Victoreen ® 960PS-200 & 960PS-210 Power Supply Modules Operators Manual March 2005 Manual No. 960PS-200-1 Rev. 2 © 2004, 2005 Fluke Corporat i on, All rights reserved .
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 120 Andrews Road Hicksville, New York 1180 1 516.870.
i Table of Contents Section 1: Introductio n ................................................................................................ 1-1 1.1 General De scription ................................................................................
(Blank Page).
Introduction General Description 1 1-1 Section 1 Introduction 1.1 General Description The 960PS Power Supply is used to su pply power to the modules that are pl ugged into the 960MB/961MB motherboards. The power supply is mou nted outside of the 960CC Ca rd Cage Module, the wall mounted Local Cont rol Unit enclosure, or control room cabinet.
Victoreen 960PS-200 & 960PS-210 Operators Manual 1-2 NOTE Figure 1-1. Power Supply Functional Bloc k Diagram 1.4 Receiving Inspection and Storage Receiving Inspection Upon receipt of the unit: 1.
Introduction Receiving Inspection and Storage 1 1-3 2. Requirements for proper storag e must be documented and written pro c edures or instructions must be established. 3. In the event of fire, post-fire evaluation must be in accordance with ANSI N45.
Victoreen 960PS-200 & 960PS-210 Operators Manual (Blank page).
Operation 960PS-200 Power Supply 2 2-1 Section 2 Operation 2.1 960PS-200 Power Supply The 960PS-200 supplies thre e different DC voltage outputs to the 960 modules through the 960MB/961MB motherboard. 115 VAC is connecte d to te rminal strip A3, part of the front panel.
Victoreen 960PS-200 & 960PS-210 Operators Manual 2-2 2.5 960PS-210 Power Supply The 960PS-210 Power Supply is similar in operati on to the 960PS-200, with the exception that it incorporates an additional 24 VDC output, an input voltage sensing circuit, and a comm on and differential mode choke.
Maintenance, Calibration, and Troubleshooting Maintenance 3 3-1 Section 3 Maintenance, Calibration and Troubleshooting 3.1 Maintenance No periodic maintenan ce is required for the modul e. The output voltage levels, howe ver, should be verified periodically.
Victoreen 960PS-200 & 960PS-210 Operators Manual 3-2 If a problem develops, verify that the volt ages at connection point inputs and out puts are present and that all wiring is secure. Ensure fuse F1 is n ot open for the 960PS-210 Power Supply. Refer to Appendi x B for drawings.
Appendix Connector Designations A A-1 Appendix A Connector Designations A.1 Connector Designations Terminal Designations, 960PS-200 Pin Description 1 GND 2 AC Line Voltage 3 AC Line Voltage 4 -15 @ 2 .
Victoreen 960PS-200 & 960PS-210 Operators Manual (Blank page).
Appendix Applicable Drawings and Bill of Materials B B-1 Appendix B Applicable Drawings and Bill of Materials B.1 Applicable Drawings Drawing Number De scription 960PS-200-50 Power Supply Assembly 960PS-200-53 Schematic Diagram 960PS-210-5 Power Supply Assembly 960PS-210-53 Schematic Diagram B.
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 120 Andrews Road Hicksville, New York 1180 1 516.870.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Fluke 960PS-200 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Fluke 960PS-200 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Fluke 960PS-200, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Fluke 960PS-200 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Fluke 960PS-200, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Fluke 960PS-200.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Fluke 960PS-200. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Fluke 960PS-200 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.