Manuel d'utilisation / d'entretien du produit E du fabricant Fluke
Aller à la page of 10
® 80T -IR/E Extended Range Infrared Temperature Probe Instr ucti on Sheet Int roduct ion The Fluke 80T-IR/ E Extended Range Inf rared Temperature Probe (t he probe) is a nonc ontact temperature m eas urement acces s ory for us e with a t es t instrum ent capable of meas uring DC volts in the millivolt range such as a digital multimeter (DMM).
Safety I nfo rm ati on The 80T-IR/E c omplies with IEC Publ ication 1010-1- 1990 including A mendment 1, CSA C22.2 No. 231, ANSI/ I SA-S82.01 and .03 S af ety Standards . W WARNING IF TARGET EMMISSIVITY IS LESS T HAN 0.95, THE PROBE CAN I NDICATE A TEM PERATURE LOWER THAN THE ACT UAL TARGET TEM PERATURE.
Compa tibilit y The probe is compatible wit h all DC millivolt measuring instrum ents that have a m inimum of 1 M W input impedance and accept s afety shrouded, standard diameter 0. 16 i n. (4 m m ) banana pl ugs . Operati on To take a meas urement, perform t he f ollowing steps : 1.
300mV V V OFF 0 1 02 03 0 300mV V V OFF A A ! 1000V 750V FUSED 10A RANGE PRESS AUTORANGE 300 mA COM V 1 SEC TOUCH HOLD Avoid touching or holding cone ON/OFF switch Battery clip Battery 9V DC Battery door Coiled cable Banana plugs 3 2 1 4 ° C ° F ° C/ ° F switch Figure 1.
Table 1. Display Codes Condition Action 1041 ° F Target temperature is over range. Select target within probe's specified temperature range. -135 ° F The temperature of the probe is near the low ambient operating range limit.** Ensure that the probe is within the specified ambient operating range.
2.5 0.1 14 0.6 7.5 0.3 21 0.8 33 1.3 mm Inch 0 0 25 1 50 2 76 3 130 5 mm Inch Distance to object Diameter of spot Spot size increases with distance from the probe tip, as shown Figure 2.
Measurem ent Consi d erati ons · If t he surf ace t o be measured i s sm all (13 mm (1/2 in.) or less), hol d the probe as cl ose as pos sible to the surf ac e (no m ore t han 50 mm (2 in.) away). · If t he surface to be meas ured is covered by frost or other materi al , c lean it to expos e t he surface.
Output: 1 mV/degree ° F Relative Hum idity: 95% RH or less @ 86° F noncondensing, Temp. Coef. appli es Storage Temperature: -13° to 158° F wi t hout batter y Power: 9V b attery; (N EDA 1604A, 6F22, 006P ) Batt ery life ( Alkaline): 50 hours typical, @ 73° F 33% du ty cycle Dimensions: (L x W x H) 180 m m x 30 mm x 50 m m (7.
LIMI T E D WARRANTY & LIMI T AT I ON OF LIABILI T Y Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. P arts, product repairs, and services are warranted for 90 days.
Servic e For servic e information in t he U.S.A., call 1-800-526- 4711. Outs ide the U.S.A. , contact the nearest Fluke Servic e Center. To locat e an authorized servic e center, visi t us on the World Wi de Web: www.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Fluke E c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Fluke E - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Fluke E, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Fluke E va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Fluke E, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Fluke E.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Fluke E. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Fluke E ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.