Man ua l P ar t No: 1 10 26 33 9 D MO DE L VX462B USER'S MANUAL VXI PULSE GENERA TOR MO DU L E.
COPYRIGHT C&H T echn ologies, Inc. ( C& H) pr ovi des t hi s man ual "as i s" wi tho ut warr anty o f an y ki nd, either expressed o r i m pl i ed, in cl udi ng but not l i m i ted t o t he i m pl i ed warrant i es o f merch antabi l i t y and fit n e ss fo r a p a r t ic u la r p u r p o s e .
iii AMENDM ENT NOTICE C&H T echn ol ogies, Inc. m a kes ever y at t em p t t o pr ovi de up - to - dat e m a n uals wi t h the associ ated equi pm en t. O c ca si o n a l l y , c h a n g es ar e m ad e to th e eq ui p m en t wh e re i n i t i s n ec e s s ar y to pro v i d e am endm ents to the m an ual .
iv INTRO DUCTION Th i s m an ua l desc ri be s t he f un cti o nal o perati on of t h e C& H Model VX46 2B V XI Pu l s e Ge ne rator (Part No. 1 10263 35). Th i s m odul e i s o n e o f a num b er o f tes t an d da ta a cq ui s i ti on /c on trol mo d u le s in t h e V M E a n d V X I fo r ma t p r o v id e d b y C & H .
v TAB LE OF CONTENTS 1.0 GENE RA L DES CRIPTI ON ............................................................................................. 1 1.1 PURPO SE OF EQUIPMENT ............................................................................... 1 1.
vi LIST OF FIGUR ES Figur e 1 . E l ect r ical T i ming a n d O u t p u t C ha r a c t e r is t ic s ........................................................... 2 Figur e 2 . Sim pli f i ed Blo ck D i ag ra m ..........................................
1 1.0 GENERAL DESCRIP TION Th e VX 462 B i s a B - si z e V XIbus co m pat ib l e pr o gra m mabl e 20 MHz pulse g enerat o r with an addi t i onal 40 MHz square wav e f uncti on. The m o dul e ca n output si n gl e or doub l e pu l se p attern s th at ca n b e co n ti n uou s s tre am s or e x ter n a l l y tr i gg e re d or g ate d.
2 Fig ure 1 . El ect r ical Tim ing and Ou tput C har acter ist ics 1.2.4 En vi ronm en t a l Th e en vi ron m en t al spe ci f i cati o n s of t he m o dul e are : Operat i ng Temperatur e: 0 C t.
3 1.2.5 Bus Co m pli ance Th e m o dul e com pl i es wi t h t he VXIb us Speci f i ca t i on Rev i si on 1.4 f o r B - size r egister based mo d u le s a n d w it h V M E bu s S p e c if ic a t io n A N S I / I E E E S T D 1 0 1 4 - 1987 and IEC 821.
4.
5 2.0 INS TALLATIO N 2.1 UNPACK ING AND INSP ECTION In m o st cas es the VX4 62B i s i n di vi dual l y se al ed an d packa ged f or sh i pm en t. Veri f y t h at t h ere h as bee n no d amag e t o t he ship ping co nt a i ner .
6.
7 3.0 FUN CTION AL D ESCR IPTI ON 3.1 GE NERAL Th e VX4 62B i s con f i gure d, con t ro l l ed, an d status ed thr o ugh o n - bo ar d r e gist e r s ac ce ss i bl e th rou gh th e VX I b ac kp l a n e .
8 3. 3 INDIC A TORS Two L ED i ndi ca to rs are provi ded on the f ron t pa ne l . On e i n di cates th e M OD ID s ta tus a n d th e oth er i n di c a te s th e b oa rd s tatu s . MODI D: Thi s f ront pan el LED i l l um i na tes wh en ev er th e h ost pro cess o r appl i es t he MODID si gnal t o t he sl o t t h e mo d u le i s oc cup y i n g.
9 3.5 CONFI GURA TION REG IST ER S Th ere are se v eral t y pe s of regi s t ers us ed to conf i gure and con tro l the VX4 62B . The VXI co n f i gurat i on r egi st ers pr ovi de for co ntr ol and stat us as requi red by t he VXIbus speci f i cat i on.
10 00 VXI ID Bit 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 Write No t Used No t Used Read De vi ce Class Addre ss Space Manuf act ure r I D Dev ic e Class Dev ice Cla ss (Re gist er B ase d = b inar y 11) Addre ss Space Addr ess Space (A16 Only = binary 11) Manuf .
11 3.5.2 Pul se Con f i gurati on Regi sters Pul se Repe t i t i o n Interva l Regi ster (08) T h i s r e a d / w r i t e r e g i s t e r c o n t r o l s t h e p u l s e r e p e t i t i o n int e r v a l a nd t h e s t at e o f t he o utpu t rel ay . See F i gur e 6 f o r detai l s.
12 Del ay Regis t er (0A) T h i s r e a d / w r i t e r e g i s t e r c o n t r ol s d e l a y of th e p r i m a r y p u l s e f r o m th e tr i g g e r out or the de la y o f t h e s e c o n d p u ls e o f a d o u ble p u ls e fr o m t he t r ig g e r o u t .
13 Pul se Contro l R egi s te r (0 E) T h i s r e a d / w r i t e r e g i s t e r c o n f i g u r e s t h e o p e r a t i n g m o d e o f t h e p u l s e .
14 Lo w/H i gh Level Ref erence Regist er ( 10 & 12) Tw o 1 2 b i t D i g i ta l to A n a l og C on v er te rs ( D A C ) are used t o c reate t h e l o w and hi gh l ev el ref eren ces t o t h e o utput l o gi c. The se regi sters are wri te on l y .
15 4.0 OPERAT ING INSTR UCTIO NS Th e VX46 2B provi de s t hre e n orm al m odes, thre e tri gg ered m odes , an d t hre e spec i al m o des of ope rati on , al on g w i t h v ol tag e l ev el p rog ra m m i n g, ou tpu t di sc on n ec t, a n d pu l s e en ab l e con trol s .
16 4.2 PROGRA MMING SE QUENCE Th e RU N b i t i n th e Pul se Con tro l Reg i ster i s th e b as i c On /Of f con tro l f or pul se g en era t i on. A separ at e ou t put relay connect s/d i sc onnect s t he gener at ed pulse t o/fro m t he BNC co nn ect or .
17 4.4 TRIGG ERED MO DES OF OPERA TION Th e p ul se stream c an b e program m ed to run c onti nuous l y , out put a s i ng l e f un cti on wh en tri gg er e d, or out put co nti nuousl y as long as the gat e (t ri gger i n) i s hi gh (i.e., f ree run, t ri gge r, o r gat ed r epeti t i o n.
18 4.6 OUT PUT VOLT A GE LEVEL PRO GR A MM ING Two 12 bi t di gi tal t o an al og con v erters (DA C' s) provi de the out put swi tch ref ere nc e vol t ages .
19 5.0 TRO UBLE ANALYSIS 5.1 BUI LT IN TES T AND DI AG NO S T I CS Bui l t i n t est f unct i o n s ar e pro vi ded for t he VX 462B i n t he f orm o f r ead back r egis ters.
20.
A -1 APPENDI X A - BOAR D L AYO UT.
A -2.
B-1 APPENDI X B - C ONN EC TOR S PIN C B A 1 2 3 4 5 D08 D09 D10 D11 D12 - - - BG0IN* BG0OUT* D00 D01 D02 D03 D04 6 7 8 9 10 D13 D14 D15 GND - BG1IN* BG10UT* BG2IN* BG20UT* BG3IN* D05 D06 D07 GND - 11.
B-2 PIN C B A 1 2 3 4 5 - - - - - +5V GND - - - - - - GND - 6 7 8 9 10 - GND - - GND - - - - - - - - - G ND 11 12 13 14 15 - - - - - - GND +5V - - - - - - - 16 17 18 19 20 GND - - - - - - - - - GND - - - - 21 22 23 24 25 - GND - - - - GND - - - - GND - - - 26 27 28 29 30 - - GND - GND - - - - - - - - MODID 31 32 - - GND +5V GND - Fig ure B - 2.
B-3 BNC FRON T VIEW PULSE PULSE OUT TRIGGE R OUTP UT TRIG OU T TRIGGE R/GATE IN PUT TRIG IN Fig ure B - 3. Fr ont Panel Connector.
B-4.
N O T E S.
.
READER'S CO MM ENT FORM Yo ur co m m en t s assi st us in im pro vi ng t he usef ul nes s of C&H' s publ icati ons; t hey are an i m po rt ant part o f the i nputs used f o r revi si o n.
INSTR UCTION S In i t s con t i nu i n g ef f o rt to i m prov e do cum en t at i o n, C& H Tech nol ogi es , In c . p r ov i d e s th i s f or m f or use in sub m i t ti ng any comm ents o r suggesti o n s t h at t h e user m ay hav e.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté C&H Technologies Pulse Generator c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du C&H Technologies Pulse Generator - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation C&H Technologies Pulse Generator, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le C&H Technologies Pulse Generator va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le C&H Technologies Pulse Generator, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du C&H Technologies Pulse Generator.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le C&H Technologies Pulse Generator. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei C&H Technologies Pulse Generator ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.