Manuel d'utilisation / d'entretien du produit CS512 du fabricant Campbell Manufacturing
Aller à la page of 16
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe Revision: 12/06 Copyright © 2001-2006 Campbell Scientific, Inc..
Warranty and Assistance The CS512 OXYGUARD TYPE III DISSOLVED OXYGEN PROBE is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materi als and workmanship under norm al use and service for t welve (12) months from date of shipm ent unless specified otherwi se.
i CS512 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. General Information ................................................
CS512 Table of Contents ii Examples 1. Sample Program using P1 .......................................................................... 5 2. Sample Program using P2 .......................................................................... 5 3. Sample Program using VoltSE .
1 Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe An accurate and reliable sensor is a critical element in any measurement system. The Oxyguard Stationary Probe meets these criteria for the measurement of dissolved oxygen.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 2 2. Specifications Principle of Measurem ent: Membrane covered galvanic oxygen probe Output Signal: 2.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 3 3. Optional Probe Accessories PT4-L — Agitator for stagnant condi tions 18026 — DO electrolyte, 1 liter 18027 — Membrane kit , 10 mem branes 4.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 4 FIGURE 4-2. CS512 with PT4 Agitator Optional Repeat Cycle Timer for Agitators : Reciprocating act ion may be controlled by the opti onal solid state Repeat Cycle Tim er. It sends 12 VDC pulses to the agitator coil.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 5 7. Programming In the CR510, CR10X, and CR 23X datalogger, Instruction P1 or P2 can be used. Example 1 uses a single-ended m easurement for the CS512 probe; exampl e 2 uses a differential measurement.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 6 8. Calibration The multiplier is used to calibrate the CS512 probe. To calculate the m ultiplier: 1) Program the datalogger using a multiplier of one. 2) Place the OxyGuard probe in the air, shad ed from the sun.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 7 9. Maintenance The CS512 probe needs little maintenance. Regular cleaning of the m embrane is all that is required. The membrane is ve ry durable and can be cleaned with a cloth or soft paper.
Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 8 EXAMPLE 8. Sample Program using P86 and P22 Instructions 45: Do (P86) 1: 41 Set Port 1 Hi gh 46: Excitation with Delay (P22 ) 1: 1 Ex Channel 2: 20 Delay W/Ex (uni ts = 0.01 sec) 3: 0 Delay After Ex (uni ts = 0.
A-1 Appendix A. Dissolved Oxygen in Fresh Water Table.
Appendix A. Dissolved Oxygen in Fre sh Water Table A-2.
This is a bl ank page ..
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 Nort h Logan, Utah 84 321 UNITED STATES www.campbellsci.com info@campbellsci.com Campbell Scientific Afri ca Pty. Lt d. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Campbell Manufacturing CS512 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Campbell Manufacturing CS512 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Campbell Manufacturing CS512, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Campbell Manufacturing CS512 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Campbell Manufacturing CS512, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Campbell Manufacturing CS512.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Campbell Manufacturing CS512. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Campbell Manufacturing CS512 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.