MAINTENANCE MANUAL
P/N: 10370-10290
REVISION: H PRINTED: AUGUST 2003 COPYRIGHT 2003 ANRITSU CO.

WARRANTY The ANRITSU product(s) listed on the title page is (are) warranted against defects in materials and workmanship for one year from the date of shipment.
ANRITSU’s obligation covers repairing or replacing products which prove to be defective during the warranty period. Buyers shall prepay transportation charges for equipment returned to ANRITSU for warranty repairs. Obligation is limited to the original purchaser. ANRITSU is not liable for con- sequential damages.
LIMITATION OF WARRANTY The foregoing warranty does not apply to ANRITSU connectors that have failed due to normal wear. Also, the warranty does not apply to defects resulting from improper or inadequate maintenance by the Buyer, unauthorized modification or misuse, or operation outside of the environmental specifi- cations of the product. No other warranty is expressed or implied, and the remedies provided herein are the Buyer’s sole and exclusive remedies.
TRADEMARK ACKNOWLEDGEMENTS Adobe Acrobat is a registered trademark of Adobe Systems Incorporated.
NOTICE ANRITSU Company has prepared this manual for use by ANRITSU Company personnel and cus- tomers as a guide for the proper installation, operation, and maintenance of ANRITSU Company equipment and computor programs. The drawings, specifications, and information contained herein are the property of ANRITSU Company, and any unauthorized use or disclosure of these drawings, specifications, and information is prohibited; they shall not be reproduced, copied, or used in whole or in part as the basis for manufacture or sale of the equipment or software programs without the prior writtten consent of ANRITSU Company.






1-1 SCOPE OF MANUAL This manual provides service information for all models of the Series 682XXB/683XXB Synthesized Signal Generators. The service informa- tion includes replaceable parts information, functional circuit descrip- tions, block diagrams, performance verification tests, and procedures for calibration, troubleshooting, and assembly/subassembly removal and replacement. (Throughout this manual, the terms 682XXB/ 683XXB and signal generator will be used interchangeably to refer to the instrument.) Manual organization is shown in the table of con- tents.
1-2 INTRODUCTION This chapter provides a general description of the Series 682XXB/ 683XXB Synthesized Signal Generators, identification numbers, related manuals, and options. Information is included concerning level of maintenance, replaceable subassemblies and RF components, exchange assembly program, and preventive maintenance. Static- sensitive component handling precautions and lists of exchangeable subassemblies and recommended test equipment are also provided.
1-3 DESCRIPTION The Series 682XXB Synthesized Signal Generator and the Series 683XXB Synthesized Signal Generator are microprocessor-based, synthesized signal sources with high resolution phase-lock capability. They generate both discrete CW frequencies and broad (full range) and narrow band sweeps across the frequency range of 10 MHz to 65 GHz. All functions of the signal generators are fully controllable locally from the front panel or remotely (except for power on/standby) via the IEEE-488 General Purpose Interface Bus (GPIB).
The Series 682XXB Synthesized Signal Generator and the Series 683XXB Synthesized Signal Generator each presently consists of 15 models covering a variety of frequency ranges and power levels. Table 1-1, pages 1-4 and 1-5, lists models, frequency ranges, and maximum leveled output.

Table 1-1. Series 682XXB/683XXB Models (2 of 2)
(GHz) Output Power Output Power w/Step Attenuator
0.01 – 2.0 GHz 2.0 – 20.0 GHz 20.0 – 40.0 GHz 40.0 – 50.0 GHz 50.0 – 65.0 GHz
+12.0 dBm +10.0 dBm
+2.5 dBm
0.0 dBm –2.0 dBm
With Option 15B (High Power) Installed
68X37B 2.0 – 20.0 GHz +17.0 dBm +15.0 dBm
68X45B 0.5 – 2.2 GHz 2.2 – 20.0 GHz +13.0 dBm +17.0 dBm
+11.0 dBm +15.0 dBm
68X47B 0.01 – 2.0 GHz 2.0 – 20.0 GHz +13.0 dBm +17.0 dBm
+11.0 dBm +15.0 dBm
68X53B 2.0 – 20.0 GHz 20.0 – 26.5 GHz +13.0 dBm +10.0 dBm
+11.0 dBm
+7.5 dBm
68X55B 0.5 – 2.2 GHz 2.2 – 20.0 GHz 20.0 – 26.5 GHz
+13.0 dBm +13.0 dBm +10.0 dBm
+11.0 dBm +11.0 dBm
+7.5 dBm
68X59B 0.01 – 2.0 GHz 2.0 – 20.0 GHz 20.0 – 26.5 GHz
+13.0 dBm +13.0 dBm +10.0 dBm
+11.0 dBm +11.0 dBm
+7.5 dBm
68X63B 2.0 – 20.0 GHz 20.0 – 40.0 GHz +13.0 dBm
+11.0 dBm
+6.0 dBm
+3.0 dBm
68X65B 0.5 – 2.2 GHz 2.2 – 20.0 GHz 20.0 – 40.0 GHz
+13.0 dBm +13.0 dBm
+11.0 dBm +11.0 dBm
+6.0 dBm
+3.0 dBm
68X69B 0.01 – 2.0 GHz 2.0 – 20.0 GHz 20.0 – 40.0 GHz
+13.0 dBm +13.0 dBm
+11.0 dBm +11.0 dBm
+6.0 dBm
+3.0 dBm
68X75B 0.5 – 50.0 GHz Standard Standard
68X77B 0.01 – 50.0 GHz Standard Standard
68X85B 0.5 – 60.0 GHz Standard Standard
68X87B 0.01 – 60.0 GHz Standard Standard
68X95B 0.5 - 65.0 GHz Standard Not Available
68X97B 0.01 – 65.0 GHz Standard Not Available
All ANRITSU instruments are assigned a unique six-digit ID number, such as “403002.” The ID number is imprinted on a decal that is af- fixed to the rear panel of the unit. Special-order instrument configura- tions also have an additional special serial number tag attached to the rear panel of the unit.
68XXXB Model · Frequency · Not Available

When ordering parts or corresponding with ANRITSU Customer Serv- ice, please use the correct serial number with reference to the specific instrument’s model number (i.e., Model 68347B Synthesized Signal Generator, Serial No. 403002).
1-5 ELECTRONIC MANUAL This manual is available on CD ROM as an Adobe Acrobat Portable Document Format (.pdf) file. The file can be viewed using Acrobat Reader, a free program that is also included on the CD ROM. The file is “linked” such that the viewer can choose a topic to view from the displayed “bookmark” list and “jump” to the manual page on which the topic resides. The text can also be word-searched. Contact ANRITSU Customer Service for price and availability.
1-6 RELATED MANUALS This is one of a four manual set that consists of an Operation Manual, a GPIB Programming Manual, a SCPI Programming Manual, and a Maintenance Manual.
This manual provides information and instructions for operation of the series 682XXB/683XXB signal generators using the front panel controls. It also in- cludes general information, performance specifica- tions, installation instructions, and operation verification procedures. The ANRITSU part number for the Operation Manual is 10370-10284.
This manual provides information for remote opera- tion of the series 682XXB/683XXB signal generators using Product Specific commands sent from an ex- ternal controller via the IEEE 488 General Purpose Interface Bus (GPIB). It contains a complete listing and description of all 682XXB/683XXB GPIB Prod- uct Specific commands and several programming examples. The ANRITSU part number for the GPIB Programming Manual is 10370-10286.
This manual provides information for remote opera- tion of the series 682XXB/683XXB signal generators using Standard Commands for Programmable In- struments (SCPI) commands sent from an external controller via the IEEE 488 General Purpose Inter- face Bus (GPIB). It contains a complete listing and description of each command in the 682XXB/ 683XXB SCPI command set and examples of com- mand usage. The ANRITSU part number for the SCPI Programming Manual is 10370-10288.

1-7 OPTIONS The following instrument options are available.
q Option 1, Rack Mounting. Rack mount kit containing a set of
track slides (90 tilt capability), mounting ears, and front panel handles for mounting the instrument in a standard 19-inch equipment rack.
q Option 2A, 110 dB Step Attenuator. Adds a 10 dB per step
attenuator with a 110 dB range for models having a high-end frequency of 26.5 GHz. Output power is selected directly in dBm on the front panel (or via GPIB). Rated output power is reduced.
q Option 2B, 110 dB Step Attenuator. Adds a 10 dB per step
attenuator with a 110 dB range for models having a high-end frequency of 40 GHz. Output power is selected directly in dBm on the front panel (or via GPIB). Rated output power is reduced.
q Option 2C, 90 dB Step Attenuator. Adds a 10 dB per step
attenuator with a 90 dB range for models having a high-end frequency of 50 GHz. Output power is selected directly in dBm on the front panel (or via GPIB). Rated output power is reduced.
q Option 2D, 90 dB Step Attenuator. Adds a 10 dB per step
attenuator with a 90 dB range for models having a high-end frequency of 60 GHz. Output power is selected directly in dBm on the front panel (or via GPIB). Rated output power is reduced.
q Option 6, Phase Modulation (M). Adds phase modulation ca-
pability. The internal FM generator becomes the FM/M genera- tor. (Not available in combination with Option 7.)
q Option 7, Delete AM/FM Generator. Deletes the internal AM
and FM generators. External AM and FM capability remains un- changed. (Not available in combination with Option 8 or Option 20.)
q Option 8, Internal Power Meter. Adds an internal power
meter that is compatible with ANRITSU 560-7, 5400-71, or 6400- 71 series detectors. (Not available in combination with Option 7.)
q Option 9, Rear Panel RF Output. Moves the RF output con-
nector to the rear panel.
q Option 11, 0.1 Hz Frequency Resolution. Provides frequency
resolution of 0.1 Hz.
q Option 14, ANRITSU 360B VNA Compatibility. Modifies rack
mounting hardware to mate unit in a ANRITSU 360B VNA con- sole.
q Option 15B, High Power Output. Adds high-power RF compo-
nents to the instrument providing increased RF output power in the 2–26.5 GHz frequency range. Option 15B is standard in mod- els having a high-end frequency that is >40 GHz.
q Option 16, High-Stability Time Base. Adds an ovenized,
10 MHz crystal oscillator with

q Option 17A, No Front Panel. Deletes the front panel for use in
remote control applications where a front panel display or key- board control are not needed.
q Option 18, mmWave Module Bias Output. Provides bias out-
put for 54000-xWRxx Millimeter Wave Source Modules. BNC Twinax connector, rear panel.
q Option 19, SCPI Programmability. Adds GPIB command
mnemonics complying with Standard Commands for Program- mable Instruments (SCPI), Version 1993. SCPI programming complies with IEEE 488.2-1987.
q Option 20, SCAN Modulator. Adds an internal SCAN modula-
tor for simulating high-depth amplitude modulated signals in models 68237B/68337B, 68245B/68345B, and 68247B/68347B only. Requires an external modulating signal. (Not available in combination with Option 7.)
Maintenance of the 682XXB/683XXB consists of:
q Troubleshooting the instrument to a replaceable subassembly or
RF component. q Repair by replacing the failed subassembly or RF component. q Calibration. q Preventive maintenance.
The 682XXB/683XXB firmware includes internal diagnostics that self-test most of the internal assem- blies of the instrument. When the signal generator fails self-test, one or more error messages are dis- played to aid in troubleshooting the failure to a replaceable subassembly or RF component. Chapter 5 lists and describes the self-test error messages and provides procedures for isolating 682XXB/ 683XXB failures to a replaceable subassembly or RF component.
Repair Most instrument failures are field repairable by re- placing the failed subassembly or RF component. Detailed instructions for removing and replacing failed subassemblies and components are provided in Chapter 6.
Calibration The signal generator may require calibration after repair. Refer to Chapter 4 for a listing of calibration requirements and calibration procedures.
Preventive maintenance on the 682XXB/683XXB consists of cleaning the fan honeycomb filter, de- scribed in paragraph 1-8.

1-9 PREVENTIVE MAINTENANCE
The 682XXB/683XXB must always receive adequate ventilation. A blocked fan filter can cause the instrument to overheat and shut down. Check and clean the rear panel fan honeycomb filter periodically. Clean the fan honeycomb filter more frequently in dusty environ- ments. Clean the filter as follows.
Step 1 Remove the filter guard from the rear panel by pull- ing out on the four panel fasteners holding them in place (Figure 1-2).
Step 2 Vacuum the honeycomb filter to clean it.
Step 3 Install the filter guard back on the rear panel.
Step 4 Press in on the panel fasteners to secure the filter guard to the rear panel.
Figure 1-2. Removing/Replacing the Fan Filter Guard
The 682XXB/683XXB contains components that can be damaged by static electricity. Figure 1-3 illustrates the precautions that should be followed when handling static-sensitive subassemblies and compo- nents. If followed, these precautions will minimize the possibilities of static-shock damage to these items.
·




Figure 1-3. Static-Sensitive Component Handling Precautions

1-11 STARTUP CONFIGURATIONS
The 682XXB/683XXB comes from the factory with a jumper across pins 2 and 3 of front panel connector J12 (Figure 1-4). In this configu- ration, connecting the instrument to line power automatically places it in operate mode (front panel OPERATE LED on).
The startup configuration can be changed so that the signal generator comes up in standby mode (front panel STANDBY LED on) when it is connected to line power. Change the startup configuration as follows:
Step 1 Disconnect the instrument from line power.
Step 2 Remove the top cover from the 682XXB/683XXB. (Refer to Chapter 6 for instructions).
Step 3 Locate front panel connector J12 and remove the jumper from across pins 2 and 3. It is located on the A2A1 PCB which plugs into the Front Panel Assem- bly.
Step 4 Install the jumper across pins 1 and 2 of connector J12.
Step 5 Install the top cover and connect the signal genera- tor to line power. The instrument should come up in standby mode.

Figure 1-4. Startup Configuration of Connector J12

1-12 RECOMMENDED TEST EQUIPMENT
Table 1-2 provides a list of recommended test equipment needed for the performance verification, calibration, and troubleshooting proce- dures presented in this manual.
Table 1-2. Recommended Test Equipment (1 of 3)
RECOMMENDED MANUFACTURER/MODEL USAGE (1)
Spectrum Analyzer, with External Mixers and Diplexer Assy
Tektronix, Model 2794, with External Mixers:
WM780K (18 to 26.5 GHz) WM780A (26.5 to 40 GHz) WM780U (40 to 60 GHz) WM780E (60 to 90 GHz) Diplexer Assy: 015-0385-00
Hewlett-Packard, Model 3585B P
Frequency Counter with Cable Kit and External Mixer
Frequency Range: 0.01 to 65 GHz Input Impedance: 50W Resolution: 1 Hz Other: External Time Base Input
EIP Microwave, Inc. Models 538B, 548B, or 578B, with Cable Kit: Option 590 and External Mixer:
Option 91 (26.5 to 40 GHz) Option 92 (40 to 60 GHz) Option 93 (60 to 90 GHz)
Power Meter, with Power Sensor
Hewlett-Packard Model 437B, with Power Sensor:
(1mW to 100mW)
HP 8487A (0.01 to 50 GHz)
Power Meter, with Power Sensor
ANRITSU ML4803A, with Power Sensor:
(1mW to 100mW)
MP716A4 (50 to 75 GHz)
Digital Multimeter Resolution: 4-1/2 digits (to 20V) DC Accuracy: 0.002% +2 counts DC Input Impedance: 10 MW AC Accuracy: 0.07% +100 counts
John Fluke, Inc., Model 8840A, with Option 8840A-09K (True RMS AC)
(to 20 kHz) AC Input Impedance: 1 MW
Absolute Time Corp., Model 300 P
Spectracom Corp., Model 8161 C
Hewlett-Packard, Model 8116A C
Hewlett-Packard, Model 33120A C

Table 1-2. Recommended Test Equipment (2 of 3)
RECOMMENDED MANUFACTURER/MODEL USAGE (1)
Hewlett-Packard, Model 8901A P
(or the IF of the Spectrum Analyzer) AM Depth: 0% to 90% AM Modulation Rates: DC to 100 kHz Filters: 20 kHz lowpass, 300 Hz highpass
Tektronix, Inc. Model TAS485 P, T
Scalar Network Analyzer, with RF Detector
560-7K50 (0.01 to 40 GHz) 560-7VA50 (0.01 to 50 GHz) SC5198 (40 to 60 GHz)
Adapter K (male) to 2.4 mm (female) Adapts the Power Sensor, HP 8487A, to the 682XXB/683XXB RF OUTPUT con- nector (£40 GHz models)
Adapter Adapts the MP716A4 Power Sensor to the ML4803A Power Meter
Adapter WR15 to V (male) Adapts the MP716A4 Power Sensor to the 682XXB/683XXB RF OUTPUT connector (>40 GHz models)
ANRITSU, Model 35WR15V C, P
ANRITSU, Model 41KC-10 C, P
ANRITSU, Model 41KC-20 P
ANRITSU, Model 41V-10 C, P
ANRITSU, Model 41V-20 P
ANRITSU, Model 75KC50 (K input/BNC output connectors)
ANRITSU, Model 75VA50 (V input/BNC output connectors)

Table 1-2. Recommended Test Equipment (3 of 3)
RECOMMENDED MANUFACTURER/MODEL USAGE (1)
ANRITSU, Model SC3800 (K input/BNC output connectors)
High Pass Filter 150 MHz High-Pass Filter (DC Block) Narda, Model 4564 P
Low Pass Filter 450 MHz Low-Pass Filter Mini-Circuits, Model SLP-450 P
Any common source C
Serial Interface Assy Provides serial interface between the PC and the 682XXB/683XXB.
ANRITSU maintains an exchange assembly program for selected 682XXB/683XXB subassemblies and RF components. If a malfunction occurs in one of these subassemblies, the defective unit can be ex- changed. Upon receiving your request, ANRITSU will ship the ex- change subassembly or RF component to you, typically within 24 hours. You then have 45 days in which to return the defective item. All exchange subassemblies or RF components are warranted for 90 days from the date of shipment, or for the balance of the original equipment warranty, whichever is longer.
Please have the exact model number and serial number of your unit available when requesting this service, as the information about your unit is filed according to the instrument’s model and serial number. For more information about the program, contact your local sales rep- resentative or call your local ANRITSU service center. Refer to Table 1-5, on page 1-18, for a list of current ANRITSU service centers.
Table 1-3 lists those replaceable subassemblies and RF components of the 682XXB/683XXB that are presently covered by the ANRITSU ex- change assembly program. Table 1-4, on page 1-16, lists common re- placeable parts for the 682XXB/683XXB that are not presently on the exchange assembly program.
All parts listed in Tables 1-3 and 1-4 may be ordered from your local ANRITSU service center.
Microwave (Pulse) Detector

Table 1-3. Replaceable Subassemblies and RF Components (1 of 2)
Printed Circuit Board Assemblies
Front Panel Assy 683XXB ND40832 Front Panel Assy 682XXB ND40833 A3 Reference Loop PCB Assy D37436-3 A4 Coarse Loop PCB Assy D40604-3 A4 Coarse Loop PCB Assy D40634-3 A5 Fine Loop PCB Assy D37452-3 A5 Fine Loop PCB Assy D40605-3 A5 Fine Loop PCB Assy (w/Option 11) ND39917 A5 Fine Loop PCB Assy D40635-3 A5 Fine Loop PCB Assy (w/Option 11) D40635-7 A6 Pulse Generator PCB Assy D40606-3 A6 Pulse Generator PCB Assy D40626-3 A7 YIG Loop PCB Assy D40607-3 A7 YIG Loop PCB Assy D40637-3 A8 Function Generator PCB Assy D37408-3 A8 Function Generator PCB Assy D40608-3 A9 PIN Control PCB Assy D40609-3 A9 PIN Control PCB Assy D40629-3 A9 PIN Control PCB Assy D40639-3 A10 ALC PCB Assy D40610-3 A11 FM PCB Assy D40611-3 A11 FM PCB Assy D40641-3 A11 FM PCB Assy (w/Option 6) D40641-4 A12 Analog Instruction PCB Assy D37448-3 A12 Analog Instruction PCB Assy (682XXB) D40612-3 A13 YIG Driver PCB Assy D40613-3 A14 SDM Driver PCB Assy, 26.5 GHz D40614-4 A14 SDM, SQM Driver PCB Assy D40614-3 A15 Regulator PCB Assy D40615-3 A16 CPU Interface PCB Assy D37416-3 A17 CPU PCB Assy D37444-3 A18 Power Supply PCB Assy D40618-3 A19 Line Conditioner PCB Assy D40619-3 A21 Line Filter/Rectifier PCB Assy ND39918 A21-1 BNC/AUX I/O Connector PCB Assy ND39919 10 MHz Crystal Oscillator Assy D37332

Table 1-3. Replaceable Subassemblies and RF Components (2 of 2)
YIG-Tuned Oscillator, 2 to 20 GHz C27327 Down Converter D27330 Pulsed Digital Down Converter D27370 Switched Doubler Module, 20 to 26.5 GHz ND40843 Switched Doubler Module, 20 to 40 GHz D28535 Source Quadrupler Module, 40 to 50 GHz D28185 Source Quadrupler Module, 40 to 60 GHz 60-141 Source Quadrupler Module, 40 to 65 GHz 60-142 Forward Coupler, 60 GHz and 65 GHz C27184 Coupler, 40 GHz D27115 Coupler, 60 GHz D27350 Pulsed Switched Filter D25015 Pulsed Switched Filter D29391 Pulsed Switched Filter (w/Option 15B) D26350 Pulsed Switched Filter (w/Option 15B) D29392 Output Connector Assy, K female ND39077 Output Connector Assy, V female ND40835 Step Attenuator, 110 dB, 26.5 GHz D27152 Step Attenuator, 110 dB, 40 GHz D25080 Step Attenuator, 90 dB, 50 GHz D27315 Step Attenuator, 90 dB, 60 GHz D28957
Table 1-4. Common Replaceable Subassemblies and Parts (1 of 2)
Cap, Protective (for RF Output Connector) A20304 Cover, Top D37131 Cover, Bottom D37135 Cover, Side D37133 Cover, Side Handle 761-67
Cover, CPU Housing C37063 Cover, Main Card Cage D37064 Cover, Power Supply Housing C37062
EMI Gasket for side covers 790-437 EMI Gasket for side covers 790-390 EMI Gasket for Front Panel Assy 790-223
RF Components

Table 1-4. Common Replaceable Subassemblies and Parts (2 of 2)
Fan Assembly A40513 Fan Mount 790-425 Fan Membrane (Honey Comb Filter) C37137 Fan Grill 790-426 Fasteners (for Fan Grill) 790-433
Fuse, 5A, 3AG Slow Blow (110 Vac) 631-33 Fuse, 2.5A, 3AG Slow Blow (230 Vac) 631-14 Gasket, RFI (“O”rings for MCX connectors) 790-393
Handle, Side Carrying 783-830 Screw, Handle Side Carrying 900-714
Line Module 260-11 Shield Cover D37228 Shield D37229 Standoff 785-922 Tape 850-70
Foot, Rear Bottom Left 2000-548 Foot, Rear Bottom Right 2000-549 Foot, Rear Top Left 2000-552 Foot, Rear Top Right 2000-553 Screw, Green Head 2000-560
682XXB/683XXB without Front Handles Installed
Foot, Front Bottom Left 2000-546 Foot, Front Bottom Right 2000-547 Foot, Front Top Left 2000-550 Foot, Front Top Right 2000-551
682XXB/683XXB with Front Handles Installed
Upper Insert B37147 Foot, Bottom Left C37170 Foot,Bottom Right C37171 Handle, Left D37168-2 Handle, Right D37169-2 Tilt Bail 790-435

Table 1-5. ANRITSU Service Centers
UNITED STATES ANRITSU COMPANY 490 Jarvis Drive Morgan Hill, CA 95037-2809 Telephone: (408) 776-8300 1-800-ANRITSU FAX: 408-776-1744
FRANCE ANRITSU S.A 9 Avenue du Quebec Zone de Courtaboeuf 91951 Les Ulis Cedex Telephone: 016-09-21-550 FAX: 016-44-61-065
JAPAN ANRITSU CUSTOMER SERVICE LTD. 1800 Onna Atsugi-shi Kanagawa-Prf. 243 Japan Telephone: 0462-96-6688 FAX: 0462-25-8379
GERMANY ANRITSU GmbH Grafenberger Allee 54-56 D-40237 Dusseldorf, Germany Telephone: 0211-968550 FAX: 0211-9685555
SINGAPORE ANRITSU (SINGAPORE) PTE LTD. 10, Hoe Chiang Road #07-01/02 Keppel Towers Singapore 089315 Telephone: 282-2400 FAX: 282-2533
ANRITSU COMPANY 10 New Maple Ave., Unit 305 Pine Brook, NJ 07058 Telephone: (973) 227-8999 1-800-ANRITSU FAX: 973-575-0092
INDIA MEERA AGENCIES PVT. LTD. 23 Community Centre Zamroodpur, Kailash Colony Extension, New Delhi, India 110 048 Phone: 011-2-6442700/6442800 FAX : 011-2-644250023
SOUTH AFRICA ETECSA 12 Surrey Square Office Park 330 Surrey Avenue Ferndale, Randburt, 2194 South Africa Telephone: 011-27-11-787-7200 FAX: 011-27-11-787-0446
ANRITSU COMPANY 1155 E. Collins Blvd Richardson, TX 75081 Telephone: 1-800-ANRITSU FAX: 972-671-1877
AUSTRALIA ANRITSU PTY. LTD. Unit 3, 170 Foster Road Mt Waverley, VIC 3149 Australia Telephone: 03-9558-8177 FAX: 03-9558-8255
ISRAEL TECH-CENT, LTD. 4 Raul Valenberg St Tel-Aviv 69719 Telephone: (03) 64-78-563 FAX: (03) 64-78-334
SWEDEN ANRITSU AB Botivid Center Fittja Backe 13A 145 84 Stockholmn Telephone: (08) 534-707-00 FAX: (08) 534-707-30
BRAZIL ANRITSU ELECTRONICA LTDA. Praia de Botafogo, 440, Sala 2401 CEP22250-040, Rio de Janeiro, RJ, Brasil Telephone: 021-527-6922 FAX: 021-53-71-456
ITALY ANRITSU Sp.A Roma Office Via E. Vittorini, 129 00144 Roma EUR Telephone: (06) 50-99-711 FAX: (06) 50-22-4252
TAIWAN ANRITSU CO., INC. 7F, No. 316, Section 1 NeiHu Road Taipei, Taiwan, R.O.C. Telephone: 886-2-8751-1816 FAX: 886-2-8751-2126
CANADA ANRITSU INSTRUMENTS LTD. 700 Silver Seven Road, Suite 120 Kanata, Ontario K2V 1C3 Telephone: (613) 591-2003 FAX: (613) 591-1006
KOREA ANRITSU CORPORATION LTD. Head Office: 14F, Hyunjuk Building 832-41 Yeoksam-Dong, Kangnam-Ku Seoul 135-080, South Korea Telephone: 02-553-6603 FAX: 02-553-6604
UNITED KINGDOM ANRITSU LTD. 200 Capability Green Luton, Bedfordshire LU1 3LU, England Telephone: 015-82-433200 FAX: 015-82-731303
CHINA ANRITSU ELECTRONICS (SHANGHAI) CO. LTD. 2F, Rm B, 52 Section Factory Building No. 516 Fu Te Rd (N) Shanghai 200131 P.R. China Telephone:21-58680226, 58680227, 58680228 FAX: 21-58680588
Service Center: 8F Hyunjuk Building 832-41, Yeoksam Dong Kangnam-Gu Seoul, South Korea 135-080 Telephone: 82-2-553-6603 FAX: 82-2-553-6605

Table of Contents
2-1 INTRODUCTION 2-3
2-2 682XXB/683XXB MAJOR SUBSYSTEMS 2-3 Digital Control 2-3 Front Panel 2-4 Frequency Synthesis 2-4 Analog Instruction 2-5 YIG Driver 2-5 Function Generator 2-5 ALC/AM/Pulse Modulation 2-8 RF Deck 2-8 Power Supply 2-8 Inputs/Outputs 2-8 Motherboard/Interconnections 2-9
2-3 FREQUENCY SYNTHESIS 2-9
Phase-Lock Loops 2-9 Overall Operation 2-10 RF Outputs 0.01 to 65 GHz 2-13 Frequency Modulation 2-14 Phase Modulation (Option 6) 2-14 Analog Sweep Mode (683XXB only) 2-14 Step Sweep Mode 2-15
2-4 ALC/AM/PULSE MODULATION 2-15
ALC Loop Operation 2-15 Pulse Generator Operation 2-18
2-5 RF DECK ASSEMBLIES 2-19
RF Deck Configurations 2-20 YIG-tuned Oscillator 2-20 Power Level Control and Modulation 2-23 RF Signal Filtering 2-23

0.01 to 2 GHz Down Converter 2-24 0.5 to 2.2 GHz Digital Down Converter 2-25 Switched Doubler Module 2-26 Source Quadrupler Module 2-29 Power Level Detection/ALC Loop 2-30 Step Attenuator 2-31

2-1 INTRODUCTION This chapter provides brief functional descriptions of the major sub- systems that are contained in each model of the Series 682XXB/ 683XXB Synthesized Signal Generators. In addition, the operation of the frequency synthesis, automatic level control (ALC), and RF deck subsystems is described so that the reader may better understand the overall operation of the instrument. Block diagrams are included to supplement the written descriptions.
The 682XXB/683XXB circuitry consists of various distinct subsystems that are contained on one or more printed circuit board (PCB) assem- blies or in microwave components located on the RF deck. The follow- ing paragraphs identify the subsystems that make up the instrument and provide a brief description of each. Figure 2-1 (page 2-6) is an overall block diagram of a typical 682XXB/683XXB.
This circuit subsystem consists of the A17 CPU and A16 CPU Interface PCBs. The central processor unit (CPU) is the main controller for the 682XXB/ 683XXB. This controller directly or indirectly con- trols all functions of the instrument. The CPU con- tains memory that stores the main operating system components and instrument firmware, instrument calibration data, and front panel setups in the power-off condition. It has a GPIB interface that al- lows it to communicate with external devices over the GPIB and it has a serial interface to a serial ter- minal port on the rear panel. The CPU is directly linked via a dedicated data and address bus to the A2 Front Panel PCB, the A8 Function Generator PCB, the A9 PIN Control PCB, the A10 ALC PCB, the A11 FM PCB, the A12 Analog Instruction PCB, and the A16 CPU Interface PCB.

The CPU is indirectly linked via the A16 CPU Inter- face PCB to the A3 Reference Loop PCB, the A4 Coarse Loop PCB, the A5 Fine Loop PCB, and the A6 Pulse Generator PCB. The A16 PCB contains cir- cuitry to perform parallel-to-serial and serial-to- parallel data conversion. It also contains circuitry for many of the rear panel signals, a 13-bit resolu- tion DVM, and decoder circuitry for the front panel rotary data knob optical encoder.
Front Panel This circuit subsystem consists of the the A1 Front Panel PCB, the A2 Front Panel Control PCB, and the Liquid Crystal Display (LCD). This subsystem interfaces the front panel LCD, LEDs, and keys to the CPU via the dedicated data and address bus. The front panel rotary data knob is indirectly linked to the CPU via the A16 CPU Interface PCB.
The A1 Front Panel PCB contains the keyboard ma- trix of conductive rubber switches. The A2 Front Panel Control PCB has circuitry to control the LCD dot-matrix display, turn the front panel LEDs on and off, and convert keyboard switch matrix signals to parallel keycode. It also contains the standby/op- erate line switch logic circuit and the optical encoder for the rotary data knob.
The frequency synthesis subsystem consists of the A3 Reference Loop PCB, the A4 Coarse Loop PCB, the A5 Fine Loop PCB, the A7 YIG Loop PCB, and the A11 FM PCB. It provides the reference frequen- cies and phase lock circuits for precise control of the YIG-tuned oscillator frequencies, as follows:
q The A3 Reference Loop PCB supplies the sta-
ble 10 MHz and 500 MHz reference frequency signals for the rest of the frequency synthesis system. q The A4 Coarse Loop PCB generates coarse
tuning frequencies of 219.5 to 245 MHz for use by the YIG Loop. q The A5 Fine Loop PCB provides fine tuning
frequencies of 21.5 to 40 MHz for use by the YIG Loop. q The A7 YIG Loop PCB performs phase detec-
tion of the YIG-tuned oscillator’s output fre- quency and provides a YIG loop error voltage to the A11 FM PCB. q The A11 FM PCB processes the YIG loop error
voltage, producing a correction signal that is

used to fine tune and phase lock the YIG-tuned oscillator. The A11 PCB also contains circuitry for frequency modulation of the YIG-tuned os- cillator RF output.
The CPU sends control data to the A3 Reference Loop PCB, the A4 Coarse Loop PCB, and the A5 Fine Loop PCB via the A16 PCB as serial data words. The CPU controls the A11 FM PCB via the dedicated data and address bus. Refer to paragraph 2-3 for a functional overview of the frequency syn- thesis subsystem.
The A12 Analog Instruction PCB provides the fre- quency tuning voltages and frequency band select signals to the A13 YIG Driver PCB. For models with a frequency range greater than 20 GHz, it supplies frequency band select signals to the A14 SDM, SQM Driver PCB. In addition, it provides a 0V to +10V ramp signal to the rear panel HORIZ OUT connector, a V/GHz signal to the rear panel V/GHz OUT connec- tor, and a SLOPE signal to the A10 ALC PCB for slope-vs-frequency correction of the RF output power. The CPU controls the A12 Analog Instruc- tion PCB via the dedicated data and address bus.
YIG Driver The A13 YIG Driver PCB supplies the main tuning current and bias voltages for the 2 to 20 GHz YIG- tuned oscillator. It also provides bias voltages for the Down Converter assembly and the amplifiers lo- cated in the Switched Filter assembly.
The A12 Analog Instruction PCB provides (1) fre- quency tuning voltages for the main tuning coil driver of the YIG-tuned oscillator and (2) frequency band select signals to activate the appropriate bias voltage supply.
The A8 Function Generator PCB provides internally generated waveforms for amplitude and frequency modulation. The waveforms available are sine wave, square wave, triangle wave, positive ramp, negative ramp, Gaussian noise, and uniform noise. The AM waveforms are supplied to the A10 ALC PCB and the rear panel AM OUT connector; the FM wave- forms are provided to the A11 FM PCB and the rear panel FM OUT connector. The CPU controls the A8 Function Generator PCB via the dedicated data and address bus.


Outputs Inputs Power Input
Figure 2-1. Block Diagram of a Typical 682XXB/683XXB Synthesized Signal Generator (Sheet 1 of 2)
Rear Panel Pulse Signals
Data and Address Bus
A21 Rear Panel PCB
-165V A18 Power Supply A15 Regulator
Power Supply
+24V (From A2 PCB) +24V (To A2 PCB)
Digital Control
-15V +24V -18V -43V
+5V +9V +15V
(Continued on Sheet 2)
D
B


AM IN Front Panel BNC Internal AM Data and Address Bus
(Continued From Sheet 1)
Step Attenuator Control Bias Switch Control
0.01 - 40 GHz
Figure 2-1. Block Diagram of a Typical 682XXB/683XXB Synthesized Signal Generator (Sheet 2 of 2)
110 dB Step Attenuator (Option)
RF Output 0.01 - 40 GHz
Serial Data · Directional Coupler

This circuit subsystem consists of the A6 Pulse Gen- erator PCB, the A9 PIN Control PCB, the A10 ALC PCB, and the A14 SDM, SQM Driver PCB. It pro- vides the following:
q Level control of the RF output power. q AM and pulse modulation of the RF output. q Current drive signals to the PIN switches lo-
cated in the Switched Filter assembly and Switched Doubler Module (SDM). q Bias voltages for the SDM and Source Quadru-
pler Module (SQM). q Drive signals for the optional Step Attenuator.
The CPU controls the A9 Pin Control PCB and the A10 ALC PCB via the dedicated data and address bus. It sends control data to the A6 Pulse Generator PCB via the A16 PCB as serial data words. Refer to paragraph 2-4 for a functional overview of the ALC and modulation subsystem.
RF Deck This subsystem contains those elements related to the generation, modulation, and control of the sweep- and CW-frequency RF signals. These ele- ments include; the 2 to 20 GHz YIG-tuned oscillator, the 0.01 to 2 GHz (0.5 to 2.2 GHz) Down Converter assembly, the Switched Filter assembly, the Switched Doubler Module (SDM), the Source Quad- rupler Module (SQM), the Directional Coupler/Level Detector, and the optional 110 dB (90 dB) Step Attenuator. Refer to paragraph 2-5 for a functional overview of the RF deck subsystem.
Power Supply The power supply subsystem consists of the A15 Regulator PCB, the A18 Power Supply PCB, the A19 Line Conditioner PCB, and part of the A21 Rear Panel PCB and Rear Casting Assembly. It sup- plies all the regulated DC voltages used by the sig- nal generator circuits. The voltages are routed throughout the instrument via the A20 Mother- board PCB.
The A21-1 BNC/AUX I/O Connector PCB and the A16 CPU Interface PCB contain the interface cir- cuitry for the majority of the rear panel input and output connectors, including the AUX I/O connector.
The front panel AM and External ALC inputs are routed via the A20 Motherboard PCB to the A10 ALC PCB. The front panel FM and Pulse Trigger

inputs are coupled directly via coaxial cables to their destination PCBs—the FM input to A11 FM PCB and the Pulse/Trigger input to the A6 Pulse Generator PCB.
The rear panel connectors, 10 MHz REF OUT, 10 MHz REF IN, and FM IN, are coupled directly to PCBs via coaxial cables—10 MHz REF OUT and 10 MHz REF IN to the A3 Reference Loop PCB and FM IN to the A11 FM PCB. The rear panel connec- tors, PULSE TRIGGER IN, PULSE SYNC OUT, and PULSE VIDEO OUT, are coupled via the A21-1 PCB and coaxial cables directly to the A6 Pulse Genera- tor PCB. The rear panel IEEE-488 GPIB and SERIAL I/O connectors are connected to the A17 CPU PCB via the A20 Motherboard PCB.
The A20 Motherboard PCB and associated cables provide the interconnections for the flow of data, signals, and DC voltages between all internal com- ponents and assemblies throughout the 682XXB/ 683XXB.
2-3 FREQUENCY SYNTHESIS The frequency synthesis subsystem provides phase-lock control of the 682XXB/683XXB output frequency. It consists of four phase-lock loops, the Reference Loop, the Coarse Loop, the Fine Loop, and the YIG Loop. The four phase-lock loops, operating together, produce an accu- rately synthesized, low-noise RF output signal. Figure 2-2 (page 2-11) is an overall block diagram of the frequency synthesis subsystem. The following paragraphs describe phase-lock loops and the overall opera- tion of the frequency synthesis subsystem.
The purpose of a phase-lock loop is to control the frequency of a variable oscillator in order to give it the same accuracy and stability as a fixed reference oscillator. It works by comparing two frequency in- puts, one fixed and one variable, and by supplying a correction signal to the variable oscillator to reduce the difference between the two inputs. For example, suppose we have a 10 MHz reference oscillator with a stability of 1 x 10–7/day, and we wish to transfer that stability to a voltage controlled oscillator (VCO). The 10 MHz reference signal is applied to the reference input of a phase-lock loop circuit. The signal from the VCO is applied to the variable input. A phase detector in the phase-lock loop circuit com- pares the two inputs and determines whether the variable input waveform is leading or lagging the reference. The phase detector generates a correction

signal that (depending on polarity) causes the VCO frequency to increase or decrease to reduce any phase difference. When the two inputs match, the loop is said to be locked. The variable input from the VCO then equals the reference input in phase, fre- quency, accuracy, and stability.
In practical applications a frequency divider is placed between the output of the variable oscillator and the variable input to the phase-lock loop. The circuit can then be used to control a frequency that is an exact multiple of the reference frequency. In this way, the variable oscillator acquires the stabil- ity of the reference without equaling its frequency. In the A3 Reference Loop, the 100 MHz oven- controlled crystal oscillator (OCXO) can be con- trolled by the phase-lock loop using a 10 MHz refer- ence. This is because a divide-by-ten circuit is between the OCXO’s output and the variable input to the phase-lock loop. Both inputs to the phase de- tector will be 10 MHz when the loop is locked.
If a programmable frequency divider is used, a number of frequencies can be phase-locked to the same reference. The limitation is that all must be exact multiples of the reference. The A4 Coarse Loop and A5 Fine Loop both use programmable fre- quency dividers.
The YIG-tuned oscillator generates a high-power RF output signal that has low broadband noise and low spurious content. The frequency of the YIG-tuned oscillator is controlled by means of (1) its main tun- ing coil and (2) its FM (fine tuning) coil. Main tun- ing coil current from the YIG Driver PCB coarsely tunes the YIG-tuned oscillator to within a few megahertz of the final output frequency. The YIG phase-lock loop is then used to fine tune the YIG- tuned oscillator to the exact output frequency and to reduce FM noise close to the carrier.
One input to the YIG Loop is the 219.5 to 245 MHz signal from the Coarse Loop. This signal is ampli- fied to drive the step-recovery diode. The step- recovery diode produces harmonics of the coarse loop signal (1.9755 to >20 GHz). These harmonics are used by the sampler.
The other input to the sampler is the RF output sig- nal from the YIG-tuned oscillator. Mixing this RF

10 MHz REF In

FM Sweep (From A12 PCB)
Internal FM (From A8 PCB)
FM In (Front Panel or Rear Panel)
Tune (From A12 PCB)

21.5 - 40 MHz Frequency Divider
26.84 MHz VCXO
219.5 - 245 MHz
10 MHz A5 Fine Loop
1.9755 to 20 GHz Harmonics
Sampled 2 - 20 GHz RF
21.5 - 40 MHz
21.5 - 40 MHz IF
Figure 2-2. Block Diagram of the Frequency Synthesis Subsystem

output signal sample with the adjacent coarse-loop harmonic produces a low frequency difference signal that is the YIG IF signal (21.5 to 40 MHz).
The 682XXB/683XXB CPU programs the coarse-loop oscillator’s output frequency so that one of its har- monics will be within 21.5 to 40 MHz of the desired YIG-tuned oscillator’s output frequency. The YIG Loop phase detector compares the YIG IF signal to the 21.5 to 40 MHz frequency reference signal from the Fine Loop. If there is a difference, the YIG phase detector fine tunes the YIG-tuned oscillator (via the FM circuitry and the FM coil driver) to eliminate any frequency difference between the two signals.
Table 2-1. RF Output and Loop Frequencies
Phase locking the instrument’s output frequency over a broad frequency range is accomplished by programming the coarse-loop oscillator’s output to various frequencies that have harmonics close to the desired operating frequencies. Exact frequency tun- ing for each desired operating frequency is accom- plished by programming the fine-loop oscillator. (In each case, the YIG-tuned oscillator is first tuned via the main tuning coil to the approximate desired op- erating frequency.) Table 2-1 shows the coarse-loop and fine-loop frequencies for some specific RF out- put frequencies.
2000 219.5 24.5
7000 240.5 25.5
8000 241.5 30.5
9000 242.5 27.5
The coarse-loop oscillator has a programming (tun- ing) range of 219.5 to 245 MHz and a resolution of 1 MHz. This provides harmonics from 1.9755 GHz to >20 GHz. This allows any YIG-tuned oscillator output frequency to be down converted to a YIG IF signal of 21.5 to 40 MHz.
11000 238.5 29
12000 239.5 25
The YIG Loop is fine tuned by varying the 21.5 to 40 MHz reference frequency signal applied to the YIG loop phase detector. By programming the fine- loop oscillator, this signal can be adjusted in 1 kHz increments over the 21.5 to 40 MHz range. The reso- lution of the fine-loop oscillator (hence the resolu- tion of the RF output signal) is 1 kHz, which is much finer than is available from the coarse loop alone. For applications requiring a resolution finer than 1 kHz, an optional tuning resolution of 0.1 Hz is available.
The Coarse Loop and Fine Loop outputs are derived from high-stability 10 MHz and 500 MHz signals generated by the Reference Loop. For applications

requiring even greater stability, the 100 MHz refer- ence oscillator can be phase locked to an optional 10 MHz reference (internal or external).
RF Outputs 0.01 to 65 GHz Refer to the RF Deck block diagram shown in Fig- ure 2-1 (page 2-7) for the following description. The 682XXB/683XXB uses one 2 to 20 GHz YIG-tuned oscillator. All other frequencies output by the signal generator are derived from the fundamental fre- quencies generated by this YIG-tuned oscillator.
0.01 to 2 GHz RF output frequencies of 0.01 to 2 GHz are gener- ated by down converting the fundamental frequen- cies of 6.51 to 8.5 GHz. This is achieved using a 6.5 GHz local oscillator signal that is phase locked to the 500 MHz output of the Reference Loop. Pre- cise control of the 0.01 to 2 GHz frequencies to 1 kHz (0.1 Hz with Option 11) accuracy is accom- plished by phase-lock control of the 6.51 to 8.5 GHz fundamental frequencies prior to down conversion.
20 to 40 GHz RF output frequencies of 20 to 40 GHz are produced by doubling the 10 to 20 GHz fundamental frequen- cies. Phase-lock control of the 10 to 20 GHz funda- mental frequencies—accomplished prior to doubling —ensures precise control of the 20 to 40 GHz fre- quencies to 1 kHz (0.1 Hz with Option 11) resolu- tion.
40 to 65 GHz RF output frequencies of 40 to 65 GHz are devel- oped by quadrupling the 10 to 16.25 GHz fundamen- tal frequencies (refer to Figure 2-5, page 2-22). Precise control of the 40 to 65 GHz to 1 kHz (0.1 Hz with Option 11) resolution is achieved by phase-lock control of the 10 to 16.25 GHz fundamental frequen- cies prior to quadrupling.
0.5 to 2.2 GHz RF output frequencies of 0.5 to 2.2 GHz for the 682X5B/683X5B models are developed by down con- verting the fundamental frequencies of 2 to 4.4 GHz (refer to Figure 2-6, page 2-27). Phase-lock control of the 2 to 4.4 GHz fundamental frequencies, achieved prior to down converting, ensures precise control of the 0.5 to 2.2 GHz frequencies to 1 kHz (0.1 Hz with Option 11) resolution.

Frequency modulation (FM) of the YIG-tuned oscil- lator RF output is achieved by summing an external or internal modulating signal into the FM control path of the YIG loop. Refer to Figures 2-1 and 2-2. The external modulating signal comes from the front panel or rear panel FM IN input; the internal modulating signal comes from the A8 Function Gen- erator PCB. Circuits on the A11 FM PCB adjust the modulating signal for the FM sensitivity selected, then sum it into the YIG loop FM control path. There, it frequency modulates the RF output signal by controlling the YIG-tuned oscillator’s FM (fine tuning) coil current.
Phase modulation (M) of the YIG-tuned oscillator RF output is achieved by summing an external or internal modulating signal into the FM control path of the YIG loop. The external modulating signal comes from the front panel or rear panel FM IN/M IN input; the internal modulating signal comes from the A8 Function Generator PCB. Circuits on the A11 FM PCB adjust the modulating signal for the M sensitivity selected, convert the modulating sig- nal to a M signal by differentiation, and then sum it into the YIG loop FM control path. There, it phase modulates the RF output signal by controlling the YIG-tuned oscillator’s FM (fine tuning) coil current.
Broad-band analog frequency sweeps (>100 MHz wide) of the YIG-tuned oscillator RF output are ac- complished by applying appropriate analog sweep ramp signals, generated by the A12 Analog Instruc- tion PCB, to the YIG-tuned oscillator’s main tuning coil (via the A13 YIG Driver PCB). In this mode, the start, stop, and bandswitching frequencies are phase-lock-corrected during the sweep.
Narrow-band analog frequency sweeps (100 MHz wide) of the YIG-tuned oscillator RF output are ac- complished by summing appropriate analog sweep ramp signals, generated by the A12 Analog Instruc- tion PCB, into the YIG-tuned oscillator’s FM tuning coil control path. The YIG-tuned oscillator’s RF out- put is then swept about a center frequency. The cen- ter frequency is set by applying a tuning signal (also from the A12 PCB) to the YIG-tuned oscillator’s main tuning coil (via the A13 YIG Driver PCB). In this mode, YIG loop phase locking is disabled except during center frequency correction, which occurs during sweep retrace.

Step (digital) frequency sweeps of the YIG-tuned os- cillator RF output consist of a series of discrete, syn- thesized steps between a start and stop frequency. Each frequency step is generated by applying the tuning signal (from the A12 Analog Instruction PCB) to the YIG-tuned oscillator’s main tuning coil, then phase-locking the RF output. Every frequency step in the sweep range is phase-locked.
The ALC, AM, and pulse modulation subsystem provides automatic level control (ALC), amplitude modulation (AM), and pulse modulation of the signal generator’s RF output signal. The ALC loop consists of circuits located on the A10 ALC PCB, the A9 PIN Control PCB, and the A14 SDM, SQM Driver PCB. These circuits interface with the Switched Filter assembly, the Down Converter assembly, the Source Quadrupler Module (SQM), and the Directional Coupler/Level Detec- tor (all located on the RF deck). AM modulation circuits (located on the A10 ALC PCB) are included in this loop.
Pulse modulation of the RF output signal is provided by circuits lo- cated on the A6 Pulse Generator PCB. These circuits interface directly with the Switched Filter assembly located on the RF deck. (In 682X5B/ 683X5B models, these circuits interface directly to the Digital Down Converter assembly and from there to the Switched Filter assembly.) The overall block diagram of the ALC and modulation subsystem is shown in Figure 2-3, page 2-17. The following paragraphs describe the operation of the subsystem components.
In the 682XXB/683XXB, a portion of the RF output is detected and coupled out of the Directional Cou- pler/Level Detector as the feedback input to the ALC loop. The feedback signal from the detector is routed to the A10 ALC PCB where it is compared with a reference voltage that represents the desired RF power output level. If the two voltages do not match, an error correction signal is fed from the A10 ALC PCB to the modulator shaper amplifier circuits located on the A9 PIN Control PCB and the A14 SDM, SQM Driver PCB, if installed. The resulting modulator control voltage output causes the modu- lators, located in the Switched Filter assembly, the SQM, and the Digital Down Converter assembly (682X5B/683X5B models only), to adjust the RF out- put level. Thus, the feedback signal from the detec- tor will be set equal to the reference voltage.

NOTE The instrument uses two internal level de- tection circuits. For frequencies 2.2 GHz for 682X5B/683X5B models), the level de- tector is part of the main Directional Cou- pler. The signal from this detector is routed to the A10 ALC PCB as the Detector 1 in- put.
The Level Reference DAC, under the control of the CPU, provides the RF level reference voltage. By setting the output of this DAC to the appropriate voltage, the CPU adjusts the RF output power to the level selected by the user. Leveled output power can be set over a maximum range of up to 28 dB (up to 131 dB with the optional 110 dB step attenuator) using front panel controls or the GPIB. Instruments with Option 15B (High Power) provide leveled out- put power over a maximum range of up to 22 dB (up to 125 dB with the optional 110 dB step attenuator).
External Leveling In the external leveling mode, an external detector or power meter monitors the RF output level of the 682XXB/683XXB instead of an internal level detec- tor. The signal from the external detector or power meter goes to the A10 ALC PCB from the front or rear panel inputs. The ALC controls the RF power output level as previously described.
ALC Slope During analog sweeps (683XXB only), a slope-vs- frequency signal, from the A12 Analog Instruction PCB, is summed with the level reference and detec- tor inputs into the ALC loop. The Slope DAC, under the control of the CPU, adjusts this ALC slope sig- nal to compensate for an increasing or decreasing output power-vs-frequency characteristic caused by the level detectors and (optional) step attenuator. In addition, the Slope DAC lets the user adjust for the slope-vs-frequency characteristics of external com- ponents.

FUNCTIONAL ALC/AM/PULSE DESCRIPTION MODULATION
ALC Slope (From A12 PCB)
Internal AM (From A8 PCB)
External ALC (From Front / Rear Panel)
External AM (From Front / Rear Panel)

10 MHz (From A5 PCB) Pulse Generator External Pulse Inputs (From Front / Rear Panel)
D0 - D15 L. SEL 2 A01 - A02 (From CPU)
Address / Data Latches
Level Detector Switched Filter
0.01 - 2 GHz Down Converter
0.01 - 20 GHz
Source Quadrupler Module 0.01 - 65 GHz
Figure 2-3. Block Diagram of the ALC/AM/Pulse Modulation Subsystem
0.01 - 40 GHz
To Step Attenuator or RF Output

Power Sweep In this mode, the CPU has the ALC step the RF out- put through a range of levels specified by the user. This feature can be used in conjunction with the sweep mode to produce a set of identical frequency sweeps, each with a different RF power output level.
Amplitude Modulation Amplitude modulation (AM) of the RF output signal is accomplished by summing an external or internal modulating signal into the ALC loop. External modulating signals come from the front panel or rear panel AM IN inputs; the internal modulating signal comes from the A8 Function Generator PCB. The AM Input Sensitivity DAC and the AM Calibra- tion DAC, under the control of the CPU, adjust the modulating signal for the proper amount of AM in both the linear (log amp in) and the log (log amp by- passed) modes of operation. The adjusted modulat- ing signal is summed with the level reference, slope, and detector inputs into the ALC loop. This pro- duces an ALC control signal that varies with the modulating signal. The action of the ALC loop then causes the envelope of the RF output signal to track the modulation signal.
Pulse Modulation Operation During pulse modulation, the ALC level amplifier (A10 ALC PCB) is operated as a sample/hold ampli- fier. The level amplifier is synchronized with the modulating pulses from the A6 Pulse Generator PCB so that the ALC loop effectively operates only during the ON portion of the pulsed modulated RF output.
The A6 Pulse Generator PCB provides the internal pulse generating function for the 682XXB/683XXB. It also interfaces external pulse inputs from the front or rear panel connectors to the pulse modula- tor driver in the external mode.
The pulse generator produces a pulse modulation waveform consisting of single, doublet, triplet, or quadruplet pulse trains at variable pulse rates, widths, and delays. It operates at two selectable clock rates—10 MHz and 40 MHz. In addition, the pulse generator produces a sync pulse and video pulse output that goes to the rear panel and a sam- ple/hold signal that goes to the A10 ALC PCB. The sync pulse output is for synchronizing auxiliary in-

struments to the internally generated pulse; the video pulse is a TTL level copy of the RF output pulse; and the sample/hold signal synchronizes the ALC loop to the ON portion of the pulse modulating waveform.
The 682XXB/683XXB has five pulse modulation modes:
q Internal pulse modulation mode – The pulse
modulation waveform is generated and timed internally. q External pulse mode – The external pulse
source signal from the front or rear panel con- nectors is interfaced by the pulse generator to the pulse modulation driver. q External triggered mode – The pulse generator
is triggered by the external pulse source signal to produce the pulse modulation waveform. q External gated mode – The external pulse
source signal gates the internal pulse genera- tor on and off. q Composite mode – The external pulse source
signal triggers the internal pulse generator and also modulates the RF output signal. The pulse generator produces a delayed, single pulse waveform that also modulates the RF output signal.
2-5 RF DECK ASSEMBLIES The primary purpose of the RF deck assembly is to generate CW and swept frequency RF signals and route these signals to the front panel RF OUTPUT connector. It is capable of generating RF signals in the frequency range of 0.01 to 65 GHz.
The series 682XXB/683XXB synthesized signal generators use a single 2 to 20 GHz YIG-tuned oscillator. All other frequencies are derived from the fundamental frequencies generated by this oscillator, as fol- lows:
q RF output frequencies of 0.01 to 2 GHz are developed by down-
converting the fundamental frequencies of 6.51 to 8.5 GHz. q RF output frequencies of 0.5 to 2.2 GHz are developed by down-
converting the fundamental frequencies of 2 to 4.4 GHz. q RF output frequencies of 20 to 40 GHz are produced by doubling
the fundamental frequencies of 10 to 20 GHz. q RF output frequencies of 40 to 65 GHz are produced by quadru-
pling the fundamental frequencies of 10 to 16.25 GHz.
The following paragraphs briefly describe the operation of the RF deck assembly.

All 682XXB/683XXB RF deck assemblies contain a 2 to 20 GHz YIG-tuned oscillator, a switched filter as- sembly, and a directional coupler. Beyond that, the configuration of the RF deck assembly varies accord- ing to the particular instrument model. Block dia- grams of the various RF deck configurations are shown in the following figures:
q Figure 2-4, page 2-21, is a block diagram of the
RF deck assembly for all 40 GHz models ex- cept for 682X5B/683X5B models. q Figure 2-5, page 2-22, is a block diagram of the
RF deck assembly for all >40 GHz models ex- cept for 682X5B/683X5B models. q Figure 2-6, page 2-27, is a block diagram of the
RF deck assembly for all 40 GHz 682X5B/ 683X5B models. q Figure 2-7, page 2-28, is a block diagram of the
RF deck assembly for all >40 GHz 682X5B/ 683X5B models.
The block diagram of the RF deck shown in Figure 2-4 (page 2-21) includes all of the common RF com- ponents found in the 682XXB/683XXB RF deck as- semblies. Refer to this block diagram during the descriptions of RF deck operation presented in the following paragraphs.
The 2 to 20 GHz YIG-tuned oscillator actually con- tains two oscillators—one covering the frequency range of 2 to 8.4 GHz and one covering the fre- quency range of 8.4 to 20 GHz. Both oscillators use a common internal amplifier.
The YIG-tuned oscillator generates RF output sig- nals that have low broadband noise and low spuri- ous content. It is driven by the Main tuning coil current and bias voltages from the A13 YIG Driver PCB and the FM tuning coil current from the A11 FM PCB. During CW mode, the main tuning coil current tunes the oscillator to within a few mega- hertz of the final output frequency. The phase-lock circuitry of the YIG loop then fine adjusts the oscil- lator’s FM tuning coil current to make the output frequency exact. Frequency modulation of the RF output is also accomplished by summing the modu- lating signal into the oscillator’s FM tuning coil con- trol path.

BIAS
FM

>+4dBm 0.01 - 40 GHz
Switched Filter Assy. - D25015 (Standard) - D26350 (Option 15B)
Switched Doubler Module - ND40843 (26.5 GHz Models)
Switched Doubler Module - D28535 (40 GHz Models)
RF Path for 20 GHz Models
Figure 2-4. Block Diagram of the RF Deck Assembly for all 40 GHz Models (except 682X5B/683X5B Models)
J2
110 dB Step Attenuator (Option)


% ) % * 2 * 3 "
/ ! 0 ) % ) %
- $ ) ' $ *
! ! " " # $ ) % - &
! ! " " # % & &
& $ * + ,
- $ ' * + , 9 - ,
+ ! !
Figure 2-5. Block Diagram of the RF Deck Assembly for all >40 GHz Models (except 682X5B/683X5B Models)
! 5 ! ! ) % ' &
! 6 / ! !
& % ) ( * ,
% ) & % * ,
% ) % * ,
$ ) ( *
8 3 / / !
< ! - $ ' * + , - , = , ; % ' - % = " " ! 6 , ; " - ) ( - ) ( %
6 , ; % ' ' " / ! / ! , ; % ' ( " " ! 6 , ; " - ) ( - ) ( %
! 6 / ! ! , 5 ! " % ' ' 0 ( ) * - ) ( 0 ( ) - * - ) ( % 0 ( ) - *
+ ! !
! / !
7 ! ! 0 3 /
$ ) - $ *

When the 683XXB is generating broad-band analog frequency sweeps (>100 MHz wide), the main tun- ing coil current tunes the oscillator through the sweep frequency range. Phase locking to fine adjust the oscillator’s output frequency is only done at the bottom and top of the sweep ramp and on both sides of each band switch point. Narrow-band analog fre- quency sweeps (100 MHz wide) in the 683XXB are accomplished by summing the appropriate sweep ramp signal into the oscillator’s FM tuning coil con- trol path. The YIG-tuned oscillator’s RF output is then swept about a center frequency that is set by the main tuning coil current. Phase locking to fine tune the output frequency is done at the center fre- quency of the sweep.
The RF output signal from the YIG-tuned oscillator is fed to connector J6 of the switched filter assembly. In the switched filter assembly, the RF signal is am- plified then goes to the modulator. A portion of the RF signal to the modulator is picked off and coupled out via connector J5 to the Sampler for use by the YIG loop circuitry. The modulator provides for power level control, AM, and pulse modulation of the RF output signals.
In addition to the amplified RF signal from the YIG-tuned oscillator, the modulator has two other inputs—the modulator control input and the pulse input. The modulator control input adjusts the gain of the modulator to control the power level of the RF output signals. The modulator control signal is re- ceived from the A9 PIN Control PCB where it is de- veloped from the ALC control signal. Amplitude modulation is accomplished by varying the modula- tor control signal with the modulating signal. Pulse modulation is achieved by switching the modulator on and off at a rate determined by the modulating pulse input. The modulating pulse input is received via connector J7 from the A6 Pulse Generator PCB.
The RF signal from the modulator is fed via PIN switches to the switched low-pass filters. PIN switch drive current is received from the A9 PIN Control PCB. A coupler in the switched filter path provides the RF signal for the down converter. Whenever an instrument is generating RF signals of

coupler in the switched filter path of high power switched filter assemblies provides the RF signal for the source quadrupler module (refer to Figure 2-5). Whenever an instrument is generating RF signals of >40 GHz, a RF signal is coupled out via J4 to the source quadrupler module.
The switched low-pass filters provide rejection of the harmonics that are generated by the YIG-tuned os- cillator. The 2 to 20 GHz (>2.2 to 20 GHz for 682X5B/683X5B models) RF signal from the modu- lator has four filtering paths and a through path. The four filtering paths are 3.3 GHz, 5.5 GHz, 8.4 GHz, and 13.5 GHz. Signals above 13.5 GHz are routed via the through path.
After routing through the appropriate path, the 2 to 20 GHz (>2.2 to 20 GHz for 682X5B/683X5B models) RF signal is multiplexed by the PIN switches and goes via a 20 GHz low-pass filter to the switched fil- ter assembly output connector J2. The 0.01 to 2 GHz (0.5 to 2.2 GHz for 682X5B/683X5B models) RF sig- nal, from the down converter, is received at connec- tor J1, then multiplexed through the same path to the switched filter output.
From J2, the RF signal goes to either the directional coupler (20 GHz models) or the input connector J1 on the switched doubler module (>20 GHz models).
The 0.01 to 2 GHz Down Converter assembly (Fig- ures 2-4 and 2-5) contains a 6.5 GHz VCO that is phase-locked to the 500 MHz reference signal from the A3 Reference Loop PCB. The 6.5 GHz VCO’s phase-lock condition is monitored by the CPU. The 6.5 GHz VCO is on at all times; however, the down converter amplifier is powered on by the A13 YIG Driver PCB only when the 0.01 to 2 GHz frequency range is selected.
During CW or swept frequency operations in the 0.01 to 2 GHz frequency range, the 6.51 to 8.5 GHz RF signal output from J3 of the switched filter as- sembly goes to input connector J1 of the down con- verter. The 6.51 to 8.5 GHz RF signal is then mixed with the 6.5 GHz VCO signal resulting in a 0.01 to 2 GHz RF signal. The resultant RF signal is fed through a 2 GHz low-pass filter, then amplified and routed to the output connector J3. A portion of the down converter’s RF output signal is detected,

amplified, and coupled out for use in internal level- ing. The detected RF sample is routed to the A10 ALC PCB.
The 0.01 to 2 GHz RF output from the down con- verter goes to input connector J1 of the switched fil- ter assembly. There, the 0.01 to 2 GHz RF signal is multiplexed into the switched filter’s output path.
The 0.5 to 2.2 GHz Digital Down Converter assem- bly (Figures 2-6 and 2-7), found in the 682X5B/ 683X5B models, provides improved phase noise across the 0.5 to 2.2 GHz frequency range. Power is applied to the down converter assembly at all times; however, the down converter amplifier is powered on by the A13 YIG Driver PCB only when the 0.5 to 2.2 GHz frequency range is selected.
During CW or swept frequency operations in the 0.5 to 2.2 GHz frequency range, the 2 to 4.4 GHz RF signal output from J3 of the switched filter assem- bly goes to the input connector J1 of the digital down converter. In the down converter, the 2 to 4.4 GHz RF signal is divided by 2 to produce fre- quencies of 1 to 2.2 GHz and divided by 2 again to develop frequencies of 0.5 to 1 GHz. From the fre- quency dividers, the 0.5 to 2.2 GHz RF signal then goes to the modulator which provides power level control, AM, and pulse modulation of the RF sig- nals.
In addition to the 0.5 to 2.2 GHz RF signal from the frequency dividers, the modulator has two other in- puts—the modulator control input and the pulse in- put. The modulator control input adjusts the gain of the modulator to control the power level of the RF output signals. The modulator control signal is re- ceived from the A9 PIN Control PCB where it is de- veloped from the ALC control signal. Amplitude modulation is accomplished by varying the modula- tor control signal with the modulating signal. Pulse modulation is achieved by switching the modulator on and off at a rate determined by the modulating pulse input. The modulating pulse input is received via connector J4 from the A6 Pulse Generator PCB. The modulating pulse signal is also routed out via connector J5 to the switched filter assembly.
The RF signal from the modulator is amplified and fed via PIN switches to the switched low-pass fil-

ters. PIN switch drive current is generated by inter- nal drivers that are controlled by signals received from the A12 Analog Instruction PCB. The switched low-pass filters provide rejection of unwanted har- monics. The 0.5 to 2.2 GHz RF signal has four filter- ing paths—700 MHz, 1000 MHz, 1400 MHz, and 2200 MHz. After routing through the appropriate filter path, the 0.5 to 2.2 GHz RF signal goes to con- nector J2. A portion of the down converter’s RF out- put signal is detected, amplified, and coupled out for use in internal leveling. The detected RF sample is routed to the A10 ALC PCB.
The 0.5 to 2.2 GHz RF output from the down con- verter goes to input connector J1 of the switched fil- ter assembly. There, the 0.5 to 2.2 GHz RF signal is multiplexed into the switched filter’s output path.
The switched doubler module (SDM), found in >20 GHz models, is used to double the fundamental frequencies of 10 to 20 GHz to produce RF output frequencies of 20 to 40 GHz.
26.5 GHz Models The RF signal from the switched filter assembly is input to the SDM at J1. During CW or swept fre- quency operations in the 20 to 26.5 GHz frequency range, the 10 to 13.25 GHz RF signal input is routed by PIN switches to the doubler/amplifier. PIN switch drive current is provided by the A9 PIN Con- trol PCB and bias voltage for the doubler/amplifier is supplied by the A14 SDM Driver PCB. The RF signal is amplified, then doubled in frequency. From the doubler, the 20 to 26.5 GHz RF signal goes to the bandpass filter. After passing through the band- pass filter, the 20 to 26.5 GHz RF signal is multi- plexed by the PIN switches to the SDM output at connector J2. RF signals input to the SDM of 20 GHz are multiplexed through by the PIN switches to output connector J2.
40 GHz Models The RF signal from the switched filter assembly is input to the SDM at J1. During CW or swept fre- quency operations in the 20 to 40 GHz frequency range, the 10 to 20 GHz RF signal input is routed by PIN switches to the doubler/amplifiers. PIN switch drive current is provided by the A9 PIN Control PCB and bias voltage for the doubler/amplifiers is supplied by the A14 SDM, SQM Driver PCB. The

BIAS
FM
2-8.4 GHz
8.4-20 GHz
2-20 GHz YIG Oscillator

J4 NOTE Switched Doubler Module not installed in Models 68245B/68345B.
Switched Doubler Module - ND40843 (26.5 GHz Models)
Switched Doubler Module - D28535 (40 GHz Models)
RF Path for 20 GHz Models
J2
Figure 2-6. Block Diagram of the RF Deck Assembly for all 40 GHz 682X5B/683X5B Models
J2
110 dB Step Attenuator (Option)


" ! " # $ % 0 $ 1
, - ! " # ! "
2 ! 3 " / ' )
" # # # ! # # %
' ) $ % & (
) ) $ % & (
' ' $ % & (
/ $ % & ( : /
3 9 3 * 2 ! 3 " '
" # $ % & (
Figure 2-7. Block Diagram of the RF Deck Assembly for all >40 GHz 682X5B/683X5B Models
# ) ! " " $ %
3 5 ! 3 " ) '
6 ,
# # % & (
# # # % & (
# # % & (
" " # # % & (
( *
' " ! # $ % &
" ) ! ' " $ % &
" # ! " ) $ % &
/
# ) ! # $ % #
8 1 , ,
= / $ % & ( / 3 > < " / " > 6 < / # ! / # ! "
6 < 3 " ) , , < " 6 < / # ! / # ! "
6 , 5 3 " ) - # ! ) # $ % & / # ! - # ! / # $ % & / # ! " - # ! / ) $ % &
( *
3 ,
7 # - 1 ,
# ) ! / ) $ %

RF signal is amplified, then doubled in frequency. From the doubler, the 20 to 40 GHz RF signal is routed by PIN switches to the bandpass filters. There are three bandpass filter paths to provide good harmonic performance. The frequency ranges of the three paths are 20 to 25 GHz, 25 to 32 GHz, and 32 to 40 GHz.
After routing through the appropriate bandpass fil- ter path, the 20 to 40 GHz RF signal is multiplexed by the PIN switches to the SDM output at connector J2. RF signals input to the SDM of 20 GHz are multiplexed through by the PIN switches to output connector J2.
The source quadrupler module (SQM), found in >40 GHz models, is used to quadruple the funda- mental frequencies of 10 to 16.25 GHz to produce RF output frequencies of 40 to 65 GHz. The RF sig- nal inputs for the SQM come from the switched fil- ter assembly. The modulator control signal for the SQM is received from the A14 SDM, SQM Driver PCB where it is developed from the ALC control sig- nal. The A14 PCB also supplies the amplifier bias voltage(s) for the SQM.
50 GHz Models (SQM P/N D28185) During CW or swept frequency operations in the 40 to 50 GHz frequency range, the 10 to 12.5 GHz RF signal input is quadrupled and amplified, then goes to the modulator. The modulator provides for power level control and amplitude modulation of the RF output signal. From the modulator, the 40 to 50 GHz RF signals goes via a band-pass filter to out- put connector J3 of the forward coupler. Note that on the 40 to 50 GHz SQM (P/N D28185), the for- ward coupler is an integral part of the SQM. The 0.01 to 40 GHz (0.5 to 40 GHz for 682X5B/683X5B models) RF output signals from the SDM are routed to input connector J2 of the SQM forward coupler. The 0.01 to 50 GHz (0.5 to 50 GHz for the 682X5B/ 683X5B models) RF output signals go from J3 of the SQM forward coupler to the directional coupler.
60 GHz Models (SQM P/N 60-141) During CW or swept frequency operations in the 40 to 60 GHz frequency range, the 10 to 15 GHz RF signal input is quadrupled and amplified, then goes to the modulator. The modulator provides for power level control and amplitude modulation of the RF output signal. From the modulator, the 40 to

60 GHz RF signal goes via a band-pass filter to the output connector of the SQM.
From the SQM, the 40 to 60 GHz RF output signal goes to input connector J1 of the forward coupler, P/N C27184. The other input to the forward coupler at connector J2 is the 0.01 to 40 GHz (0.5 to 40 GHz for 682X5B/683X5B models) RF output signals from the SDM. From forward coupler output connector J3, the 0.01 to 60 GHz (0.5 to 60 GHz for the 682X5B/683X5B models) RF output signals go to the directional coupler.
65 GHz Models (SQM P/N 60-142) During CW or swept frequency operations in the 40 to 65 GHz frequency range, the 10 to 16.25 GHz RF signal input is quadrupled and amplified, then goes to the modulator. The modulator provides for power level control and amplitude modulation of the RF output signal. From the modulator, the 40 to 65 GHz RF signal goes via a band-pass filter to the output connector of the SQM.
From the SQM, the 40 to 65 GHz RF output signal goes to input connector J1 of the forward coupler, P/N C27184. The other input to the forward coupler at connector J2 is the 0.01 to 40 GHz (0.5 to 40 GHz for 682X5B/683X5B models) RF output signals from the SDM. From forward coupler output connector J3, the 0.01 to 65 GHz (0.5 to 65 GHz for the 682X5B/683X5B models) RF output signals go to the directional coupler.
The RF signal output from either the switched filter assembly (20 GHz models), the SDM (26.5 GHz or 40 GHz models), or forward coupler (>40 GHz mod- els) goes to the directional coupler for transfer to the RF OUTPUT connector. A portion of the RF output signal is detected, amplified, and coupled out as feedback to the ALC circuitry on the A10 ALC PCB. In these circuits, the signal from the detector is summed with the reference voltage that represents the desired RF output power level. The resulting voltage is fed from the A10 ALC PCB to the ALC modulator driver circuit on the A9 PIN Control PCB (and the ALC modulator driver circuit on the A14 SDM, SQM Driver PCB for >40 GHz models). The resulting modulator control signals go to the modu- lators in the switched filter assembly, the SQM (for >40 GHz models), and the digital down converter

(for 682X5B/683X5B models) to adjust the RF out- put power level.
The optional step attenuator provides up to 110 dB (90 dB for 50 GHz and 60 GHz models) attenuation of the RF output in 10 dB steps. The step attenuator drive current is supplied by the A9 PIN Control PCB.



Table of Contents
3-1 INTRODUCTION 3-3
3-3 TEST RECORDS 3-3
3-5 682XXB/683XXB POWER LEVELS 3-6
Test Setup 3-9 Test Procedure 3-9 EXAMPLE: 3-10
3-7 FREQUENCY SYNTHESIS TESTS 3-11
Test Setup 3-11 Coarse Loop/YIG Loop Test Procedure 3-12 Fine Loop Test Procedure 3-13
3-8 SPURIOUS SIGNALS TEST 3-14
Test Setup 3-14 0.01 - 2 GHz Test Procedure 3-14 0.5 - 2.2 GHz Test Procedure 3-17
3-9 HARMONIC TEST: 2 TO 20 GHz 3-18
Test Setup 3-18 Test Procedure (2 to 10 GHz) 3-19 Test Procedure (11 to 20 GHz) 3-20
Test Setup 3-22 Test Procedure 3-23

3-11 POWER LEVEL TESTS 3-26
Test Setup 3-26 Power Level Accuracy Test Procedure 3-27 Power Level Flatness Test Procedure 3-27
3-12 AMPLITUDE MODULATION TEST 3-30
Test Setup 3-30 AM Input Sensitivity Procedure 3-31
Test Setup 3-33 FM Input Sensitivity Procedure 3-33
3-14 PULSE MODULATION TESTS 3-39
Test Setup 3-39 Rise/Fall Time and Overshoot 3-40 Pulse Leveling Accuracy 3-41
FEEDTHROUGH 3-43 Test Setup 3-43 Test Procedure 3-43
RATIO 3-45 Test Setup 3-45 Test Procedure 3-45
3-17 PHASE MODULATION TESTS 3-48
Test Setup 3-48 Μ Input Sensitivity Procedure 3-48

3-1 INTRODUCTION This chapter contains tests that can be used to verify the performance of the Series 682XXB/683XXB Synthesized Signal Generators to speci- fications. These tests support all instrument models having any ver- sion of firmware. Units with Option 2A, 2B, 2C, or 2D (110 dB or 90 dB step attenuators), Option 11 (0.1 Hz frequency resolution), and Option 15B (high power output) are also covered.
Table 3-1 (page 3-4) provides a list of the recommended test equipment for the performance verification tests.
The test procedures refer to specific test equipment front panel control settings when the test setup is critical to making an accurate meas- urement. In some cases, the user may substitute test equipment hav- ing the same critical specifications as those on the recommended test equipment list.
Contact your local ANRITSU service center (refer to Table 1-5 on page 1-18) if you need clarification of any equipment or procedural refer- ence.
3-3 TEST RECORDS A blank copy of a sample performance verification test record for each 682XXB/683XXB model is provided in Appendix A. Each test record contains the model-specific variables called for by the test procedures. It also provides a means for maintaining an accurate and complete record of instrument performance. We recommend that you copy these pages and use them to record the results of your initial testing of the instrument. These initial test results can later be used as benchmark values for future tests of the same instrument.
The test procedures include many references to equipment intercon- nections and control settings. For all 682XXB/683XXB references, spe- cific labels are used to denote the appropriate menu key, data entry key, data entry control, or connector (such as CW/SWEEP SELECT or RF OUTPUT). Most references to supporting test equipment use gen- eral labels for commonly used controls and connections (such as Span or RF Input). In some cases, a specific label is used that is a particular feature of the test equipment listed in Table 3-1.

Table 3-1. Recommended Test Equipment for Performance Verification Tests (1 of 2)
RECOMMENDED MANUFACTURER/MODEL
TEST NUMBER
Spectrum Analyzer, with External Mixers and Diplexer Assy
Tektronix, Model 2794, with External Mixers:
3-8, 3-9, 3-12, 3-13, 3-16, 3-17
WM780K (18 to 26.5 GHz) WM780A (26.5 to 40 GHz) WM780U (40 to 60 GHz) WM780E (60 to 90 GHz) and Diplexer Assy: 015-0385-00
Hewlett-Packard, Model 3585B 3-10
Frequency Counter with Cable Kit and External Mixer
Frequency Range: 0.01 to 65 GHz Input Impedance: 50Ω Resolution: 1 Hz Other: External Time Base Input
EIP Microwave, Inc. Models 538B, 548B, or 578B, with Cable Kit: Option 590 and External Mixer:
Option 91 (26.5 to 40 GHz) Option 92 (40 to 60 GHz) Option 93 (60 to 90 GHz)
Power Meter, with Power Sensor
Hewlett-Packard Model 437B, with Power Sensor:
(1mW to 100mW)
HP 8487A (0.01 to 50 GHz)
Power Meter with Power Sensor
ANRITSU ML4803A, with Power Sensor:
(1mW to 100mW)
MP716A4 (50 to 75 GHz)
Absolute Time Corp., Model 300 3-6
Hewlett-Packard, Model 8901A 3-12
(or the IF of the spectrum analyzer) AM Depth: 0% to 90% AM Modulation Rates: dc to 100 kHz Filters: 20 kHz lowpass, 300 Hz highpass
Tektronix, Inc. Model TAS485 3-6, 3-11, 3-14, 3-15
Adapter K (male) to 2.4 mm (female) Adapts the Power Sensor, HP 8487A, to the 682XXB/683XXB RF OUTPUT con- nector (£40 GHz models)
Adapter Adapts the MP716A4 Power Sensor to the ML4803A Power Meter

Table 3-1. Recommended Test Equipment for Performance Verification Tests (2 of 2)
RECOMMENDED MANUFACTURER/MODEL
TEST NUMBER
Adapter WR15 to V (male) Adapts the MP716A4 Power Sensor to the 682XXB/683XXB RF OUTPUT connector (>40 GHz)
ANRITSU, Model 35WR15V 3-11
ANRITSU, Model 41KC-10 3-9, 3-10
ANRITSU, Model 41KC-20 3-9
ANRITSU, Model 41V-10 3-9, 3-10
ANRITSU, Model 41V-20 3-9
ANRITSU, Model SC3800 (K input/BNC output connectors)
High Pass Filter 150 MHz High-Pass Filter (DC Block) Narda, Model 4564 3-14
Low Pass Filter 450 MHz Low-Pass Filter Mini-Circuits, Model SLP-450 3-15
Microwave (Pulse) Detector

3-5 682XXB/683XXB POWER LEVELS
Table 3-2 is a listing of the Series 682XXB and 683XXB Synthesized Signal Generator models and their maximum leveled ouput power lev- els. Certain test procedures will refer you to this table for the maxi- mum leveled output power level setting of the instrument model being tested.
Table 3-2. 682XXB/683XXB Maximum Leveled Output Power (1 of 2)
Max Leveled Output Power w/Step Attenuator
Max Leveled Output Power
68X37B 2.0 – 20.0 GHz +13.0 dBm +11.0 dBm
68X45B 0.5 – 20.0 GHz +13.0 dBm +11 dBm
68X47B 0.01 – 20.0 GHz +13.0 dBm +11.0 dBm
68X53B 2.0 – 20.0 GHz 20.0 – 26.5 GHz +9.0 dBm +6.0 dBm
+7.0 dBm +3.5 dBm
68X55B 0.5 – 2.2 GHz 2.2 – 20.0 GHz 20.0 – 26.5 GHz
+13.0 dBm
+11.0 dBm
+9.0 dBm +6.0 dBm
+7.0 dBm +3.5 dBm
68X59B 0.01 – 2.0 GHz 2.0 – 20.0 GHz 20.0 – 26.5 GHz
+13.0 dBm
+11.0 dBm
+9.0 dBm +6.0 dBm
+7.0 dBm +3.5 dBm
68X63B 2.0 – 20.0 GHz 20.0 – 40.0 GHz +9.0 dBm +6.0 dBm
+7.0 dBm +3.0 dBm
68X65B 0.5 – 2.2 GHz 2.2 – 20.0 GHz 20.0 – 40.0 GHz
+13.0 dBm
+11.0 dBm
+9.0 dBm +6.0 dBm
+7.0 dBm +3.0 dBm
68X69B 0.01 – 2.0 GHz 2.0 – 20.0 GHz 20.0 – 40.0 GHz
+13.0 dBm
+11.0 dBm
+9.0 dBm +6.0 dBm
+7.0 dBm +3.0 dBm
0.5 – 2.2 GHz 2.2 – 20.0 GHz 20.0 – 40.0 GHz 40.0 – 50.0 GHz
+11.0 dBm +10.0 dBm
+10.0 dBm
+8.5 dBm
+2.5 dBm +2.5 dBm
0.0 dBm –1.0 dBm
0.01 – 2.0 GHz 2.0 – 20.0 GHz 20.0 – 40.0 GHz 40.0 – 50.0 GHz
+12.0 dBm +10.0 dBm
+10.0 dBm
+8.5 dBm
+2.5 dBm +2.5 dBm
0.0 dBm –1.0 dBm
0.5 – 2.2 GHz 2.2 – 20.0 GHz 20.0 – 40.0 GHz 40.0 – 50.0 GHz 50.0 – 60.0 GHz
+11.0 dBm +10.0 dBm
+10.0 dBm
+8.5 dBm
+2.5 dBm +2.0 dBm +2.0 dBm
0.0 dBm –1.5 dBm –2.0 dBm
0.01 – 2.0 GHz 2.0 – 20.0 GHz 20.0 – 40.0 GHz 40.0 – 50.0 GHz 50.0 – 60.0 GHz
+12.0 dBm +10.0 dBm
+10.0 dBm
+8.5 dBm
+2.5 dBm +2.0 dBm +2.0 dBm
0.0 dBm –1.5 dBm –2.0 dBm
68XXXB Model · Frequency · (GHz)

3-6 INTERNAL TIME BASE AGING RATE TEST (Optional)
The following test can be used to verify that the 682XXB/683XXB 10 MHz time base is within its aging specification. The instrument de- rives its frequency accuracy from an internal 100 MHz crystal oscilla- tor standard. (With Option 16 installed, frequency accuracy is derived from an internal high-stability 10 MHz crystal oscillator.) An inherent characteristic of crystal oscillators is the effect of crystal aging within the first few days to weeks of operation. Typically, the crystal oscilla- tor’s frequency increases slightly at first, then settles to a relatively constant value for the rest of its life. The 682XXB/683XXB reference oscillator aging is specified as
For greatest absolute frequency accuracy, allow the 682XXB/683XXB to warm up until its RF output frequency has stabilized (usually 7 to 30 days). Once stabilized, the change in reference oscillator frequency should remain within the aging rate if; (1) the time base oven is not al- lowed to cool, (2) the instrument orientation with respect to the earth’s magnetic field is maintained, (3) the instrument does not sustain any mechanical shock, and (4) ambient temperature is held constant. This test should be performed upon receipt of the instrument and again af- ter a period of several days to weeks to fully qualify the aging rate.

! "
Figure 3-1. Equipment Setup for Internal Time Base Aging Rate Test
Test Setup Connect the 682XXB/683XXB rear panel 10 MHz REF OUT to the Frequency Reference front panel input connector labeled 10 MHz when directed to do so during the test procedure.

The frequency error is measured at the start and finish of the test time period of 24 hours. The aging rate is the difference between the two error read- ings.
a. Press the ESC key until the MAIN MENU is
displayed.
b. At the MAIN MENU display, press 1 to select
CONFIGURATION.
c. At the CONFIGURATION MENU display,
press 8 to select MEAS.
d. Press the MOD key and use the Up/Down Ar-
row keys to get to the menu display: MEASUREMENT = FREQ.
e. Press the ENTER key.
f. Press the ESC key until the MAIN MENU is
displayed.
g. At the MAIN MENU display, press 3 to select
the REVIEW MENU.
h. At the REVIEW MENU display, press 8 to se-
lect TFM.
The frequency error in the signal under test is displayed in ps/s (Picoseconds/Second). For exam- ple, an error of –644681 ps/s is –644681 10–12 or –6.44681 10–7 away from the 10 MHz internal reference of the Frequency Reference.
The frequency error display is continuously up- dated as a running 5000-second average. The av- eraging smooths out the short-term instability of the oscillator.


3-7 FREQUENCY SYNTHESIS TESTS
The following tests can be used to verify correct operation of the fre- quency synthesis circuits. Frequency synthesis testing is divided into two parts—coarse loop/YIG loop tests and fine loop tests.

Figure 3-2. Equipment Setup for Frequency Synthesis Tests
Test Setup Connect the equipment, shown in Figure 3-2, as fol- lows:
a. For measuring frequencies of 0.01 to 1.0 GHz,
connect to the Band 2 input (Connection A).
b. For measuring frequencies of 1.0 to 26.5 GHz,
connect to the Band 3 input (Connection A).
c. For measuring frequencies of 26.5 to 40.0 GHz,
connect to the Band 4 input via the Option 91 waveguide mixer (Connection B).
d. For measuring frequencies of 40.0 to 60.0 GHz,
connect to the Band 4 input via the Option 92 waveguide mixer (Connection B).
e. For measuring frequencies of 60.0 to 65.0 GHz,
connect to the Band 4 input via the Option 93 waveguide mixer (Connection B).

The following procedure tests both the coarse loop and YIG loop by stepping the signal generator through its full frequency range in 1 GHz steps and measuring the RF output at each step.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to the first test frequency indicated on
the Test Record for the model being tested.

The following procedure tests the fine loop by step- ping the instrument through ten 1 kHz steps (ten 100 Hz steps for instruments with Option 11) and measuring the RF output at each step.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to the first test frequency indicated on
the Test Record.

The following test can be used to verify that the signal generator meets its spurious signals specifications for RF output signals from 0.01 to 2 GHz (0.5 to 2.2 GHz for 682X5B/683X5B Models). This test is applicable only to instruments which cover the frequency range 10 MHz to 2 GHz (500 MHz to 2.2 GHz for 682X5B/683X5B models). The 0.01 to 2 GHz test procedure begins on this page; the 0.5 to 2.2 GHz test procedure begins on page 3-17.

Figure 3-3. Equipment Setup for Spurious Signals Test: RF Output Signals £2 GHz
Test Setup Connect the equipment, shown in Figure 3-3, as fol- lows:
The following procedure lets you measure the worst case spurious signals (harmonic and non-harmonic) of the 0.01 to 2 GHz RF output to verify that they meet specifications.
clearly)

a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit L1 to open the current power level
parameter for editing.
c. Set L1 to the lesser of +10 dBm or the maxi-
mum leveled power level for the instrument being tested (refer to Table 3-2, page 3-6).
d. Press Edit F1 to open the current frequency
Table 3-3. Spurious Signals Specifications
parameter for editing.
Harmonic and Harmonic Related:
500 MHz to £2.2 GHz (68XX5B):
e. Set F1 to 10 MHz.
to £20 GHz:
500 MHz to £2.2 GHz (68XX5B):
to £20 GHz:
500 MHz to £2.2 GHz (68XX5B):

presence by entering their levels on the Test Re- cord.

The following procedure lets you measure the 0.5 to 2.2 GHz RF output harmonic levels to verify that they meet specifications.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit L1 to open the current power level
parameter for editing.
c. Set L1 to the lesser of +10 dBm or the maxi-
mum leveled power level for the instrument being tested (refer to Table 3-2, page 3-6).
d. Press Edit F1 to open the current frequency
parameter for editing.
e. Set F1 to the frequency indicated on the Test
Record.
value.

3-9 HARMONIC TEST: RF OUTPUT SIGNALS FROM 2 TO 20 GHz
The following test can be used to verify that the 682XXB/683XXB meets its harmonic specifications for RF output signals from 2 to 20 GHz (2.2 to 20 GHz for 68XX5B Models). Test record entries are supplied for harmonics up to a frequency limit of 40 GHz. Additional harmonic checks may be made at any frequency of interest up to the RF output frequency limit of the 682XXB/683XXB model being tested. These additional harmonic checks can be accomplished through the use of waveguide mixers to extend the frequency range of the spec- trum analyzer.

Figure 3-4. Equipment Setup for Harmonic Test: RF Output Signals from 2 to 20 GHz
Test Setup Connect the equipment, shown in Figure 3-4, as fol- lows:

The following procedure lets you measure the 2 to 10 GHz (2.2 to 10 GHz for 68XX5Bs) RF output har- monic levels to verify that they meet specifications.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit L1 to open the current power level
Table 3-4. Spurious Signals Specifications
parameter for editing.
Harmonic and Harmonic Related:
500 MHz to £2.2 GHz (68XX5B):
c. Set L1 to the lesser of +10 dBm or the maxi-
mum leveled power level for the instrument being tested (refer to Table 3-2, page 3-6).
to £20 GHz:
d. Press Edit F1 to open the current frequency
parameter for editing.
e. Set F1 to the frequency indicated on the Test
Record.
500 Mhz to £2.2 GHz (68XX5B):
to £20 GHz:
value.
500 MHz to £2.2 GHz (68XX5B):

The following procedure lets you measure the 11 to 20 GHz RF output harmonic levels to verify that they meet specifications.
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to the frequency indicated on the Test
Record.
d. Press Edit L1 to open the current power level
parameter for editing.
e. Set L1 to –30 dBm output power.
value.

slightly to accomplish this; however, do not exceed –-20 dBm output power.

3-10 SINGLE SIDEBAND PHASE NOISE TEST
The following test can be used to verify that the signal generator meets its single sideband phase noise specifications. For this test, a second 682XXB/683XXB is required. This additional instrument acts as a local oscillator (LO). The CW RF output of the 682XXB/683XXB under test (DUT) is mixed with the CW RF output from the 682XXB/ 683XXB LO which is offset by 1 MHz. Single sideband phase noise is measured at offsets of 100 Hz, 1 kHz, 10 kHz, and 100 kHz away from the resultant 1 MHz IF.

Figure 3-5. Equipment Setup for Single Sideband Phase Noise Test
Test Setup Connect the equipment, shown in Figure 3-5, as fol- lows:

The following procedure lets you measure the RF output single sideband phase noise levels to verify that they meet specifications.
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to the frequency indicated on the Test
Record.
d. Press Edit L1 to open the current power level
parameter for editing.
e. Set L1 to the maximum leveled power level for
the instrument being tested (refer to Table 3-2, page 3-6).
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to a frequency that is 1 MHz lower than
the 682XXB/683XXB DUT frequency set in step 1.c.
d. Press Edit L1 to open the current power level
parameter for editing.
e. Set L1 to the maximum leveled power level for
the instrument model (refer to Table 3-2).

Table 3-5. Single Sideband Phase Noise Test Specification
100 Hz 1 kHz 10 kHz 100 kHz
0.6 GHz
d. Position the Marker to the peak of the signal.
e. Select OFFSET, ENTER OFFSET, and
MKRCF.
100 Hz 1 kHz 10 kHz 100 kHz
0.6 GHz (68XX5B)
f. Adjust the marker for a 100 Hz offset.
g. Select NOISE LVL.
100 Hz 1 kHz 10 kHz 100 kHz
2.0 GHz
100 Hz 1 kHz 10 kHz 100 kHz
2.0 GHz (68XX5B)
b. Set Frequency Span to 20 kHz.
100 Hz 1 kHz 10 kHz 100 kHz
c. Set RBW to 100 Hz.
6.0 GHz
d. Adjust the Marker for a 1 kHz offset.
e. Select NOISE LVL.
100 Hz 1 kHz 10 kHz 100 kHz
10.0 GHz
100 Hz 1 kHz 10 kHz 100 kHz
20.0 GHz
b. Set Frequency Span to 100 kHz.
100 Hz 1 kHz 10 kHz 100 kHz
c. Adjust the Marker for a 10 kHz offset.
26.5 GHz
d. Select NOISE LVL.
* 3 dB difference from 682XXB/683XXB single side- band phase noise specifications to account for LO phase noise.

a. Deselect NOISE LVL.
b. Set Frequency Span to 300 kHz.
c. Adjust the Marker for a 100 kHz offset.
d. Select NOISE LVL.

3-11 POWER LEVEL ACCURACY AND FLATNESS TESTS
The following tests can be used to verify that the 682XXB/683XXB meets its power level specifications. Power level verification testing is divided into two parts—power level accuracy tests and power level flatness tests.

Figure 3-6. Equipment Setup for Power Level Accuracy and Flatness Tests
Test Setup Connect the equipment, shown in Figure 3-6, as fol- lows:

Power level accuracy is tested by stepping the out- put power level down in 1 dB increments from its maximum rated power level and measuring the out- put power level at each step.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to the CW frequency indicated on the
Test Record.
d. Press Edit L1 to open the current power level
parameter for editing.
e. Set L1 to the power level indicated on the Test
Record.
Power level flatness is tested by measuring the out- put power level variation during a full band sweep; first in the step sweep mode (682XXB/683XXB mod- els), then in analog sweep mode (683XXB models only).
a. Reset the instrument by pressing SYSTEM,
then Reset . The CW Menu is displayed.
b. Press Step to place the signal generator in the
step sweep frequency mode and display the Step Sweep Menu.

c. With the Step Sweep Menu displayed, press
the main menu key
CONTROL
The Sweep Frequency Control Menu is then displayed.
d. Press Full to select a full range frequency
sweep.
e. Press Edit L1 to open the current power level
parameter for editing.
f. Set L1 to the power level indicated on the test
record.
g. Now, return to the Step Sweep Menu by press-
ing the main menu key
SELECT
h. At the Step Sweep Menu, press Sweep Ramp
to go to the Step Sweep Ramp Menu.
i. At this menu, press Dwell Time to open the
dwell time-per-step parameter for editing.
j. Set the dwell time to 1 second.

a. Reset the instrument by pressing SYSTEM,
then Reset . The CW Menu is displayed.
b. Press Analog to place the 683XXB in the ana-
log sweep frequency mode and display the Analog Sweep Menu.
c. With the Analog Sweep Menu displayed, press
the main menu key
CONTROL
The Sweep Frequency Control Menu is then displayed.
d. Press Full to select a full range frequency
sweep.
e. Press Edit L1 to open the current power level
parameter for editing.
f. Set L1 to the power level noted on the test rec-
ord.
g. Now, return to the Analog Sweep Menu by
pressing the main menu key
SELECT
h. At the Analog Sweep Menu, press the menu
soft-key Sweep Ramp to go to the Analog Sweep Ramp Menu.
i. At this menu, press Sweep Time to open the
sweep time parameter for editing.
j. Set the sweep time to 99 seconds.

3-12 AMPLITUDE MODULATION TEST
This procedure verifies the operation of the 682XXB/683XXB ampli- tude modulation input sensitivity circuit.
The RF OUTPUT of the 682XXB/683XXB is monitored on a Spectrum Analyzer display. The (modulated) IF Output of the Spectrum Ana- lyzer is monitored with a Modulation Analyzer. A 50% AM (default value) signal is set to a reference point on the Spectrum Analyzer. The actual modulation value is then computed from the indicated Modula- tion Meter values. (The absolute values of the AM PK+ and AM PK– readings are used in the given formula to compensate for non-linearity errors in the test equipment.)

Figure 3-7. Equipment Setup for Amplitude Modulation Tests
Test Setup Connect the equipment, shown in Figure 3-7, as fol- lows:

The following procedure lets you measure the abso- lute peak AM values for a 50% AM signal and calcu- late the actual modulation value.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
d. Press Edit L1 to open the current power level
parameter for editing.
e. Set L1 to a power level that is 6 dB below
maximum rated power (refer to Table 3-2, page 3-6).
f. Press MODULATION, then AM . At the result-
ing AM Status menu, press On/Off to turn AM on.
b. 300 Hz High-Pass Filter
c. 20 kHz Low-Pass Filter

% | ( )| | (–)| | ( )| | (–)| AM AM PK AM PK AM PK AM PK 100 200
· ·

3-13 FREQUENCY MODULATION TESTS
This procedure verifies the operation of the 682XXB/683XXB fre- quency modulation input sensitivity circuitry.
The RF OUTPUT of the 682XXB/683XXB is monitored on a Spectrum Analyzer display. In Unlocked Wide FM mode, the Spectrum Analyzer is used as a frequency counter to measure the actual FM deviation and determine FM accuracy. In Unlocked Narrow, Locked, and Locked Low-Noise FM modes, the Spectrum Analyzer is used to monitor the carrier level. FM accuracy is determined by measuring the actual FM deviation necessary to reduce the carrier level to its lowest level. These tests qualify how the FM input signal affects the signal genera- tor’s RF output.

Figure 3-8. Equipment Setup for Frequency Modulation Tests
Test Setup Connect the equipment, shown in Figure 3-8, as fol- lows:
The following procedure lets you measure the FM deviation for a specific input sensitivity in Unlocked Wide, Unlocked Narrow, Locked, and Locked Low- Noise FM modes and determine FM accuracy.

Unlocked Wide FM Mode 1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
ing FM Status menu, press More to go to the additional FM Status menu.
b. At the additional FM Status menu, press
Unlocked Wide , then press Previous Menu to return to the main FM Status menu.
c. At the FM Status menu, press Mod. Wave to
go to the Modulation Waveform Selection menu. Use the cursor control key to highlight the squarewave modulation waveform, then press Select .
d. Press Previous Menu to return to the FM
Status menu, then press Edit Dev. and set the deviation to 100 MHz. Press Edit Rate and set the rate to 0.4 Hz.
e. Press On/Off to turn FM on.
the low carrier at the center of the display. Re- cord the frequency reading on the Test Record.
b. Adjust the center frequency control to position
the high carrier at the center of the display. Record the frequency reading on the Test Re- cord.
play)

Unlocked Narrow FM Mode 1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
ing FM Status menu, press More to go to the additional FM Status menu.
NULL
b. At the additional FM Status menu, press
Unlocked Narrow , then press Previous Menu to return to the main FM Status menu.
c. At the FM Status menu, press Edit Dev. and
set the deviation to 0.240 MHz.
d. Press Edit Rate and set the rate to 99.8 kHz.
e. Press On/Off to FM on.
Figure 3-9. Typical Spectrum Analyzer Display of Bessel Null on FM Waveform

Accuracy in FM Deviation inkHz ( %) ( )
240 100
Locked FM Mode 1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
Status menu is displayed.
b. At the FM Status menu, press Edit Dev. and
set the deviation to 0.240 MHz.
c. Press Edit Rate and set the rate to 99.8 kHz.
d. Press On/Off to FM on.

Accuracy in FM Deviation inkHz ( %) ( )
240 100
Locked Low-Noise FM Mode 1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
ing FM Status menu, press More to go to the additional FM Status menu.
b. At the additional FM Status menu, press
Locked LN , then press Previous Menu to return to the main FM Status menu.
c. At the FM Status menu, press Edit Dev. and
set the deviation to 0.240 MHz.
d. Press Edit Rate and set the rate to 99.8 kHz.
e. Press On/Off to FM on.

Accuracy in FM Deviation inkHz ( %) ( )
240 100

3-14 PULSE MODULATION TESTS: RISE TIME, FALL TIME, OVERSHOOT, AND LEVEL
Pulse modulation tests verify correct operation as well as rise time, fall time, overshoot, and level. The pulse leveling accuracy test compares the pulsed RF output level to verify the performance of the level detector(s), detector amplifiers, and ALC sample/ hold circuits.

Figure 3-10. Equipment Setup for Pulse Modulation Tests: Rise Time, Fall Time, Overshoot, and Level
Test Setup Connect the equipment, shown in Figure 3-10, as follows:

The following procedure lets you measure the rise time, fall time, and overshoot parameters of the pulse modulation waveform.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
d. Press Edit L1 to open the current power level
parameter for editing.
e. Set L1 to the maximum leveled output power
level for the instrument being tested (refer to Table 3-2, page 3-6).
f. Press MODULATION, then Pulse . The Inter-
nal Pulse Status menu is displayed.
g. At the Internal Pulse Status menu, press
Edit Width and set pulse width W1 to 5.0 s.
h. Press Edit Period and set the PRI to 0.01 ms.
i. Press On/Off to turn pulse modulation on.

Percentage of RF Voltage
OVERSHOOT
Figure 3-11. Measurement Parameters for a Pulse Modulation Waveform
The following procedure lets you measure the pulsed RF output level and CW output level and compare them to verify pulse leveling accuracy.
turn off pulse modulation.
b. Press CW/SWEEP SELECT to return to the
CW Menu display.
c. At the CW menu, press Edit F1 and set F1 to
the frequency noted in the Test Record.
actly on the center graticule; this is used as the CW-level reference line.
b. Use the Auto Triggering mode to continue
sweeping the display when the pulse is off.
c. Adjust the vertical sensitivity to the most sen-
sitive setting (mV/div) possible while keeping the trace at the center graticule reference line.

a. Press MODULATION, then Pulse to go to the
Internal Pulse Status Menu display.
b. Press Edit Width and set W1 to the pulse
width noted in the Test Record.
c. Press On/Off to turn on pulse modulation.
CW Menu display.
b. Record the value of the output level L1, shown
in the Level display area, on the Test Record.
c. Press Edit L1 and adjust the power level until
the nominal peak level is evenly centered on the display centerline reference.
d. Record this value of the output level L1 on the
Test Record.
eled output power level for the instrument be- ing tested (refer to Table 3-2, page 3-6).
b. Press MODULATION, then Pulse to return to
the Internal Pulse Status Menu display.
c. Press On/Off to turn off pulse modulation.

3-15 PULSE MODULATION TEST: VIDEO FEEDTHROUGH
This pulse modulation test verifies that video feedthrough is within specifications.

450 MHz Low Pass Filter (For Video Feedthrough Measurement) CH. 4 Trigger
Figure 3-12. Equipment Setup for Pulse Modulation Test: Video Feedthrough
Test Setup Connect the equipment, shown in Figure 3-12, as follows:

The following procedure lets you measure the video feedthrough that occurs during pulse modulation.
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
d. Press Edit L1 to open the current power level
parameter for editing.
e. Set L1 to the maximum leveled output power
level for the instrument being tested (refer to Table 3-2, page 3-6).
f. Press MODULATION, then Pulse . The Inter-
nal Pulse Status menu is displayed.
g. At the Internal Pulse Status menu, press
Edit Width and set pulse width W1 to 5.0 s.
h. Press Edit Period and set the PRI to 0.01 ms.
i. Press On/Off to turn pulse modulation on.
Channel 1 signal input

3-16 PULSE MODULATION TEST: RF ON/OFF RATIO
This pulse modulation test verifies that the ratio of RF on power to RF off power is within specifications.

Figure 3-13. Equipment Setup for Pulse Modulation Test: RF On/Off Ratio
Test Setup Connect the equipment, shown in Figure 3-13, as follows:
The following procedure lets you determine the ratio of RF on power to RF off power during pulse modu- lation.
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to the frequency noted on the Test Re-
cord.
d. Press MODULATION, then Pulse . The Inter-
nal Pulse Status menu is displayed.

e. Press More to go to the additional Internal
Pulse Status Menu display.
g. Press Internal/External to select the external
source for the modulation signal. The External Pulse Status Menu is displayed.
h. Press Front/Rear to select the rear panel
PULSE TRIGGER IN connector.
i. Press On/Off to turn pulse modulation on.
quency noted in step 1.c.
the top graticule of the screen.


3-17 PHASE MODULATION TESTS
This procedure verifies the operation of the phase modulation (M) in- put sensitivity circuits in 682XXB/683XXBs with Option 6.
The carrier level of the 682XXB/683XXB RF output is monitored on a Spectrum Analyzer display. M accuracy, in both Wide and Narrow M modes, is determined by measuring the actual M deviation nec- essary to reduce the carrier level to its lowest level. These tests qualify how the FM input signal affects the signal generator’s RF output.

Figure 3-14. Equipment Setup for Phase Modulation Tests
Test Setup Connect the equipment, shown in Figure 3-14, as follows:
The following procedure lets you measure the M deviation for a specific input sensitivity in Wide and Narrow M modes and determine M accuracy.

WIDE M Mode 1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
sulting M Status menu, press More to go the additional M Status menu.
b. At the additional M Status menu, press
Wide , then press Previous Menu to return to the main M Status menu.
NULL
c. At the M Status menu, press Edit Dev. and
set the deviation for 2.40 rad.
d. Press Edit Rate and set the rate to 99.8 kHz.
e. Press On/Off to turn M on.
Figure 3-15. Typical Spectrum Analyzer Display
of Bessel Null on FM Waveform
Accuracy in M Deviation inrad ( %) ( ) .
· · 240 100

Narrow M Mode 1. Set up the 682XXB/683XXB as follows:
a. Reset the instrument by pressing SYSTEM,
then Reset . Upon reset, the CW Menu is dis- played.
b. Press Edit F1 to open the current frequency
parameter for editing.
c. Set F1 to 5.0 GHz.
sulting M Status menu, press More to go to the additional M Status menu.
b. At the additional M Status menu, press
Narrow , then press Previous Menu to return to the main M Status menu.
c. At the M Status menu, press Edit Dev. and
set the deviation to 2.40 rad.
d. Press Edit Rate and set the rate to 99.8 kHz.
e. Press On/Off to turn M on.
Accuracy in M Deviation inrad ( %) ( ) .
· · 240 100

Table of Contents
4-1 INTRODUCTION 4-3
4-3 TEST RECORDS 4-3
4-4 CALIBRATION FOLLOWING SUBASSEMBLY REPLACEMENT 4-4
4-6 INITIAL SETUP 4-7
Interconnection 4-7 PC Setup — Windows 3.1 4-8 PC Setup — Windows 95 4-10
4-7 PRELIMINARY CALIBRATION 4-13
Equipment Setup 4-13 Calibration Steps 4-14
Equipment Setup 4-17 Log Amplifier Zero Calibration 4-18 Limiter DAC Adjustment 4-18 Shaper DAC Adjustment 4-19
4-9 RF LEVEL CALIBRATION 4-22
VERSION 1.01 TO 2.00) 4-23 Equipment Setup 4-23 ALC Slope DAC Adjustment 4-24
VERSION 2.00 AND ABOVE) 4-27 Equipment Setup 4-27

ALC Slope DAC Adjustment 4-28
4-11 ALC BANDWIDTH CALIBRATION 4-31
Equipment Setup 4-31 Bandwidth Calibration 4-31
4-12 AM CALIBRATION 4-33
Equipment Setup 4-33 AM Calibration Procedure 4-34
TO 2.10 (VERSION 1.00 FOR 682X5B/683X5B) . 4-37 Equipment Setup 4-37 FM Calibration Procedure 4-38
AND ABOVE FOR 682XXBs/VERSION 2.20 AND ABOVE FOR 683XXBs (VERSION 1.07 AND ABOVE FOR 682X5B/683X5B) 4-44 Equipment Setup 4-44 FM Calibration Procedure 4-45
Equipment Setup 4-51 Μ Calibration Procedure 4-52

4-1 INTRODUCTION This chapter contains procedures for calibrating the Series 682XXB/ 683XXB Synthesized Signal Generators. These procedures are typi- cally performed because out-of-tolerance conditions have been noted during performance verification testing (see Chapter 3) or as a result of replacement of subassemblies or RF components.
Table 4-1 (page 4-4) provides a list of the recommended test equipment for these calibration procedures.
The procedures refer to specific test equipment front panel control settings when the test setup is critical to making accurate measure- ments. In some cases, the user may substitute test equipment having the same critical specifications as those on the recommended test equipment list.
Contact your local ANRITSU service center (Refer to Table 1-5 on page 1-18) if you need clarification of any equipment or procedural refer- ence.
4-3 TEST RECORDS A blank copy of a sample calibration test record for each 682XXB/ 683XXB model is provided in Appendix A. Each test record contains model-specific variables called for by the calibration procedures. It also provides a means for maintaining an accurate and complete record of instrument calibration. We recommend that you copy these pages and use them to record the results from (1) your initial calibra- tion of out-of-tolerance 682XXB/683XXB circuits, or (2) your initial calibration of the 682XXB/683XXB following replacement of subassem- blies or RF components. These initial readings can be used later as benchmark values for future tests of the same serial-numbered instruments.

4-4 CALIBRATION FOLLOWING SUBASSEMBLY REPLACEMENT
Table 4-2 (page 4-6) lists the calibration that should be performed fol- lowing the replacement of 682XXB/683XXB subassemblies or RF com- ponents.
The calibration procedures include many references to equipment in- terconnections and control settings. For all 682XXB/683XXB refer- ences, specific labels are used to denote the appropriate menu key, data entry key, data entry control, or connector (such as CW/SWEEP SELECT or RF OUTPUT). Most references to supporting test equip- ment use general labels for commonly used controls and connections (such as Span or RF Input). In some cases, a specific label is used that is a particular feature of the test equipment listed in Table 4-1.
Table 4-1. Recommended Test Equipment for Calibration Procedures (1 of 2)
RECOMMENDED MANUFACTURER/MODEL
PROCEDURE
NUMBER
EIP Microwave, Inc. Model 578B 4-7
Tektronix, Model 2794 4-13, 4-14
Power Meter, with Power Sensor
Hewlett-Packard Model 437B, with Power Sensor:
(1mW to 100mW)
HP 8487A (0.01 to 50 GHz)
Power Meter with Power Sensor
Anritsu ML4803A, with Power Sensor:
(1mW to 100mW)
MP716A4 (50 to 75 GHz
Spectracom Corp., Model 8161 4-7
Hewlett-Packard, Model 8116A 4-12, 4-13, 4-14
Hewlett-Packard, Model 33120A 4-12. 4-13. 4-14
Digital Multimeter Resolution: 4-1/2 digits (to 20V) DC Accuracy: 0.002% +2 counts DC Input Impedance: 10 MW AC Accuracy: 0.07% +100 counts
John Fluke, Inc.,Model 8840A, with Option 8840A-09K (True RMS AC)
(to 20 kHz) AC Input Impedance: 1 MW
4-12, 4-13, 4-14

Table 4-1. Recommended Test Equipment for Calibration Procedures (2 of 2)
RECOMMENDED MANUFACTURER/MODEL
PROCEDURE
NUMBER
Scalar Network Analyzer, with RF Detector
560-7K50 (0.01 to 40 GHz) 560-7VA50 (0.01 to 50 GHz) SC5198 (40 to 60 GHz)
Adapter K (male) to 2.4 mm (female) Adapts the Power Sensor, HP 8487A, to the 682XXB/683XXB RF OUTPUT connector (£40 GHz models)
Adapter Adapts the MP716A4 Power Sensor to the ML4803A Power Meter
Adapter WR15 to V (male) Adapts the MP716A4 Power Sensor to the 682XXB/683XXB RF OUTPUT connector (>40 GHz)
ANRITSU, Model 35WR15V 4-12
ANRITSU, Model 41KC-10 4-8, 4-9
ANRITSU, Model 41V-10 4-8, 4-9
Any common source All procedures
Serial Interface Assy Provides serial interface between the PC and the 682XXB/683XXB.
4-8, 4-10 · Anritsu MA4002B 4-12

Table 4-2. Calibration Following Subassembly/RF Component Replacement
Subassembly/RF Component Replaced Perform the Following Calibration(s) in Paragraph(s):
A1, A2 Front Panel Assy None
A3 Reference Loop PCB 4-7
A4 Coarse Loop PCB 4-7
A5 Fine Loop PCB 4-7
A6 Pulse Generator PCB None
A7 YIG Loop PCB None
A8 Function Generator PCB 4-12
A9 PIN Control PCB 4-8, 4-9, 4-10, 4-11
A11 FM PCB 4-13 (and 4-14 if Option 6 is installed)
A12 Analog Instruction PCB 4-7
A13 YIG Driver PCB 4-7
A14 SDM, SQM Driver PCB 4-9, 4-10, 4-11
A15 Regulator PCB None
A16 CPU Interface PCB 4-7
A17 CPU PCB 4-7 thru 4-13. None, if calibration EEPROM reused.
A18 Power Supply PCB None
A19 AC Line Conditioner PCB None
A21 Line Filter/Rectifier PCB None
A21-1 BNC/Aux IO Connector PCB None
0.01 to 2 GHz Down Converter Assy 4-8, 4-9, 4-10, 4-11
0.5 to 2.2 GHz Digital Down Converter Assy 4-8, 4-9, 4-10, 4-11
Switched Filter Assy 4-8, 4-9, 4-10, 4-11
Switched Doubler Module (SDM) 4-8, 4-9, 4-10, 4-11
Source Quadrupler Module (SQM) 4-8, 4-9, 4-10, 4-11
Forward Coupler 4-8, 4-9, 4-10, 4-11
Directional Coupler 4-8, 4-9, 4-10, 4-11
Step Attenuator 4-8, 4-9, 4-10
YIG-tuned Oscillator 4-7

4-6 INITIAL SETUP The 682XXB/683XXB is calibrated using an IBM compatible PC and external test equipment. The PC must have the Windows 3.1 or Win- dows 95 operating system installed and be equipped with a mouse. Initial setup consists of interfacing the PC to the signal generator.

Figure 4-1. PC to 682XXB/683XXB Interconnection for Calibration
Using the ANRITSU P/N T1678 serial interface as- sembly, connect the PC to the 682XXB/683XXB as follows:

Configure the PC with Windows 3.1 operating sys- tem to interface with the 682XXB/683XXB as fol- lows:




Baud Rate 9600 (See Note) Data Bits 8 Stop Bits 1 Parity None Flow Control Xon/Xoff Connector Select connection made during interconnection


Configure the PC with Windows 95 operating sys- tem to interface with the 682XXB/683XXB as fol- lows:






Bits per second 19200 (See Note) Data bits 8 Parity None Stop bits 1 Flow control Xon / Xoff

4-7 PRELIMINARY CALIBRATION
This procedure provides the steps necessary to initially calibrate the coarse loop, fine loop, frequency instruction, and internal DVM cir- cuitry and the 100 MHz reference oscillator of the 682XXB/683XXB.

Figure 4-2. Equipment Setup for Preliminary Calibration
Connect the equipment, shown in Figure 4-2, as fol- lows:
Before beginning this calibration proce- dure, always let the 682XXB/683XXB warm up for a minimum of one hour.

Each of the steps in this procedure provides initial calibration of a specific 682XXB/683XXB circuit or component. To ensure accurate instrument calibra- tion, each step of this procedure must be performed in sequence.
press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.
press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.
press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.
press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.
press .
b. Follow the instructions on the screen.
c. Record step completion on the Test Record.

press .
b. Follow the instructions on the screen. Enter
the value of the frequency counter reading as XXXX MHz.
c Record step completion on the Test Record.
a. If Option 16 (High Stability Time Base) is in-
stalled, disconnect the cable at A3J6.
b. Connect the frequency counter to the 682XXB/
683XXB RF output connector.
press .
d. Follow the instructions on the screen.
e. Reconnect the cable to A3J6, if removed.
f. If Option 16 is installed, use a Phillips screw-
driver and remove the screw on top of the 10 MHz High Stability Crystal Oscillator as- sembly to gain access to the timebase adjust- ment screw.
g. Using the Phillips screwdriver, adjust the ti-
mebase to obtain a frequency counter reading of exactly 10 GHz.
h. Record step completion on the Test Record.
press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.
press . (This calibration can take ap- proximately 2 minutes to complete.)
The $ prompt will appear on the screen when the calibration is complete.
b. Record step completion on the Test Record.

press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.
a. At the $ prompt, type: calterm 787 and
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
press . (The $ prompt will appear on the screen when the calibration data has been stored.)
b. Record step completion on the Test Record.

4-8 SWITCHED FILTER SHAPER CALIBRATION
This procedure provides the steps necessary to adjust the Switched Filter Shaper Amplifier gain to produce a more constant level ampli- fier gain with power level changes.

Figure 4-3. Equipment Setup for Switched Filter Shaper Calibration
Connect the equipment, shown in Figure 4-3, as fol- lows:

Before the Switched Filter Shaper Amplifier can be adjusted, zero calibration of the ALC Log amplifier must be performed to eliminate any DC offsets.
calterm 115 and press .
The $ prompt will appear on the screen when ALC Log amplifier zero calibration is complete. (This can take up to 3 minutes for a 40 GHz unit.)
b. Record step completion on the Test Record.
The following steps in the procedure let you adjust the Switched Filter Limiter DAC which controls the maximum gain of the Switched Filter Shaper Ampli- fier. Each frequency band will be scanned for the maximum unleveled power point before adjustment of the Limiter DAC to ensure that the Shaper Am- plifier is not driven to saturation.
b. From the System Menu display, select RESET.
c. Press CHANNEL 2 DISPLAY: OFF.
d. Press CHANNEL 1 DISPLAY: ON.
e. Press CHANNEL 1 MENU key.
f. From the Channel 1 Menu display, select
POWER.
g. Press OFFSET/RESOLUTION.
h. Set Resolution to 5 dB/Div.
i. Adjust Offset to center the display.

calterm 145 and press .
Shaper Amplifier Saturation Waveform
b. On the 562 Network Analyzer, set the Resolu-
tion to 0.2 dB and adjust the offset to center the top of the triangle waveform on the display.
c. Observe the displayed waveform to determine
whether the Shaper Amplifier is being driven to saturation. This is indicated by a dip in the top of the triangle waveform (Figure 4-4).
Shaper Amplifier Out of Sauration
If the displayed waveform indicates there is no saturation, proceed to step e. If there is a dip in the waveform, go to step d.
0.3 dB
d. On the computer keyboard, use 8, 9, or 0 to
Final Waveform After Adjustment
decrement the value of the DAC’s setting until the top of the triangle waveform starts to be- come rounded (Shaper Amplifier is no longer being driven to saturation). Continue decre- menting until the top of the waveform is 0.3 dB below this point.
Figure 4-4. Limiter DAC Adjustment Waveforms
quency band.
f. Repeat steps b thru e until the DAC has been
checked and adjusted for all frequency bands.
g. Press Q on the keyboard to exit the program.
h. Record step completion on the Test Record.
For Models 682XXB with Firmware Versions 1.01 to 2.37; Models 683XXB with Firmware Versions 1.01 to 2.44; Models 682X5B with Firmware Versions 1.00 to 1.26; and Models 683X5B with Firmware Versions 1.00 to 1.30.
The following steps in the procedure lets you adjust the Switched Filter Shaper DAC which controls the gain of the Switched Filter Shaper Amplifier. Each frequency band will be scanned for the minimum unleveled power point before adjustment of the shaper DAC.
a. Press the System Menu key.
b. From the System Menu display, select RESET.

c. Press CHANNEL 2 DISPLAY: OFF.
d. Press CHANNEL 1 DISPLAY: ON.
e. Press CHANNEL 1 MENU key.
f. From the Channel 1 Menu display, select
POWER.
g. Press OFFSET/RESOLUTION.
b. Set Resolution to 5 dB/Div.
c. Adjust Offset to center the display.
calterm 138 and press .
b. On the computer keyboard, use 1, 2, or 3 to in-
crement and 8, 9, or 0 to decrement the value of the DAC’s setting to adjust the triangle waveform being displayed on the 562 Network Analyzer (Figure 4-5) for the most linear re- sponse.
There will be a delay while the amplifier com- pression point is determined and after each key stroke as the DAC is adjusted.
c. When the DAC has been adjusted for the cur-
rent frequency band, press Q on the keyboard to go to the next frequency band.
d. Repeat steps b. and c. until the DAC has been
Figure 4-5. Shaper DAC Adjustment
adjusted for all frequency bands.
e. Press Q on the keyboard to exit the program.
f. Record step completion on the Test Record.
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
(The $ prompt will appear on the screen when the data has been stored.)
b. Record step completion on the Test Record.

For Models 682XXB with Firmware Version 2.39 and above; Models 683XXB with Firmware Version 2.46 and above; Models 682X5B with Firmware Version 1.28 and above; and Models 683X5B with Firmware Version 1.32 and above.
The following step in the procedure adjusts the Switch Filter Shaper DAC which controls the gain of the Switched Filter Shaper Amplifier. Each fre- quency band will be scanned for the minimum un- leveled power point before automatic adjustment of the shaper DAC.
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
(The $ prompt will appear on the screen when the data has been stored.)
b. Record step completion on the Test Record.

4-9 RF LEVEL CALIBRATION RF level calibration requires the use of an automated test system. A computer-controlled power meter measures the 682XXB/683XXB power output at many frequencies throughout the frequency range of the instrument. Correction factors are then calculated and stored in non-volatile memory (EEPROM) located on the A17 CPU PCB.
This calibration is required following replacement of either the A9 PIN Control PCB, the A10 ALC PCB, the A14 SDM, SQM Driver PCB, the Switched Filter Assembly, the 0.01 to 2 GHz Down Converter Assem- bly, the 0.5 to 2.2 GHz Digital Down Converter Assembly, the Switched Doubler Module (SDM), the Source Quadrupler Module (SQM), the Forward Coupler, the Directional Coupler, or the Step Attenuator.
The RF level calibration software is available from ANRITSU by order- ing:
q Part number 2300-104, Version 2.0 and above, for all 50 GHz
682XXB/683XXB models. q Part number 2300-209, Version 1.00 and above, for 60 GHz and
65 GHz 682XXB/683XXB models.
This calibration program warrants level accuracy specifications from maximum power to –70 dBm. For calibration below –70 dBm, the 682XXB/683XXB must be returned to your ANRITSU service center for calibration. The RF level calibration software comes on a 3.5-inch/ 1.44 Mbyte, MS-DOS formatted floppy disk.
For information concerning test equipment requirements and ordering the automated program, contact your ANRITSU service center (refer to Table 1-5 on page 1-18).

4-10 ALC SLOPE CALIBRATION
Models 683XXB with Firmware Version 1.01 to 2.00 This procedure provides the steps necessary to perform ALC Slope calibration. The ALC Slope DAC is calibrated to adjust for an increas- ing or decreasing output power-vs-output frequency in the analog sweep mode. The ALC Slope DAC has two calibrations—one for fre- quencies 2 GHz and one for frequencies >2 GHz.

Figure 4-6. Equipment Setup for ALC Slope Calibration
Connect the equipment, shown in Figure 4-6, as fol- lows:

The following procedure lets you adjust the ALC Slope DAC to compensate for decreasing output power-vs-frequency for frequencies 2 GHz and for frequencies >2 GHz.
a. Press the System Menu key.
b. From the System Menu display, select RESET.
c. Press CHANNEL 2 DISPLAY: OFF.
d. Press CHANNEL 1 DISPLAY: ON.
e. Press CHANNEL 1 MENU key.
f. From the Channel 1 Menu display, select
TRANSMISSION and SELECT INPUT (NON-RATIO A).
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit L1 and set L1 to 5 dB less than
maximum leveled power for the instrument be- ing calibrated (refer to Table 3-2, page 3-6).
c. Press Step . The Step Sweep Menu is dis-
played.
d. Press Edit F1 and set the F1 start frequency
to 0.01 GHz.
e. Press Edit F2 and set the F2 stop frequency to
2.1 GHz.
f. Press Sweep Ramp . At the resulting Step
Sweep Ramp Menu, press Num of Steps and set the number of steps to 200.
g. Press Step to return to the Step Sweep Menu
display.

a. Press CALIBRATION and follow the menu on
the display.
b. Press AUTOSCALE.
c. Press OFFSET/RESOLUTION and set the
Resolution to 0.2 dB.
calset and press .
b. Type: 1 and press .
c. Select item 10 (ALC Slope Het) from the menu.
Follow the instructions presented at the bot- tom of the screen to enter a new DAC setting value.
d. Enter different DAC setting values to find the
setting that adjusts the slope so that the power at the start and stop frequencies match the normalized straight line in step sweep mode.
e. When finished, exit the program.
menu. (The $ prompt will appear on the screen.)
g. Record step completion on the Test Record.
a. Reset the instrument by pressing SYSTEM
then Reset . Upon reset the CW Menu is dis- played.
b. Press Edit L1 and set L1 to 5 dB less than
maximum leveled power for the instrument be- ing calibrated (refer to Table 3-2, page 3-6).
c. Press Step . The Step Sweep Menu is dis-
played.
d. Press Edit F1 and set the F1 start frequency
to 2.0 GHz.
e. Press Edit F2 and set the F2 stop frequency to upper limit of the 683XXB.

f. Press Sweep Ramp . At the resulting Step
Sweep Ramp Menu, press Num of Steps and set the number of steps to 200.
g. Press Step to return to the Step Sweep Menu
display.
a. Press CALIBRATION and follow the menu on
the display.
b. Press AUTOSCALE.
c. Press OFFSET/RESOLUTION and set the
Resolution to 0.2 dB.
calset and press .
b. Type 1 and press .
c. Select item 11 (ALC Slope Main) from the
menu. Follow the instructions presented at the bottom of the screen to enter a new DAC set- ting value.
d. Enter different DAC setting values to find the
setting that adjusts the slope so that the power at the start and stop frequencies match the normalized straight line in step sweep mode.
e. When finished, exit the program.
menu. (The $ prompt will appear on the screen.)
g. Record step completion on the Test Record.
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
(The $ prompt will appear on the screen when the data has been stored.)
b. Record step completion on the Test Record.

Models 683XXB with Firmware Version 2.00 and above (Models 683X5B with Firmware Version 1.00 and above) This procedure provides the steps necessary to perform ALC Slope calibration. The ALC Slope is calibrated to adjust for decreasing out- put power-vs-output frequency in full band analog sweep.

Figure 4-7. Equipment Setup for ALC Slope Calibration
Connect the equipment, shown in Figure 4-7, as fol- lows:

The following procedure lets you adjust the ALC Slope over individual frequency ranges to compen- sate for decreasing output power-vs-frequency in analog sweep.
The procedure begins by letting you adjust the ALC Slope for band 0 (0.01 to 2.0 GHz), if installed. It then continues letting you adjust the ALC Slope from 2 GHz to the top frequency of the instrument in up to four bands. The band frequency ranges are:
Band 1 2.0 to 8.4 GHz Band 2 8.4 to 20 GHz Band 3 20.0 to 40.0 GHz Band 4 40.0 to 65.0 GHz
During band 1 thru 3/4 ALC Slope adjustment, the 562 Network Analyzer display (Figure 4-8) shows the response from 2 GHz to the top frequency of the model, as adjustment is done band by band.
b. From the System Menu display, select RESET.
c. Press CHANNEL 2 DISPLAY: OFF.
d. Press CHANNEL 1 DISPLAY: ON.
e. Press CHANNEL 1 MENU key.
f. From the Chanel 1 Menu Display, select
TRANSMISSION and SELECT INPUT (NON-RATIO A).
then Reset . Upon reset the CW Menu is dis- played.
b. Press Step . The Step Sweep Menu is dis-
played.
c. Press FREQUENCY CONTROL, then Full to
select the full frequency range of the unit being calibrated.

8.4 GHz 20 GHz 40 GHz
ALC Slope Before Adjustment
Figure 4-8. ALC Slope Adjustment Waveform Display
d. Press CW/SWEEP SELECT to return to the
Step Sweep Menu display.
e. Press Sweep Ramp . At the resulting Step
Sweep Ramp Menu, press Num of Steps and set the number of steps to 400.
the display.
b. Press AUTOSCALE.
c. Press OFFSET/RESOLUTION and set the Reso-
lution to 0.5 dB.
slpcal and press .
On the computer keyboard, the adjustment keys are:
Slope E (Up) D (Down) (all bands)
Offset Q (Up) A (Down) (band 1-4 only)
Normalized Step Sweep

b. Adjust the ALC Slope so that the power at the
start and stop frequencies (of the analog sweep for band 0) match as closely as possible the normalized straight line in step sweep mode. When completed, press n for the next band.
c. Using the Slope and Offset adjustment keys,
continue until the ALC Slope for all bands has been adjusted.
bration routine. (The $ prompt will appear on the screen.)
e. Record step completion on the Test Record.
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
(The $ prompt will appear on the screen when the data has been stored.)
b. Record step completion on the Test Record.

4-11 ALC BANDWIDTH CALIBRATION
This procedure provides the steps necessary to perform ALC Band- width calibration. The ALC Bandwidth is adjusted to compensate for gain variations of the modulator. The adjustment is performed for each frequency band. This provides a more consistent bandwith throughout the frequency range of the instrument.

Figure 4-9. Equipment Setup for AM Bandwidth Calibration
Connect the equipment, shown in Figure 4-9, as fol- lows:
The following procedure lets you (1) calibrate the ALC bandwith and (2) store the calibration data in non-volatile memory (EEPROMs) on the A17 CPU PCB.
calterm 110 and press .
The following will appear on the PC display: Enter calculated gain X.XX Input:
b. Type: 1.50 and press .

The $ prompt will appear on the screen when the ALC Bandwidth calibration is complete. (This can take up to 15 minutes depending on the frequency range of the 682XXB/683XXB.)
b. Record step completion on the Test Record.
a. At the $ prompt, type: calterm 787 and
press . (The $ prompt will appear on the screen when the calibration data has been stored.)
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
b. Record step completion on the Test Record.

4-12 AM CALIBRATION This procedure provides the steps necessary to perform AM calibra- tion. This consists of calibrating the AM Calibration DAC, the AM Me- ter circuit, and the AM Function Generator. The AM Calibration DAC is calibrated for input sensitivities of 100%/V (linear mode) and 25 dB/V (logarithmic mode) for frequencies 2 GHz and >2 GHz.

Figure 4-10. Equipment Setup for AM Calibration
Connect the equipment, shown in Figure 4-10, as follows:

The following procedure let you (1) adjust the AM Calibration DAC to provide the correct amount of AM in both linear (100%/V sensitivity) and log (25 dB/V sensitivity) modes of operation for frequen- cies of 2 GHz and >2 GHz, (2) calibrate the AM Me- ter circuit, (3) calibrate the AM Function Generator, and (4) store the results in non-volatile memory (EEPROM) on the A17 CPU PCB.
calterm 112 and press .
b. Set the function generator output to 0.00 volts.
When done, press any key on the keyboard to continue calibration.
c. Now, set the function generator to output
0.50 volts. Use the COMPL button on the function generator to toggle the output be- tween +0.50 volts and –0.50 volts.
d. On the computer keyboard, use 1, 2 or 3 to in-
crement and 8, 9 and 0 to decrement the value of the DAC’s setting to obtain a 9.54 dB differ- ence in the power meter’s reading when the function generator’s output is toggled.

the keyboard to exit the program. (If the in- strument has a Down Converter installed, you will be returned to the start of the program to perform this calibration for frequencies of >2 GHz.)
When the DAC has been completely adjusted, the program will exit to the $ prompt.
e. Record step completion on the Test Record.
press .
b. Set the function generator’s output 0.20 volts.
Use the COMPL button to toggle the output between –0.20 volts and +0.20 volts.
c. On the computer keyboard, use 1, 2, or 3 to in-
crement and 8, 9, and 0 to decrement the value of the DAC’s setting to obtain a 10.00 dB differ- ence in the power meter’s reading when the function generator’s output is toggled.
the keyboard to exit the program. (If the in- strument has a Down Converter installed, you will be returned to the start of the program to perform this calibration for frequencies of >2 GHz.)
When the DAC has been completely adjusted, the program will exit to the $ prompt.
e. Record step completion on the Test Record.
press .
b. Set up the Function Generator for a 1 kHz
sinewave with an output level of 0.354 volts RMS (1 volt peak to peak). When done, press any key on the keyboard to continue calibra- tion.
The $ prompt will appear on the screen when the calibration is complete.
c. Record step completion on the Test Record.

press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.
a. At the $ prompt, type: calterm 787 and
press . (The $ prompt will appear on the screen when the calibration data has been stored.)
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
b. Record step completion on the Test Record.

4-13 FM CALIBRATION Models 682XXB/683XXB with Firmware Versions 1.01 to 2.10 Models 682X5B/683X5B with Firmwave Version 1.00 This procedure provides the steps necessary to perform FM calibra- tion. This consists of calibrating the FM Meter circuit, the Offset Bal- ance DAC, and the FM Gain Control DAC. The FM Gain Control DAC is calibrated for input sensitivities in both narrow and wide FM modes.

Figure 4-11. Equipment Setup for FM Calibration
Connect the equipment, shown in Figure 4-11, as follows:

The following steps in the procedure lets you (1) calibrate the FM Meter circuit, (2) calibrate the FM Offset Balance DAC, (3) adjust the FM Gain Control DAC in both narrow and wide modes, and (4) store the results in non-volatile memory (EEPROM) on the A17 CPU PCB.
calterm 123 and press .
b. Set up the Function Generator for a 83.3 kHz
sinewave with an output level of 0.707 volts RMS (2 volts peak to peak). When done, press any key to continue calibration.
The $ prompt will appear on the screen when the calibration is complete.
c. Record step completion on the Test Record.
calterm 126 and press . (The $ prompt will appear on the screen when the calibration is complete.)
b. Record step completion on the Test Record.

Perform the calibration as follows:
press .
b. Set up the Function Generator for a 0.4 Hz
square wave with an output level of 2 volts peak to peak.
c. On the Spectrum Analyzer, set the Span/Div to
50 MHz per division.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing the ‘ key.
e. While observing the Spectrum Analyzer dis-
play, adjust the value of the DAC’s setting to obtain a 200 MHz peak to peak deviation. This is the coarse adjustment.
f. On the Spectrum Analyzer, set the Span/Div to
5 MHz per division and adjust the center fre- quency control to position the low carrier at the center of the display. Note the frequency reading.
g. Now, adjust the center frequency control to po-
sition the high carrier at the center of the dis- play. Note the frequency reading.
h. The difference between these two frequencies
is the actual peak-to-peak frequency deviation. It should be 200 MHz 8 MHz. If not, fine ad- just the value of the DAC’s setting to obtain this deviation.
keyboard to go to the next calibration step (ad- justing the DAC to obtain 20 MHz deviation).
j. Start calibration by pressing the 1 key.
k. While observing the Spectrum Analyzer dis-
play, adjust the value of the DAC’s setting to obtain a 20 MHz peak to peak deviation. This is the coarse adjustment.
l. On the Spectrum Analyzer, set the Span/Div to
1 MHz per division and adjust the center fre- quency control to position the low carrier at the center of the display. Note the frequency reading.

m.Now, adjust the center frequency control to po-
sition the high carrier at the center of the dis- play. Note the frequency reading.
n. The difference between these two frequencies
is the actual peak-to-peak frequency deviation. It should be 20 MHz 0.8 MHz. If not, fine ad- just the value of the DAC’s setting to obtain this deviation.
NULL
o. When finished setting the DAC, press Q on the
keyboard to go to the next calibration step (ad- justing the DAC to obtain 200 MHz deviation at 15 GHz).
When the DAC has been completely adjusted, the program will exit to the $ prompt.
p. Record step completion on the Test Record.
Figure 4-12. Typical Spectrum Analyzer Display
of Bessel Null on FM Waveform
press .
b. Set up the Function Generator for a 83.3 kHz
sine wave with an output level of 0.707 volts RMS (2 volts peak to peak).
c. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
e. While observing the first Bessel null (Figure
4-12) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.

keyboard to go to the next calibration step.
When the DAC has been completely adjusted, the program will exit to the $ prompt.
g. Record step completion on the Test Record.
press .
The screen will display the following menu: 1) Toggle polarity (this lets you go from a positive deviation to negative deviation.) 2) Change DAC value. 3) Quit.
b. On the Spectrum Analyzer, set the Span/Div to
50 MHz per division.
calibration.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing the ‘ key.
e. While observing the Spectrum Analyzer dis-
play, adjust the value of the DAC’s setting to obtain a 100 MHz change in frequency.
menu, then enter 1 to toggle the polarity.
g. Measure the change in frequency on the Spec-
trum Analyzer. The change in frequency should be 200 MHz. If not, enter 2 and adjust the DAC.
h. Repeat steps 5.f and 5.g as necessary to obtain
a 200 MHz frequency change. When finished adjusting the DAC, press Q to return to the menu.
i. On the Spectrum Analyzer, set the Span/Div to
5 MHz per division and adjust the center fre-

quency control to position the high (or low) car- rier at the center of the display. Note the frequency reading.
j. On the keyboard, enter 1 to toggle the polarity.
k. Now, adjust the center frequency control to po-
sition the low (or high) carrier at the center of the display. Note the frequency reading.
l. The difference between these two frequencies
should be 200 MHz 8 MHz. If not, enter 2 and fine adjust the DAC to obtain this devia- tion.
keyboard to return to the menu then enter 3 to go to the next calibration step (adjusting the DAC to obtain 20 MHz deviation).
to start calibration.
o. While oberving the Spectrum Analyzer display,
adjust the value of the DAC’s setting to obtain a 10 MHz change in frequency.
menu, then enter 1 to toggle the polarity.
q. Measure the change in frequency on the Spec-
trum Analyzer. The change in frequency should be 20 MHz. If not, enter 2 and adjust the DAC.
r. Repeat steps 5.p and 5.q as necessary to obtain
a 20 MHz frequency change. When finished ad- justing the DAC, press Q to return to the menu.
s. On the Spectrum Analyzer, set the Span/Div to
1 MHz per division and adjust the center fre- quency control to position the high (or low) car- rier at the center of the display. Note the frequency reading.
t. On the keyboard, enter 1 to toggle the polarity.
u. Now, adjust the center frequency control to po-
sition the low (or high) carrier at the center of the display. Note the frequency reading.
v. The difference between these two frequencies
should be 20 MHz 0.8 MHz. If not, enter 2 and fine adjust the DAC to obtain this devia- tion.

keyboard to return to the menu. Then, enter 3 to go to the next calibration step (adjusting the DAC to obtain 200 MHz deviation at 15 GHz).
When the DAC has been completely adjusted, the program will exit to the $ prompt.
x. Record step completion on the Test Record.
press .
b. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
c. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
d. While observing the first Bessel null (Figure
4-12, page 4-38) on the Spectrum Analyzer dis- play, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to go to the next calibration step.
When the DAC has been completely adjusted, the program will exit to the $ prompt.
f. Record step completion on the Test Record.
a. At the $ prompt, type: calterm 787 and
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
press . (The $ prompt will appear on the screen when the calibration data has been stored.)
b. Record step completion on the Test Record.

Models 682XXB with Firmware Version 2.18 and above Models 683XXB with Firmwave Version 2.20 and above Models 682X5B/683X5B with Firmware Version 1.07 and above This procedure provides the steps necessary to perform FM calibra- tion. This consists of calibrating the FM Meter circuit and the FM Variable Gain Control DAC. The FM Variable Gain Control DAC is calibrated for input sensitivities in Locked, Locked Low-Noise, Un- locked Narrow, and Unlocked Wide FM modes.

Figure 4-13. Equipment Setup for FM Calibration
Connect the equipment, shown in Figure 4-13, as follows:

The following steps in the procedure lets you cali- brate the (1) FM Meter circuit; (2) FM Variable Gain Linearity; (3) Locked, Locked Low-Noise, Unlocked Narrow, and Unlocked Wide FM Mode Sensitivity; and (4) FM Rear Panel Input Gain, and store the re- sults in non-volatile memary (EEPROM) on the A17 CPU PCB.
calterm 123 and press .
b. Set up the Function Generator for a 100.0 kHz
sine wave with an output level of 0.707 volts 0.01 volts RMS (2 volts peak to peak). Use a frequency counter to verify the output fre- quency of your function generator is set to 1%. When done, press any key on the key- board to continue calibration.
The $ prompt will appear on the screen when the calibration is complete.
c. Record step completion on the Test Record.
calterm 148 and press .
b. Set up the Function Generator for a +1.00 Vdc
output. When done, press any key on the key- board to continue calibration.
The $ prompt will appear on the screen when the calibration is complete.
c. Record step completion on the Test Record.

Perform the calibration as follows:
press .
b. Set up the Function Generator for a 0.1 Hz
square wave with an output level of 1 volt 0.002 volts (2 volts peak to peak).
c. On the Spectrum Analyzer, set the Span/Div to
50 MHz per division.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing the ‘ key.
e. While observing the Spectum Analyzer display,
adjust the value of the DAC’s setting to obtain a 200 MHz peak to peak deviation. This is the coarse adjustment.
f. On the Spectrum Analyzer, set the Span/Div to
5 MHz per division and adjust the center fre- quency control to position the low carrier at the center of the display. Note the frequency reading.
g. Now, adjust the center frequency control to po-
sition the high carrier at the center of the dis- play. Note the frequency reading.
h. The difference between these two frequencies
is the actual peak-to-peak frequency deviation. It should be 200 MHz 0.5 MHz. If not, fine ad- just the value of the DAC’s setting to obtain this deviation.
keyboard to go to the next calibration step (ad- justing the DAC to obtain a 200 MHz deviation at 15 GHz).
When the DAC has been completely adjusted, the program will exit to the $ prompt.
j. Record step completion on the Test Record.

Perform the calibration as follows:
press .
The screen will display the following menu: 1) Toggle polarity (this lets you go from a positive deviation to negative deviation.) 2) Change DAC value. 3) Quit.
b. On the Spectrum Analyzer, set the Span/Div to
50 MHz per division.
calibration.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing the ‘ key.
e. While observing the Spectrum Analyzer dis-
play, adjust the value of the DAC’s setting to obtain a 100 MHz change in frequency.
menu, then enter 1 to toggle the polarity.
g. Measure the change in frequency on the Spec-
trum Analyzer. The change in frequency should be 200 MHz. If not, enter 2 and adjust the DAC.
h. Repeat steps 4.f and 4.g as necessary to obtain
a 200 MHz frequency change. When finished adjusting the DAC, press Q to return to the menu.
i. On the Spectrum Analyzer, set the Span/Div to
5 MHz per division and adjust the center fre- quency control to position the high (or low) carrier at the center of the display. Note the frequency reading.
j. On the keyboard, enter 1 to toggle the polarity.

k. Now adjust the center frequency control to po-
sition the low (or high) carrier at the center of the display. Note the frequency reading.
l. The difference between these two frequencies
should be 200 MHz 0.5 MHz. If not, enter 2 and fine adjust the DAC to obtain this devia- tion.
keyboard to return to the menu, then enter 3 to go to the next calibration step (adjusting the DAC to obtain a 200 MHz deviation at 15 GHz).
When the DAC has been completely adjusted, the program will exit to the $ prompt.
n. Record step completion on the Test Record.
NULL
press .
b. Set up the Function Generator for a 99.8 kHz
sine wave with an output level of 0.707 volts RMS (2 volts peak to peak). Use a frequency counter to verify the output frequency of your function generator is set to 1%.
Figure 4-14. Typical Spectrum Analyzer Display
of Bessel Null on FM Waveform
c. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
e. While observing the first Bessel null (Figure
4-14) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to go to the next calibration step.

When the DAC has been completely adjusted, the program will exit to the $ prompt.
g. Record step completion on the Test Record.
press .
b. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
c. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
d. While observing the first Bessel null (Figure
4-14) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to go to the next calibration step.
When the DAC has been completely adjusted, the program will exit to the $ prompt.
f. Record step completion on the Test Record.
ial cable from the front panel FM IN connector and connect it to the rear panel FM IN connec- tor.
press .

c. Set up the Function Generator for a 99.8 kHz
sine wave with an output level of 0.707 volts RMS (2 volts peak to peak). Use a frequency counter to verify the output frequency of your function generator is set to 1%.
d. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
e. On the computer keyboard, use the ‘, 1, 2. and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
f. While observing the first Bessel null (Figure
4-14) on the Spectum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to exit the calibration routine.
h. Record step completion on the Test Record.
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
press . (The $ prompt will appear on the screen whe the calibration data has been stored.)
b. Record step completion on the Test Record.

4-14 PHASE MODULATION ( M) CALIBRATION
This procedure provides the steps necessary to perform phase modula- tion (M) calibration for 682XXB/683XXBs with Option 6. This con- sists of calibrating the FM Variable Gain Control DAC and the M Flatness DAC for input sensitivities in Narrow and Wide M modes.

Figure 4-15. Equipment Setup for FM Calibration
Connect the equipment, shown in Figure 4-15, as follows:

The following steps in the procedure lets you cali- brate the FM Variable Gain Control DAC and the M Flatness DAC in both Wide and Narrow M modes and store the results in non-volatile memory (EEPROM) on the A17 CPU PCB.
NULL
press .
b. Set up the Function Generator for a 99.8 kHz
sine wave with an output level of 0.707 volts RMS (2 volts peak to peak). Use a frequency counter to verify the output frequency of your function generator is set to 1%.
c. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
Figure 4-16. Typical Spectrum Analyzer Display
d. On the computer keyboard, use the ‘, 1, 2, and
of Bessel Null on FM Waveform
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
e. While observing the first Bessel null (Figure
4-16) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to go to the next calibration step.

When the DAC has been completely adjusted, the program will exit to the $ prompt.
g. Record step completion on the Test Record.
Perform the calibration as follows:
press .
b. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
c. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
d. While observing the first Bessel null (Figure
4-16) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to go to the next calibration step.
When the DAC has been completely adjusted, the program will exit to the $ prompt.
f. Record step completion on the Test Record.
press .
b. Set up the Function Generator for a 400 Hz
sine wave with an output level of 0.707 volts RMS (2 volts peak to peak). Use a frequency counter to verify the output frequency of your function generator is set to 1%.

c. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
e. While observing the first Bessel null (Figure
4-16) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to go to the next calibration step.
When the DAC has been completely adjusted, the program will exit to the $ prompt.
g. Record step completion on the Test Record.
press .
b. Set up the Function Generator for a 99.8 kHz
sine wave with an output level of 0.707 volts RMS (2 volts peak to peak). Use a frequency counter to verify the output frequency of your function generator is set to 1%.
c. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
e. While observing the first Bessel null (Figure
4-16) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keybaord to go to the next calibration step.

When the DAC has been completely adjusted, the program will exit to the $ prompt.
g. Record step completion on the Test Record.
Perform the calibration as follows:
press .
b. On the Spectrum Analyzer, set the Span/Div to
50 kHz per division.
c. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start calibration by pressing an increment key.
d. While observing the first Bessel null (Figure
4-16) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard to go to the next calibration step.
When the DAC has been completely adjusted, the program will exit to the $ prompt.
f. Record step completion on the Test Record.
Perform the calibration as follows:
press .
b. Set up the Function Generator for a 400 Hz
sine wave with an output level of 0.707 volts RMS (2 volts peak to peak). Use a frequency counter to verify the output of your function generator is set to 1%.

c. On the Spectrum Analyzer, set the Span/Div ot
50 kHz per division.
d. On the computer keyboard, use the ‘, 1, 2, and
3 keys to increment and the 7, 8, 9, and 0 keys to decrement the value of the DAC’s setting. Start the calibration by pressing an increment key.
e. While observing the first Bessel null (Figure
4-16) on the Spectrum Analyzer display, adjust the value of the DAC’s setting to reduce the carrier level as low as possible.
keyboard.
The program will exit to the $ prompt.
g. Record step completion on the Test Record.
a. At the $ prompt, type: calterm 787 and
When saving calibration data, turn- ing off the instrument before the $ prompt returns to the screen can cause all stored data to be lost.
press . (The $ prompt will appear on the screen when the calibration data has been stored.)
b. Record step completion on the Test Record.

Table of Contents
5-1 INTRODUCTION 5-3
5-3 ERROR AND WARNING/STATUS MESSAGES. . . 5-3
Self-Test Error Messages 5-3 Normal Operation Error and Warning/Status Messages 5-7
MESSAGE 5-10
5-5 TROUBLESHOOTING TABLES 5-10

The majority of the troubleshooting procedures presented in this chapter require the removal of the instrument covers to gain access to test points on printed circuit boards and other subassemblies.
Hazardous voltages are present inside the 682XXB/683XXB whenever ac line power is connected. Turn off the instrument and remove the line cord before removing any covers or pan- els. Troubleshooting or repair procedures should only be per- formed by service personnel who are fully aware of the potential hazards.
Many subassemblies in the signal generator contain static- sensitive components. Improper handling of these subassem- blies may result in damage to the components. Always observe the static-sensitive component handling precautions described in Chapter 1, Figure 1-3.

5-1 INTRODUCTION This chapter provides information for troubleshooting signal generator malfunctions. The troubleshooting procedures presented in this chap- ter support fault isolation to a replaceable subassembly or RF compo- nent. (Remove and replace procedures for the subassemblies and RF components are found in Chapter 6.)
The recommended test equipment for the troubleshooting procedures presented in this chapter is listed in Chapter 1, Table 1-2 (page 1-12).
During normal operation, the 682XXB/683XXB generates error mes- sages to indicate internal malfunctions, abnormal instrument opera- tions, or invalid signal inputs or data entries. It also displays warning messages to alert the operator to conditions that could result in inac- curate signal generator output. In addition, status messages are dis- played to remind the operator of current menu selections or settings.
The 682XXB/683XXB firmware includes internal di- agnostics that self-test the instrument. These self- test diagnostics perform a brief go/no-go test of most of the instrument PCBs and other internal assem- blies.
You can perform an instrument self-test at any time during normal operation by pressing SYSTEM and then the System Menu soft-key Selftest .
If the signal generator fails self-test, an error mes- sage(s) is displayed on the front panel data display. These error messages describe the malfunction and, in most cases, provide an indication of what has failed. Table 5-1 is a summary listing of the self-test error messages. Included for each is a reference to the troubleshooting table that provides a description of the probable cause(s) and a procedure for identify- ing the failed component or assembly.

Table 5-1. Self-Test Error Messages (1 of 3)
Error 100 DVM Ground Offset Failed
Error 101 DVM Positive 10V Reference
Error 102 DVM Negative 10V Reference
Error 105 Power Supply Voltage(s) out of Regulation
Error 106 Power Supply not Locked
Error 107 Sweep Time Check Failed
Error 108 Crystal Oven Cold
Error 109 The 100MHz Reference is not Locked to the External Reference
Error 110 The 100MHz Reference is not Locked to the High Stability 10MHz Crystal Oscillator
Error 111 Fine Loop Osc Failed
Error 112 Coarse Loop Osc Failed
Error 113 Yig Loop Osc Failed
Error 114 Down Converter LO not Locked
Error 115 Not Locked Indicator Failed
Error 116 FM Loop Gain Check Failed
Error 117 Linearizer Check Failed
Error 118 Switchpoint DAC Failed
Error 119 Center Frequency Circuits Failed

Table 5-1. Self-Test Error Messages (2 of 3)
Error 120 Delta-F Circuits Failed
Error 121 Unleveled Indicator Failed
Error 122 Level Reference Failed
Error 123 Detector Log Amp Failed
Error 124 Full Band Unlocked and Unleveled
Error 125 8.4 – 20 GHz Unlocked and Unleveled
Error 126 2 – 8.4 GHz Unlocked and Unleveled
Error 127 Detector Input Circuit Failed
Error 128 .01 – 2 GHz Unleveled
Error 129 Switched Filter or Level Detector Failed
Error 130 2 – 3.3 GH Switched Filter 5-20 5-47
Error 131 3.3 – 5.5 GH Switched Filter 5-20 5-47
Error 132 5.5 – 8.4 GH Switched Filter 5-20 5-47
Error 133 8.4 – 13.25 GH Switched Filter 5-20 5-47
Error 134 13.25 – 20 GH Switched Filter 5-20 5-47
Error 135 Modulator or Driver Failed
Error 142 Sample and Hold Circuit Failed
Error 143 Slope DAC Failed

Table 5-1. Self-Test Error Messages (3 of 3)
Error 144 RF was Off when Selftest started. Some tests were not performed.
Error 145 AM meter or associated circuitry failed
Error 147 Internal FM circuitry failed
Error 148 Pulse 40 MHz reference circuitry failed
682XXB/683XXB Models with SDM
Error 138 SDM Unit or Driver Failed
Error 139 32 – 40 GHz SDM Section Failed 5-22 5-50
Error 140 25 – 32 GHz SDM Section Failed 5-22 5-50
Error 141 20 – 25 GHz SDM Section Failed 5-22 5-50

When an abnormal condition is detected during op- eration, the 682XXB/683XXB displays an error mes- sage to indicate that the output is abnormal or that a signal input or data entry is invalid. It also dis- plays warning messages to alert the operator to conditions that could cause an inaccurate signal generator output. Status messages to remind the operator of current menu selections or settings are also generated.
Table 5-2 is a summary list of possible error mes- sages that can be displayed during normal opera- tions. Table 5-3 is a summary list of possible warning/status messages.
Table 5-2. Possible Error Messages during Normal Operations (1 of 2)
ERROR Displayed (on the frequency mode title bar) when (1) the output frequency is not phase-locked, (2) an invalid frequency parameter entry causes a frequency range error, or (3) an invalid pulse parameter entry causes a pulse modulation error.
LOCK ERROR Displayed (in the frequency parameters area) when the output frequency is not phase-locked. The frequency accuracy and stability of the RF output is greatly reduced. Normally caused by an internal component failure. Run self-test to verify malfunction.
RANGE Displayed (in the frequency parameters area) when (1) the analog sweep start frequency entered is greater than the stop frequency (683XXB models only), (2) the dF value entered results in a sweep outside the range of the instrument, (3) the step size value entered is greater than the sweep range, or (4) the number of steps entered results in a step size of less than 1 kHz (0.1 Hz with Option 11) or 0.1 dB. Entering valid values usually clears the error.
SLAVE Displayed (in the frequency parameters area of the Master 68XXXB) during master-slave operation in VNA mode when the slave frequency offset value entered results in a CW frequency or frequency sweep outside the range of the slave 68XXXB. Entering a valid offset value clears the error.
ERR Displayed (in the modulation status area) when one or more of the following error conditions occurrs: (1) The external AM modulating signal exceeds the input voltage range. In addition, the message “Reduce AM Input Level” appears at the bottom of the AM status display. (Continued on next page)
Error Message Description

Table 5-2. Possible Error Messages during Normal Operations (2 of 2)
ERR Continued: (2) The external FM modulating signal exceeds the input voltage range. In addition, the message “Reduce FM Input Level” appears at the bottom of the FM status display. (3) A pulse parameter setting is invalid for the current pulse modulation state, as follows:
Single Pulse Mode:
Free Run or Gated Trigger:
Width1 > PRI Delayed Trigger:
Delay1 + Width1 > PRI Doublet Pulse Mode:
Free Run Trigger:
Width1 > Delay2 or Width1 + (Delay2 – Width1) + Width2 > PRI Delayed Trigger:
Width1 > Delay2 or Delay1 + Width1 + (Delay2 – Width1) +Width2 > PRI External Trigger with or without Delay:
Width1 > Delay2 Triplet Pulse Mode:
Free Run Trigger:
Width1 > Delay2 or Width2 > Delay3 or Width1 + (Delay2 – Width1) + Width2 +
(Delay3 – Width2) + Width 3 > PRI Delayed Trigger:
Width1 > Delay2 or Width2 > Delay3 or Delay1 + Width1 + (Delay2 – Width1) + Width2 +
(Delay3 – Width2) + Width 3 > PRI External Trigger with or without Delay:
Width1 > Delay2 or Width2 > Delay3 Quadruplet Pulse Mode:
Free Run Trigger:
Width1 > Delay2 or Width2 > Delay3 or Width3 > Delay4 or Width1 + (Delay2 – Width1) + Width2 + (Delay3 –
Width2) + Width3 + (Delay4 – Width3) + Width4 > PRI Delayed Trigger:
Width1 > Delay2 or Width2 > Delay3 or Width3 > Delay4 or Delay1 + Width1 + (Delay2 – Width1) + Width2 +
(Delay3 – Width2) + Width3 + (Delay4 – Width3) + Width4 > PRI External Trigger with or without Delay:
Width1 > Delay2 or Width2 > Delay3 or Width3 > Delay4
Error Message Description

Table 5-3. Possible Warning/Status Messages during Normal Operation
OVN COLD This warning message indicates that the 100 MHz Crystal oven (or the 10 MHz Crystal oven if Option 16 is installed) has not reached operating temperature. Normally displayed during a cold start of the signal generator. If the message is displayed during normal operation, it could indicate a malfunction. Run self-test to verify.
UNLEVELED Displayed when the RF output goes unleveled. Normally caused by exceeding the specified leveled-power rating. Reducing the power level usually clears the warning message. If the warning message is displayed only when AM is selected ON, the modulating signal may be driving the RF output unleveled. Reducing the modulating signal or adjusting the power level usually clears the warning.
UNLOCKED When Unlocked/Narrow FM or Unlocked/Wide FM is selected ON, this warning message appears indicating that the instrument is not phase-locked during this FM mode of operation.
Reduce Rate This warning message is displayed when the AM Rate or FM Rate is set >100 kHz for a non-sinewave modulating waveform. Amplitude or frequency modulation of the output signal will continue but the modulating waveform may be distorted.
EXT REF This status message indicates that an external 10 MHz signal is being used as the reference signal for the 682XXB/683XXB.
OFFSET This status message indicates that a constant (offset) has been applied to the displayed power level.
SLOPE This status message indicates that a power slope correction has been applied to the ALC.
USER 1...5 This status message indicates that a user level flatness correction power-offset table has been applied to the ALC.
SS MODE This status message indicates that the 682XXB/683XXB has been placed in a source lock mode for operation with a 360B Vector Network Analyzer.
Message Description

5-4 MALFUNCTIONS NOT DISPLAYING AN ERROR MESSAGE
The 682XXB/683XXB must be operating to run self-test. Therefore, malfunctions that cause the instrument to be non-operational do not produce error messages. These problems generally are a failure of the 682XXB/683XXB to power up properly. Table 5-4, beginning on page 5-12, provides troubleshooting procedures for these malfunctions.
Malfunctions related to pulse modulating the RF output do not pro- duce error messages. In those 682XXB/683XXB models that produce RF outputs of >40 GHz, malfunctions related to quadrupling the source signal do not generate error messages. Troubleshooting proce- dures for these malfunctions are included in Table 5-4.
Tables 5-4 through 5-27, beginning on page 5-12, provide procedures for isolating malfunctions to a replaceable subassembly or RF compo- nent. In those cases where any of several subassemblies or RF compo- nents could have caused the problem, subassembly/RF component re- placement is indicated. The recommended replacement order is to replace first the subassemblies/RF components that are most likely to have failed.
Figure 5-1, on the following page, shows the location of the 682XXB/ 683XXB connectors and test points that are called out in the trouble- shooting procedures of Tables 5-4 through 5-27.
Never remove or replace a subassembly or RF component with power applied. Serious damage to the instrument may occur.


Figure 5-1. Top View of the 682XXB/683XXB Showing Connector and Test Point Locations

Table 5-4. Malfunctions Not Displaying an Error Message (1 of 6)
( OPERATE light is OFF )
Normal Operation: When the 682XXB/683XXB is connected to the power source, the OPERATE light should illuminate and the instru- ment should power up.
q If the line fuse is good, go to step 2.
q If the line fuse is defective, replace but do not apply
power. Go to step 2.
q If the voltage is correct, go to step 5.
q If the voltage is incorrect or the line fuse blows, replace
the A21 PCB.
q If the voltage is correct, go to step 8.
q If the voltage is incorrect or the line fuse blows, replace
the A19 PCB.

Table 5-4. Malfunctions Not Displaying an Error Message (2 of 6)
q If the voltage is correct, the Front Panel assembly or the
cable between Motherboard connector A20J22 and the Front Panel assembly may be defective.
q If the voltage is incorrect, the +24V standby power supply
may be loaded down by (1) a shorted oven heater for the 100 MHz reference oscillator located on the A3 PCB, (2) a shorted heater for the optional 10 MHz high stability time base (if installed), or (3) a defective Front Panel assembly.
( OPERATE light is ON )
Normal Operation: When the 682XXB/683XXB is connected to the power source, the OPERATE light should illuminate and the instru- ment should power up.
q If the voltage is correct, go to step 4.
q If the voltage is incorrect or missing, the Front Panel
assembly or the cable between Motherboard connector A20J22 and the Front Panel assembly may be defective.
q If yes, the malfunction may be caused by a failed front
panel circuit. Replace the Front Panel assembly.
q If the LEDs do not toggle or if both LEDs are lit, the prob-
lem may be caused by a malfunction on the A17 CPU PCB.

Table 5-4. Malfunctions Not Displaying an Error Message (3 of 6)
Description: The signal generator does not display any error messages during self-test; however, there is no pulse modulation of the RF out- put.
a. Press SYSTEM, then Reset .
b. Press MODULATION, then Pulse to go to the Internal Pulse Modulation Status display.
c. Press On/Off to turn on pulse modulation.
q If present, go to step 5.
q If not present, go to step 3.
q If present, replace the cable W129.
q If not present, replace the A6 PCB.
Step 7. Using the oscilloscope, verify the presence of (1) a pulse train at A6J7 (pulse signal to the Switched Filter) with an output level of +2 volts to –2 volts, a pulse period of 1 ms, and a pulse width of 200 s; and (2) a pulse train at A6J6 (sample and hold signal to the A10 ALC PCB) with an output level of 0 volts to +1 volt, a pulse period of 1 ms, and a pulse width of 200 s.

Table 5-4. Malfunctions Not Displaying an Error Message (4 of 6)
q If the signals are present and correct, go to step 8.
q If the signals are not present or are incorrect, replace the
A6 PCB.
q If present, replace the Pulse Switched Filter assembly.
q If not present, replace the cable W123.

Table 5-4. Malfunctions Not Displaying an Error Message (5 of 6)
(>40 GHz) 682XXB/683XXB Models with SQM
Description: The signal generator does not display any error message during self-test; however, there is no or low RF output above 40 GHz.
a. 682XXB Setup: CW/SWEEP SELECT: Step F1: 40.0 GHz F2: 50.0, 60.0, or 65.0 GHz (Model dependent) Number of Steps: 400 L1: –2.0 dBm
683XXB Setup: CW/SWEEP SELECT: Analog F1: 40.0 GHz F2: 50.0, 60.0, or 65.0 GHz (Model dependent) Sweep Time: 0.100 Sec L1: –2.0 dBm
a. For models having a high end frequency of 50 GHz, check the SQM bias voltages at A14TP5 and A14TP7. The bias voltage at A14TP5 should be +10 volts; the bias voltage at A14TP7 should be –5 volts.
For models having a high end frequency of 60 or 65 GHz, check for a +12 volts SQM bias voltage at A14TP6.
b. For all models, check for a –2 volt PIN switch drive volt- age at A9TP22.
q If the SQM bias and PIN switch drive voltages are
correct, go to step 4.
q If the SQM bias voltage(s) is not correct, replace the
A14 PCB.
q If the PIN switch drive voltage is not correct, replace
the A9 PCB.

Table 5-4. Malfunctions Not Displaying an Error Message (6 of 6)
a. Connect the 682XXB/683XXB AUX I/O to the 562 AUX I/O.
b. Connect the 562 DEDICATED GPIB to the 682XXB/ w683XXB IEEE-488 GPIB.
c. Connect the RF Detector to the 562 Channel A Input.
a. Press SYSTEM MENU display.
b. From System Menu display, select RESET.
c. Press CHANNEL 2 DISPLAY: OFF.
d. Press CHANNEL 1 DISPLAY: ON.
e. Press CHANNEL 1 Menu key.
f. From the Channel 1 Menu display, select POWER.
q If the amplitude of the RF signal is correct, replace the
SQM.
q If there is no RF signal or if the amplitude of the RF sig-
nal is low, replace the switched filter assembly.

Table 5-5. Error Messages 100, 101, and 102
Error 100 DVM Ground Offset Failed, or Error 101 DVM Positive 10V Reference, or Error 102 DVM Negative 10V Reference
Description: The DVM circuitry, located on the A16 CPU Interface PCB, is calibrated using the 10 volts from the reference supplies on the A12 Analog Instruction PCB. The error messages indicate a calibration-related problem or a defective 10 volt reference.
q If no error message is displayed, the problem is cleared.
q If any of the error messages, 100, 101, and 102, are dis-
played, go to step 3.
q If the 10V reference voltages are correct, go to step 5.
q If incorrect, replace the A12 PCB.
q If no error message is displayed, the problem is cleared.
q If any of the error messages, 100, 101, and 102, are dis-
played, go to step 6.
q If no error message is displayed, the problem is cleared.
q If any of the error messages, 100, 101, and 102, are dis-
played, contact your local ANRITSU service center for assistance.

Table 5-6. Error Messages 105 and 106 (1 of 10)
Voltages hazardous to life are present throughout the power supply circuits, even when the front panel LINE switch is in the STANDBY postion. When performing maintenance, use utmost care to avoid electrical shock.
Error 105 Power Supply Voltage(s) out of Regulation.
Description: The out of regulation circuit, located on the A15 Regula- tor PCB, monitors all of the regulated power supply outputs. This error message indicates that one of more of the voltages from the power sup- ply, with the exception of the 5 volt supply, is out of regulation. If the 5 volt supply is faulty, the 682XXB/683XXB will not operate.
q If incorrect for a supply, go to the referenced step.
q If incorrect for several supplies, go to step 2.
q If all voltages are correct, go to step 3.
Table 5-7. Regulated Power Supply Voltages
Point Value Refer to Step
+15VG A15 TP3 A15 TP1 +15V ±0.75V 4
–15VG A15 TP9 A15 TP1 –15V ±0.75V 4
+15VA A15 TP2 A15 TP1 +15 ±0.75V 5
–15VA A15 TP7 A15 TP1 –15V ±0.75V 5
+15VLP A15 TP12 A15 TP1 +15V ±0.75V 6
–15VLP A15 TP13 A15 TP1 –15V ±0.75V 6
+15VFM A15 TP8 A15 TP1 +15V ±0.75V 7
–15VFM A15 TP11 A15 TP1 –15V ±0.75V 7
–18VT A15 TP10 A15 TP1 –18.06V ±0.36V 8
–43VT A15 TP15 A15 TP1 –43.1V ±0.9V 9
+24VH A15 TP6 A15 TP1 +24.32V ±0.5V 10

Table 5-6. Error Messages 105 and 106 (2 of 10)
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the regulated voltages per Table 5-7.
q If the voltages are correct, the problem is cleared.
q If the voltages are incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Replace the A18 PCB.
f. Place the LINE switch to OPERATE and measure the regulated voltages per Table 5-7.
q If the voltages are correct, the problem is cleared.
q If the voltages are still incorrect, contact your local
ANRITSU service center for assistance.
q If no error message is displayed, the problem is cleared.
q If error 105 displays again, contact your local ANRITSU
service center for assistance.

Table 5-6. Error Messages 105 and 106 (3 of 10)
This supply provides 15 volts to the YIG Driver; SDM, SQM Driver; CPU; and CPU I/O circuits andw the Switched Filter and Down Con- verter assemblies.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the 15VG voltages per Table 5-7.
q If the voltages are correct, the problem is cleared.
q If the voltages are incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Remove the A13, A14, A16, and A17 PCBs as applicable for your model.
f. Place the LINE switch to OPERATE and measure the 15VG voltages.
q If the voltages are correct, go to step j.
q If the voltages are still incorrect, go to step g.
g. Place the LINE switch to STANDBY.
h. Remove the switched filter and down converter assem- blies.
i. Place the LINE switch to OPERATE and measure the 15VG voltages.
q If the voltages are correct, go to step j.
q If the voltages are still incorrect, contact your local
ANRITSU service center for assistance.
j. Place the LINE switch to STANDBY, then install one of the removed PCBs/assemblies.
k. Place the LINE switch to OPERATE and measure the 15VG voltages.
l. Continue steps j and k until the faulty PCB/assembly is located.

Table 5-6. Error Messages 105 and 106 (4 of 10)
This supply provides 15 volts to the Function Generator, PIN Control, ALC, FM, and Analog Instruction circuits.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the 15VA voltages per Table 5-7.
q If the voltages are correct, the problem is cleared.
q If the voltages are incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Remove the A8, A9, A10, A11, and A12 PCBs.
f. Place the LINE switch to OPERATE and measure the 15VA voltages.
q If the voltages are correct, go to step g.
q If the voltages are still incorrect, contact your local
ANRITSU service center for assistance.
g. Place the LINE switch to STANDBY, then install one of the removed PCBs.
h. Place the LINE switch to OPERATE and measure the 15VA voltages.
i. Continue steps g and h until the faulty PCB is located.

Table 5-6. Error Messages 105 and 106 (5 of 10)
This supply provides 15 volts to the Pulse Generator and the Refer- ence, Coarse, Fine, and YIG Loop circuits.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the 15VLP voltages per Table 5-7.
q If the voltages are correct, the problem is cleared.
q If the voltages are incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Remove the A3, A4, A5, A6, and A7 PCBs.
f. Place the LINE switch to OPERATE and measure the 15VLP voltages.
q If the voltages are correct, go to step g.
q If the voltages are still incorrect, contact your local
ANRITSU service center for assistance.
g. Place the LINE switch to STANDBY, then install one of the removed PCBs.
h. Place the LINE switch to OPERATE and measure the 15VLP voltages.
i. Continue steps g and h until the faulty PCB is located.

Table 5-6. Error Messages 105 and 106 (6 of 10)
This supply provides 15 volts to the FM circuits.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the 15VFM voltages per Table 5-7.
q If the voltages are correct, the problem is cleared.
q If the voltages are incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Remove the A11 and A13 PCBs.
f. Place the LINE switch to OPERATE and measure the 15VFM voltages.
q If the voltages are correct, go to step g.
q If the voltages are still incorrect, contact your local
ANRITSU service center for assistance.
g. Place the LINE switch to STANDBY, then install on the remove PCBs.
h. Place the LINE switch to OPERATE and measure the 15VFM voltages.
i. Continue steps g and h until the faulty PCB is located.

Table 5-6. Error Messages 105 and 106 (7 of 10)
This supply provides –18 volts to the front panel LCD contrast circuit and to drive the YIG-tuned oscillator main tuning coil during CW and slow analog sweeps.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the –18VT voltage per Table 5-7.
q If the voltage is correct, the problem is cleared.
q If the voltage is incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Remove the A13 PCB and the YIG-tuned oscillator as- sembly.
f. Place the LINE switch to OPERATE and measure the –18VT voltage.
q If the voltage is correct, go to step i.
q If the voltage is still incorrect, go to step g.
g. Replace the front panel assembly by following the steps in paragraph 6-3 of Chapter 6—Removal and Replace- ment Procedures.
h. Place the LINE switch to OPERATE and measure the –18VT voltage.
q If the voltage is correct, the problem is cleared.
q If the voltage is still incorrect, contact your local
ANRITSU service center for assistance.
i. Place the LINE switch to STANDBY.
j. Install the A13 PCB.
k. Place the LINE switch to OPERATE and measure the –18VT voltage.
q If the voltage is correct, replace the YIG-tuned oscilla-
tor assembly.
q If the voltage is incorrect, replace the A13 PCB.

Table 5-6. Error Messages 105 and 106 (8 of 10)
This supply provides –43 volts to drive the YIG-tuned oscillator main tuning coil during bandswitching, fast analog sweeps, and digital sweeps.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE.
d. Set up the 682XXB/683XXB to perform a 2 to 20 GHz step sweep.
e. Measure the –43VT voltage per Table 5-7.
q If the voltage is correct, the problem is cleared.
q If the voltage is incorrect, go to step f.
f. Place the LINE switch to STANDBY.
g. Remove the A13 PCB and the YIG-tuned oscillator assembly.
h. Place the LINE switch to OPERATE and measure the –43VT voltage.
q If the voltage is correct, go to step i.
q If the voltage is still incorrect, contact your local
ANRITSU service center for assistance.
i. Place the LINE switch to STANDBY.
j. Install the A13 PCB.
k. Place the LINE switch to OPERATE and measure the –43VT voltage.
q If the voltage is correct, replace the YIG-tuned
oscillator assembly.
q If the voltage is incorrect, replace the A13 PCB.

Table 5-6. Error Messages 105 and 106 (9 of 10)
This supply provides +24 volts for the YIG-tuned oscillator heater, the V/GHz circuit on the A12 PCB, and the coarse, fine, and YIG loop circuits. When the 682XXB/683XXB is switched to OPERATE, it also takes over the function of the 24VS supply and supplies +24 volts to the 100 MHz reference oscillator oven heater, the front panel LINE switch circuitry, and the optional 10 MHz high stability time base oven heater.
the +24VH supply and outlying load circuits.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the +24VH voltage per Table 5-7.
q If the voltage is correct, the problem is cleared.
q If the voltage is incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Remove the A4, A5, A7, and A12 PCBs and the YIG- tuned oscillator assembly.
f. Place the LINE switch to OPERATE and measure the +24VH voltage.
q If the voltage is correct, go to step g.
q If the voltage is still incorrect, contact your local
ANRITSU service center for assistance.
g. Place the LINE switch to STANDBY, then install one of the removed PCBs or the YIG-tuned oscillator assembly.
h. Place the LINE switch to OPERATE and measure the +24VH voltage.
i. Continue steps g and h until the faulty PCB/assembly is located.

Table 5-6. Error Messages 105 and 106 (10 of 10)
Description: The switching power supply is not phase locked to the 400 kHz reference signal from the A6 Pulse Generator PCB. Error 106 may be displayed alone or it may be accompanied by error 148 (Pulse 40 MHz reference circuitry failed).
q If present, replace the A18 Power Supply PCB.
q If not present, replace the A6 PCB.

Table 5-8. Error Messages 108, 109 and 110 (1 of 2)
Description: The oven of the 100 MHz crystal oscillator or the Option 16 high-stability 10 MHz crystal oscillator has not reached operating temperature.
q If error 108 is not displayed, the problem is cleared.
q If error 108 displays and Option 16 is not installed,
replace the A3 PCB.
q If error 108 displays and Option 16 is installed, go to
step 2.
q If error 108 is not displayed, replace the Option 16 crystal
oscillator assembly.
q If error 108 is still displayed, replace the A3 PCB.
Description: The reference loop is not phase-locked to the external 10 MHz reference.
q If present, replace the A3 PCB.
q If not present, replace the cable W110.

Table 5-8. Error Messages 108, 109 and 110 (2 of 2)
Error 110 The 100MHz Reference is not Locked to the High Stability 10MHz Crystal Oscillator
Description: The reference loop is not phase-locked to the Option 16 high stability 10 MHz crystal oscillator.
q If present, replace the A3 PCB.
q If not present, replace the Option 16 crystal oscillator
assembly.
Table 5-9. Error Message 111 (1 of 2)
Description: One or more of the oscillators within the fine loop is not phase-locked.
q If present, go to step 9.
q If not present, go to step 3.
q If present, replace the cable W121.
q If not present, go to step 5.

Table 5-9. Error Message 111 (2 of 2)
q If present, replace the A4 PCB.
q If not present, go to step 7.
q If present, replace the cable W105.
q If not present, replace the A3 PCB.
listed in Table 5-10.
q If the signals are correct in both frequency and amplitude,
go to step 12.
q If the signals are incorrect, replace the A5 PCB.
q If error 111 is not displayed, the problem is cleared.
q If error 111 is still displayed, contact your local ANRITSU
service center for assistance.
Table 5-10. Fine Loop Frequencies
10.102 GHz 22 MHz
10.110 GHz 30 MHz
10.120 GHz 40 MHz
682XXB/683XXB CW Frequency · Measured Frequency · at A5J1

Table 5-11. Error Message 112
Description: The coarse loop oscillator is not phase-locked.
q If present, go to step 5.
q If one or both of the signals are not present, go to step 3.
q If the 500 MHz signal is present, replace cable W103.
q If the 10 MHz signal is present, replace cable W105.
q If the signals are not present, replace the A3 PCB.
Table 5-12. Coarse Loop Frequencies
Measured Frequency at A4J3
2.000 GHz 219.5 MHz ±10 kHz
2.050 GHz 225.0 MHz ±10 kHz
q If the signals are correct in both frequency and ampli-
2.225 GHz 244.5 MHz ±10 kHz
tude, go to step 8.
q If the signals are incorrect, replace the A4 PCB.
q If error 112 is not displayed, the problem is cleared.
q If error 112 is still displayed, contact your local
ANRITSU service center for assistance.
682XXB/683XXB CW Frequency

Table 5-13. Error Messages 113 and 115 (1 of 2)
Description: Error 113 indicates that the YIG loop is not phase- locked. Error 115 indicates a failure of the not phased-lock indicator circuit.
q If the coarse loop signals are correct in both frequency
and amplitude, go to step 2.
q If the coarse loop signals are incorrect, replace the A4
PCB.
q If the fine loop signals are correct in both frequency and
amplitude, go to step 3.
q If the fine loop signals are incorrect, replace the A5 PCB.
q If the signal is correct in both frequency and amplitude,
go to step 6.
q If the signals are incorrect, replace the switched filter as-
sembly.

Table 5-13. Error Messages 113 and 115 (2 of 2)
q If error 113 or 115 is not displayed, the problem is cleared.
q If either error 113 or 115 is displayed, contact your local
ANRITSU service center for assistance.
Table 5-14. Error Message 114
Description: The local oscillator in the down converter assembly is not phase-locked.
q If present, go to step 3.
q If not present, replace the A3 PCB.
q If present, replace the down converter assembly.
q If not present, replace cable W112.

Table 5-15. Error Message 116
Description: The FM loop has failed or the loop gain is out of toler- ance.
q If error 116 is not displayed, the problem is cleared.
q If error 116 is still displayed, go to step 3.
q If error 116 is not displayed, the problem is cleared.
q If error 116 is displayed, contact your local ANRITSU serv-
ice center for assistance.

Table 5-16. Error Messages 107, 117, 118, 119, and 120
Error 107 Sweep Time Check Failed Error 117 Linearizer Check Failed Error 118 Switchpoint DAC Failed Error 119 Center Frequency Circuits Failed Error 120 Delta-F Circuits Failed
Description: Each of these error messages indicates a problem in the circuitry on the A12 Analog Instruction PCB that provides frequency tuning voltages for the YIG-tuned oscillator.
q If no error message is displayed, the problem is cleared.
q If any of the error messages, listed above, is displayed, go
to step 3.
q If no error message is displayed, the problem is cleared.
q If any of the error messages, listed above, is displayed,
contact your local ANRITSU service center for assistance.

Table 5-17. Error Messages 121, 122, 123, 127, 142, and 143 (1 of 2)
Error 121 Unleveled Indicator Failed Error 122 Level Reference Failed Error 123 Detector Log Amp Failed Error 127 Detector Input Circuit Failed
Description: Error 121 indicates a failure of the circuit that alerts the CPU whenever the RF output power becomes unleveled. Each of the other error messages indicates a problem in the circuitry on the A10 ALC PCB that provides control of the RF output power level.
q If no error message is displayed, the problem is cleared.
q If any of the error messages, listed above, is displayed,
contact your local ANRITSU service center for assistance.
Description: Error 142 indicates a failure of the sample and hold cir- cuitry on the A10 PCB.
a. Press SYSTEM, then Reset .
b. Press MODULATION, then Pulse to go to the Internal Pulse Modulation Status display.
c. Press On/Off to turn on pulse modulation.
q If the signal is present and correct, reconnect cable W122
to A6J6 and go to step 4.
q If the signal is not present or is incorrect, replace the A6
PCB.

Table 5-17. Error Messages 121, 122, 123, 127, 142, and 143 (2 of 2)
q If present, replace the A10 PCB.
q If not present, replace the cable W122.
Description: Error 143 indicates a problem with the level slope DAC circuitry on the A10 PCB.
q If error 143 is not displayed, the problem is cleared.
q If error 143 is still displayed, go to step 3.
q If error 143 is not displayed, the problem is cleared.
q If error 143 is still displayed, go to step 4.
q If error 143 is not displayed, the problem is cleared.
q If error 143 is still displayed, contact your local ANRITSU
service center for assistance.

Table 5-18. Error Messages 124, 125, and 126 (1 of 2)
Error 124 Full Band Unlocked and Unleveled Error 125 8.4-20 GHz Unlocked and Unleveled Error 126 2-8.4 GHz Unlocked and Unleveled
Description: These error messages indicate either a failure of one or both of the oscillators in the 2-20 GHz YIG-tuned oscillator assembly.
a. Connect the 682XXB/683XXB AUX I/O to the 562 AUX I/O.
b. Connect the 562 DEDICATED GPIB to the 682XXB/ 683XXB IEEE-488 GPIB.
c. Connect the RF Detector to the 562 Channel A Input.
a. Press SYSTEM MENU key.
b. From System Menu display, select RESET.
c. Press CHANNEL 2 DISPLAY: OFF.
d. Press CHANNEL 1 DISPLAY: ON.
e. Press CHANNEL 1 MENU key.
f. From the Channel 1 Menu display, select POWER.

Table 5-18. Error Messages 124, 125, and 126 (2 of 2)
q If the RF signal is correct in both frequency and ampli-
tude throughout the full sweep, go to step 8.
q If there is no RF signal for all or part of the sweep or if
the amplitude of the RF signal is low, go to step 5.
q If the ramp signal is correct, go to step 7.
q If the ramp signal is incorrect or not present, replace the
A13 PCB.
q If the YIG bias voltages are correct, replace the YIG-
tuned oscillator assembly.
q If the YIG bias voltages are incorrect, replace the A13
PCB.
q If no error message is displayed, the problem is cleared.
q If any of the error messages, listed above, are displayed,
contact your local ANRITSU service center for assistance.
Table 5-19. YIG-tuned Oscillator Bias Voltages
YIG-tuned Oscillator Bias Voltages
2 to 8.4 GHz 8.4 to 20 GHz
A13TP5 0V +8V
A13TP6 –5V –5V
A13TP7 +8V 0V
Test Point

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (1 of 8)
(0.01 to 20 GHz)
Description: Error 128 indicates a failure of of the down converter leveling circuitry. The 682XXB/683XXB may or may not produce an RF output in the 0.01 to 2 GHz frequency range. Thus, there are two trou- bleshooting paths for this problem—unleveled with output power and unleveled with no/low output power.
Unleveled with output power (The warning message UNLEVELED appears on the front panel display):
a. 682XXB Setup: CW/SWEEP SELECT: Step F1: 0.010 GHz F2: 2.000 GHz Number of Steps: 400 L1: +1.00 dBm
683XXB Setup: CW/SWEEP SELECT: Analog F1: 0.010 GHz F2: 2.000 GHz Sweep Time: 0.100 Sec L1: +1.00 dBm
q If the warning message UNLEVELED no longer appears
on the front panel display, replace the down converter.
q If the warning message UNLEVELED is still displayed,
replace the A10 PCB.

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (2 of 8)
a. 682XXB Setup: CW/SWEEP SELECT: Step F1: 0.010 GHz F2: 2.000 GHz Number of Steps: 400 L1: +1.00 dBm
683XXB Setup: CW/SWEEP SELECT: Analog F1: 0.010 GHz F2: 2.000 GHz Sweep Time: 0.100 Sec L1: +1.00 dBm
q If the detector voltage is correct, replace the A10 PCB.
q If the detector voltage is incorrect, go to step 5.
q If the bias voltage is correct, go to step 6.
q If the bias voltage is not correct, replace the A13 PCB.
q If the PIN switch drive voltage(s) is correct, go to step 7.
q If the PIN switch drive voltage(s) is not correct, replace
the A9 PCB.

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (3 of 8)
a. Connect the 682XXB/683XXB AUX I/O to the 562 AUX I/O.
b. Connect the 562 DEDICATED GPIB to the 682XXB/ 683XXB IEEE-488 GPIB.
c. Connect the RF Detector to the 562 Channel A Input.
a. Press SYSTEM MENU key.
b. From System Menu display, select RESET.
e. Press CHANNEL 1 MENU key.
f. From the Channel 1 Menu display, select POWER.
q If the amplitude of the RF signal is correct, replace the
down converter assembly.
q If there is no RF signal or if the amplitude of the RF sig-
nal is low, replace the switched filter assembly.

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (4 of 8)
Description: Error 129 indicates a failure of either the switched filter or level detector circuitry. The 682XXB/683XXB may or may not pro- duce an RF output in the 2 to 20 GHz frequency range. Thus, there are two troubleshooting paths for this problem—unleveled with output power and unleveled with no/low output power.
Unleveled with output power (The warning message UNLEVELED appears on the front panel display):
a. 682XXB Setup: CW/SWEEP SELECT: Step F1: 0.010 GHz F2: 20.000 GHz Number of Steps: 400 L1: +1.00 dBm
683XXB Setup: CW/SWEEP SELECT: Analog F1: 0.010 GHz F2: 20.000 GHz Sweep Time: 0.100 Sec L1: +1.00 dBm
q If the warning message UNLEVELED no longer appears
on the front panel display, replace the directional coupler.
q If the warning message UNLEVELED is still displayed,
replace the A10 PCB.

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (5 of 8)
a. 682XXB Setup: CW/SWEEP SELECT: Step F1: 0.010 GHz F2: 20.000 GHz Number of Steps: 400 L1: +1.00 dBm
683XXB Setup: CW/SWEEP SELECT: Analog F1: 0.010 GHz F2: 20.000 GHz Sweep Time: 0.100 Sec L1: +1.00 dBm
q If the bias and the PIN switch drive voltages are correct,
go to step 4.
q If the bias voltages are not correct, replace the A13 PCB.
q If the PIN switch drive voltage is not correct, replace the
A9 PCB.
a. Connect the 682XXB/683XXB AUX I/O to the 562 AUX I/O.
b. Connect the 562 DEDICATED GPIB to the 682XXB/ 683XXB IEEE-488 GPIB.
c. Connect the RF Detector to the 562 Channel A Input.

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (6 of 8)
a. Press SYSTEM MENU key.
b. From System Menu display, select RESET.
c. Press CHANNEL 2 DISPLAY: OFF.
d. Press CHANNEL 1 DISPLAY: ON.
e. Press CHANNEL 1 MENU key.
f. From the Channel 1 Menu display, select POWER.
q If the amplitude of the RF signal is correct, check for bad
cables.
q If there is no RF signal or if the amplitude of the RF sig-
nal is low, replace the switched filter assembly.

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (7 of 8)
Error 130 2-3.3 GH Switched Filter Error 131 3.3-5.5 GH Switched Filter Error 132 5.5-8.4 GH Switched Filter Error 133 8.4-13.25 GH Switched Filter Error 134 13.25-20 GH Switched Filter
Description: Each of these error messages indicates a failure in a switched filter path within the switched filter assembly. The 682XXB/ 683XXB may or may not produce an RF output in the frequency range of the failed switched filter path.
q If the PIN switch drive voltages are correct, replace the
switched filter assembly.
q If the PIN switch drive voltages are incorrect, replace the
A9 PCB.
Table 5-21. Switched Filter PIN Switch Drive Voltages
Test Point Active Frequency
A9TP18 2 to 3.3 GHz –2V +1V
A9TP10 3.3 to 5.5 GHz –2V +1V
A9TP12 5.5 to 8.4 GHz –2V +1V
A9TP16 8.4 to 13.25 GHz –2V +1V
A9TP21 13.25 to 20 GHz –2V +1V
A9TP17 2 to 8.4 GHz –2V +2V

Table 5-20. Error Messages 128, 129, 130, 131, 132, 133, 134, and 135 (8 of 8)
Description: Error 135 indicates a failure of the modulator in the switched filter assembly or the modulator driver circuitry on the A9 PIN Control PCB.
q If error 135 is not displayed, the problem is cleared.
q If error 135 is still displayed, go to step 2.
q If error 135 is not displayed, the problem is cleared.
q If error 135 is still displayed, contact your local ANRITSU
service center for assistance.

Table 5-22. Error Messages 138, 139, 140, and 141 (1 of 2)
(20 to 40 GHz) 682XXB/683XXB Models with SDM
Description: Error 138 indicates a failure of the SDM, a failure of the SDM bias regulator circuitry on the A14 SDM Driver PCB, or a failure of the frequency band selection circuitry on the A12 Analog Instruction PCB. The 682XXB/683XXB will not produce an RF output in the 20 to 40 GHz frequency range.
a. 682XXB Setup: CW/SWEEP SELECT: Step F1: 20.000 GHz F2: 40.000 GHz Number of Steps: 400 L1: +1.00 dBm
683XXB Setup: CW/SWEEP SELECT: Analog F1: 20.000 GHz F2: 40.000 GHz Sweep Time: 0.100 Sec L1: +1.00 dBm
q If the SDM bias voltage is correct, replace the SDM.
q If the SDM bias voltage is not correct, go to step 4.
q If error 138 is not displayed, the problem is cleared.
q If error 138 is still displayed, go to step 5.
q If error 138 is not displayed, the problem is cleared.
q If error 138 is still displayed, contact your local ANRITSU
service center for assistance.

Table 5-22. Error Messages 138, 139, 140, and 141 (2 of 2)
Error 139 32-40 GHz SDM Section Failed Error 140 25-32 GHz SDM Section Failed Error 141 20-25 GHz SDM Section Failed
Description: Each of these error messages indicates a failure in a switched doubler filter path within the SDM. The 682XXB/683XXB will not produce an RF output in the frequency range of the failed switched doubler filter path.
a. 682XXB Setup: CW/SWEEP SELECT: Step F1: 2.000 GHz F2: 40.000 GHz Number of Steps: 400 L1: +1.00 dBm
683XXB Setup: CW/SWEEP SELECT: Analog F1: 2.000 GHz F2: 40.000 GHz Sweep Time: 0.100 Sec L1: +1.00 dBm
q If the PIN switch drive voltages are correct, replace the
SDM.
q If the PIN switch drive voltages are not correct, replace
the A9 PCB.
Table 5-23. SDM PIN Switch Drive Voltages
Test Point Active Frequency
A9TP9 0.01 to 20 GHz +20V –15V
A9TP11 20 to 25 GHz +20V –15V
A9TP15 25 to 32 GHz +20V –15V
A9TP24 32 to 40 GHz +20V –15V

Table 5-24. Error Message 144
Error 144 RF was Off when Selftest started. Some tests where not performed
Description: Indicates that some self-tests where not performed be- cause the RF Output was selected OFF on the front panel.
Table 5-25. Error Message 145
Description: Indicates a failure of the internal amplitude modulation function. The 682XXB/683XXB may or may not provide amplitude modulation of the RF output signal using modulating signals from an external source.
a. Press SYSTEM, then Reset .
b. Press MODULATION, then AM to go to the Internal AM Status display.
c. Press On/Off to turn internal amplitude modulation on.
q If present, replace the A10 PCB.
q If not present, replace the A8 PCB.

Table 5-26. Error Message 147
Description: Indicates a failure of the internal frequency modulation function. The 682XXB/683XXB may or may not provide frequency modulation of the RF output signal using modulating signals from an external source.
a. Press SYSTEM, then Reset .
b. Press MODULATION, then FM to go to the Internal FM Status display.
c. Press On/Off to turn internal frequency modulation on.
q If present, replace the A11 PCB.
q If not present, replace the A8 PCB.
Table 5-27. Error Message 148 (1 of 2)
Error 148 Pulse 40 MHz reference circuitry failed.
Description: Indicates a failure of the pulse generator 40 MHz oscilla- tor circuitry. The pulse generator may still function; however, the 40 MHz oscillator is not phase locked to the 10 MHz reference time- base. The pulse modulation function may or may not operate. Error 106 (Power Supply not Phase-locked) may also be displayed.
q If present, replace the A6 PCB.
q If not present, go to step 3.
q If present, replace the cable W108.
q If not present, go to step 5.

Table 5-27. Error Message 148 (2 of 2)
q If present, replace the A5 PCB.
q If not present, go to step 7.
q If present, replace the cable W121.
q If not present, go to step 9.
signal at the end of cable W105.
q If present, replace the A4 PCB.
q If not present, go to step11.
signal at A3J4.
q If present, replace the cable W105.
q If not present, replace the A3 PCB.



Table of Contents
6-1 INTRODUCTION 6-3
COVERS 6-4 Preliminary 6-4 Procedure 6-4
PANEL ASSEMBLY 6-6 Preliminary 6-6 Procedure 6-6
OR A6 PCB 6-8 Preliminary 6-8 Procedure 6-8
Preliminary 6-10 Procedure 6-10
A11, OR A12 PCB 6-10 Preliminary 6-10 Procedure 6-10
OR A15 PCB 6-11 Preliminary 6-11 Procedure 6-11

A17 PCB 6-11 Preliminary 6-11 Procedure 6-11
A19 PCB 6-12 Preliminary 6-12 Procedure 6-12
PANEL ASSEMBLY 6-13 Preliminary 6-13 Procedure 6-13
Preliminary 6-16 Procedure 6-16
Preliminary 6-17 Procedure 6-17
ASSEMBLY 6-18 Preliminary 6-18 Procedure 6-18

6-1 INTRODUCTION This chapter provides procedures for gaining access to the major 682XXB/683XXB assemblies, subassemblies, and components for troubleshooting or replacement.
Hazardous voltages are present inside the 682XXB/683XXB whenever ac line power is connected. Turn off the unit and remove the line cord before removing any covers or panels. Troubleshooting and repair procedures should only be per- formed by service personnel who are fully aware of the po- tential hazards.
Many subassemblies in the signal generator contain static- sensitive components. Improper handling of these subas- semblies may result in damage to the components. Always observe the static-sensitive component handling procedures described in Chapter 1, Figure 1-4.

6-2 REMOVING AND REPLACING THE CHASSIS COVERS
Troubleshooting procedures require removal of the top cover. Replace- ment of some 682XXB/683XXB assemblies and parts require removal of all covers. The following procedure describes this process.
Preliminary Disconnect the power cord from the unit.
Procedure Remove and replace the chassis covers as follows:
Step 1 Using a Phillips screwdriver, remove the screws and the two feet from the top cor- ners at the rear of the instrument (Figure 6-1).
Step 2 Remove the screw that fastens the top cover to the chassis. (The screw is located at the rear of the instrument.)
Step 3 Slide the top cover out along the grooves in the chassis and set it aside.
Step 4 Turn the instrument over so that the bot- tom cover is on top.
Step 5 Remove the screws and the two feet from the bottom corners at the rear of the in- strument.
Step 6 Remove the screw that fastens the bottom cover to the chassis. (The screw is located at the rear of the instrument.)
Step 7 Slide the bottom cover out along the grooves in the chassis and set it aside.
Step 8 Turn the instrument over to return it to the upright position.
Step 9 Remove the screws and the carrying han- dle from the side handle cover. (The two screws fastening the carrying handle through the side handle cover to the chas- sis are accessable by lifting up the rubber covering at the each end of the handle.)
Step 10 Remove the screw that fastens the side handle cover to the chassis. (The screw is located at the rear of the instrument.)
Step 11 Remove the side handle cover and set it aside.

Step 12 Remove the screw that fastens the other side cover to the chassis. (The screw is lo- cated at the rear of the instrument.)
Step 13 Remove the side cover and set it aside.
Step 14 To replace the chassis covers, reverse the procedure used to remove them.

Figure 6-1. Chassis Covers Removal

6-3 REMOVING AND REPLACING THE FRONT PANEL ASSEMBLY
This paragraph provides instructions for removing and replacing the front panel assembly of the 682XXB/683XXB. The front panel assem- bly contains the A1 and A2 Front Panel PCBs. Refer to Figure 6-2 dur- ing this procedure.
Preliminary Disconnect the power cord from the unit and remove the chassis covers as described in paragraph 6-2.
Procedure Remove and replace the front panel assembly as fol- lows:
Step 1 Using a Phillips screwdriver, remove the screws and the front handle assemblies from the instrument. (For instruments not having front handles, remove the screws and the front top and bottom feet from the instrument.)
Step 2 Remove the rotary knob from the front panel by pulling straight out on it.
Step 3 Carefully pull the front panel away from the chassis until the screws attaching the front panel assembly to the chassis are accessable.
Step 4 Remove the screws attaching the front panel assembly to the chassis sides.
Step 5 Disconnect the front panel ribbon cables from connectors J1 and J22 of the Moth- erboard.
Step 6 Turn the instrument upside down.
Step 7 Remove the screw attaching the front panel assembly to the chassis pan.
Step 8 Carefully pull the front panel assembly forward until it is clear of the RF OUT- PUT connector.
Step 9 Disconnect the coaxial cables, W119 and W120, from the front panel assembly con- nectors A2J14 and A2J13 by pulling straight out on the cable connectors.
Step 10 To replace the front panel assembly, re- verse the removal process.


Figure 6-2. Front Panel Assembly Removal

6-4 REMOVING AND REPLACING THE A3, A4, A5, OR A6 PCB
This paragraph provides instructions for removing and replacing the A3 Reference Loop PCB, the A4 Coarse Loop PCB, the A5 Fine Loop PCB, or the A6 Pulse Generator PCB, all of which are located in the RF housing (see Figure 6-3).
Preliminary Disconnect the power cord from the unit and remove the top cover as described in paragraph 6-2.
Procedure Remove and replace the A3, A4, A5, or A6 PCB as follows:
Step 1 Disconnect the coaxial cables from the PCB to be removed by lifting up on the cable connectors.
Step 2 Using a Phillips screwdriver, remove the nine screws that retain the RF housing cover and set aside.
Step 3 Remove the RF housing cover and set aside.
Step 4 Lift up on the edge tabs of the PCB and lift it out of the RF housing.
Step 5 Remove the “O” rings installed on each MCX connector and retain them for use on the replacement PCB.
Step 6 To replace the PCB, reverse the removal process.


Figure 6-3. Assembly and Component Locator Diagram

6-5 REMOVING AND REPLACING THE A7 PCB
This paragraph provides instructions for removing and replacing the A7 YIG Loop PCB, which is located in the main card cage (see Figure 6-3).
Preliminary Disconnect the power cord from the unit and remove the top cover as described in paragraph 6-2.
Procedure Remove and replace the A7 PCB as follows:
Step 1 Remove the main card cage cover and set aside.
Step 2 Using a 516-inch wrench, disconnect coax- ial cable W31 from the Sampler/SRD module on the PCB.
Step 3 Disconnect the coaxial cables at A7J1 and A7J5 by lifting up on the cable connec- tors.
Step 4 Lift up on the edge tabs of the PCB and lift it out of the card cage.
Step 5 To replace the A7 PCB, reverse the re- moval process.
This paragraph provides instructions for removing and replacing the A8 Function Generator PCB, the A9 PIN Control PCB, the A10 ALC PCB, the A11 FM PCB, or the A12 Analog Instruction PCB, all of which are located in the main card cage (see Figure 6-3).
Preliminary Disconnect the power cord from the unit and remove the top cover as described in paragraph 6-2.
Procedure Remove and replace the A8, A9, A10, A11, or A12 PCB as follows:
Step 1 Remove the main card cage cover and set aside.
Step 2 Disconnect any coaxial cables from the PCB to be removed by lifting up on the cable connectors.
Step 3 Lift up on the edge tabs of the PCB and lift it out of the card cage.
Step 4 To replace the PCB, reverse the removal process.

6-7 REMOVING AND REPLACING THE A13, A14, OR A15 PCB
This paragraph provides instructions for removing and replacing the A13 YIG Driver PCB, the A14 SDM, SQM Driver PCB, or the A15 Regulator PCB, all of which are located in the main card cage (see Fig- ure 6-3). Each of these PCB assemblies consists of a PCB and a PCB Heat Sink subassembly.
Preliminary Disconnect the power cord from the unit and remove the top cover as described in paragraph 6-2.
Procedure Remove and replace the A13, A14, or A15 PCB as follows:
Step 1 Remove the main card cage cover and set aside.
Step 2 Lift up on the edge tabs of the PCB and lift it out of the card cage.
Step 3 Using a Phillips screwdriver, remove the two screws that fasten the PCB Heat Sink subassembly to the chassis pan.
Step 4 Lift the PCB Heat Sink subassembly out of the card cage.
Step 5 To replace the PCB, reverse the removal process.
This paragraph provides instructions for removing and replacing the A16 CPU Interface PCB or the A17 CPU PCB, both of which are lo- cated in the CPU housing assembly (see Figure 6-3).
Preliminary Disconnect the power cord from the unit and remove the top cover as described in paragraph 6-2.
Procedure Remove and replace the A16 or A17 PCB as follows:
Step 1 Remove the CPU cover and set aside.
Step 2 Lift up on the edge tabs of the PCB and lift it out of the CPU housing.
Step 3 To replace the PCB, reverse the removal process.

6-9 REMOVING AND REPLACING THE A18 OR A19 PCB
This paragraph provides instructions for removing and replacing the A18 Power Supply PCB or the A19 AC Line Conditioner PCB, both of which are located in the power supply housing assembly (see Figure 6-3).
Preliminary Disconnect the power cord from the unit and remove the top cover as described in paragraph 6-2.
When power is removed from the instrument, always allow five minutes for the capacitors on the A21 Line Filter/Recti- fier PCB to discharge before removing either the A18 or A19 PCB.
Procedure Remove and replace the A18 or A19 PCB as follows:
Step 1 Remove the power supply cover and set it aside.
Step 2 If the A19 PCB is being removed, (1) dis- connect the cable assembly from the A21 Line Filter/Rectifier PCB at A19P2 and (2) remove the Nylatch fastener used to secure the PCB to the housing assembly.
Step 3 Lift up the edge tabs on the PCB and lift it out of the power supply housing.
Step 4 To replace the PCB, reverse the removal process.

6-10 REMOVING AND REPLACING THE REAR PANEL ASSEMBLY
This paragraph provides instructions for removing and replacing the rear panel assembly of the 682XXB/683XXB. The rear panel assembly contains the A21 Line Filter/Rectifier PCB, the A21-1 BNC/AUX I/O Connector PCB, the line module assembly, and the fan assembly. Refer to Figure 6-4 during this procedure.
Preliminary Disconnect the power cord from the unit and remove the chassis covers as described in paragraph 6-2.
When power is removed from the instrument, always allow five minutes for the capacitors on the A21 Line Filter/Recti- fier PCB to discharge before removing the rear panel assem- bly.
Procedure Remove and replace the rear panel assembly as fol- lows:
Step 1 Using a Phillips screwdriver, remove the screw on the top of the rear panel assem- bly that fastens the rear panel assembly to the bracket attached to the power sup- ply housing.
Step 2 Turn the instrument upside down.
Step 3 Remove the three screws from the rear panel assembly. (One screw fastens the rear panel assembly to the bracket at- tached to the power supply housing; the other two screws attach the rear panel as- sembly to the Motherboard.)
Step 4 Return the instrument to the upright po- sition.
Step 5 Remove the screws attaching the rear panel assembly to the chassis sides.
Step 6 Using a 716-inch wrench, disconnect the coaxial cables (W110, W109, and W130) going to the rear panel 10 MHz REF IN, 10 MHz REF OUT, and FM IN BNC con- nectors.
Step 7 Disconnect the coaxial cables W127, W128, and W129 from the A21-1 PCB by lifting up on the cable connectors.
Step 8 Remove the power supply cover and set it aside.

Step 9 Disconnect the cable assembly from the A21 Line Filter/Rectifier PCB at connec- tor P2 on the A19 PCB.
Step 10 Carefully pull the rear panel assembly away from the signal generator chassis until the cable connections to the Mother- board are accessable.
Step 11 Disconnect the fan cable connector from J13 on the Motherboard.
Step 12 Disconnect the A21-1 PCB ribbon cable connector from J14 on the Motherboard.
Step 13 Disconnect the GPIB cable connector from J16 on the Motherboard.
Step 14 Carefully pull the rear panel assembly completely free from the 682XXB/683XXB chassis.
Step 15 To replace the rear panel assembly, re- verse the removal process.


Figure 6-4. Rear Panel Assembly Removal

6-11 REMOVING AND REPLACING THE A21 PCB
This paragraph provides instructions for removing and replacing the A21 Line Filter/Rectifier PCB, which is located on the rear panel as- sembly (see Figure 6-4).
Preliminary Disconnect the power cord from the unit. Remove the chassis covers as described in paragraph 6-2. Remove the rear panel assembly as described in paragraph 6-10.
When power is removed from the instrument, always allow five minutes for the capacitors on the A21 Line Filter/Recti- fier PCB to discharge before removing the rear panel assem- bly.
Procedure Remove and replace the A21 PCB as follows:
Step 1 Using a Phillips screwdriver, remove the four screws that fasten the A21 shield to the A21 PCB.
Step 2 Remove the shield and set it aside.
Step 3 Disconnect the cables connected to P1, P2, and P6 on the PCB.
Step 4 Using a Phillips screwdriver, remove the four screws that fasten the PCB to the rear panel assembly and set aside.
Step 5 Remove the PCB from the rear panel as- sembly.
Step 6 To replace the PCB, reverse the removal process.

6-12 REMOVING AND REPLACING THE A21-1 PCB
This paragraph provides instructions for removing and replacing the A21-1 BNC/AUX I/O Connector PCB, which is located on the rear panel assembly (see Figure 6-4).
Preliminary Disconnect the power cord from the unit. Remove the chassis covers as described in paragraph 6-2. Remove the rear panel assembly as described in paragraph 6-10.
When power is removed from the instrument, always allow five minutes for the capacitors on the A21 Line Filter/Recti- fier PCB to discharge before removing the A21-1 PCB.
Procedure Remove and replace the A21-1 PCB as follows:
Step 1 Using a Phillips screwdriver, remove the four screws that fasten the A21 shield to the A21 PCB.
Step 2 Remove the shield and set it aside.
Step 3 Disconnect the ribbon cable connector from the A21-1 PCB.
Step 4 Using a ANRITSU P/N T1451 tool, remove the dress nuts from the 12 rear panel BNC connectors.
Step 5 Carefully remove the A21-1 PCB from the rear panel assembly.
Step 6 To replace the PCB, reverse the removal process.

6-13 REMOVING AND REPLACING THE FAN ASSEMBLY
This paragraph provides instructions for removing and replacing the fan assembly, which is located on the rear panel assembly (see Figure 6-4).
Preliminary Disconnect the power cord from the unit. Remove the chassis covers as described in paragraph 6-2. Remove the rear panel assembly as described in paragraph 6-10.
When power is removed from the instrument, always allow five minutes for the capacitors on the A21 Line Filter/Recti- fier PCB to discharge before removing the rear panel assem- bly.
Procedure Remove and replace the fan assembly as follows:
Step 1 With the rear panel laying flat, use a Phillips screwdriver to remove the four screws and flat washers that fasten the fan mount to the rear panel.
Step 2 Lift the fan mount, containing the fan as- sembly, from the rear panel assembly.
Step 3 Remove the fan assembly from the fan mount.
Step 4 Clean the honeycomb fan filter as re- quired before replacing the fan assembly.
Step 5 To replace the fan assembly, reverse the removal process.

A-1 INTRODUCTION This appendix provides test records for recording the results of the Performance Verification tests (Chapter 3) and the Calibration proce- dures (Chapter 4). They jointly provide the means for maintaining an accurate and complete record of instrument performance. Test records are provided for all models of the Series 682XXB/683XXB Synthesized Signal Generators. Table A-1provides the location of each test record in this appendix.
Each test record has been customized to cover a particular 682XXB/ 683XXB model. It contains specific references to frequency parameters and power levels that are applicable only to that instrument model and its available options.
We recommend that you make a copy of these pages each time the test procedures are performed. By dating each Test Record copy, a detailed history of instrument performance can be accumulated.
Table A-1. Test Record Index
Model Number Performance Verification Calibration
68237B/68337B A-3 thru A-14 A-15 thru A-18
68245B/68345B A-19 thru A-32 A-33 thru A-36
68247B/68347B A-37 thru A-50 A-51 thru A-54
68253B/68353B A-55 thru A-67 A-69 thru A-72
68255B/68355B A-73 thru A-87 A-89 thru A-92
68259B/68359B A-93 thru A-107 A-109 thru A-112
68263B/68363B A-113 thru A-125 A-127 thru A-130
68265B/68365B A-131 thru A-145 A-147 thru A-150
68269B/68369B A-151 thru A-165 A-167 thru A-170
68275B/68375B A-171 thru A-183 A-185 thru A-188
68277B/68377B A-189 thru A-201 A-203 thru A-206
68285B/68385B A-207 thru A-219 A-221 thru A-224
68287B/68387B A-225 thru A-237 A-239 thru A-242
68295B/68395B A-243 thru A-254 A-255 thru A-258
68297B/68397B A-259 thruA-270 A-271 thru A-274



Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure Fine Loop Test Procedure (Standard 68X37B)
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value **
2.000 000 000
2.000 001 000
3.000 000 000
2.000 002 000
4.000 000 000
2.000 003 000
5.000 000 000
2.000 004 000
6.000 000 000
2.000 005 000
7.000 000 000
2.000 006 000
8.000 000 000
2.000 007 000
9.000 000 000
2.000 008 000
10.000 000 000
2.000 009 000
11.000 000 000
2.000 010 000
12.000 000 000
** Specification for all frequencies listed above is ±100 Hz
13.000 000 000
14.000 000 000
Test Frequency (in GHz) Measured Value ***
15.000 000 000
2.000 000 100
16.000 000 000
2.000 000 200
17.000 000 000
2.000 000 300
18.000 000 000
2.000 000 400
19.000 000 000
2.000 000 500
20.000 000 000
2.000 000 600
* Specification for all frequencies listed above is ±100 Hz
2.000 000 700
2.000 000 800
2.000 000 900
2.000 001 000
*** Specification for all frequencies listed above is ±10 Hz

This test is not applicable to the 68237B/68337B model.
Test Procedure (2 to 10 GHz) Measured Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure (11 to 20 GHz) Measured Value Upper Limit
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure Measured Value Upper Limit
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68237B/68337B without Option 2A Step Attenuator)
Set F1 to 5.0 GHz
* Specification is ±1.0 dB.
+13.0 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+13 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68237B/68337B with Option 2A Step Attenuator)
Set F1 to 5.0 GHz
* Specification is ±1.0 dB.
+11 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+11 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68237B/68337B with Option 15B High Power & without Option 2A Step Attenuator)
Set F1 to 5.0 GHz
* Specification is ±1.0 dB.
+17 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+17 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (typical, not a specification).

3-11 Power Level Accuracy And Flatness Tests (Continued) (Model 68237B/68337B with Option 15B High Power & Option 2A Step Attenuator)
Set F1 to 5.0 GHz
* Specification is ±1.0 dB.
+15 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+15 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68237B/68337B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68237B/68337B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68237B/68337B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s)
Set F1 to 5.0 GHz Set pulse width W1 to 0.5 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68237B/68337B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68237B/68337B rad
Calculate and record the FM accuracy 93.7% % 106.3%

Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68237B with Firmware Version 2.18 and above; 68337B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure Fine Loop Test Procedure (Standard 68X45B)
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value **
1.000 000 000
1.000 001 000
2.000 000 000
1.000 002 000
3.000 000 000
1.000 003 000
4.000 000 000
1.000 004 000
5.000 000 000
1.000 005 000
6.000 000 000
1.000 006 000
7.000 000 000
1.000 007 000
8.000 000 000
1.000 008 000
9.000 000 000
1.000 009 000
10.000 000 000
1.000 010 000
11.000 000 000
** Specifications for all frequencies listed above is ±100 Hz
12.000 000 000
13.000 000 000
Test Frequency (in GHz) Measured Value ***
14.000 000 000
1.000 000 100
15.000 000 000
1.000 000 200
16.000 000 000
1.000 000 300
17.000 000 000
1.000 000 400
18.000 000 000
1.000 000 500
19.000 000 000
1.000 000 600
20.000 000 000
1.000 000 700
* Specification for all frequencies listed above is ±100 Hz.
1.000 000 800
1.000 000 900
1.000 001 000
*** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 500 MHz Record the level of all harmonics of the 500 MHz carrier
1.0 GHz (2nd harmonic)
1.5 GHz (3rd harmonic)
2.0 GHz (4th harmonic)
2.5 GHz (5th harmonic)
Set F1 to 800 MHz Record the level of all harmonics of the 800 MHz carrier:
1.6 GHz (2nd harmonic)
2.4 GHz (3rd harmonic)
3.2 GHz (4th harmonic)
4.0 GHz (5th harmonic)
Set F1 to 1.3 GHz Record the level of all harmonics of the 1.3 GHz carrier:
2.6 GHz (2nd harmonic)
3.9 GHz (3rd harmonic)
5.2 GHz (4th harmonic)
6.5 GHz (5th harmonic)
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.2 GHz (4th harmonic)
10.2 GHz (5th harmonic)

Test Procedure (2.2 to 10 GHz) Measured Value Upper Limit
Set F1 to 2.4 GHz Record the level of all harmonics of the 2.4 GHz carrier:
4.8 GHz (2nd harmonic)
7.2 GHz (3rd harmonic)
9.6 GHz (4th harmonic)
12.0 GHz (5th harmonic)
14.4 GHz (6th harmonic)
16.8 GHz (7th harmonic)
19.2 GHz (8th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure (11 to 20 GHz) Measured Value Upper Limit
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68245B/68345B without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+13 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+13 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68245B/68345B with Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+11 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+11 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68245B/68345B with Option 15B High Power & without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+13 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+13 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68245B/68345B with Option 15B High Power & Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+11 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+11 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68245B/68345B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68245B/68345B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 69245A/69345A MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68245B/68345B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68245B/68345B rad
Calculate and record the FM accuracy 93.7% % 106.3%

Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure Fine Loop Test Procedure (Standard 68X47B)
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value **
1.000 000 000
1.000 001 000
2.000 000 000
1.000 002 000
1.000 003 000
3.000 000 000
4.000 000 000
1.000 004 000
5.000 000 000
1.000 005 000
6.000 000 000
1.000 006 000
7.000 000 000
1.000 007 000
8.000 000 000
1.000 008 000
9.000 000 000
1.000 009 000
10.000 000 000
1.000 010 000
11.000 000 000
** Specifications for all frequencies listed above is ±100 Hz
12.000 000 000
13.000 000 000
Test Frequency (in GHz) Measured Value ***
14.000 000 000
1.000 000 100
15.000 000 000
1.000 000 200
16.000 000 000
1.000 000 300
17.000 000 000
1.000 000 400
18.000 000 000
1.000 000 500
19.000 000 000
1.000 000 600
20.000 000 000
1.000 000 700
* Specification for all frequencies listed above is ±100 Hz.
1.000 000 800
1.000 000 900
1.000 001 000
*** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 10 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 20 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 30 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 40 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 350 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 1.6 GHz Record the presence of the worst case non-harmonic dBc –40 dBc
Set F1 to 1.6 GHz Record the level of the harmonics of the 1.6 GHz carrier:
3.2 GHz (2nd harmonic)
4.8 GHz (3rd harmonic)

Test Procedure (2 to 10 GHz) Measured Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure (11 to 20 GHz) Measured Value Upper Limit
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68247B/68347B without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+13 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 20 GHz).
+13 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68247B/68347B with Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+11 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 20 GHz).
+11 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68247B/68347B with Option 15B High Power & without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+13 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 20 GHz).
+13 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68247B/68347B with Option 15B High Power & Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+11 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz);1.6 dB (0.05 to 20 GHz).
+11 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68247B/68347B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68247B/68347B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68247B/68347B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68247B/68347B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68247B/68347B rad
Calculate and record the FM accuracy 93.7% % 106.3%
Value Upper Limit

Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68247B with Firmware Version 2.18 and above; 68347B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
2.000 000 000
21.000 000 000
3.000 000 000
22.000 000 000
4.000 000 000
23.000 000 000
5.000 000 000
24.000 000 000
6.000 000 000
25.000 000 000
7.000 000 000
26.000 000 000
8.000 000 000
9.000 000 000
10.000 000 000
11.000 000 000
12.000 000 000
13.000 000 000
14.000 000 000
15.000 000 000
16.000 000 000
17.000 000 000
18.000 000 000
19.000 000 000
20.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X53B) Fine Loop Test Procedure (68X53B with Option 11)
Test Frequency (in GHz) Measured Value ** Test Frequency (in GHz) Measured Value ***
2.000 001 000
2.000 000 100
2.000 002 000
2.000 000 200
2.000 003 000
2.000 000 300
2.000 004 000
2.000 000 400
2.000 005 000
2.000 000 500
2.000 006 000
2.000 000 600
2.000 007 000
2.000 000 700
2.000 008 000
2.000 000 800
2.000 009 000
2.000 000 900
2.000 010 000
2.000 001 000
** Specification for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.
This test is not applicable to the 68253B/68353B model.

Test Procedure (2 to 10 GHz) Measured Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure (11 to 20 GHz) Measured Value Upper Limit
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc is Option 15B (High Power) installed.

Test Procedure Measured Value Upper Limit
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68253B/68353B without Option 2A Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 22.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 6 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (2 to 20 GHz); 4.0 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68253B/68353B with Option 2A Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 22.0 GHz
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3.5 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 3.5 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (2 to 20 GHz); 8.2 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68253B/68353B with Option 15B High Power & without Option 2A Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 22.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 10 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 10 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (2 to 20 GHz); 4.0 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68253B/68353B with Option 15B High Power & Option 2A Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 22.0 GHz
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 7.5 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 7.5 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (2 to 20 GHz); 8.2 dB (20 to 26.5 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68253B/68353B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68253B/68353B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68253B/68353B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s)
Set F1 to 5.0 GHz Set pulse width W1 to 0.5 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68253B/68353B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68253B/68353B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68253B with Firmware Version 2.18 and above; 68353B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
21.000 000 000
2.000 000 000
22.000 000 000
3.000 000 000
23.000 000 000
4.000 000 000
24.000 000 000
5.000 000 000
25.000 000 000
6.000 000 000
26.000 000 000
7.000 000 000
8.000 000 000
9.000 000 000
10.000 000 000
11.000 000 000
12.000 000 000
13.000 000 000
14.000 000 000
15.000 000 000
16.000 000 000
17.000 000 000
18.000 000 000
19.000 000 000
20.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X55B) Fine Loop Test Procedure (68X55B with Option 11)
Test Frequency (in GHz) Measured Value ** Test Frequency (in GHz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 500 MHz Record the level of all harmonics of the 500 MHz carrier
1.0 GHz (2nd harmonic)
1.5 GHz (3rd harmonic)
2.0 GHz (4th harmonic)
2.5 GHz (5th harmonic)
Set F1 to 800 MHz Record the level of all harmonics of the 800 MHz carrier:
1.6 GHz (2nd harmonic)
2.4 GHz (3rd harmonic)
3.2 GHz (4th harmonic)
4.0 GHz (5th harmonic)
Set F1 to 1.3 GHz Record the level of all harmonics of the 1.3 GHz carrier:
2.6 GHz (2nd harmonic)
3.9 GHz (3rd harmonic)
5.2 GHz (4th harmonic)
6.5 GHz (5th harmonic)
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.2 GHz (4th harmonic)
10.2 GHz (5th harmonic)

Test Procedure (2.2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.4 GHz Record the level of all harmonics of the 2.4 GHz carrier:
4.8 GHz (2nd harmonic)
7.2 GHz (3rd harmonic)
9.6 GHz (4th harmonic)
12.0 GHz (5th harmonic)
14.4 GHz (6th harmonic)
16.8 GHz (7th harmonic)
19.2 GHz (8th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.
Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68255B/68355B without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 6 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (0.5 to 20 GHz); 4.0 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68255B/68355B with Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3.5 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 3.5 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (0.5 to 20 GHz); 8.2 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68255B/68355B with Option 15B High Power & without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 10 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 10 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (0.5 to 20 GHz); 4.0 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68255B/68355B with Option 15B High Power & Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 7.5 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 7.5 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (0.5 to 20 GHz); 8.2 dB (20 to 26.5 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68255B/68355B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68255B/68355B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68255B/68355B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68255B/68355B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68255B/68355B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
21.000 000 000
2.000 000 000
22.000 000 000
3.000 000 000
23.000 000 000
4.000 000 000
24.000 000 000
5.000 000 000
25.000 000 000
6.000 000 000
26.000 000 000
7.000 000 000
8.000 000 000
9.000 000 000
10.000 000 000
11.000 000 000
12.000 000 000
13.000 000 000
14.000 000 000
15.000 000 000
16.000 000 000
17.000 000 000
18.000 000 000
19.000 000 000
20.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X59B) Fine Loop Test Procedure (68X59B with Option 11)
Test Frequency (in GHz) Measured Value ** Test Frequency (in GHz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 10 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 20 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 30 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 40 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 350 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 1.6 GHz Record the presence of the worst case non-harmonic dBc –40 dBc
Set F1 to 1.6 GHz Record the level of the harmonics of the 1.6 GHz carrier:
3.2 GHz (2nd harmonic)
4.8 GHz (3rd harmonic)

Test Procedure (2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.
Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68259B/68359B without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 26.5 GHz).
+ 6 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68259B/68359B with Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3.5 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 26.5 GHz).
+ 3.5 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68259B/68359B with Option 15B High Power & without Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 10 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 26.5 GHz).
+ 10 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz);2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 26.5 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68259B/68359B with Option 15B High Power & Option 2A Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 7.5 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 26.5 GHz).
+ 7.5 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 26.5 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68259B/68359B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68259B/68359B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68259B/68359B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68259B/68359B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68259B/68359B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68259B with Firmware Version 2.18 and above; 68359B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
2.000 000 000
21.000 000 000
3.000 000 000
22.000 000 000
4.000 000 000
23.000 000 000
5.000 000 000
24.000 000 000
6.000 000 000
25.000 000 000
7.000 000 000
26.000 000 000
8.000 000 000
27.000 000 000
9.000 000 000
28.000 000 000
10.000 000 000
29.000 000 000
11.000 000 000
30.000 000 000
12.000 000 000
31.000 000 000
13.000 000 000
32.000 000 000
14.000 000 000
33.000 000 000
15.000 000 000
34.000 000 000
16.000 000 000
35.000 000 000
17.000 000 000
36.000 000 000
18.000 000 000
37.000 000 000
19.000 000 000
38.000 000 000
20.000 000 000
39.000 000 000
40.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X63B) Fine Loop Test Procedure (68X63Bwith Option 11)
Test Frequency (in GHz) Measured Value ** Test Frequency (in GHz) Measured Value ***
2.000 001 000
2.000 000 100
2.000 002 000
2.000 000 200
2.000 003 000
2.000 000 300
2.000 004 000
2.000 000 400
2.000 005 000
2.000 000 500
2.000 006 000
2.000 000 600
2.000 007 000
2.000 000 700
2.000 008 000
2.000 000 800
2.000 009 000
2.000 000 900
2.000 010 000
2.000 001 000
** Specification for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.
This test is not applicable to the 68263B/68363B model.

Test Procedure (2 to 10 GHz) Measured Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure (11 to 20 GHz) Measured Value Upper Limit
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure Measured Value Upper Limit
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68263B/68363B without Option 2B Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 6 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (2 to 20 GHz); 4.0 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68263B/68363B with Option 2B Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 3 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (2 to 20 GHz); 8.2 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68263B/68363B with Option 15B High Power & without Option 2B Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 6 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (2 to 20 GHz); 4.0 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68263B/68363B with Option 15B High Power & Option 2B Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz
* Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 3 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (2 to 20 GHz); 8.2 dB (20 to 40 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68263B/68363B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68263B/68363B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68263B/68363B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s)
Set F1 to 5.0 GHz Set pulse width W1 to 0.5 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68263B/68363B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68263B/68363B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68263B with Firmware Version 2.18 and above; 68363B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
21.000 000 000
2.000 000 000
22.000 000 000
3.000 000 000
23.000 000 000
4.000 000 000
24.000 000 000
5.000 000 000
25.000 000 000
6.000 000 000
26.000 000 000
7.000 000 000
27.000 000 000
8.000 000 000
28.000 000 000
9.000 000 000
29.000 000 000
10.000 000 000
30.000 000 000
11.000 000 000
31.000 000 000
12.000 000 000
32.000 000 000
13.000 000 000
33.000 000 000
14.000 000 000
34.000 000 000
15.000 000 000
35.000 000 000
16.000 000 000
36.000 000 000
17.000 000 000
37.000 000 000
18.000 000 000
38.000 000 000
19.000 000 000
39.000 000 000
20.000 000 000
40.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X65B) Fine Loop Test Procedure (68X65B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 500 MHz Record the level of all harmonics of the 500 MHz carrier
1.0 GHz (2nd harmonic)
1.5 GHz (3rd harmonic)
2.0 GHz (4th harmonic)
2.5 GHz (5th harmonic)
Set F1 to 800 MHz Record the level of all harmonics of the 800 MHz carrier:
1.6 GHz (2nd harmonic)
2.4 GHz (3rd harmonic)
3.2 GHz (4th harmonic)
4.0 GHz (5th harmonic)
Set F1 to 1.3 GHz Record the level of all harmonics of the 1.3 GHz carrier:
2.6 GHz (2nd harmonic)
3.9 GHz (3rd harmonic)
5.2 GHz (4th harmonic)
6.5 GHz (5th harmonic)
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.2 GHz (4th harmonic)
10.2 GHz (5th harmonic)

Test Procedure (2.2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.4 GHz Record the level of all harmonics of the 2.4 GHz carrier:
4.8 GHz (2nd harmonic)
7.2 GHz (3rd harmonic)
9.6 GHz (4th harmonic)
12.0 GHz (5th harmonic)
14.4 GHz (6th harmonic)
16.8 GHz (7th harmonic)
19.2 GHz (8th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure (11 to 20 GHz) Measure Value Upper Limit
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.
Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68265B/68365B without Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 6 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (0.5 to 20 GHz); 4.0 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68265B/68365B with Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 3 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (0.5 to 20 GHz); 8.2 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68265B/68365B with Option 15B High Power & without Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 6 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (0.5 to 20 GHz); 4.0 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68265B/68365B with Option 15B High Power & Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3 dBm dBm dBm dB
** Maximum variation is 1.6 dB.
+ 3 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (0.5 to 20 GHz); 8.2 dB (20 to 40 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68265B/68365B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68265B/68365B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68265B/68365B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 µs
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 µs
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68265B/68365B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68265B/68365B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
21.000 000 000
2.000 000 000
22.000 000 000
3.000 000 000
23.000 000 000
4.000 000 000
24.000 000 000
5.000 000 000
25.000 000 000
6.000 000 000
26.000 000 000
7.000 000 000
27.000 000 000
8.000 000 000
28.000 000 000
9.000 000 000
29.000 000 000
10.000 000 000
30.000 000 000
11.000 000 000
31.000 000 000
12.000 000 000
32.000 000 000
13.000 000 000
33.000 000 000
14.000 000 000
34.000 000 000
15.000 000 000
35.000 000 000
16.000 000 000
36.000 000 000
17.000 000 000
37.000 000 000
18.000 000 000
38.000 000 000
19.000 000 000
39.000 000 000
20.000 000 000
40.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X69B) Fine Loop Test Procedure (68X69B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 10 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 20 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 30 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 40 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 350 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 1.6 GHz Record the presence of the worst case non-harmonic dBc –40 dBc
Set F1 to 1.6 GHz Record the level of the harmonics of the 1.6 GHz carrier:
3.2 GHz (2nd harmonic)
4.8 GHz (3rd harmonic)

Test Procedure (2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.

Test Procedure (11 to 20 GHz) Measure Value Upper Limit
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –60 dBc*
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –60 dBc*
* –50 dBc if Option 15B (High Power) installed.
Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68269B/68369B without Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz).
+ 6 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68269B/68369B with Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz).
+ 3 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68269B/68369B with Option 15B High Power & without Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 6 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz).
+ 6 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 40 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68269B/68369B with Option 15B High Power & Option 2B Step Attenuator)
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 25.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.0 dB.
+ 3 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz).
+ 3 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 40 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68269B/68369B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68269B/68369B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68269B/68369B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68269B/68369B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68269B/68369B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68269B with Firmware Version 2.18 and above 68369B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
26.000 000 000
2.000 000 000
27.000 000 000
3.000 000 000
28.000 000 000
4.000 000 000
29.000 000 000
5.000 000 000
30.000 000 000
6.000 000 000
31.000 000 000
7.000 000 000
32.000 000 000
8.000 000 000
33.000 000 000
9.000 000 000
34.000 000 000
10.000 000 000
35.000 000 000
11.000 000 000
36.000 000 000
12.000 000 000
37.000 000 000
13.000 000 000
38.000 000 000
14.000 000 000
39.000 000 000
15.000 000 000
40.000 000 000
16.000 000 000
41.000 000 000
17.000 000 000
42.000 000 000
18.000 000 000
43.000 000 000
19.000 000 000
44.000 000 000
20.000 000 000
45.000 000 000
21.000 000 000
46.000 000 000
22.000 000 000
47.000 000 000
23.000 000 000
48.000 000 000
24.000 000 000
49.000 000 000
25.000 000 000
50.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X75B) Fine Loop Test Procedure (68X75B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 500 MHz Record the level of all harmonics of the 500 MHz carrier
1.0 GHz (2nd harmonic)
1.5 GHz (3rd harmonic)
2.0 GHz (4th harmonic)
2.5 GHz (5th harmonic)
Set F1 to 800 MHz Record the level of all harmonics of the 800 MHz carrier:
1.6 GHz (2nd harmonic)
2.4 GHz (3rd harmonic)
3.2 GHz (4th harmonic)
4.0 GHz (5th harmonic)
Set F1 to 1.3 GHz Record the level of all harmonics of the 1.3 GHz carrier:
2.6 GHz (2nd harmonic)
3.9 GHz (3rd harmonic)
5.2 GHz (4th harmonic)
6.5 GHz (5th harmonic)
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.2 GHz (4th harmonic)
10.2 GHz (5th harmonic)

Test Procedure (2.2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.4 GHz Record the level of all harmonics of the 2.4 GHz carrier:
4.8 GHz (2nd harmonic)
7.2 GHz (3rd harmonic)
9.6 GHz (4th harmonic)
12.0 GHz (5th harmonic)
14.4 GHz (6th harmonic)
16.8 GHz (7th harmonic)
19.2 GHz (8th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –50 dBc

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68275B/68375B without Option 2C Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 45.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 2.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 1.5 dBm
+ 0.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 1.5 dBm
– 2.5 dBm
– 2.5 dBm
– 3.5 dBm
– 3.5 dBm
– 4.5 dBm
– 4.5 dBm
– 5.5 dBm
– 5.5 dBm
– 6.5 dBm
– 6.5 dBm
– 7.5 dBm
– 7.5 dBm
– 8.5 dBm
– 8.5 dBm
– 9.5 dBm
– 9.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
+ 2.5 dBm dBm dBm dB
** Maximum variation is 1.6 dB (0.5 to 40 GHz); 2.2 dB (40 to 50 GHz).
+ 2.5 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (0.5 to 20 GHz); 4.0 dB (20 to 40 GHz); 5.0 dB (40 to 50 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68275B/68375B with Option 2C Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 45.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 8.5 dBm
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
– 1 dBm dBm dBm dB
** Maximum variation is 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 50 GHz).
– 1 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (0.5 to 20 GHz); 8.2 dB (20 to 40 GHz); 10.2 dB (40 to 50 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68275B/68375B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68275B/68375B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68275B/68375B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68275B/68375B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68275B/68375B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
26.000 000 000
2.000 000 000
27.000 000 000
3.000 000 000
28.000 000 000
4.000 000 000
29.000 000 000
5.000 000 000
30.000 000 000
6.000 000 000
31.000 000 000
7.000 000 000
32.000 000 000
8.000 000 000
33.000 000 000
9.000 000 000
34.000 000 000
10.000 000 000
35.000 000 000
11.000 000 000
36.000 000 000
12.000 000 000
37.000 000 000
13.000 000 000
38.000 000 000
14.000 000 000
39.000 000 000
15.000 000 000
40.000 000 000
16.000 000 000
41.000 000 000
17.000 000 000
42.000 000 000
18.000 000 000
43.000 000 000
19.000 000 000
44.000 000 000
20.000 000 000
45.000 000 000
21.000 000 000
46.000 000 000
22.000 000 000
47.000 000 000
23.000 000 000
48.000 000 000
24.000 000 000
49.000 000 000
25.000 000 000
50.000 000 000
* Specification for all frequencies listed above is ±100 Hz.

Fine Loop Test Procedure (Standard 68X77B) Fine Loop Test Procedure (68X77B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 10 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 20 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 30 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 40 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 350 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 1.6 GHz Record the presence of the worst case non-harmonic dBc –40 dBc
Set F1 to 1.6 GHz Record the level of the harmonics of the 1.6 GHz carrier:
3.2 GHz (2nd harmonic)
4.8 GHz (3rd harmonic)

Test Procedure (2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –50 dBc

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68277B/68377B without Option 2C Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 45.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 2.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 1.5 dBm
+ 0.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 1.5 dBm
– 2.5 dBm
– 2.5 dBm
– 3.5 dBm
– 3.5 dBm
– 4.5 dBm
– 4.5 dBm
– 5.5 dBm
– 5.5 dBm
– 6.5 dBm
– 6.5 dBm
– 7.5 dBm
– 7.5 dBm
– 8.5 dBm
– 8.5 dBm
– 9.5 dBm
– 9.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
+ 2.5 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 50 GHz).
+ 2.5 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 40 GHz); 5.0 dB (40 to 50 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68277B/68377B with Option 2C Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 45.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 8.5 dBm
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
– 1 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 50 GHz).
– 1 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 40 GHz); 10.2 dB (40 to 50 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68277B/68377B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68277B/68377B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68277B/68377B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 µs
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 µs
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68277B/68377B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68277B/68377B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68277B with Firmware Version 2.18 and above; 68377B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
26.000 000 000
2.000 000 000
27.000 000 000
3.000 000 000
28.000 000 000
4.000 000 000
29.000 000 000
5.000 000 000
30.000 000 000
6.000 000 000
31.000 000 000
7.000 000 000
32.000 000 000
8.000 000 000
33.000 000 000
9.000 000 000
34.000 000 000
10.000 000 000
35.000 000 000
11.000 000 000
36.000 000 000
12.000 000 000
37.000 000 000
13.000 000 000
38.000 000 000
14.000 000 000
39.000 000 000
15.000 000 000
40.000 000 000
16.000 000 000
41.000 000 000
17.000 000 000
42.000 000 000
18.000 000 000
43.000 000 000
19.000 000 000
44.000 000 000
20.000 000 000
45.000 000 000
21.000 000 000
46.000 000 000
22.000 000 000
47.000 000 000
23.000 000 000
48.000 000 000
24.000 000 000
49.000 000 000
25.000 000 000
50.000 000 000

Coarse Loop/YIG Loop Test Procedure (Continued)
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
51.000 000 000
56.000 000 000
52.000 000 000
57.000 000 000
53.000 000 000
58.000 000 000
54.000 000 000
59.000 000 000
55.000 000 000
60.000 000 000
* Specification for all frequencies listed above is ±100 Hz.
Fine Loop Test Procedure (Standard 68X85B) Fine Loop Test Procedure (68X85B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 500 MHz Record the level of all harmonics of the 500 MHz carrier
1.0 GHz (2nd harmonic)
1.5 GHz (3rd harmonic)
2.0 GHz (4th harmonic)
2.5 GHz (5th harmonic)
Set F1 to 800 MHz Record the level of all harmonics of the 800 MHz carrier:
1.6 GHz (2nd harmonic)
2.4 GHz (3rd harmonic)
3.2 GHz (4th harmonic)
4.0 GHz (5th harmonic)
Set F1 to 1.3 GHz Record the level of all harmonics of the 1.3 GHz carrier:
2.6 GHz (2nd harmonic)
3.9 GHz (3rd harmonic)
5.2 GHz (4th harmonic)
6.5 GHz (5th harmonic)
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.2 GHz (4th harmonic)
10.2 GHz (5th harmonic)

Test Procedure (2.2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.4 GHz Record the level of all harmonics of the 2.4 GHz carrier:
4.8 GHz (2nd harmonic)
7.2 GHz (3rd harmonic)
9.6 GHz (4th harmonic)
12.0 GHz (5th harmonic)
14.4 GHz (6th harmonic)
16.8 GHz (7th harmonic)
19.2 GHz (8th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –50 dBc

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68285B/68385B without Option 2D Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 55.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
– 9.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
+ 2 dBm dBm dBm dB
** Maximum variation is 1.6 dB (0.5 to 40 GHz); 2.2 dB (40 to 60 GHz).
+ 2 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (0.5 to 20 GHz); 4.0 dB (20 to 40 GHz); 5.0 dB (40 to 60 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68285B/68385B with Option 2D Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 55.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 8.5 dBm
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
– 2 dBm dBm dBm dB
** Maximum variation is 1.6 dB (0.5 to 40 GHz); 2.2 dB (40 to 60 GHz).
– 2 dBm dBm dBm dB
*** Maximum variation is 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 40 GHz); 10.2 dB (40 to 60 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68285B/68385B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68285B/68385B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68285B/68385B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68285B/68385B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68285B/68385B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
26.000 000 000
2.000 000 000
27.000 000 000
3.000 000 000
28.000 000 000
4.000 000 000
29.000 000 000
5.000 000 000
30.000 000 000
6.000 000 000
31.000 000 000
7.000 000 000
32.000 000 000
8.000 000 000
33.000 000 000
9.000 000 000
34.000 000 000
10.000 000 000
35.000 000 000
11.000 000 000
36.000 000 000
12.000 000 000
37.000 000 000
13.000 000 000
38.000 000 000
14.000 000 000
39.000 000 000
15.000 000 000
40.000 000 000
16.000 000 000
41.000 000 000
17.000 000 000
42.000 000 000
18.000 000 000
43.000 000 000
19.000 000 000
44.000 000 000
20.000 000 000
45.000 000 000
21.000 000 000
46.000 000 000
22.000 000 000
47.000 000 000
23.000 000 000
48.000 000 000
24.000 000 000
49.000 000 000
25.000 000 000
50.000 000 000

Coarse Loop/YIG Loop Test Procedure (Continued)
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
51.000 000 000
56.000 000 000
52.000 000 000
57.000 000 000
53.000 000 000
58.000 000 000
54.000 000 000
59.000 000 000
55.000 000 000
60.000 000 000
* Specification for all frequencies listed above is ±100 Hz.
Fine Loop Test Procedure (Standard 68X87B) Fine Loop Test Procedure (68X87B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 10 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 20 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 30 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 40 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 350 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 1.6 GHz Record the presence of the worst case non-harmonic dBc –40 dBc
Set F1 to 1.6 GHz Record the level of the harmonics of the 1.6 GHz carrier:
3.2 GHz (2nd harmonic)
4.8 GHz (3rd harmonic)

Test Procedure (2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –50 dBc

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

3-11 Power Level Accuracy and Flatness Tests (Model 68287B/68387B without Option 2D Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 55.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
– 9.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
+ 2 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 60 GHz).
+ 2 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 40 GHz); 5.0 dB (40 to 60 GHz)(typical, not a specification).

3-11 Power Level Accuracy and Flatness Tests (Continued) (Model 68287B/68387B with Option 2D Step Attenuator)
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 55.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 8.5 dBm
+ 7.5 dBm
+ 6.5 dBm
+ 5.5 dBm
+ 4.5 dBm
+ 3.5 dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
– 2 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 60 GHz).
– 2 dBm dBm dBm dB
*** Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 40 GHz); 10.2 dB (40 to 60 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68287B/68387B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68287B/68387B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68287B/68387B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68287B/68387B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68287B/68387B rad
Calculate and record the FM accuracy 93.7% % 106.3%



Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68287B with Firmware Version 2.18 and above; 68387B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
26.000 000 000
2.000 000 000
27.000 000 000
3.000 000 000
28.000 000 000
4.000 000 000
29.000 000 000
5.000 000 000
30.000 000 000
6.000 000 000
31.000 000 000
7.000 000 000
32.000 000 000
8.000 000 000
33.000 000 000
9.000 000 000
34.000 000 000
10.000 000 000
35.000 000 000
11.000 000 000
36.000 000 000
12.000 000 000
37.000 000 000
13.000 000 000
38.000 000 000
14.000 000 000
39.000 000 000
15.000 000 000
40.000 000 000
16.000 000 000
41.000 000 000
17.000 000 000
42.000 000 000
18.000 000 000
43.000 000 000
19.000 000 000
44.000 000 000
20.000 000 000
45.000 000 000
21.000 000 000
46.000 000 000
22.000 000 000
47.000 000 000
23.000 000 000
48.000 000 000
24.000 000 000
49.000 000 000
25.000 000 000
50.000 000 000

Coarse Loop/YIG Loop Test Procedure (Continued)
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
51.000 000 000
59.000 000 000
52.000 000 000
60.000 000 000
53.000 000 000
61.000 000 000
54.000 000 000
62.000 000 000
55.000 000 000
63.000 000 000
56.000 000 000
64.000 000 000
57.000 000 000
65.000 000 000
58.000 000 000
* Specification for all frequencies listed above is ±100 Hz.
Fine Loop Test Procedure (Standard 68X95B) Fine Loop Test Procedure (68X95B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 500 MHz Record the level of all harmonics of the 500 MHz carrier
1.0 GHz (2nd harmonic)
1.5 GHz (3rd harmonic)
2.0 GHz (4th harmonic)
2.5 GHz (5th harmonic)
Set F1 to 800 MHz Record the level of all harmonics of the 800 MHz carrier:
1.6 GHz (2nd harmonic)
2.4 GHz (3rd harmonic)
3.2 GHz (4th harmonic)
4.0 GHz (5th harmonic)
Set F1 to 1.3 GHz Record the level of all harmonics of the 1.3 GHz carrier:
2.6 GHz (2nd harmonic)
3.9 GHz (3rd harmonic)
5.2 GHz (4th harmonic)
6.5 GHz (5th harmonic)
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.2 GHz (4th harmonic)
10.2 GHz (5th harmonic)

Test Procedure (2.2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.4 GHz Record the level of all harmonics of the 2.4 GHz carrier:
4.8 GHz (2nd harmonic)
7.2 GHz (3rd harmonic)
9.6 GHz (4th harmonic)
12.0 GHz (5th harmonic)
14.4 GHz (6th harmonic)
16.8 GHz (7th harmonic)
19.2 GHz (8th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –50 dBc

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Power Level Accuracy Test Procedure
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 60.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
– 9.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
– 2 dBm dBm dBm dB
** Maximum variation is 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 65 GHz).
– 2 dBm dBm dBm dB
*** Maximum variation is 2.0 dB (0.5 to 20 GHz); 4.0 dB (20 to 40 GHz); 5.0 dB (40 to 65 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68295B/68395B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68295B/68395B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68295B/68395B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.10 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68295B/68395B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68295B/68395B rad
Calculate and record the FM accuracy 93.7% % 106.3%

Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion

Serial Number __________________ Tested By: __________________________
3-6 Internal Time Base Aging Rate Test
Test Procedure Measured Value Upper Limit
Record frequency error value
Record frequency error value (after 24 hours)
Record the computed aging rate per day 2x10–8 per day (5x10–10 per day with Option 16)

Coarse Loop/YIG Loop Test Procedure
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
1.000 000 000
26.000 000 000
2.000 000 000
27.000 000 000
3.000 000 000
28.000 000 000
4.000 000 000
29.000 000 000
5.000 000 000
30.000 000 000
6.000 000 000
31.000 000 000
7.000 000 000
32.000 000 000
8.000 000 000
33.000 000 000
9.000 000 000
34.000 000 000
10.000 000 000
35.000 000 000
11.000 000 000
36.000 000 000
12.000 000 000
37.000 000 000
13.000 000 000
38.000 000 000
14.000 000 000
39.000 000 000
15.000 000 000
40.000 000 000
16.000 000 000
41.000 000 000
17.000 000 000
42.000 000 000
18.000 000 000
43.000 000 000
19.000 000 000
44.000 000 000
20.000 000 000
45.000 000 000
21.000 000 000
46.000 000 000
22.000 000 000
47.000 000 000
23.000 000 000
48.000 000 000
24.000 000 000
49.000 000 000
25.000 000 000
50.000 000 000

Coarse Loop/YIG Loop Test Procedure (Continued)
Test Frequency (in GHz) Measured Value * Test Frequency (in GHz) Measured Value *
51.000 000 000
59.000 000 000
52.000 000 000
60.000 000 000
53.000 000 000
61.000 000 000
54.000 000 000
62.000 000 000
55.000 000 000
63.000 000 000
56.000 000 000
64.000 000 000
57.000 000 000
65.000 000 000
58.000 000 000
* Specification for all frequencies listed above is ±100 Hz.
Fine Loop Test Procedure (Standard 68X97B) Fine Loop Test Procedure (68X97B with Option 11)
Test Frequency (in Ghz) Measured Value ** Test Frequency (in Ghz) Measured Value ***
1.000 001 000
1.000 000 100
1.000 002 000
1.000 000 200
1.000 003 000
1.000 000 300
1.000 004 000
1.000 000 400
1.000 005 000
1.000 000 500
1.000 006 000
1.000 000 600
1.000 007 000
1.000 000 700
1.000 008 000
1.000 000 800
1.000 009 000
1.000 000 900
1.000 010 000
1.000 001 000
** Specifications for all frequencies listed above is ±100 Hz. *** Specification for all frequencies listed above is ±10 Hz.

Test Procedure Measured Value Upper Limit
Set F1 to 10 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 20 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 30 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 40 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 350 MHz Record the presence of the worst case harmonic
Record the presence of the worst case non-harmonic
Set F1 to 1.6 GHz Record the presence of the worst case non-harmonic dBc –40 dBc
Set F1 to 1.6 GHz Record the level of the harmonics of the 1.6 GHz carrier:
3.2 GHz (2nd harmonic)
4.8 GHz (3rd harmonic)

Test Procedure (2 to 10 GHz) Measure Value Upper Limit
Set F1 to 2.1 GHz Record the level of all harmonics of the 2.1 GHz carrier:
4.2 GHz (2nd harmonic)
6.3 GHz (3rd harmonic)
8.4 GHz (4th harmonic)
10.5 GHz (5th harmonic)
12.6 GHz (6th harmonic)
14.7 GHz (7th harmonic)
16.8 GHz (8th harmonic)
18.9 GHz (9th harmonic)
Set F1 to 3.6 GHz Record the level of all harmonics of the 3.6 GHz carrier:
7.2 GHz (2nd harmonic)
10.8 GHz (3rd harmonic)
14.4 GHz (4th harmonic)
18.0 Ghz (5th harmonic)
Set F1 to 7.0 GHz Record the level of all harmonics of the 7 GHz carrier:
14.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 10.0 GHz Record the level of all harmonics of the 10 GHz carrier:
20.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 12.4 GHz Record the level of all harmonics of the 12.4 GHz carrier:
24.8 GHz (2nd harmonic)
37.2 GHz (3rd harmonic)
Set F1 to 16.0 GHz Record the level of all harmonics of the 16.0 GHz carrier:
32.0 GHz (2nd harmonic) dBc –50 dBc
Set F1 to 20.0 GHz Record the level of all harmonics of the 20.0 GHz carrier:
40.0 GHz (2nd harmonic) dBc –50 dBc

Test Procedure Measured Value Upper Limit
Set F1 to 0.6 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 2.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 6.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 10.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz
Set F1 to 20.0 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Test Procedure Measured Value Upper Limit
Set F1 to 26.5 GHz Record the phase noise levels at these offsets:
100 Hz
1 kHz
10 kHz
100 kHz

Power Level Accuracy Test Procedure
Set F1 to 5.0 GHz Set F1 to 25.0 GHz Set F1 to 60.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
– 9.5 dBm
* Specification is ±1.0 dB. * Specification is ±1.0 dB. * Specification is ±1.5 dB.
– 2 dBm dBm dBm dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 40 GHz); 2.2 dB (40 to 65 GHz).
– 2 dBm dBm dBm dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz); 4.0 dB (20 to 40 GHz); 5.0 dB (40 to 65 GHz)(typical, not a specification).

AM Imput Sensitivity Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the Modulation Analyzer AM PK(+) reading
Measure and record the Modulation Analyzer AM PK(–) reading
Calculate and record the actual AM input sensitivity 45% % 55%
FM Imput Sensitivity Procedure
(Unlocked Wide FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the low carrier frequency
Measure and record the high carrier frequency
Calculate and record the actual peak-to-peak frequency deviation . 190 MHz MHz 210 MHz
(Unlocked Narrow FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68297B/68397B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
(Locked FM Mode)
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68297B/68397B MHz
Calculate and record the FM accuracy 93.7% % 106.3%

FM Imput Sensitivity Procedure
(Locked Low-Noise FM Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation frequency displayed on the 68297B/68397B MHz
Calculate and record the FM accuracy 93.7% % 106.3%
Rise Time, Fall Time, and Overshoot Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Measure and record the following: Rise Time
Pulse Width
PRI (period)
Overshoot
Pulse Level Accuracy (1 GHz, Pulse Width = 1.0 s)
Set F1 to 1.0 GHz Set pulse width W1 to 1.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 1.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB

Pulse Level Accuracy (5 GHz, Pulse Width = 0.5 s) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –1.0 dB dB +1.0 dB
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms
Record the displayed output level L1 (step 5.b)
Record the displayed output level L1 (step 5.d)
Calculate and record the pulse level error –0.5 dB dB +0.5 dB
Test Procedure Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz Set pulse width W1 to 5.0 ms Set PRI (period) to 0.01 ms
Measure and record the Video Feedthrough voltage spikes –10 mV mV +10 mV

Test Procedure (1 GHz) Lower Limit Measured Value Upper Limit
Set F1 to 1.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
Set F1 to 5.0 GHz
Measure and record the peak of the signal on the Spectrum Analyzer. (Measured signal must be >60 dB below top graticule to meet specification; this represents an On/Off Ratio of >80 dB.) . 60 dB dB N/A
M Imput Sensitivity Procedure
(Wide M Mode) Lower Limit Measured Value Upper Limit
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68297B/68397B rad
Calculate and record the FM accuracy 93.7% % 106.3%
(Narrow M Mode)
Set F1 to 5.0 GHz
Record the FM deviation displayed on the 68297B/68397B rad
Calculate and record the FM accuracy 93.7% % 106.3%

Serial Number __________________ Calibrated By: __________________________
4-7 Preliminary Calibration
Procedure Step Step Completion
Log Amplifier Zero Calibration Step Completion

This calibration is performed using an automatic test system. Contact ANRITSU Customer Service for further information.
Procedure Step Step Completion
Procedure Step Step Completion
Procedure Step Step Completion

Procedure Step Step Completion
4-13 FM Calibration (68297B with Firmware Version 2.18 and above; 68397B with Firmware Version 2.20 and above)

Procedure Step Step Completion

Test Record, 68295B/68395B, A-255 - A-258 Test Record, 68297B/68397B, A-271 - A-274 Chassis Covers
682XXB/683XXB Assembly and Component Locator Diagram, 6-9 General Description, 1-3 Major Subsystems Functional Description, 2-3 Manuals, Related, 1-6 Models, List of, 1-4 - 1-5 Options, List of, 1-7 Startup Configurations, 1-11 System Block Diagram, 2-6 - 2-7
Electrostatic Discharge Precautions, 1-9 Error Messages
Operation Related, 5-7 Self-Test, 5-3 - 5-6 Exchange Assembly Program
F
Block Diagram, 2-17 Functional Description, 2-15 ANRITSU Service Centers, 1-18
Remove/Replace Procedures, 6-18 Frequency Synthesis Subsystem
Block Diagram, 2-11 Functional Description, 2-9 Front Panel Assembly
ALC Bandwidth Calibration, 4-31 ALC Slope Calibration, 4-23 AM Calibration, 4-33 FM Calibration, 4-37 Following Subassembly Replacement, 4-4 Initial Setup, 4-7 Phase Modulation Calibration, 4-51 Preliminary Calibration, 4-13 RF Level Calibration, 4-22 Switched Filter Shaper Calibration, 4-17 Test Equipment, 4-3 Test Record, 68237B/68337B, A-15 - A-18 Test Record, 68245B/68345B, A-33 - A-36 Test Record, 68247B/68347B, A-51 - A-54 Test Record, 68253B/68353B, A-69 - A-72 Test Record, 68255B/68355B, A-89 - A-92 Test Record, 68259B/68359B, A-109 - A-112 Test Record, 68263B/68363B, A-127 - A-130 Test Record, 68265B/68365B, A-147 - A-150 Test Record, 68269B/68369B, A-167 - A-170 Test Record, 68275B/68375B, A-185 - A-188 Test Record, 68277B/68377B, A-203 - A-206 Test Record, 68285B/68385B, A-221 - A-224 Test Record, 68287B/68387B, A-239 - A-242
Remove/Replace Procedures, 6-6 Functional Description
682XXB/683XXB Major Subsystems, 2-3 ALC/AM/Pulse Modulation Subsystem, 2-15 Frequency Synthesis Subsystem, 2-9 RF Deck Assemblies, 2-19 System Block Diagram, 2-6 - 2-7
M
Maintenance, Level of , 1-8 Manual, Electronic, 1-6 Manual, GPIB Programming, 1-6 Manual, Operation, 1-6 Manual, SCPI Programming, 1-6 Models, List of, 1-4 - 1-5

N
T
Normal Operation Error/Warning Messages, 5-7
Test Equipment, Recommended, 1-12 Testing, Performance Verification
Amplitude Modulation Test, 3-30 Frequency Modulation Tests, 3-33 Frequency Synthesis Tests, 3-11 Harmonic Test: RF Output 2 to 20 GHz, 3-18 Internal Time Base Aging Rate Test, 3-8 Maximum Leveled Power Listing, 3-6 - 3-7 Phase Modulation Tests, 3-48 Power Level Accuracy and Flatness Tests, 3-26 Pulse Modulation Test: Overshoot, 3-39 Pulse Modulation Test: Pulse Level
Options, List of, 1-7
Parts and Subassemblies, Replaceable, 1-14 Parts Ordering Information, 1-14
ANRITSU Service Centers, 1-14 Performance Verification Tests, 3-3 Preventive Maintenance
Fan Filter Cleaning, 1-9
Remove/Replace Procedures, 6-13 Recommended Test Equipment, 1-12 Removal and Replacement Procedures
Feedthrough, 3-43 Single Sideband Phase Noise Test, 3-22 Spurious Signals Test: RF Output
A13, A14, or A15 PCB, 6-11 A16 or A17 PCB, 6-11 A18 or A19 PCB, 6-12 A21 Line Filter/Rectifier PCB, 6-16 A21-1 BNC/AUX I/O Connector PCB, 6-17 A3, A4, A5, or A6 PCBs, 6-8 A7 YIG Loop PCB, 6-10 A8, A9, A10, A11, or A12 PCB, 6-10 Assembly and Component Locator Diagram, 6-9 Chassis Covers, 6-4 Fan Assembly, 6-18 Front Panel Assembly, 6-6 Rear Panel Assembly, 6-13 Replaceable Subassemblies and Parts, 1-14 RF Deck Assemblies
Block Diagram, 2-21 - 2-22, 2-27 - 2-28 Functional Description, 2-19
Connector/Test Points Locator Diagram, 5-11 Malfunctions Not Displaying an Error
Scope of Manual, 1-3 Self-Test Error Messages, 5-3 Startup Configurations, 1-11 Static-Sensitive Component Precautions, 1-9
Message, 5-10 Normal Operation Error/Warning
Messages, 5-7 Self-Test Error Messages, 5-3 Troubleshooting Tables, 5-10, 5-12 - 5-53
Rear Panel Assembly