency in the offset table is used.
The procedure for setting, saving, recalling, and applying the offset table is as follows:

Figure 3-14. Offset Entry Screen
The word Offset appears in the display window indicating that an offset table correction is applied to the current measurement. Also, a check mark is applied in front of the Offset Table On selection in the OffsetTable menu.

Using the Power Sensor Menu Bar

Figure 3-15. Save As Dialog
Similarly, S2P files can be imported as shown below:

Figure 3-16. Open Dialog

Menu Bar Using the Power Sensor

Figure 3-17. Offset Table Menu (Offset Table Off Status)
The Anritsu Power Meter application retains the set up information of a session, even if the USB power sensor becomes disconnected from the PC. When the USB power sensor is reconnected, the changed properties (if different from default) will be highlighted for five seconds as a reminder of the changed set up. The set up information is lost once the application is closed.

Figure 3-18. Session Restore

Using the Power Sensor Making Measurements
This section presents common procedures for using the MA24106A power sensor. These procedures refer to the MA24106A buttons and menus that were previously described. You should be familiar with the Anritsu Power Meter PC application before attempting these procedures.
To perform a power measurement:

Figure 3-19. Measurement Setup
RF signal connections are made to the Type N male RF connector, which has a 50 Ω characteristic impedance.
result.
When connecting to the Type N connector of the MA24106A to a Type N female connector, observe the following proper practice for tightening the connection:
programmed over this USB interface. In addition to programming, the MA24106A is powered by the USB. The interface is USB 2.0 compatible, but with an interface speed of 12 Mbps.

Making Measurements Using the Power Sensor
Zero the sensor before making power measurements, particularly when operating within the lower 20 dB dynamic range of the power sensor. If frequent low-level measurements are being made, it is advised to check the sensor zeroing often and repeat as necessary. Before zeroing the sensor, connect it to the DUT (device under test) test port and remove RF power from the connection to a level 20 dB below the noise floor of the power sensor. For the MA24106A power sensor, this level is less than –60 dBm. It is preferable to leave the sensor connected to the DUT test port so that ground noise and thermal EMF (electro-magnetic fields) are zeroed out of the measurement. The sensor may also be connected to a grounded connector on the DUT or disconnected from any signal source.
To zero the sensor, press the Zero button on the application. If the sensor fails the zeroing operation, the message box states “Sensor zero failed” and “ZERO_ERROR” will be displayed on the application screen until the problem is corrected. If RF is detected, a reminder message will pop up asking to remove the RF source.
The signal channel/analog signal acquisition hardware is integrated along with the RF front end of the power sensor. All of the necessary frequency and temperature corrections take place within the sensor. Therefore, there is no need for a reference calibration (at 50 MHz and 1 mW) with the MA24106A.
The MA24106A power sensor has an internal EEPROM containing correction and calibration factors that were programmed into the sensor at the factory. The power sensor has an internal temperature sensor that reports its readings periodically to the microprocessor. The sensor makes all of the required calculations on the measurement once the measurement frequency has been entered by the user.
This section presents information on how to get the fastest readings from the MA24106A power sensor when using the Anritsu Power Meter application or operating under remote control (refer to Chapter 4 for specific remote programming command descriptions). Measurement speed depends greatly on the type of measurement, the power level, and stability of the signal. Stability of a measurement is influenced by noise and signal modulation. If high resolution is required, averaging must be increased.

Using the Power Sensor Making Measurements
Table 3-1, describes the number of averages needed to attain a certain noise level for a particular power level measurement with the Low Aperture Time mode setting.
Table 3-1. MA24106A Averaging Table (Low Aperture Time, Default Mode)
< ±0.20 dB
< ±0.15 dB
< ±0.10 dB
< ±0.05 dB
< ±0.01 dB
20 100 1
15 31.6 1
10 10.0 1
5 3.16 1 2
0 1.00 1 1 1 4 16
-5 0.316 1 1 1 20 78
-10 0.100 1
-15 0.0316 1
-20 0.0100 1 7
-25 0.00316 1 1 1 3 61
-30 0.00100 2 3 7 25 –
-35 0.000316 16 28 62 245 –
-40 0.000100 158 –
Table 3-2, describes the number of averages needed to attain a certain noise level for a particular power level measurement with the High Aperture Time mode setting.
Table 3-2. MA24106A Averaging Table (High Aperture Time)
< ±0.20 dB
< ±0.15 dB
< ±0.10 dB
< ±0.05 dB
< ±0.01 dB
20 100 1
15 31.6 1
10 10.0 1
5 3.16 1
0 1.00 1
-5 0.316 1 1 1 2 5
-10 0.100 1
-15 0.0316 1
-20 0.0100 1
-25 0.00316 1 4
-30 0.00100 1 1 1 2 38
-35 0.000316 1 2 4 16 –
-40 0.000100 10 18 39 153 –

Measurement Considerations Using the Power Sensor
This section details some of the error messages that may appear on the application screen. In most cases, the error condition can be easily corrected. The status LED will light yellow when an error state occurs. If not, note the error message and contact an Anritsu Service Center.
Table 3-3. Error Messages
Zero invalid as temperature changed by more than 10 Degrees C
The sensor’s ambient temperature has changed by more than 10 ºC since the last zero operation.
Perform the zero operation again.
Temperature out of operating range
Operating range of the sensor is 0 ºC to 55 ºC. Re-examine the ambient conditions.
Sensor zero failed This message box appears if the zero operation is unsuccessful. The reason could be the presence of RF power at the input of the sensor.
Turn off the RF input to the sensor or disconnect the sensor from the RF source and try the zero operation again.
ZERO_ERROR This message appears on the application screen if the zero operation is unsuccessful. The reason could be the presence of RF power at the input of the sensor.
Turn off the RF input to the sensor or disconnect the sensor from the RF source and try the zero operation again.
ADC_TEMP_OVERRNGE This message appears on the application screen if the sensor is being operated in extremely high temperatures and has overheated.
Remove the sensor from the USB connection and allow to cool to the operating range of the sensor: 0 ºC to 55 ºC
Case 1: Modulated signals with pulse or pattern repetition times ≤1 ms (PRF ≥1 KHz)
If you obtain a steady power reading of a modulated signal (no significant fluctuations of the displayed power) with no averaging, then it is likely that the pulse or pattern repetition rate is greater than 1 KHz. In this case, most of the averaging of the envelope power is performed in the front end of the sensor (before being digitized). When this is the case, the MA24106A will provide an accurate indication of the average power with no special considerations.
Case 2: Modulated signals with pulse or pattern repetition times between 1 ms and 50 ms (100 Hz < PRF < 1 KHz)
In this case, the signal is varying too slowly to be averaged in the front end of the sensor, so averaging must be performed after digitalization by increasing the averaging number in the power meter application (or calculating the average of several measurements if controlling the sensor over the bus). A large amount of averaging must be used for some pulse/pattern repetition frequencies to get a steady reading. If Low Aperture Time (LAT) mode is selected, the maximum recommended pulse repetition time is about 10 ms. If High Aperture Time (HAT) mode is selected, signals with pulse repetition periods as long as 50 ms can usually be measured.
In this case, it can be difficult to get an accurate average power reading even by averaging many readings. The sample rate of the sensor and the pulse repetition rate of the signal may be close enough that they can “beat” together resulting in low frequency modulation of the power indication. If averages are not calculated over many of these beats, or an integer number of beats, errors can result. This is not unique to the MA24106A and can be an issue with any power sensor/meter and any sampled data system.
Message Description Resolution

Using the Power Sensor Measurement Considerations
High crest factor signals, such as CDMA/WCDMA, may have crest factors as high as 10 dB. To ensure the most accurate power measurement, the statistically-low peak signals should not exceed +30 dBm.
For example, if a signal has an expected crest factor of 10 dB, then the highest average power measured should not exceed +20 dBm. A sensor’s linearity graph of a WCDMA (TestModel_5_8HSPDSCH) signal with 10 dB crest factor is shown below:
-40 -30 -20 -10 0 10 20
Figure 3-20. Sensor Linearity Graph
The MA24106A is a True-RMS sensor that can measure very wide bandwidth modulation without much restriction. The only limitation is the frequency flatness of the sensor. Because the sensor’s sensitivity is not identical for all frequencies and when measuring multi-tone signals, the frequency entered into the sensor’s application should be the average frequency of all significant tones. The MA24106A has an error of 0.01 dB for every 100 MHz bandwidth at frequencies below 3 GHz, and an error of 0.03 dB for every 100 MHz bandwidth at frequencies above 3 GHz.
For example, a dual tone signal of 2.0 GHz and 2.2 GHz may have an additional measurement error of 0.02 dB (0.01 dB × 2) when the application frequency is set to 2.1 GHz.
When there is a need to achieve a required reading resolution, particularly at low power levels, averaging is often needed to reduce noise and steady the displayed power reading. Use the noise vs. resolution table in the sensor manual (Using the Power Sensor|Making Measurements|Optimizing the Readings) to determine the number of averages that will typically be required for a given resolution. Alternatively, determine the number of averages through calculation by using the noise specifications and the fact that noise will be proportional to the square root of N, where N is the number of averages.

Measurement Considerations Using the Power Sensor
For example, a CW tone at –30 dBm is to be measured to 0.01 dB resolution. Using the table in the sensor manual, the required number of averages is 38 averages using High Aperture Time mode (the same measurement would require more than 256 averages in Low Aperture Time mode).
The MA24106A samples power continuously every 70 ms in the Low Aperture Time (LAT) mode and 700 ms in the High Aperture Time (HAT) mode. The sensor’s front end and digitizer settles completely to a step change in power in this amount of time. However, there is no way to synchronize the sensor’s sampling to any other event, such as a power step or bus request for a measurement. Therefore, the first measurement requested from the sensor after a power step may not be fully settled. To ensure a fully settled measurement when operating the sensor over the bus, wait 70 ms (700 ms if in HAT) after a power step before requesting the measurement from the sensor. Alternatively, request two measurements from the sensor and discard the first.
If averaging is required as described above, settling time increases by N × sample period, where N is the number of averages and the sample period is the time is milliseconds. The measurement sample period is 70 ms for LAT and 700 ms for HAT. When operating the sensor over the bus, request N+1 measurements from the sensor, discard the first, and then average the subsequent readings. The settling time is approximately (N+1) × sample period.

Using the Power Sensor Uncertainty of a Measurement
Power measurements have many component parts that affect overall measurement uncertainty when measuring power with the MA24106A sensor:
% Mismatch Uncertainty = 100[|1 + Γ1Γ2|2 – 1]
dB Mismatch Uncertainty = 10log|1 + Γ1Γ2|
Γ1 and Γ2 are the reflection coefficients of the power sensor and the device under test
where

Uncertainty of a Measurement Using the Power Sensor
An example of a measurement uncertainty calculation (for Low Aperture Time mode) is detailed for the MA24106A in Table 3-4 below. The MA24106A is used to measure the power of a 3 GHz, +12.0 dBm and –35 dBm CW signal from a signal source with 1.5:1 VSWR.
Table 3-4. Measurement Uncertainty Example
3.0 3.0 Rectangular 1.8 1.8
Noise 0.0 0.8 Normal at 2σ 2 0.0 0.4
Zero Set 0.0 3.2 Rectangular 0.0 1.8
Zero Drift 0.0 0.9 Normal at 2σ 2 0.0 0.5
1.4 1.4 Normal at 2σ 2 0.7 0.7
4.0 4.0 Rectangular 2.3 2.3
Combined Uncertainty (RSS), Room
3.0 3.6
6.0 7.1
1.4 1.4 Rectangular 0.8 0.8
Combined Uncertainty (RSS, 0 to 50 °C)
3.1 3.6
6.2 7.3
with K=2 (RSS, 0 to 50 °C)
Table 3-5. Noise Measurement Uncertainty Calculations
Noise Calculations at 12 dBm (16 mW):
Noise 400 nW/16 mW = 0.0%
Zero Set 1700 nW/16 mW = 0.0%
Zero Drift 500 nW/16 mW = 0.0%
Noise Calculations at –35 dBm (316 nW):
Noise 2.5 nW/316 nW = 0.8%
Zero Set 10 nW/316 nW = 3.2%
Zero Drift 3 nW/316 nW = 0.9%

Once connected to a PC using a USB cable, the MA24106A shows up as an RS-232 Serial COM port on the PC. You can check the COM port number from the Tools drop-down menu or by using the Windows control panel. The COM port number and following settings are needed in order to control the sensor remotely:
When the power sensor is connected to a PC using a USB cable, it will be configured in idle mode. While in idle mode, the sensor is waiting for a START command from the host application. The application developer must send the START command to put the sensor in measurement mode. Once the sensor is in measurement mode, various commands can be sent to communicate with the power sensor. When the application is being closed, the STOP command should be sent to put sensor in idle mode.
exception of the ZERO command, which takes about 19 seconds.
In High Aperture Time mode, the sensor takes about 700 milliseconds to respond to all commands with the exception of the ZERO command, which takes about 19 seconds.
The following is a list of commands available to control the MA24106A:
Table 4-1. Remote Operation Commands
START Puts the power sensor in measurement mode
STOP Stops the measurement mode and puts the sensor in idle mode
IDN? Gets identification information from the sensor
PWR? Gets the power reading from the power sensor.
NPWR? Gets a new power reading from the sensor
FREQ? Gets the current cal factor frequency value from the sensor
FREQ Sets the cal factor frequency value
ZERO Zeros the power sensor
TMP? Gets the current temperature from the sensor
STATUS? Gets the error status byte from the sensor
HAT Sets High Aperture Time mode
LAT Sets Low Aperture Time mode
Command Description

Remote Operation Command Details Remote Operation
Each command needs to be followed by a Line Feed (0x0A, ASCII 10) termination character. Response from the sensor will have a Line Feed (0x0A, ASCII 10) termination character attached at the end.
Description: Puts the power sensor in measurement mode.
Return Value: None for the first time, OK for any subsequent command sent.
START commands, the sensor will return OK. This is helpful, if the user wants to know if the sensor is in measurement mode.
Description: Stops the measurement mode and puts the sensor in idle mode.
Remarks: This command should be sent before exiting the user application.
Description: Gets identification information from the sensor.
Description: Gets the power reading from the power sensor.
letter “E” and the sensor’s LED turns yellow. The STATUS? command can then be issued to find details about the error.
Description: Gets a new power reading from the power sensor.
in the sensor’s buffer. A new measurement of the current power is initiated to get a new power reading.
If an error condition exists, the returned power reading values are pre-tagged with the letter “E” and the sensor’s LED turns yellow. The STATUS? command can then be issued to find details about the error.

Remote Operation Remote Operation Command Details
Description: Gets the current cal factor frequency value from the sensor.
Description: Sets the current cal factor frequency value for the sensor.
Remarks: “fghz” is the cal factor frequency value in GHz. “fghz” must be between 0.05 GHz to
6 GHz. Available resolution is 3 digits after the decimal point.
Description: Zeros the power sensor.
Return Value: OK if successful and ERR if zeroing failed.
Remarks: In case of zero failure, the STATUS? command can be used to retrieve more detail about
the error. See the STATUS? command for more detail. It takes 19 seconds to zero a sensor, please wait at least this long to get a response from the sensor.
Description: Gets the current temperature reading from the sensor.
Return Value: Current temperature in degrees C.
Description: Get error status byte from the sensor.
Status.b0 -> ZERO_TEMP_ERROR (Temperature changed more than allowable limit after zeroing sensor)
Status.b1 -> Not Used
Status.b2 -> ADC_CH2_OR (Temperature over range)
Status.b3 -> ADC_CH3_OR (Detector A over ranged)
Status.b4 -> ZERO_ERROR_DET_A
Status.b5 -> ZERO_ERROR_DET_B
Status.b6 -> TEMP_ERROR (Temperature beyond operating range)
Status.b7 -> Not Used

Remote Operation Command Details Remote Operation
Description: Sets the high aperture time mode.
Remarks: This command will put the sensor in high aperture time mode. In this mode, the A to D
converter integration time is about 160 milliseconds.
Description: Sets the low aperture time mode.
Remarks: This command will put the sensor in low aperture time mode. In this mode, the A to D
converter integration time is about 10 milliseconds. This mode is the default mode for the sensor when powered up.

The test methodology and equipment described herein can be used to gain some confidence in the measurement accuracy of the MA24106A Power Sensor. This is accomplished by comparing the sensor to another sensor with a specified cal factor and linearity performance or uncertainty. General commercially available equipment is used for these tests; however, these procedures are not sufficiently accurate to verify sensors to factory specification. Therefore, sensor test limits in these procedures are set appropriately to the specified comparison equipment. All tests should be performed at an ambient temperature of 20ºC to 25ºC.
and equipment. For calibration, calibration verification, and to maintain the factory specifications of your power sensor, please send sensors to qualified Anritsu Customer Service Centers.
Refer to the following sections for required equipment and test procedures:
Table 5-1. Required Equipment
Vector Network Analyzer (Pretest) Anritsu MS4624B or equivalent
Reflection Coefficient Uncertainty ≤0.013 to 2 GHz Uncertainty ≤0.020 to 6 GHz
Synthesizer (Cal. Factor and Linearity Tests)
Anritsu MG3692 or equivalent
Output Power: >+20 dBm 0.05 to 6 GHz Output Power Setting Resolution: 0.01 dBm Harmonics: ≤–40 dBc Source VSWR ≤2
Reference Power Meter (Cal. Factor and Linearity Tests)
Anritsu ML2438 or equivalent
Instrumentation Accuracy ≤0.5%
Reference Power Sensor (Cal. Factor and Linearity Tests)
Anritsu MA2422B or equivalent
NIST Calibration or equivalent
Anritsu 41KC-10 VSWR ≤1.15 to 6 GHz
10 dB K Attenuator (Linearity Test)
6 dB K Attenuator (Cal. Factor Test)
Anritsu 41KC-6 VSWR ≤1.15 to 6 GHz
Adapter N(f) to K(f) (Cal. Factor and Linearity Tests)
Anritsu 34ANF50 and 34AS50 VSWR ≤1.05 to 6 GHz
Power Splitter (Linearity Tests)
Anritsu K241B Effective Output VSWR
Power Sensor Under Test MA24106A See Datasheet, 11410-00424
Personal Computer Any See Chapter 2
Manufacturer and · Equipment Description · Model Critical Specifications

VSWR Pretest Sensor Operational Tests
The most common cause of power sensor failure is excess input power. Applying power exceeding the damage level shown on the label will damage the sensor’s sensing element resulting in impedance change. Input match will be degraded when element impedance is changed. If you suspect that a senor is damaged, you should start with an input match pretest.
The maximum VSWR values are listed in the Performance Specification section of this manual. The uncertainty of the VSWR test equipment will affect actual measurement values. See the following Table 5-2 for an example of how measurement system uncertainty can affect the Expected Maximum Reflection Coefficient when using the Anritsu MS4624B Vector Network Analyzer.
Follow the manufacturers S11 (or return loss) calibration procedure to perform calibration on a network analyzer. Connect the power sensor to the network analyzer test port and measure power sensor input match. Typically, matches are measured in terms of return loss in dB. Return loss and magnitude of the reflection coefficient conversion equations are as follows:
ρ = 10–RL/20
RL = –20logρ
RL = Return Loss in dB
ρ = Magnitude of the Reflection Coefficient
VSWR and magnitude of the reflection coefficient conversion equations are as follows:
VSWR = (1 + ρ) / (1 – ρ)
ρ = (VSWR – 1) / (VSWR + 1)
VSWR = Voltage Standing Wave Ratio
ρ = Magnitude of the Reflection Coefficient
Record the measured data into Table 5-2 under the Actual Measurement column. The Actual Measurement should be smaller than the Maximum Reflection coefficient. The Maximum Reflection Coefficient is equal to the measurement system uncertainty added to the sensor’s reflection coefficient specification. If the Actual Measurement reflection coefficient is larger than the Maximum Reflection Coefficient, then the power sensor may be defective. If the actual reflection coefficient is significantly larger than the maximum values in Table 5-2, then the sensor is damaged and it is not necessary to perform further testing.
and return defective sensors with a detailed description of the observed problem.
Table 5-2. Pretest Measurement Result
MS4624B Reflection Coefficient Uncertainty
50 MHz to 2 GHz 0.013 0.050 + 0.013 = 0.063
2 GHz to 6 GHz 0.020 0.100 + 0.020 = 0.120

Sensor Operational Tests Calibration Factor Test
In this test the calibration factors of the MA24106A are compared against another sensor (referred to in this procedure as the “reference sensor”) with known calibration factor uncertainties. This reference sensor should be calibrated by a reputable standards laboratory which has low published measurement uncertainty. To perform the comparison, both sensors are used to measure the output power of a synthesizer with a high quality attenuator, such as the 41KC-6, on the output. The attenuator improves the source match of the synthesizer by lowering the mismatch ripples, thereby lowering the uncertainty in the comparison.
a. Connect the reference power sensor to the reference power meter using the appropriate cables.
b. Connect the MA24106A USB cable between the personal computer with the Power Meter
application installed and the MA24106A power sensor under test.
c. Launch the Power Meter application.
d. Turn the power on to all of the instruments and allow them to warm up for the amount of time
specified in their respective manuals.
e. Reset or Preset all of the instruments.
f. Configure the reference meter and sensor to measure a CW signal.
g. Perform a sensor Zero and a 1 mW reference calibration on the reference sensor and meter per the
manufacturer’s instructions.
h. Perform a low level Zero of the MA24106A as follows:
With the MA24106A disconnected from the synthesizer, click the Zero Sensor button on the Power Meter application and wait for the Zeroing Sensor message to close.
i. Connect the attenuator to the output of the synthesizer with the appropriate adapter to the output
of the attenuator.
j. Set the synthesizer to +6 dBm and 50 MHz.

Figure 5-1. Cal Factor Test Set Up

Calibration Factor Test Sensor Operational Tests
Press the Frequency button on the Power Meter application, and then enter the frequency of the measurement in GHz.
Table 5-3. Calfactor Test Measurement Results
|A-B| Absolute Value of Difference in Power
B
0.05 0.26
0.1 0.26
0.3 0.26
0.5 0.26
1.0 0.26
2.0 0.31
3.0 0.31
4.0 0.31
5.0 0.33
6.0 0.33

Sensor Operational Tests Linearity Test
The linearity correction of the MA24106A is compared to a thermal power sensor, which has very good inherent linearity over a power range of about –20 to +10 dBm. For this reason, the MA24106A will be compared to the thermal sensor in two ranges, keeping the power levels to the thermal sensor in the range of –17 dBm to +5 dBm, while the power to the MA24106A will vary from about –26 dBm to about +14 dBm.
a. Connect the reference power sensor to the reference power meter using the appropriate cables.
b. Connect the MA24106A USB cable between the personal computer with the Power Meter
application installed and the MA24106A power sensor under test.
c. Launch the Power Meter application.
d. Turn the power on to all of the instruments and allow them to warm up for the amount of time
specified in the instrument’s respective manuals.
e. Reset or Preset all of the instruments.
f. Configure the reference meter and sensor to measure a CW signal.
g. Perform a sensor Zero and a 1 mW reference calibration on the reference sensor and meter per the
manufacturer’s instructions.
h. Perform a low-level Zero of the MA24106A as follows:
With the MA24106A disconnected from the synthesizer, click the Zero Sensor button on the Power Meter application and wait for the Zeroing Sensor message to close.
i. Connect the power splitter to the output of the synthesizer and connect the 10 dB attenuator to
one of the splitter outputs.
j. Connect an N(f) to K adapter to each power sensor.
k. Connect the reference sensor and adapter to the 10 dB attenuator.
l. Connect the MA24106A and adapter to the other splitter output.
m. Set the synthesizer to 50 MHz and +20 dBm.

Linearity Test Sensor Operational Tests
n. Increase averaging on the MA24106A by clicking the Averages button, enter “16” and then click
Apply.

Figure 5-2. Linearity Test Setup
Click the Frequency button on the Power Meter application, and then enter the frequency of the measurement in GHz.
a. Record the power reading by the reference meter in the appropriate space in Table 5-4.
b. Record the power reading by the MA24106A in the appropriate space in Table 5-4.
c. Set the synthesizer power to +15 dBm.
d. Record the reference meter and the MA24106A power sensor readings in the appropriate places in
Table 5-4.
e. Repeat the measurement for synthesizer output levels of +10, +5, and 0 dBm.

Sensor Operational Tests Linearity Test
a. Remove the 10 dB attenuator from in between the reference sensor and splitter and connect the
reference sensor directly to the splitter.
b. Remove the MA24106A from the splitter and connect the 10 dB attenuator between the splitter
and the MA24106A power sensor (see Figure 5-3).
c. Turn Off the synthesizer’s RF output and perform a low-level Zero of both the Reference sensor
and the MA24106A.
d. Turn On the synthesizer’s RF output.

Figure 5-3. Linearity Test Setup
e. Set the synthesizer output level to +10 dBm and then adjust it until the sensor/meter under test
reads as close as possible to the value obtained above in step 6e.
a. Read and record the power indicated by the reference meter in the appropriate place in Table 5-4.
b. Lower the output power level of the synthesizer to +5 dBm.
c. Record the reference meter and the MA24106A power sensor readings in the appropriate place in
Table 5-4.

Linearity Test Sensor Operational Tests
d. Repeat the measurement for synthesizer output levels of 0, –5, and –10 dBm.
Table 5-4. Measurement Results (50 MHz)
B = (A6–A5)
C = (A+B) Corrected Reference Power
E = (C–D)
1 +20 10 0
2 +15 10 0
3 +10 10 0
4 +5 10 0
5 0 10 0
7 +5 0 0 10
8 0 0 0 10
9 -5 0 0 10
10 -10 0 0 10
11 Max:
12 Min:
13 Delta (E11 – E12):
a. Subtract the Reference Power Measurement of row 5 from the Reference Power Measurement of
row 6. Record this value in the Correction column of rows 1 through 5.
The Correction column of rows 6 through 10 have values of 0.
b. Add the Reference Power Measurement and Correction values of row 1 and record the result in the
Corrected Reference Power Measurement column of row 1.
c. Repeat step 9b for rows 2 through 10.
d. Subtract the MA24106A Measurement of row 1 from the Corrected Reference Power Measurement
of row 1 and record the result in the Difference Calculation column of row 1.
e. Repeat step 9d for rows 2 through 10.
f. Find the largest (most positive) value in the Difference Calculation column and record this value
next to the word Max in row 11.
g. Find the smallest (least positive or most negative) value in the Difference Calculation column and
record this value next to the word Min in row 12.
h. Subtract the Min value from step 9g from the Max value from step 9f and record the result next to
the word Delta in row 13.
i. The Delta result should be less than 0.3 dB. If it is larger, contact Anritsu customer service.

Sensor Operational Tests Linearity Test
Table 5-5. Measurement Results (2 GHz)
B = (A6–A5)
C = (A+B) Corrected Reference Power
C–D
Attenuation in Test Arm
1 +20 10 0
2 +15 10 0
3 +10 10 0
4 +5 10 0
5 0 10 0
7 +5 0 0 10
8 0 0 0 10
9 –5 0 0 10
10 –10 0 0 10
11 Max:
12 Min:
13 Delta (E11 – E12):
Table 5-6. Measurement Results (4 GHz)
B = (A6–A5)
C = (A+B) Corrected Reference Power
E = (C–D)
Attenuation in Test Arm
1 +20 10 0
2 +15 10 0
3 +10 10 0
4 +5 10 0
5 0 10 0
7 +5 0 0 10
8 0 0 0 10
9 –5 0 0 10
10 –10 0 0 10
11 Max:
12 Min:
13 Delta (E11 – E12):

Linearity Test Sensor Operational Tests
Table 5-7. Measurement Results (6 GHz)
B = (A6–A5)
C = (A+B) Corrected Reference Power
E = (C–D)
1 +20 10 0
2 +15 10 0
3 +10 10 0
4 +5 10 0
5 0 10 0
7 +5 0 0 10
8 0 0 0 10
9 –5 0 0 10
10 –10 0 0 10
11 Max:
12 Min:
13 Delta (E11 – E12):

Anritsu MA24106A Power Sensors are high-quality precision laboratory instruments and should receive the same care and respect afforded to such instruments. Follow the precautions listed below when handling or connecting these devices. Complying with these precautions will guarantee longer component life and less equipment down time due to connector or device failure. Also, such compliance will ensure that Power Sensor failures are not due to misuse or abuse – two failure modes not covered under the Anritsu warranty.
Destructive pin depth of mating connectors is the major cause of failure in the field. When an RF component is mated with a connector having a destructive pin depth, damage will usually occur to the RF component connector. A destructive pin depth is one that is too long in respect to the reference plane of the connector (Figure A-1, left).
The center pin of a precision RF component connector has a precision tolerance. The mating connectors of various RF components may not be precision types. Consequently, the center pins of these devices may not have the proper pin depth. The pin depth of DUT connectors should be measured to assure compatibility before attempting to mate them with Power Sensor connectors. An Anritsu Pin Depth Gauge or equivalent can be used for this purpose (Figure A-1, right)


Figure A-1. Pin Depth
If the measured connector is out of tolerance in the “+” region, the center pin is too long (see Table A-1, below). Mating under this condition will likely damage the precision RF component connector. If the test device connector measures out of tolerance in the “–” region, the center pin is too short. This should not cause damage, but it will result in a poor connection and a consequent degradation in performance.
Table A-1. Allowable DUT connector pin depth
Connector Type Pin Depth Gage Model # Pin Depth (inches) Pin Depth Gage Readout
N-Male SC3825 0.207 +0.030 –0.000
0.207 +0.000 –0.030
N-Female SC3825 0.207 +0.000 –0.030
0.207 +0.000 –0.030

Connector Care and Handling
Over torquing connectors is destructive; it may damage the connector center pin. Always use a connector torque wrench when tightening connectors. Never use pliers to tighten connectors.
Precision connectors are designed to withstand years of normal bench handling. However, do not drop or otherwise treat them roughly. Mechanical shock will significantly reduce their service life.
The MA24106A sensor is rated at +33 dBm maximum continuous input power. Exceeding the maximum input power level permanently damages the internal components, rendering the power sensor useless.
The precise geometry that makes the RF component's high performance possible can easily be disturbed by dirt and other contamination adhering to the connector interfaces. When not in use, keep the connectors covered. To clean the connector interfaces, use a clean cotton swab that has been dampened with denatured alcohol.
off most of the cotton and then twist the remaining cotton tight. Be sure that the remaining cotton does not get stuck in the connector. Cotton swabs of the appropriate size can be purchased through a medical laboratory-type supply center.
The following are some important tips on connector care:

The MA24106A CD contains a demo application that allows you to interface with the power sensor using the remote programming protocol. The sample code is written in Microsoft® Visual Basic® 6.0 and is given at the end of this appendix. The complete project, DempApp.vbp, is available on the CD that shipped with the sensor. The Demo Application’s main form is shown below:

Figure B-1. Demo Application
It is recommended to use the source code and project available on the CD that shipped with the sensor to minimize typing errors. You may need to add Microsoft® Comm Control 6.0 manually, which can be added from Visual Basic® 6.0 IDE by navigating to: Project | Components.
Once connected to the PC using the USB cable, the MA24106A shows up as a Serial port device on the PC. You can check the COM port number using the device manager in the Windows® control panel.
Type the COM port number in the ComPortNo: text box and click Initialize. Once the COM port is initialized, you can type commands in the Command text box, and then click the Send button. Any responses from the sensor will be displayed in the Received text box. You can use the Functions group buttons to exercise the sensor for power readings, frequency readings and settings, and for zeroing sensor.
(COM3 to COM16).

Using the Demo Application Sample Visual Basic Code
************************************************************************ // This sample program shows how to control Anritsu MA24106A USB power sensor using //Microsoft Visual basic 6.0
Option Explicit Public gstrInputBuffer As String
'Event handler for InitializeComPort button Private Sub btnInitializeComPort_Click() Call SetCommPort(Val(Trim(txtCOMPORTNo.Text))) End Sub
'Subroutine to set the com port Public Sub SetCommPort(portNo As Integer) On Error GoTo errHndler 'Setup MSComm control MSComm1.Settings = "115200,n,8,1" MSComm1.CommPort = Trim(txtCOMPORTNo.Text) MSComm1.PortOpen = True MSComm1.RThreshold = 1 MSComm1.SThreshold = 1 ' 'Wait for half a second before sending START command Delay (0.5) 'Arm sensor to start making measurements txtCommand.Text = "START" Call btnSend_Click ' Exit Sub
'Event handler for ResetComPort button Private Sub btnResetComPort_Click() 'Close com port If MSComm1.PortOpen = True Then MSComm1.PortOpen = False End If End Sub
'Event handler for Send button Private Sub btnSend_Click() Dim strResult As String 'Clear buffer & receive text window before sending command gstrInputBuffer = "" txtReceived.Text = "" 'Send command and appeand Termination character, 0x0A(10)with it. MSComm1.Output = UCase(txtCommand.Text) & Chr(10) 'Display received result on the Received text box txtReceived.Text = strResult ' End Sub

Sample Visual Basic Code Using the Demo Application
' 'Event handler for MSComm1 event Private Sub MSComm1_OnComm() 'Get data from Input buffer gstrInputBuffer = MSComm1.Input 'Display received result on the Received text box txtReceived.Text = gstrInputBuffer End Sub
'Event handler for GetFreq button Private Sub btnGetFreq_Click() txtCommand.Text = "FREQ?" Call btnSend_Click End Sub
'Event handler for GetPower button Private Sub btnGetPower_Click() txtCommand.Text = "PWR?" Call btnSend_Click End Sub
'Event handler for SetFreq button Private Sub btnSetFreq_Click() txtCommand.Text = "FREQ " & txtFreq.Text Call btnSend_Click End Sub
'Event handler for ZeroSensor button Private Sub btnZeroSensor_Click() txtCommand.Text = "ZERO" Call btnSend_Click 'Sensor will return OK after about 19 Seconds End Sub
'Event handler for Close button Private Sub btnClose_Click() ' 'Make sure we close the com port before we exit the app If MSComm1.PortOpen = True Then 'Stop sensor from making measurements txtCommand.Text = "STOP" Call btnSend_Click ' 'Wait for half a second after sending START command Delay (0.5) ' MSComm1.PortOpen = False End If 'Close the app End End Sub

Using the Demo Application Sample Visual Basic Code
'Delay routine Public Sub Delay(ByVal Seconds As Single) ' Dim fStartTimer As Single Dim fFinish As Single ' fStartTimer = Timer ' Do DoEvents fFinish = Timer If Abs(fFinish - fStartTimer) > Seconds Then Exit Do End If Loop ' End Sub ************************************************************************

The Anritsu Power Meter 1.0 application requires that the serial ports between COM1 and COM16 be used. You can find out the COM port number by going to Start | Settings | Control Panel | System | Hardware | Device Manager | Ports (COM & LPT). Disconnect and reconnect the power sensor’s USB cable from the computer and notice the new COM port number that appears in the Ports list.

Figure C-1. Device Manager
If this number is less than 16, then the PC application will work fine. However, in some cases when the power sensor is connected, Windows may map your serial port to a port number greater than 16, such as COM17 or COM29 depending on which USB port that is being used. If you connect your power sensor and the Anritsu Power Meter application displays No Sensor, a port number above COM16 may be assigned. To correct this problem, you will need to follow one of the two steps outlined below. Whichever method you use, you should only need to perform the procedure once.

Method 1–Download Updated Software Serial Port Compatibility
The preferred method for resolving serial port compatibility issues is to download software updates for your product from www.us.anritsu.com
A serial port may be remapped to a different number, such as changing a serial port from COM19 to COM5. This may be needed if method 1 above does not result in a serial port assignment between COM1 and COM16 or if you prefer to use a USB hub or a specific USB port on your computer.

Serial Port Compatibility Method 3–Remapping a Serial Port

Figure C-2. System Properties

Method 3–Remapping a Serial Port Serial Port Compatibility

Figure C-3. Device Manager

Serial Port Compatibility Method 3–Remapping a Serial Port

Figure C-4. Advanced Settings for COM Port

Method 3–Remapping a Serial Port Serial Port Compatibility

Figure C-5. Device Manager

The Anritsu Power Meter application provides the necessary software to upgrade the MA24106A power sensor’s firmware. The current sensor firmware can be determined from the Help | About menu in the Anritsu Power Meter application. To upgrade the firmware, proceed as follows:

Figure D-1. Firmware Upgrade Dialog
upgrade driver. Follow the driver installation procedure as outlined in Section 2-3, starting with step 9, and select the MA24106A Upgrade Mode driver when prompted from: C:\Program Files\Anritsu\AnritsuPowerMeter\UpgradeModeDriver\AnritsuMA24106AUpgradeMode.inf

Figure D-2. Upgrade Firmware Driver

Upgrading the Firmware Upgrading the Firmware

Figure D-3. Firmware Upgrade Dialog

Figure D-4. Firmware Upgrade Application

Upgrading the Firmware Upgrading the Firmware

Figure D-5. Open File Dialog
sequence as this will cause an unrecoverable programming error.

Upgrading the Firmware Upgrading the Firmware
MESSAGE - Programming FLASH Completed
MESSAGE - Erasing and Programming FLASH...

Figure D-6. Firmware Upgrade Application

Figure D-7.

excessive A-2 measured 3-3 measurement 3-11 relative 3-3 units 3-3 y-axis of graph 3-7 power units button 3-2 powergraph menu 3-6 powergraph, setup 3-7 pulse signal measurements 3-14 PWR? 4-2
contents 1-1 default settings 3-1 description 1-1 impedance 3-11 inspection 1-1 installation 2-1 operation 2-1 serial number 1-1 mapping serial port C-2 measured power 3-3 measurements 3-11
R rate, baud 4-1 reconnecting sensor 3-10 relative power 3-3 relative/normal button 3-2 remapping serial port C-2 remote commands 4-1 required equipment, for test 5-1 requirements
of high crest factor signals 3-15 of modulated signals 3-14 of multitone signals 3-15 of power 3-11 mechanical shock A-2 menu
datalogging 3-5 file 3-4 offsettable 3-8 powergraph 3-6 tools 3-4 menu bar 3-4 mismatch uncertainty 3-17 modulated signal measurements 3-14 multitone signals 3-15
hardware 2-1 software 2-1 reset averages 3-4 restore session 3-10 run/hold button 3-2
S sample VB code B-1 sensor
N noise 3-15, 3-17 normal/relative button 3-2 NPWR? 4-2 number, serial 1-1
calibrating 3-12 linearity 3-17 zeroing 3-12 sensor zero failed 3-14 serial number 1-1 serial port C-1
fixed 3-3, 3-4 table 3-3 offsettable menu 3-8 operation 2-1 operational test 5-1 optimizing 3-12
configuration C-2 remapping C-2 session restore 3-10 settings
COM port 4-1 default 3-1 settling time 3-16 setup, of graph 3-7 shipment 1-2 shipping contents 1-1 software update C-2 software, part number 1-1 START 4-2 starting a measurement 3-1 status bar 3-3 STATUS? 4-3 STOP 4-2 stop bits 4-1 storage 1-2
P parity 4-1 part number
CD 1-1 software 1-1 USB cable 1-1 pin depth A-1

units of power 3-3 update software C-2 upgrading firmware 3-4, D-1 USB
averaging 3-13 error messages 3-14 offset 3-3 remote commands 4-1 required equipment 5-1 temperature
cable 1-1 serial port C-1 user interface, layout 3-1
V VB code B-1 VSWR test 5-2
compensation 3-17 out of range 3-14 test
calibration factor 5-3 linearity 5-5 operational 5-1 VSWR 5-2 tightening, connection 3-11 time
W window, display 3-3
X x-axis of graph 3-7
Y y-axis of graph 3-7
aperture 3-4 x-axis of graph 3-7 time, settling 3-16 TMP? 4-3 tools menu 3-4 torquing connectors A-2
Z ZERO 4-3 zero
button 3-2 drift 3-17 failed 3-14 invalid 3-14 set 3-17 ZERO_ERROR 3-14 zeroing sensor 3-12
U uncertainty 3-17
calibration factor 3-17 example 3-18 linearity 3-17 mismatch 3-17 noise 3-17 temperature compensation 3-17 zero drift 3-17 zero set 3-17
T to Z · T table · MA24106A UG Index-3