© 1998–2003 National Instruments Corp. All rights reserved. LabVIEW™, National Instru ments™, NI™, ni.com™, NI-DAQ ™, and SCXI™ are t rademarks of National Instrument s Corporation. Product and com pany names mentioned herein are trademarks or trade names of their respective companies.
SCXI-1321 T erminal Block Installation Guide 2 ni.com italic Italic text denotes variables, emphasis, a cross reference, or an int roduction to a key concept. This font also denote s te xt that is a placeholder for a word or v alue that you must supply .
© National Instruments Cor poration 3 SCXI-1321 T erminal Block Installa tion Guide Connecting the Signals Note Refer to the Read Me First: Safety and Radio-Frequency Interference document before removing equipment covers, or connecting or disconnecting any signal wires.
SCXI-1321 T erminal Block Installation Guide 4 ni.com Figure 1. SCXI-132 1 Parts Locator Diagram 1 Strain-Relief Bar 2 Strain-Relief Screws 3 Safety-Ground Lug 4 Mating Connector 5 Thumbscrew 6 Top-Co.
© National Instruments Cor poration 5 SCXI-1321 T erminal Block Installa tion Guide Figure 2. SCXI-1321 Circuit Parts Locator Diagram Installing the T erminal Block To connect the terminal block to the SCXI mo dule front connector, complete the follow ing steps: 1.
SCXI-1321 T erminal Block Installation Guide 6 ni.com Note T o minimize the temperature gradient inside the termin al block, move the SCXI chassis aw ay from any extr eme temperature dif ferential. Specifications All specifications are typical at 25 °C unless otherwise specified .
© National Instruments Cor poration 7 SCXI-1321 T erminal Block Installa tion Guide Environmental Operating temperature........................ .... 0 to 50 °C Storage temperature ............. ................. . –20 to 70 °C Humidity ..........
SCXI-1321 T erminal Block Installation Guide 8 ni.com Note Refer to the Declaration of Conformity (DoC) for this product fo r any additional regulatory compliance in formation. T o obtain th e DoC for this p roduct, click Declarations of Conformity Information at ni.
© National Instruments Cor poration 9 SCXI-1321 T erminal Block Installa tion Guide The factory default v alue of all the nulling resistors on the terminal blo ck is 39 k Ω . Note These resistors are socketed for easy re placement. These sockets best f it a 1/4 W resistor lead size.
SCXI-1321 T erminal Block Installation Guide 10 ni.com Assuming a strain-gauge range with a gauge factor of GF = 2 and a quarter-bridge conf iguration, this range corresponds to ±1,498 µε , as given by the strain formula for a quarter-bridge strain-gauge configuration: where Figure 3.
© National Instruments Cor poration 11 SCXI-1321 T erminal Bloc k Installa tion Guide Figure 4. Half-Bridge Nulling Circuit CH+ CH+ CH– EX+ EX– + – + – R 1 R 2 4.
SCXI-1321 T erminal Block Installation Guide 12 ni.com Figure 5. Full-Bridge Nulling Cir cuit Shunt Calibration Shunt calibration prov ides a method of adjusti ng for gain error to improve accuracy.
© National Instruments Cor poration 13 SCXI-1321 T erminal Bloc k Installa tion Guide z is the module slot of the SCXI-1121 w is the channel o f the module that you want to engage th e shunt and take.
SCXI-1321 T erminal Block Installation Guide 14 ni.com gauge with a gauge factor of GF = 2, V exc = 3.3 V , and R = 120 Ω prod uces the following: V change = 0.3321 mV Replacing the strained voltage by V change in the quarter-bridge strain equation produces an equ i valent 199 µε of change.
© National Instruments Cor poration 15 SCXI-1321 T erminal Bloc k Installa tion Guide Y ou can read a temperature at the begi nning of the test and use that value with the data that follows. Using this method assumes that there are no temperature v ariations during the measurement period.
321924C-01 J an03 *321924C-01* W3 Nulling circuit of Channel 2 is enabled (factory default setting) Nulling circuit of Channel 2 is disabled W4 Nulling circuit of Channel 3 is enabled (factory default setting) Nulling circuit of Channel 3 is disabled Table 4.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté National Instruments SCXI-1321 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du National Instruments SCXI-1321 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation National Instruments SCXI-1321, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le National Instruments SCXI-1321 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le National Instruments SCXI-1321, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du National Instruments SCXI-1321.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le National Instruments SCXI-1321. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei National Instruments SCXI-1321 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.