MPR-513 MPR-513R MPR-1013 MPR-1013R Pharmaceutical Refrigerator INSTRUCTION MANUAL MPR-513R MPR-1013R.
1 Note: 1. No part of this manual may be reprodu ced in any form without the expressed w ritten permission of SANYO. 2. The contents of this manual are subject to change without notice. 3. Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies.
2 PRECAUTIONS FOR SAFE OPERATION It is imperative that the user compli es with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person.
3 PRECAUTIONS FOR SAFE OPERATION Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. Only qualified engineers or serv ice personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire.
4 PRECAUTIONS FOR SAFE OPERATION Do not pull out more than 2 draw ers at the same time when storing or taking out of the stocked material. This may cause injury by falling or ti pping of the unit if something heavy items are stored.
5 PRECAUTIONS FOR SAFE OPERATION Fix the shelves securely . Incomplete installation may cause injury or damage. When removing the plug from the power supply outlet, grip the pow er supply plug, not the cord. Pulling the cord may result in electr ic shock or fire by short circuit.
6 CAUTIONS FOR USAGE 1. If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor. 2. This inner cabinet is refrigerated by forced circul ation of cooled air inside the chamber.
7 CAUTIONS FOR USAGE Circuit breaker This unit is equipped with a circuit breaker on the back. Make sure to switch ON this breaker before the unit starts to run.
8 REFRIGERATOR COMPONENTS 10 11 5 6 3 13 4 12 9 14 8 2 7 1 14 14 11 5 9 3 13 15 16 8 2 15 16 14 1 12 10 7 6 MPR-513R MPR-1013R 17 10 4.
9 REFRIGERATOR COMPONENTS 1. Door: Sliding type. The recessed portion on the rail enables the self-closing of the door. The glass is pair construction. 2. Control panel: Panel can be opened when upper right corner of the lower part cover is pushed. Refer to page 10.
10 REFRIGERATOR COMPONENTS Control panel and keypad 1. Buzzer stop key (BZ): To silence the audible alarm, press this key. Press it once again to reactivate the alarm. 2. Set key (SET): Temperature setting mode is initiated by pre ssing this key. Once the key is pressed, the digit to be changed will flash.
11 INSTALLATION Installation site T o operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: 1. A location not subjected to direct sunlight Installation in a location subjected to dire ct sunlight may lead to inadequate cooling.
12 INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, cl ean them with a neutral detergent and wipe it off with a soft cloth washed in clean water.
13 OPTIONAL COMPONENTS Automatic temperature recorder An automatic temperature recorder is available fo r this refrigerator as an optional accessory. If an automatic temperature recorder is required, contact your Sanyo dealer. For the proper usage of temperature recorder, refer to an instruction manual included with the recorder.
14 OPTIONAL COMPONENTS Attachment of recorder MTR-G04 WARNING Alw ays disconnect the pow er supply plug before installing an automatic temperature recorder in order to prevent electric shock or injury. 1. Follow the instruction manual provided with the temperature recorder and attach the recorder to the fixture MPR-S7.
15 START-UP OF UNIT Follow the procedures for the initia l and consequent operations of the unit. 1. Connect the power cord to the dedicated outlet with appropriate rating. 2. On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the alarm stop key (BZ).
16 TEMPERATURE SETTING Chamber temperature Table 1 shows the basic procedure for setting the cham ber temperature. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the desired temperature is 4 o C.
17 TEMPERATURE SETTING Alarm temperature setting This unit is provided with both high and low temperatur e alarms. The temperature at which the alarm is activated may be changed. The available set range for high temperature alarm is between +2 o C and +14 o C and -2 o C and -14 o C for low temperature alarm against the chamber temperature.
18 ALARMS & SAFETY FUNCTIONS This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions. Table 5. Alarms and safety functions Alarm & Safety Situation Indication Buzzer Safety operation High temperature alarm If the chamber temperature deviates from the set temperature +2 o C or up to +14 o C.
19 ALARMS & SAFETY FUNCTIONS Remote alarm terminal WARNING Alw ay s disconnect the pow er supply to the unit pr ior to connection of alarm equipment to the terminal in order to prevent el ectric shock or injury. The terminal to which a lead wire of remote alarm is connected is positioned in t he box at the rear of the cabinet.
20 ROUTINE MAINTENANCE WARNING Alw ay s disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electr ic shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance.
21 ROUTINE MAINTENANCE Clean the evaporating tray once a m onth following the procedure below. Cleaning of evaporating tray 1. Remove the screws on the bottom of the unit cover and remove the cover. See Fig. 1. 2. For model MPR-513 and MPR-513R, remove the evaporating tray mounting plate by pulling out the clips towards front.
22 ROUTINE MAINTENANCE Replacement of fluorescent lamp 1. The fluorescent lamp is placed vertically at the center of the frame (MPR1013 and MPR-1013R) and horizontally at the top front of the cabinet (MPR-513 and MPR-513R) 2. Turn off the fluorescent lamp switch and disconnect the power supply plug .
23 TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. If nothing operates even when sw itched on 1. There is a power failure. 2. The circuit breaker is inactivated. 3. The unit is not connected to the power supply.
24 SPECIFICATIONS Name Pharmaceutical Refrigerator Model MPR-1013 MPR-1013R MPR-513 MPR-513R External dimensions W1800 x D600 x H1790 (mm) W900 x D600 x H1790 (mm) Internal dimensions W1700 x D465 x H1300 (mm) W800 x D465 x H1300 (mm) Effective capacity 1037 L 1034 L 489 L 486 L Exterior Painted steel Interior Stainless steel Door 2 pcs.
25 PERFORMANCE Model MPR-513, MPR-513R Temperature control range +2 o C to +14 o C Usable ambient temperature -5 o C to +35 o C Noise level 45dB (A scale) Maximum pressure 1500 kPa Rated voltage AC 11.
26 CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Refrigerator contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: (List all potentially hazardous materials that have been stored in this unit.
7FB6P101277006 (5 April 2005) Recycl ed paper SANYO Electric Biomedical Co., Ltd. 10-15, Hongo 3-Chome Bunkyoku, T okyo 1 13-8434 Jap an Printed in Japan.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Sanyo MPR-1013 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Sanyo MPR-1013 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Sanyo MPR-1013, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Sanyo MPR-1013 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Sanyo MPR-1013, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Sanyo MPR-1013.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Sanyo MPR-1013. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Sanyo MPR-1013 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.