INSTRUCTION MANUAL Pharmaceutical Refrigerator =MPR-311 (II) MPR-3110(H) ---MPR-314DR{II} MPR-311 DR(H).
Note: 1. No part of this manual may be reproduced in any form with the expressed written permission of SANYO. 2. The contents of this manual are subject to change without notice. 3. Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies.
CONTENTS PRECAUTIONS FOR SAFE OPERATION P. 1 CAUTIOINS FOR USAGE P.5 NAME AND FUNCTION OF PARTS P. 6 BEFORE COMMENCING OPERATION P.8 AUTOMATIC TEMPERATURE RECORDER P.10 START-UP OF UNIT P. 11 OPERATING INSTRUCTIONS P. 12 MAINTENANCE P. 14 TROUBLE SHOOTING P.
PRECAUTIONS FOR SAFE OPERATION o o o o e ~WARNING Do not use the unit outdoors. Current leakage or electric shOck may result if the unit is exposed to rainwater. Only qualified engineers or service personnel should install the unit. The installation by unqualified personnel may cause gas or liquid leakage, electric shock or fire.
PRECAUTIONS FOR SAFE OPERATION I o o o ~WARNING Never disassemble, repair, or modify the unit yourself. Any such work carried out by an unauthorized person may result in fire or injury due to a malfunction. Make the power supply to the unit is disconnected when the fluorescent light is replaced as this will prevent electric S110ck.
PRECAUTIONS FOR SAFE OPERATION o ~CAUTION Always hold the handle when closing the door. This will prevent the likelihood of a trapped finger. Never lean or press on the glass. Intentional force may cause injury if the glass breaks. Do not lean on the door.
CAUTIONS FOR USAGE G) If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor. ® This inner cabinet is refrigerated by the forced circulation of cooled air inside the chamber.
NAME AND FUNCTION OF PARTS .-. ii~r . ..""''' - -~ --- @ (Q) @ @ - IIIII il ' .. 1 i "- ~ I , .. . MPR-311 D(H) ([) Door Consists of double pair glass with heat ray reflection film. ® Frame (Product unit) @ Fluorescent lamp 20W white light.
NAME AND FUNCTION OF PARTS I Control panel ® 0' 0' II II El 0 0 0 ,0 " 0 0" DEFROST "" ~" " . - _:.All .. ._. TeMP. CONTROL UGHT MPR·311(H) ®@@ @ ® 0' II sll AA El 0 0 0 0 0 0" ~~ 'ow "~ ,~ DEFROST w,'m.
BEFORE COMMENCING OPERATION Installation site To operate this unit property and to obtain maximum performance, install the unit in a location with the following conditions. <D A location not subjected to direct sunlight Installation in a location subjected to direct sunlight may lead· to inadequate cooling.
BEFORE COMMENCING OPERATION I Installation <D Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a mild detergent and rinse with water.
AUTOMATIC TEMPERATUER RECORDER I Automatic temperature recorder (optional) To record the chamber temperature, an optional automatic temperature recorder is available. Please consult with our sales department or agency for recorder installation. The available recorder model is MTR-0620LH.
START-UP OF UNIT The following procedure should be adhered to for initial start-up and continuous operation. <D Connect unit to dedicated power supply. Do not put any product in the unit at this time. ® When the power is initially applied, the audible alarm may sound.
.-; OPERATING INSTRUCTIONS I Temperature set Set up of chamber tern perature Table 1 and 2 show the basic operation method. Follow the key operation sequence indicated in the table. The example in the table is based on the assumption that the chamber temperature set at 1 GOC.
OPERATING INSTRUCTIONS I Alarm, Safety function and Self diagnostic function This unit has the alarm and safety functions shown in Table 3 and Table 4, and also a self diagnostic function. Table 3 Alarm and safety functions (MPR-311 H)) Alarm or safety Situation Indication Buzzer Safety operation High temp.
MAINTENANCE &WARNING .. Disconnect Always ~ supply to the unit maintenance of the disconnect the power prior to any repair or unit in order to prevent electric shock or injury. Cleaning G) Clean this unit once a month. Regular cleaning keeps the unit looking new.
TROUBLE SHOOTING If the unit malfunctions, check out the following before calling for service. 1. If nothing operates even when switched on 1) Is unit connected to the power supply? 2) Is there a power failure? 3) Is the fuse blown.or the circuit breaker inactivated? 2.
SPECIFICATIONS Name Pharmaceutical Refrigerator Model MPR-311(H) MPR-311 D(H) • MPR-311DR(H) External dimensions W800 x D450x H1800 (mm) Internal dimensions W720 x D350x H1435 (mrn) Effective capaci.
& CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1.
" i t 'j ! "-j 'I 7FB6P10124000 SANYO Electric Co., Ltd. Refrigerating Products Division 1-1-1, Sakata Oizumi-Machi, Ora-Gun, Gunma 370-0596 Japan Printed in Japan.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Sanyo MPR-311D(H) c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Sanyo MPR-311D(H) - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Sanyo MPR-311D(H), vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Sanyo MPR-311D(H) va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Sanyo MPR-311D(H), mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Sanyo MPR-311D(H).
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Sanyo MPR-311D(H). Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Sanyo MPR-311D(H) ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.