Manuel d'utilisation / d'entretien du produit HASAxE-IM-HS du fabricant Emerson
Aller à la page of 298
Gas Analyzer Series Instruction Manual www.EmersonProcess.com Instruction Manual HASAxE-IM-HS 0 7 /2006.
ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards.
Instruction Manual HASAxE-IM-HS 04/2006 X-STREAM Emerson Process Management GmbH & Co. OHG S-1 Safety Instructions This instruction manual provides information about X-STREAM series gas analyzers concerning subassemblies, functions, procedures, installation, operation and maintenance.
X-STREAM Instruction Manual HASAxE-IM-HS 04/2006 Emerson Process Management GmbH & Co. OHG S-2 IMPORT ANT Safety Instructions Wiring and Inst allation of this Ap paratus The following safety instructions apply specifically to all EU member states.
Instruction Manual HASAxE-IM-HS 04/2006 X-STREAM Emerson Process Management GmbH & Co. OHG S-3 Safety Instructions Safety Instructions Operating and Maint aining this Apparatus This instrument has left the factory in com- pliance with all applicable safety regulations.
X-STREAM Instruction Manual HASAxE-IM-HS 04/2006 Emerson Process Management GmbH & Co. OHG S-4 Safety Instructions ELECTRICAL SHOCK HAZARD ! Do not operate without covers secure. Do not open while energized. Installation requires access to live parts which can cause death or serious injury .
Instruction Manual HASAxE-IM-HS 04/2006 X-STREAM Emerson Process Management GmbH & Co. OHG S-5 Safety Instructions Safety Instructions HIGH TEMPERA TURES ! While working with photometers and/or th.
X-STREAM Instruction Manual HASAxE-IM-HS 04/2006 Emerson Process Management GmbH & Co. OHG S-6 Safety Instructions T ake care of the safety instructions applicable for the gases (sample gases and .
Instruction Manual HASAxE-IM-HS 04/2006 X-STREAM Emerson Process Management GmbH & Co. OHG S-7 Safety Instructions Safety Instructions POWER SUPPL Y V erify the mains voltage at site of inst allat.
X-STREAM Instruction Manual HASAxE-IM-HS 04/2006 Emerson Process Management GmbH & Co. OHG S-8 These instrument s provide a protective earth terminal.
Instruction Manual HASAxE-IM-HS 04/2006 X-STREAM Emerson Process Management GmbH & Co. OHG S-9 Safety Instructions General Operating Instructions DANGER TO LIFE ! EXPLOSION HAZARD ! V erify all ga.
X-STREAM Instruction Manual HASAxE-IM-HS 04/2006 Emerson Process Management GmbH & Co. OHG S-10 Magnetically Operated Front Panel Magnetically Operated Front Panel DANGER TO LIFE ! Persons with ca.
T-1 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG T able of Contents SHORT FORM GUIDE FOR THIS MANUAL T o find information about see chapter Safety measures .............................................
T-2 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG TABLE OF CONTENTS Preamble S-1 Definitions S- 1 Safety Instructions S- 2 General Operating Instructions ................................................
T-3 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG T able of Contents T able of Contents Chapter 2 T echnical Data 2 - 1 2-1 Common T echnical Data .......................................................
T-4 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG Chapter 5 User Interface and Sof tware Menus 5 - 1 5-1 Abstract ........................................................................................
T-5 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG T able of Contents 5-4-3-4-1-1 Analog Output Signal Setup .................................................................. 5-35 5-4-3-4-1-2 Analog Output Range Setup .
T-6 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 7-3 Calibration Procedures ............................................................................................ 7-3 7-3-1 Preparing a Calibration .
T-7 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG T able of Contents Chapter 8 T roubleshooting 8 -1 8-1 Abstract ........................................................................................
T-8 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG Index of Figures Index of Figures Fig. 1-1: X-STREAM GP , GP S, front si de view .......................................................................
T-9 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG T able of Contents Index of Figures Fig. 4-9: IEC mains input plug .....................................................................................
T-10 X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG Index of T ables Index of T ables Table 1-1: Digital Inputs IN4-IN7, array ............................................................................
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-1 Emerson Process Management GmbH & Co. OHG 1 T echnical Description Chapter 1 T echnical Description 1-1 Overview Emerson Process Management´s .
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-2 Emerson Process Management GmbH & Co. OHG 1-1 Overview The X-STREAM series analyzers offer a wide range of available configurations and options.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-3 Emerson Process Management GmbH & Co. OHG 1 T echnical Description 1-1 Overview The different X-STREAM series models and their appearance. The following sections 1-2 to 1-3 give detailed descriptions for all available configurations.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-4 Emerson Process Management GmbH & Co. OHG 1-2 X-STREAM General Purpose T abletop or Rack Mount V ersion 1-2 X-STREAM GP , GPS This basic genera.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-5 Emerson Process Management GmbH & Co. OHG 1 T echnical Description Fig. 1-2: X-STREAM GP, terminals version, rear side view 1 23 4 5 6 Fig.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-6 Emerson Process Management GmbH & Co. OHG 1-3 X-STREAM F This IP66 / NEMA 4X protected housing is intended for outside wall mounting: The housing (fig. 1-4) is made of painted stainless steel. Gaskets protect against water and dust.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-7 Emerson Process Management GmbH & Co. OHG 1 T echnical Description Fig. 1-4: X-STREAM F, front side view & rear side view at carrying handl.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-8 Emerson Process Management GmbH & Co. OHG 1-3 X-STREAM F Fig. 1-5: X-STREAM F, front panel 1 4x20 characters alphanumeric display 2 LED (red) 3.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-9 Emerson Process Management GmbH & Co. OHG 1 T echnical Description 1-3 X-STREAM F Fig. 1-6: X-STREAM F, bottom view 1 Cable glands for mains an.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-10 Emerson Process Management GmbH & Co. OHG 1-3 X-STREAM F 1 T erminals for signal cables 2 Mains EMI filter 3 Cable glands for mains and signal cables 4 Mains terminals with integrated fuses Fig.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-11 Emerson Process Management GmbH & Co. OHG 1 T echnical Description 1-4 Reserved For Future Use 1-4 Reserved For Future Use.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-12 Emerson Process Management GmbH & Co. OHG 1-5 Gaspath Design 1-5 Gaspath Design V arious materials are available to provide a best possible analyzer adaption to the appli- cation. Materials are selected taking into account e.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-13 Emerson Process Management GmbH & Co. OHG 1 T echnical Description 1-5 Gaspath Design Depending on the application and the selected options several gas paths configurations are 1-5-5 Gas Path V ariations available, as shown in the following figures (examples): Fig.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-14 Emerson Process Management GmbH & Co. OHG Fig. 1-10: Single channel or serial tubing, with options V1 V2 Sample gas (V3) V4 Analyzer Solenoid .
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-15 Emerson Process Management GmbH & Co. OHG 1 T echnical Description 1-6 Optional Gas Path Components 1-6 Optional Gas Path Components Optionall.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-16 Emerson Process Management GmbH & Co. OHG 1-6 Optional Gas Path Components 1-6-5 Optional Heated Compartment T wo optional internal heated com.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-17 Emerson Process Management GmbH & Co. OHG 1 T echnical Description 1-7 Standard Interfaces 1-7-2 Modbus Interface The analyzers are equipped with a serial interface supporting Modbus RTU protocol to provide communication with external hosts (e.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-18 Emerson Process Management GmbH & Co. OHG 1-8 Optional Interfaces 1-8 Optional Interfaces Optionally available are the following interfaces (i.
Instruction Manual HASAxE-IM-HS 05/2006 X-STREAM 1-19 Emerson Process Management GmbH & Co. OHG 1 T echnical Description 1-8 Optional Interfaces Fig. 1-14: Digital Inputs - Priorities Example 1: Input A activates a zero calibration. Input B starts during the zero calibration.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 1-20 Emerson Process Management GmbH & Co. OHG 1-8 Optional Interfaces Inputs IN4 to IN7 are evaluated in decreasing order of priority: highest priority within this group is assigned to IN4, lowest priority to IN7.
X-STREAM 2-1 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2 T echnical Dat a Chapter 2 T echnical Dat a This chapter lists all the analyzers technical data, separated into common and model specific data, therefore the user has to select the appropriate section depending on his analyzer .
X-STREAM 2-2 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2-1 Common T echnical Data 2-1 Common T echnical Dat a Site of installation Hum idity (non conde nsing ) < 90 % r . h. at 68 F (+20 °C) < 70 % r .
X-STREAM 2-3 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2 T echnical Dat a Interfaces, signal inputs / outputs 2 analog outputs channel 4 (0) - 20 mA (R B &l.
X-STREAM 2-4 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2-1 Common T echnical Dat a Purge Option Purge medium (e.g. for minimizing CO 2 inter- ference, safety in events of internal leakage of flammable or/and aggressive gas etc.
X-STREAM 2-5 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2 T echnical Dat a 2-2 Model Specific T echnical Data 2-2 Model Specific T echnical Data 2-2-1 X-STREAM GP , GPS T abletop or Rack Mount V ersion Fig. 2-1: X-STREAM GP, GPS - dimensions [approx.
X-STREAM 2-6 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG Housing Permissible ambient temperature range 32 F to 122 F ( 0 °C to +50 °C) Weight: approx.
X-STREAM 2-7 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2 T echnical Dat a 2-2-1 X-STREAM GP , X-STREAM GP S Signal inputs / outputs All signal lines need to be connected to screw terminals located at the analyzer‘s rear panel.
X-STREAM 2-8 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2-2-1-2 Socket s V ersion Power Connection Connection via IEC power appliance, provided at the analyzer's rear panel (fig. 2-3). Power fuses T wo fuses are located in the power appliance.
X-STREAM 2-9 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2 T echnical Dat a Signal inputs / outputs All signal lines need to be connected to 9 pole and 25 pole submin-D-plugs and sockets, located at the analyzer‘s rear panel (fig.
X-STREAM 2-10 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2-2-2 X-STREAM F 2-2-2 X-STREAM F Field Housing Fig. 2-4: X-STREAM F - dimensions (approx. mm [inch]) Housing Permissible ambient temperature range -4 F to 122 F ( -20 °C to +50 °C) Weight: approx.
X-STREAM 2-11 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2 T echnical Dat a Power Connection Connection via internal screw terminals near cable glands, (fig.
X-STREAM 2-12 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2-2-2 X-STREAM F Signal inputs / outputs All signal lines need to be connected to internal screw terminals (fig.
X-STREAM 2-13 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 2 T echnical Dat a 2-3 Information on the Nameplate Label The analyzer nameplate label gives importa.
X-STREAM 2-14 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG.
X-STREAM 3-1 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles Chapter 3 Measuring Principles 3-1 Infrared Measurement (IR) Ultraviolet Measurement (UV) X-STREAM series analyzers support several measuring principles or combinations of principles, depending on measured gas components.
X-STREAM 3-2 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3-1 Infrared (IR) and Ultraviolet (UV) Measurement Fig.
X-STREAM 3-3 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles 3-1-2 Opto Pneumatic Measuring Principle This high performance principle is used for IR and UV measurements and utilizes a separated analysis cell whose one side is passed by the sample gas.
X-STREAM 3-4 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG The detector is filled with the gas to be measured and therefore sensitive to the related characteristic wavelength band only . The absorption chamber is sealed with a window transparent for IR radiation [CaF 2 (Calziumfluorid)] .
X-STREAM 3-5 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles Filter Analysis cell (undivided) Pyroelectrical detector (solid state detector) Adaptor .
X-STREAM 3-6 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG Fig. 3-4 shows the mechanical design of an pyro detector assembly: Attached to the chopper are IR source (07) as well as analysis cell (09) with signal detector unit (filter cell 14 / 15) and pyroelectrical detector with integrated pre amplifier (16).
X-STREAM 3-7 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles Fig. 3-4: Photometer Assembly with Pyroelectrical Detector 03 Chopper 04 / 05 Filter ape.
X-STREAM 3-8 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG The design of a gas detector photometer assembly is on principle similiar to the pyro detector design. One main difference is the analysis cell, that now is lengthwise divided and both ends sealed with windows.
X-STREAM 3-9 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles 3-2 Oxygen Measurement 3-2 Oxygen Measurement T wo different principles are used for measuring oxygen concentrations.
X-STREAM 3-10 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3-2 Oxygen Measurement Fig. 3-6: Paramagnetic Oxygen Detector, Assembly Principle 1 Permanentmagnet .
X-STREAM 3-11 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles 3-2 Oxygen Measurement 3-2-2 Electrochemical Measurement This sensor uses the principle of galvanic cells, fig. 3-7 shows the design. Fig.
X-STREAM 3-12 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3-2 Oxygen Measurement Resistance (6) Thermistor (5) (11) ( re d ) (black) (+) Lead anode (1) (-) Go.
X-STREAM 3-13 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles 3-2 Oxygen Measurement 3-2-3 Special Hints on Oxygen Sensors Paramagnetic Sensor The table below shows how accompanying gases interfere the paramagnetical oxygen measurement.
X-STREAM 3-14 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3-3 Thermal Conductivity Measurement Thermal Conductivity Measurement primarily is used for measuring concentrations of hydrogen (H 2 ) and helium (He).
X-STREAM 3-15 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles 3-3-2 T echnical Implementation A block made either of aluminum, stainless steel or hastelloy contains two gas paths. Both, the volume of the block and the mass of the sensors have been minimized in order to obtain short response times.
X-STREAM 3-16 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3-4 Measurement Specification 3-4 Measurement Specification Sample gas components and measuring ranges (standard configurations *) ) Note! The following table shows generic data.
X-STREAM 3-17 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 3 Principles 3-4 Measurement Specification Performance Specifications Table 3-7: Measurement Performance Specifications 1 Related to full scale 2 Related to measuring value; 1 psi = 68.
X-STREAM 3-18 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG.
X-STREAM 4-1 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation Chapter 4 Inst allation This chapter describes how to install the diffe- rent analyzer models in a safe manner and gives instructions on what to care about.
X-STREAM 4-2 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-1 Inst allation - Abstract T o stay in compliance with regulations regar- ding electromagnetic compatibility it is recom- mended to use only shielded cables, as optionally available from Emerson Process Management or equivalent.
X-STREAM 4-3 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-2 Installation - Gas Conditioning Do not interchange gas inlet and outlet! All gases must be conditioned before supplying! When supplying corrosive gases ensure that gas path components are not affected! Max.
X-STREAM 4-4 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG Internal Solenoid V alve Block Supply overpressure for all gases is limited to 0.7 to 7.2 psi (50 to 500 hPa) when the analyzer is equipped with an internal solenoid valve block.
X-STREAM 4-5 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-3 Electrical Connections 4-3 Installation - Electrical Connections ELECTRICAL SHOCK .
X-STREAM 4-6 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG The analyzer variations X-STREAM GP and GPS are intended for horizontal orientation during operation. X-STREAM GP / GPS with mounting frames beside the front panel may be mounted into a rack (rack mount version).
X-STREAM 4-7 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Fig.
X-STREAM 4-8 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Preparation of signal cables All signal cables are to be connected via submin-d connectors.
X-STREAM 4-9 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Fig.
X-STREAM 4-10 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Relay status Signals Design: dry relay contacts Electrical specification: max. 30 V , 1 A, 30 W Note! T ake care of the special installation instructions in section 4-5! Fig.
X-STREAM 4-11 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Digital Inputs & Outputs Design: Open collector (outputs) Electrical specification: outputs: max.
X-STREAM 4-12 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG Preparation of power cable The power power inlet appliance is of type IEC. Use a standard power cable with IEC connector complying with your local requirements to supply the instrument with power .
X-STREAM 4-13 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS The number and assignment of gas inlet and outlet fittings depends on the application and is given on a label attached to the analyzer's rear panel adjacent to the fittings.
X-STREAM 4-14 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG All signal cables are to be connected via screw terminals, located at the analyzer's rear panel. The strain reliefs provide a metal strip to connect Preparation of signal cables Supported wire cross sections: 24 to 14 A WG (0.
X-STREAM 4-15 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation Modbus interface S pecification and driving the interface: Chapter 9 The lower 5 term.
X-STREAM 4-16 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Relay status Signals Design: dry relay contacts Electrical specification: max.
X-STREAM 4-17 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Digital Inputs & Outputs T o connect the terminals remove the cover at the analyzer's rear panel (4 screws).
X-STREAM 4-18 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS Connecting the power cable The power cable is connected by screw termi- nals, located at the analyzer's rear panel.
X-STREAM 4-19 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-4-1 Inst allation - X-STREAM GP , X-STREAM GPS ELECTRICAL SHOCK HAZARD! Before comp.
X-STREAM 4-20 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-4-2 Installation - X-STREAM F 4-4-2 X-STREAM F The X-STREAM F analyzer variation is intended for o.
X-STREAM 4-21 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-4-2 Installation - X-STREAM F The instrument provides internal screw termi- nals for connecting power and electrical signal cables. This requires opening the instrument during installation utilizing the sash fasteners.
X-STREAM 4-22 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG All signal cables are to be connected via screw terminals, located inside the analyzer . The instruments inner components are accessible Preparation of signal cables Supported wire cross sections: 24 to 14 A WG (0.
X-STREAM 4-23 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation Signal inputs / outputs Analog signal outputs 4-4-2 Installation - X-STREAM F ELECTRI.
X-STREAM 4-24 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-4-2 Installation - X-STREAM F Modbus interface S pecification and driving the interface: Chapter 7.
X-STREAM 4-25 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation Relay status Signals Design: dry relay contacts Electrical specification: max.
X-STREAM 4-26 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4-4-2 Installation - X-STREAM F Digital Inputs & Outputs For accessing the terminals open the instrument's front door . The leftmost terminal strip is reserved for the digital inputs and outputs.
X-STREAM 4-27 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation Connecting the power cable The power cable is connected by screw terminals, located inside analyzer . Supported wire cross sections: 24 to 12 A WG (0.
X-STREAM 4-28 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG After all connections are established in correct manner and verified, • close the front door and secure it utilizing the two sash fasteners.
X-STREAM 4-29 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-5 Hints on Wiring Signal Inputs and Outputs 4-5 Installation - Hints on Wiring Emer.
X-STREAM 4-30 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG Local on-site conditions usually differ from test conditions and may require special measures. This is when strong fields are expected, potentially generating high parasitic currents on the cable shield .
X-STREAM 4-31 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG 4 Installation 4-5-2 Wiring Inductive Loads Switching inductive loads is a standard application generating electromagnetic disturbances: The moment an inductive load (e.
X-STREAM 4-32 Instruction Manual HASAxE-IM-HS 07/2006 Emerson Process Management GmbH & Co. OHG • It‘s better to apply "parallel" wiring at which each single load is supplied by a se.
X-STREAM 5-1 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface Chapter 5 User Interface and Sof tware Menus This chapter describes the elements of the X-STREAM user i nterface, the structure and the contents of the software menus.
X-STREAM 5-2 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-2 User Interface 5-2-1 The Display The display is either a alphanumeric 4x20 character liquid cristal or vacuum fluorescence display . The information shown depends on the currently selected menu.
X-STREAM 5-3 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-2 User Interface 5-2-3 The Keys Six keys are provided to operate the menu system.
X-STREAM 5-4 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-2 User Interface LEFT key: Mode Function Measure Exit measuring screen Browse One page/level back i.
X-STREAM 5-5 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-3 Software 5-3 Software 5-3-1 Power-On Sequence After Power-On a sequence of start-up messa- ges is displayed. During this time all keys are disabled.
X-STREAM 5-6 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-3 Software 4 different line types distinguish between diffe- rent functionalities: T ype Description TEXT Simple text (not selectable by cursor) V ARIABLE This line has 2-3 fields: 1.
X-STREAM 5-7 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-3 Software Editing An editing session allows changing a variable's value. It is initiated by pressing the key . The cursor is now placed over the last character of the current value.
X-STREAM 5-8 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-3 Software 5-3-5 Access Levels Access levels allow to prevent unauthorized personnel from changing parameters.
X-STREAM 5-9 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-3-6 Special Screens Depending on the latest operator action one of the following m.
X-STREAM 5-10 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System 5-4 Menu System This section describes all menus of the X- STREAM gas analyzer software and how the menus are linked. Fig. 5-3 on page 5-12 gives a graphical over- view of the menu structure.
X-STREAM 5-11 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Power-On and Measuring Screen 5-4-1 Power-On and Measuring Screen.
X-STREAM 5-12 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Control Menu Fig. 5-3: X-STREAM Software menu structure.
X-STREAM 5-13 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Control Menu 5-4-2 Control Menu Control.
X-STREAM 5-14 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-2-1 Zero Calibration Menu 5-4 Menu System - Control Menu Dual channel instrument: Pressing the key returns to the optional gas component selection menu to perform the same action for the other measuring channel.
X-STREAM 5-15 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-2-2 Span Calibration Menu 5-4 Menu System - Control Menu Dual channel instrument: Pressing the key returns to the optional gas component selection menu to perform the same action for the other measuring channel.
X-STREAM 5-16 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG Control.. Adv. calibration.. 5-4-2-3 Advanced Calibration Menu 5-4 Menu System - Control Menu When t.
X-STREAM 5-17 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Control Menu 5-4-2-4 Calibration Status Screen The 1 st line shows the currently applied gas This line gives the measured concentration Gasflow Sample CO.
X-STREAM 5-18 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-2-5 Apply Gas Menu T oggle between Sample , ZeroGas , Sp anGas and None using the and keys. Entering the selected value using the key opens the related valve and closes all other valves (except for None : all valves are closed).
X-STREAM 5-19 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-2-6 Acknowledgements Menu 5-4 Menu System - Control Menu Control.. Acknowledgements.. This page allows to acknowledge (=reset) all status messages:Just press the key to per- form an acknowledgement.
X-STREAM 5-20 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Setup Menu 5-4-3 Setup Menu Setup.
X-STREAM 5-21 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Display Setup Menu 5-4-3-1 Display Setup Menu If system is setup .
X-STREAM 5-22 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG Current temperature sensor 1 value Set temperature unit:: °C, F Set temperature value's decima.
X-STREAM 5-23 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-3-1-1 Display Language Setup Menu 5-4 Menu System - Display Setup Menu Language Language: English Select your prefferred language for the analyzer software.
X-STREAM 5-24 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-3-1-2 Menu Access Setup Menu Use lines 1 to 3 to specify if the related access level is protected by code or not: Off : Menu access is not protected.
X-STREAM 5-25 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Display Setup Menu 5-4-3-1-3 Component Setup Menu Dual channel in.
X-STREAM 5-26 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Calibration Setup Menu 5-4-3-2 Calibration Setup Menu This parameter defines if the tolerance check during calibration procedures is activated or not.
X-STREAM 5-27 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Calibration Setup Menu 5-4-3-2-1 Calibration Gases Setup Menu Setup.. Calibration.. Calibration gases.. Enter the concentration value for the zero gas to be used during zero calibration.
X-STREAM 5-28 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Calibration Setup Menu 5-4-3-2-2 Interval Time Setup Menu Enter the time between two zero calibrations when in autocal mode. If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM 5-29 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Measurement Setup Menu 5-4-3-3 Measurement Setup Menu If system is setup accordingly access level 2 code must be entered to access this menu.
X-STREAM 5-30 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-3-3-1 Signal Damping Setup Menu Enter the electronic signal damping time. Lower values give faster updated measuring results, higher values may suppress noise from varying gas concentrations.
X-STREAM 5-31 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Measurement Setup Menu 5-4-3-3-2 Cross Interference Setup Menu T .
X-STREAM 5-32 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-3-3-3 AD Conversion Setup Menu Select the first line to setup the AD converters offset for the first channel: The offset's unit are counts. Setup.
X-STREAM 5-33 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-3-4 In/Outputs Setup Menu Select how the internal pump status is controlled. Available options: Menu: This allows to manually switch the pump via the related entry in the "Control.
X-STREAM 5-34 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Analog Output Setup Menu 5-4-3-4-1 Analog Output Setup Menu Signal: C1-C2 SignalRange: 0-20mA Scaling.. Trim output.. Select the signal range.
X-STREAM 5-35 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface Enter this line to select which values to output via the analog outputs.
X-STREAM 5-36 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Analog Output Setup Menu Enter this line to select the analog outputs ranges. In addition this entry allows to setup the analog outputs to follow the NAMUR NE43 recommendations.
X-STREAM 5-37 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Analog Output Setup Menu Table 5-2: Analog Output Signal Setting & Operation Modes Signal R a nge Sett ing Ope ra tion Mode Failur e Signal Le ve l a cc.
X-STREAM 5-38 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-3-4-1-3 Analog Output Signal Scaling Setup Dual channel instrument: O ptional gas component selection menu - Select the component to be setup 5-4 Menu System - Analog Output Setup Menu Setup.
X-STREAM 5-39 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Analog Output Setup Menu 5-4-3-4-1-4 Analog Output Signal T rimmi.
X-STREAM 5-40 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Analog Output Setup Menu 5-4-3-4-2 V alve Assignment Setup.
X-STREAM 5-41 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-3-5 Installed Options Setup Menu If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM 5-42 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-3-6 Communication Setup Menu Protocol: MODB RTU MODB mode: 32Bit ID number: 2 Interface: RS485/2w If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM 5-43 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Alarms Setup Menu 5-4-3-7 Alarms Setup Menu If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM 5-44 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-3-8 Save-Load Menu If system is setup accordingly access level 3 code must be entered to access this menu. Save-Load CfgData>COMport! COMport>CfgData.
X-STREAM 5-45 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 2 nd menu page FactData>CfgData.. CfgData>UserData.. UserData>CfgData.. 5-4 Menu System - Safe-LoadMenu Explanation of terms: FactData This is the factory setup analyzer configuration.
X-STREAM 5-46 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Miscellaneous Menu 5-4-3-9 Miscellaneous Screen This entry specifies if the analog.
X-STREAM 5-47 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Status Menu 5-4-4 Status Menu Status.
X-STREAM 5-48 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Status Menu 5-4-4-1 Failure Status Screen Status.. 1 Failures.. The number starting the "Failures.." line in the previous menu shows how much failures currently are active (here: 1).
X-STREAM 5-49 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-4-2 Check Requests Status Screen Status.. 0 Check requests.. 5-4 Menu System - Status Menu The number starting the "Check requests.
X-STREAM 5-50 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-4-3 Function Check Status Screen Status.. 1 Function checks.. 5-4 Menu System - Status Menu The number starting the "Function checks.." line in the previous menu shows how much such messages currently are upcoming (here: 1).
X-STREAM 5-51 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface Status.. 0 Off specs.. 5-4-4-4 Off Spec Status Screen 5-4 Menu System - Status Menu The number starting the "Off spec.." line in the previous menu shows how much such mes- sages currently are upcoming (here: 0).
X-STREAM 5-52 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Status Menu 5-4-4-5 Calibration Status Screen Status.. Calibration.. If a calibration is ongoing, this screens gives some information about the status.
X-STREAM 5-53 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-4-6 Measurement Status Screen Shows the detectors raw signal value. This is the signal without being linearized, compensated or converted to the expected unit.
X-STREAM 5-54 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4-4-7 Alarms Status Screen Dual channel instrument: Pressing the key returns to the optional gas component selection menu to open the same menu for the other measuring channel.
X-STREAM 5-55 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4-5 Info Menu Serial 1 Ax123 Serial 2 4567890 Revision 0.16 05 Range.. Installed options info menu 5-4-5-1, page 5-58 1 st menu page 5-4 Menu System - Info Menu Europe.
X-STREAM 5-56 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Info Menu 5-4-5-1 Measuring Range Info Screen Dual channel instrument: O ptional gas component selection menu - Select the component of interest.
X-STREAM 5-57 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5 User Interface 5-4 Menu System - Info Menu 5-4-5-2 Factory Settings Screen Access level 4 code must be entered to access this menu. Info.. Factory settings.
X-STREAM 5-58 Instruction Manual HASAxE-IM-HS 05/2006 Emerson Process Management GmbH & Co. OHG 5-4 Menu System - Info Menu 5-4-5-3 Installed Options Info Screen Installed options Valves Int+Ext C.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-1 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup Chapter 6 Initial St artup After unpacking and installing the instrument it's .
X-STREAM Emerson Process Management GmbH & Co. OHG 6-2 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup Language.. If system is setup accordingly access level 3 code must be entered to access this menu. Factory default setting is "no code required".
X-STREAM Emerson Process Management GmbH & Co. OHG 6-3 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup Sequence of keys to be entered for chan- ging the language of an analyzer , if the .
X-STREAM Emerson Process Management GmbH & Co. OHG 6-4 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup Next check the display settings for the mea- surement screen, temperature & pressure units and menu access using the DOWN and ENTER keys to access the related menu pages.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-5 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup 6-2 Checking the Instrument's Setup Press the LEFT key to return to the CALIBRA- TION setup menu and concentrate on the "T ol.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-6 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup S pecifying an interval time is required for unattended autocalibration: This means that the analyzer can initiate calibrations based on a given time interval.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-7 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup 6-2 Checking the Instrument's Setup Analog output.. PumpControl: Menu Note! If an internal pump is installed, the menu shows two lines, without pump the lower line does not show .
X-STREAM Emerson Process Management GmbH & Co. OHG 6-8 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup Signal: C1-C2 SignalRange: 4-20mA Scaling.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-9 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup 6-2 Checking the Instrument's Setup Normally it is expected that a gas concentration of "0" generates an similiar analog output of 0 mA.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-10 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup Si gna l Range Sett in g Ope ra tion Mode Fa i l ure Si gna l Le ve l a cc.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-11 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup 6-2 Checking the Instrument's Setup Press the LEFT key several times to return to the SETUP MENU. The cursor is now placed in the IN/OUTPUTS.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-12 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup Protocol: MODB RTU MODB mode: 32Bit ID number: 2 Interf.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-13 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup 6-2 Checking the Instrument's Setup The "S panRange" p arameter is .
X-STREAM Emerson Process Management GmbH & Co. OHG 6-14 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup Note! "Function" parameter factory setting is Off FS (fail-safe) unless otherwise specified at time of order .
X-STREAM Emerson Process Management GmbH & Co. OHG 6-15 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup S tandard mode: If an alarm is activated, the corresponding alarm outputs a "High" level signal.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-16 Instruction Manual HASAxE-IM-HS 05/2006 • Setting "HIGH" and "HIGH-HIGH" alarms S tandard mode: If an alarm is activated, the corresponding alarm outputs a "High" level signal.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-17 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup • Setting "LOW" and "LOW-LOW" alarms S tandard mode: If an alarm is activated, the corresponding alarm outputs a "High" level signal.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-18 Instruction Manual HASAxE-IM-HS 05/2006 6-2 Checking the Instrument's Setup The line "Save-Load-Update.." is used for saving/reloading a configuration and for resetting the instrument to the factory settings.
X-STREAM Emerson Process Management GmbH & Co. OHG 6-19 Instruction Manual HASAxE-IM-HS 05/2006 6 Initial Startup Now the most important parameter settings are checked and the instrument is setup to operate to your needs. This is the moment where you could save for backup reasons the changed data into a special memory .
X-STREAM Emerson Process Management GmbH & Co. OHG 6-20 Instruction Manual HASAxE-IM-HS 05/2006 Press the HOME key to return to the MEASURING SCREEN, because the setup checking procedure is finished.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-1 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance Chapter 7 Maintenance Maintenance carried out on a regular basis ensures long-term effi.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-2 Instruction Manual HASAxE-IM-HS 05/2006 7-2 Performing a Leak T est T o achieve best and proper measuring results you must ensure the gas path system does not have leaks. The following procedure describes how to per- form a leak test with focus on the instrument.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-3 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3 Calibration Procedures 7-3 Calibration Procedures Note! T o achieve best and proper measuring result s it is recommended to perform zero and span calibrations on a regular weekly basis.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-4 Instruction Manual HASAxE-IM-HS 05/2006 7-3 Calibration Procedures 7-3-1 Preparing a Calibration Before starting calibrations it is required to tell the instrument the calibration gas concen- trations.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-5 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3 Calibration Procedures Calibration gases... Tol.Check: Off Purge time: 15 s Interval time... By default the option TOL. CHECK (tolerance check) is disabled ( OFF ).
X-STREAM Emerson Process Management GmbH & Co. OHG 7-6 Instruction Manual HASAxE-IM-HS 05/2006 7-3 Calibration Procedures Calibration gases... Tol.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-7 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3 Calibration Procedures 7-3-2 Manual Calibration S tarting from the measurement screen press the DOWN key to open the MAIN MENU, enter the CONTROL.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-8 Instruction Manual HASAxE-IM-HS 05/2006 7-3 Calibration Procedures After having started a calibration watch the screen for information about .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-9 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance When finished either press the LEFT key two times to return to the component selection .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-10 Instruction Manual HASAxE-IM-HS 05/2006 7-3-3 Advanced Calibration 7-3-3 Advanced Calibration S tandard manual calibration procedures offer limited funtionality: T o zero and span calibrate a dual channel instrument the operator has to manually start 4 procedures in proper sequence.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-11 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance As described earlier , advanced calibration is possible only with installed internal and/or external valves.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-12 Instruction Manual HASAxE-IM-HS 05/2006 7-3-3 Advanced Calibration Setup.. In/Outputs.. Valve assignment.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-13 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance Note for dual channel instruments: The combinations may be assigned for channel 1 and 2 independently .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-14 Instruction Manual HASAxE-IM-HS 05/2006 7-3-3 Advanced Calibration T o start a ZERO calibration for AL L channels select the second line. S tarting from the measurement screen press the DOWN key to open the MAIN MENU and enter the CONTROL - ADV .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-15 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3-3 Advanced Calibration When the prepurge time has elapsed, the instrument starts t.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-16 Instruction Manual HASAxE-IM-HS 05/2006 7-3-3 Advanced Calibration T o start a SP AN calibration for ALL channels select the third line. S tarting from the measurement screen press the DOWN key to open the MAIN MENU and enter the CONTROL - ADV .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-17 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3-3 Advanced Calibration After span calibration has finished, the instru- ment closes the last open valve and opens the sample gas valve.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-18 Instruction Manual HASAxE-IM-HS 05/2006 7-3-3 Advanced Calibration T o start a ZERO & SP AN calibration for AL L channels select the last line. S tarting from the measurement screen press the DOWN key to open the MAIN MENU and enter the CONTROL - ADV .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-19 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance Gasflow V4 Procedure Zero 1 Time 29 s When the prepurge time has elapsed, the instrume.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-20 Instruction Manual HASAxE-IM-HS 05/2006 After span calibration has finished, the instru- ment closes the last open valve and opens the sample gas valve.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-21 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3-4 Auto Calibration 7-3-4 Auto Calibration 7-3-4-1 Additional Preparations for Auto Calibration As described in an earlier section autocal- ibration is possible only with installed internal and/or external valves.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-22 Instruction Manual HASAxE-IM-HS 05/2006 Setup.. In/Outputs.. Valve assignment.. Dual channel instrument: O ptional gas component selection menu - Select the component to be setup If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-23 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3-4 Auto Calibration Note for dual channel instruments: The combinations may be assigned for channel 1 and 2 independently .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-24 Instruction Manual HASAxE-IM-HS 05/2006 T o start any calibration select one of the two upper lines to perform the related calibration. Select the ZERO CALIBRA TION.. line. Dual channel instrument: O ptional gas component selection menu - Select the component to be calibrated.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-25 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3-4 Auto Calibration Gasflow ZeroGas CO2.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-26 Instruction Manual HASAxE-IM-HS 05/2006 7-3-4 Auto Calibration 7-3-4-3 Sp an Auto Calibration Dual channel instrument: O ptional gas component selection menu - Select the component to be calibrated.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-27 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3-4 Auto Calibration The span calibration procedure has finished when the last time interval shows remaining 0 seconds and the gas flow is sample .
X-STREAM Emerson Process Management GmbH & Co. OHG 7-28 Instruction Manual HASAxE-IM-HS 05/2006 7-3-4 Auto Calibration 7-3-4-4 Unattended Auto Calibration Before selecting any further line make sure the required calibra- tion gases are applied! Supply all calibration gases with the same flow as the sample gas (recommeded approx.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-29 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance Within the SETUP CALIBRA TION menu the "INTERV AL TIME.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-30 Instruction Manual HASAxE-IM-HS 05/2006 7-3 Calibration Procedures 7-3-5 Resetting a Calibration In case a wrong configuration was detected after calibration was performed (e.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-31 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance Changing the APPL Y GAS parameter opens the related valve. Available options are SpanGas, ZeroGas, Sample, None . The TIME line starts counting down the pre- purge time.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-32 Instruction Manual HASAxE-IM-HS 05/2006 7-3 Calibration Procedures CO2.1 0.000 ppm Procedure Zero 1 Time 0 s Gasflow Sample CO2.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-33 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-3 Calibration Procedures 7-3-8 Cross Interference Compensation Note! The following is valid for dual channel instruments only! Several gases interfere the measuring results when appearing in the gas stream at the same time.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-34 Instruction Manual HASAxE-IM-HS 05/2006 7-3 Calibration Procedures Now perform a calibration considering the following procedure (only possible by manual calibration!): 1. Zero cal both channels 2.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-35 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-4 Replacing the Electrochemical Sensor 7-4 Replacing the ElectrochemicalSensor In consequence of it‘s design the sensor‘s lifetime is limited and depends on theoretical designed life and Oxygen concentration.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-36 Instruction Manual HASAxE-IM-HS 05/2006 7-4 Replacing the Electrochemical Sensor EXPLOSIVE, FLAMMABLE AND HARMFUL GASES HAZARD ! Before open.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-37 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance Do not expose the sensor to a temperature other than the temperature range of -20 to +60'C (-4 to +140 F).
X-STREAM Emerson Process Management GmbH & Co. OHG 7-38 Instruction Manual HASAxE-IM-HS 05/2006 7-4 Replacing the Electrochemical Sensor If the electrolyte leaks due to sensor damage, put the sens.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-39 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-4 Replacing the Electrochemical Sensor 7-4-2 Opening the Analyzer ELECTRICAL SHOCK H.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-40 Instruction Manual HASAxE-IM-HS 05/2006 7-4 Replacing the Electrochemical Sensor 7-4-3 Locating the Sensor There are two options for the sensor to be placed (fig.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-41 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-4 Replacing the Electrochemical Sensor 7-4-4 Sensor Unit Disassembly The sensor unit consists of a mounting support, an electronics board and the sensor itself (fig.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-42 Instruction Manual HASAxE-IM-HS 05/2006 7-4 Replacing the Electrochemical Sensor 7-4-5 Sensor Amplifier Adjustment Strip P2 for Oxygen sensor signal Tp 1 Tp 2 Potentiometer R4 Fig. 7-11: OXS Board, top view T o do so, • power on the opened instrument.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-43 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance • Power off the analyzer and close the housing. In a next step now perform a zero and span calibration at least for the channel the replaced sensor is related to.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-44 Instruction Manual HASAxE-IM-HS 05/2006 7-5 Cleaning the Instrument's Outside 7-5 Cleaning the Instrument's Outside Do NOT drench the cloth, just moisten it to prevent liquid ente- ring the housing! • Clean the analyzer housing outside with the moistened cloth.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-45 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-6 Save / Restore Configuration Data Sets 7-6 Save / Restore Configuration Data Sets After a couple of days operating the instrument one can assume the operator has setup the instrument to his needs.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-46 Instruction Manual HASAxE-IM-HS 05/2006 If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-47 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-6-2 Restore UserData to CfgData If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-48 Instruction Manual HASAxE-IM-HS 05/2006 7-6 Save / Restore Configuration Data Sets 7-6-3 Restore FactData to Cfg Data If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-49 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance Before starting any of these procedures an external device (e.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-50 Instruction Manual HASAxE-IM-HS 05/2006 7-6 Save / Restore Configuration Data Sets 7-6-4-1 Save CfgData to COMPort If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-51 Instruction Manual HASAxE-IM-HS 05/2006 7 Maintenance 7-6 Save / Restore Configuration Data Sets T o verify , the saved data on the external device is not corrupted during transmission, you can now select V erify! from the SA VE-LOAD.
X-STREAM Emerson Process Management GmbH & Co. OHG 7-52 Instruction Manual HASAxE-IM-HS 05/2006 7-6-4-2 Restore COMPort to Cfg Data 7-6 Save / Restore Configuration Data Sets If system is setup accordingly access level 3 code must be entered to access this menu.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-1 Instruction Manual HASAxE-IM-HS 05/2006 Message / Level Potential Cause How to Solve 8 T roubleshooting Chapter 8 T roubleshooting 8-1 Abstra.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-2 Instruction Manual HASAxE-IM-HS 05/2006 Message / Level Potential Cause How to Solve 8-2 Solving Problems Indicated by Status Messages 8-2 So.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-3 Instruction Manual HASAxE-IM-HS 05/2006 Message / Level Potential Cause How to Solve 8 T roubleshooting 8-2 Solving Problems Indicated by Sta.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-4 Instruction Manual HASAxE-IM-HS 05/2006 Message / Level Potential Cause How to Solve Analyzer related messages 8-2 Solving Problems Indicated.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-5 Instruction Manual HASAxE-IM-HS 05/2006 Message / Level Potential Cause How to Solve 8 T roubleshooting Channel related messages (preceded by a channel tag, e.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-6 Instruction Manual HASAxE-IM-HS 05/2006 Message / Level Potential Cause How to Solve Channel related messages (preceded by a channel tag, e.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-7 Instruction Manual HASAxE-IM-HS 05/2006 Message / Level Potential Cause How to Solve 8 T roubleshooting Channel related messages (preceded by a channel tag, e.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-8 Instruction Manual HASAxE-IM-HS 05/2006 Effect Potential Cause How to Solve 8-3 Solving Problems Not Indicated by Status Messages The followi.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-9 Instruction Manual HASAxE-IM-HS 05/2006 Effect Potential Cause How to Solve 8 T roubleshooting 8-3 Solving Problems Not Indicated by Status Messages Fluctuation or invalid readout 1. Leak in gas path 2.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-10 Instruction Manual HASAxE-IM-HS 05/2006 Effect Potential Cause How to Solve Fluctuation or invalid readout (continued) 6. IR channel: Source not connected or defective 7. Analog preamplifier of affected channel defective 8.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-11 Instruction Manual HASAxE-IM-HS 05/2006 Effect Potential Cause How to Solve 8 T roubleshooting 8-3 Solving Problems Not Indicated by Status Messages Readout damping time too long 1. Wrong signal damping settings 2.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-12 Instruction Manual HASAxE-IM-HS 05/2006 8-4 T roubleshooting on Component s This section shows how to check and replace internal components, assumed to be the source of trouble.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-13 Instruction Manual HASAxE-IM-HS 05/2006 8 T roubleshooting 8-4 T roubleshooting on Component s 8-4-3 Measuring Points 8-4-3-1 Measuring Points at BKS Board (Pin 1 of all connectors are marked within the silk screen or by use of square pads) Fig.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-14 Instruction Manual HASAxE-IM-HS 05/2006 8-4 T roubleshooting on Component s Measuring Point: X 8 Measuring instrument: D V M Signal: approx.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-15 Instruction Manual HASAxE-IM-HS 05/2006 8 T roubleshooting 8-4 T roubleshooting on Component s Potential cause: - IR measurement: Signal level set point: U = 6 V SS (± 0.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-16 Instruction Manual HASAxE-IM-HS 05/2006 8-4 T roubleshooting on Component s 8-4-3-1-6 Analog Preamplifier a) Paramagnetic Oxygen Measurement Measuring Point: X 25 (channel 1) Measuring instrument: D V M Signal when connected to zero gas: 0 V dc (± 50 mV) ambient air (approx.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-17 Instruction Manual HASAxE-IM-HS 05/2006 8 T roubleshooting 8-4 T roubleshooting on Component s Fig. 8-4: OXS board, assembled, top view 8-4-3-2-1 Sensor Signal Measuring Points: Tp 1 (Signal) Tp 2 Measuring instrument: D V M Signal when connected to ambient air (approx.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-18 Instruction Manual HASAxE-IM-HS 05/2006 8-4 T roubleshooting on Component s 8-4-4 BKS 20 Board Jumper Configuration There is only one single jumper on the BKS board: J7 is for enabling SRAM battery buffe- ring.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-19 Instruction Manual HASAxE-IM-HS 05/2006 8 T roubleshooting 8-4-5 Fuse on BKS 20 Board Fig. 8-6: Allocation of fuse on BKS board Before replacing the fuse ensure the instrument is de-energized! The fuse is placed in a socket.
X-STREAM Emerson Process Management GmbH & Co. OHG 8-20 Instruction Manual HASAxE-IM-HS 05/2006.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-1 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus Chapter 9 Modbus Functions This chapter lists all Modbus functions and registers supported by X-STREAM gas analyzers. Refer to the www .Modbus-IDA.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-2 Instruction Manual HASAxE-IM-HS 05/2006 9-2 Supported Functions Modbus Description Function Code Remark 1 decimal (hex) ReadCoils 01 (0x01) f.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-3 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus This list is ordered alphabetically by the parameter name („Mnemonic“) column. Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-4 Instruction Manual HASAxE-IM-HS 05/2006 9-3 Modbus - List of Parameters and Registers This list is ordered alphabetically by the parameter name („Mnemonic“) column. Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-5 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-3 Modbus - List of Parameters and Registers This list is ordered alphabetically by the parameter name („Mnemonic“) column.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-6 Instruction Manual HASAxE-IM-HS 05/2006 9-3 Modbus - List of Parameters and Registers Attribute 1 Daniel / Enron Modicon Type Range / Enum va.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-7 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-3 Modbus - List of Parameters and Registers Attribute 1 Daniel / Enron Modicon Type Range .
X-STREAM Emerson Process Management GmbH & Co. OHG 9-8 Instruction Manual HASAxE-IM-HS 05/2006 9-3 Modbus - List of Parameters and Registers This list is ordered alphabetically by the parameter name („Mnemonic“) column. Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-9 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-3 Modbus - List of Parameters and Registers This list is ordered alphabetically by the parameter name („Mnemonic“) column.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-10 Instruction Manual HASAxE-IM-HS 05/2006 9-3 Modbus - List of Parameters and Registers This list is ordered alphabetically by the parameter name („Mnemonic“) column. Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-11 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-3 Modbus - List of Parameters and Registers This list is ordered alphabetically by the parameter name („Mnemonic“) column.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-12 Instruction Manual HASAxE-IM-HS 05/2006 9-3 Modbus - List of Parameters and Registers Attribute 1 Daniel / Enron Modicon Type Range / Enum v.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-13 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-3 Modbus - List of Parameters and Registers This list is ordered alphabetically by the parameter name („Mnemonic“) column.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-14 Instruction Manual HASAxE-IM-HS 05/2006 9-3 Modbus - List of Parameters and Registers Attribute 1 Daniel / Enron Modicon Type Range / Enum v.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-15 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus This list is ordered alphabetically by the parameter name („Mnemonic“) column. Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-16 Instruction Manual HASAxE-IM-HS 05/2006 9-4 Comparison of Registers and Parameters 9-4 M odbus - Comparison of Parameters and Registers Dani.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-17 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-4 M o d b u s - C o m p a r i s o n o f P a r a m e t e r s a n d Registers Daniel / Enro.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-18 Instruction Manual HASAxE-IM-HS 05/2006 9-4 M odbus - Comparison of Parameters and Registers Daniel / Enron Modicon Mnemonic Modbus-Register.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-19 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-4 M o d b u s - C o m p a r i s o n o f P a r a m e t e r s and Registers Daniel / Enron .
X-STREAM Emerson Process Management GmbH & Co. OHG 9-20 Instruction Manual HASAxE-IM-HS 05/2006 9-4 M o d b u s - C o m p a r i s o n o f P a r ameters and Registers Daniel / Enron Modicon Mnemonic Modbus-Register 4002 4002 CalibrationCountDB 4003 4003 CalValveStateDB 4004 4005 4004 4005 LinearizerError#DB 4006 4006 ParamAccessModeDB 4011.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-21 Instruction Manual HASAxE-IM-HS 05/2006 9 Modbus 9-4 M o d b u s - C o m p a r i s o n o f P a r a m e t e r s and Registers Daniel / Enron Modicon Mnemonic Modbus-Register 9044..9046 6087..6092 ADC_Temp2CoeffsDB (A0, A1, A2) 9047.
X-STREAM Emerson Process Management GmbH & Co. OHG 9-22 Instruction Manual HASAxE-IM-HS 05/2006 9-4 M o d b u s - C o m p a r i s o n o f P a r ameters and Registers.
X-STREAM Emerson Process Management GmbH & Co. OHG 10-1 Instruction Manual HASAxE-IM-HS 05/2006 10 Service Information Chapter 10 Service Information If factory repair of defective equipment is re- quired, proceed as follows: 1 .
X-STREAM Emerson Process Management GmbH & Co. OHG 10-2 Instruction Manual HASAxE-IM-HS 05/2006 10-2 Customer Service For order administration, replacement parts, applicaton assistance, on-site or factory repair , service or maintenance contract information, contact: In Europe: Emerson Process Management GmbH & Co.
X-STREAM Emerson Process Management GmbH & Co. OHG A-1 Instruction Manual HASAxE-IM-HS 05/2006 Appendix Appendix This chapter contains an extract from the Modbus publication "Modbus_over_seri.
X-STREAM Emerson Process Management GmbH & Co. OHG A-2 Instruction Manual HASAxE-IM-HS 05/2006 A- 1 Modbus Implementation A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-3 Instruction Manual HASAxE-IM-HS 05/2006 Appendix A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-4 Instruction Manual HASAxE-IM-HS 05/2006 A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-5 Instruction Manual HASAxE-IM-HS 05/2006 Appendix A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-6 Instruction Manual HASAxE-IM-HS 05/2006 A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-7 Instruction Manual HASAxE-IM-HS 05/2006 Appendix A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-8 Instruction Manual HASAxE-IM-HS 05/2006 A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-9 Instruction Manual HASAxE-IM-HS 05/2006 Appendix A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-10 Instruction Manual HASAxE-IM-HS 05/2006 A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-11 Instruction Manual HASAxE-IM-HS 05/2006 Appendix A-1 Modbus Implementation.
X-STREAM Emerson Process Management GmbH & Co. OHG A-12 Instruction Manual HASAxE-IM-HS 05/2006 A- 2 EC Declaration of Conformity A-2 EC Declaration of Conformity.
X-STREAM Emerson Process Management GmbH & Co. OHG A-13 Instruction Manual HASAxE-IM-HS 0 5 /2006 A-3 Block Diagrams A- 3 Block Diagrams X-STREAM GP T erminals version.
X-STREAM Emerson Process Management GmbH & Co. OHG A-14 Instruction Manual HASAxE-IM-HS 0 5 /2006 A-3 Block Diagrams X-STREAM GPS Sockets version.
X-STREAM Emerson Process Management GmbH & Co. OHG A-15 Instruction Manual HASAxE-IM-HS 0 5 /2006 X-STREAM F A-3 Block Diagrams.
X-STREAM Emerson Process Management GmbH & Co. OHG A-16 Instruction Manual HASAxE-IM-HS 0 5 /2006.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006.
X-STREAM Instruction Manual HASAxE-IM-HS 05/2006 ASIA - P ACIFIC Emerson Process Management Asia Pacific Pte Ltd 1 Pandan Crescent Singapore 128461 T +65 6777 821 1 F +65 6777 0947 Internet: www .
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Emerson HASAxE-IM-HS c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Emerson HASAxE-IM-HS - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Emerson HASAxE-IM-HS, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Emerson HASAxE-IM-HS va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Emerson HASAxE-IM-HS, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Emerson HASAxE-IM-HS.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Emerson HASAxE-IM-HS. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Emerson HASAxE-IM-HS ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.