Manuel d'utilisation / d'entretien du produit NI 9217 du fabricant National Instruments
Aller à la page of 28
OPERATING INSTRUCTIONS AND S PECIFICATIONS NI 9217 4-Channel, 24-Bit, 100 Ω RTD Analog Input Module ni.com/manuals Deutsch Français.
NI 9217 Operating I nstruction s and Specificati ons 2 ni.com This doc u ment describes ho w to u se the National Instr u ments 9 217 and incl u des specif ications and term inal assignments for the NI 9217. V isit ni.com/info and enter rd s oftwarever s ion to determine which software yo u need for the mod u les yo u are u sing.
© National Inst ruments Corp. 3 NI 9217 Operating Instructi ons and Specific ations Hot Surface This icon denotes that the component may be hot. T o u ching this component may res u lt in bodily inj u ry . Safety Guidelines for Hazardous V oltages If hazardo u s voltages are connected to the mod u le, ta ke the follow ing preca u tions.
NI 9217 Operating I nstruction s and Specificati ons 4 ni.com Fig u re 1 sho ws the NI 9939 connector backshell. Figure 1. NI 9939 Connector Bac kshell Safety Guidelines fo r Hazardous Loca tions The .
© National Inst ruments Corp. 5 NI 9217 Operating Instructi ons and Specific ations only . Follo w these g u ideline s if yo u are installing the NI 9217 in a potentially explosi ve environm e nt. Not follo wing these g u idelines may res u lt in serio u s inj u ry or death.
NI 9217 Operating I nstruction s and Specificati ons 6 ni.com Special Conditions for Hazard ous Locations Use in Europe This eq u ipment has been e val u ated as EEx nC IIC T4 eq u ipment u nder DEMK O Certificate No. 03 A TEX 0324020X. Each mod u le is marked II 3G and is s u itable for u se in Zone 2 hazardo u s locations.
© National Inst ruments Corp. 7 NI 9217 Operating Instructi ons and Specific ations Connecting the NI 9217 The NI 9217 has a 16-terminal, de tachable scre w-terminal connector that provides con nections for fo u r R TD channels.
NI 9217 Operating I nstruction s and Specificati ons 8 ni.com Yo u can connect any combination of 3-wire or 4-wire R TDs to the NI 9217. Each NI 92 17 channel contains circ u itry that a u tomatically detect s the type of R TD connected to the channel.
© National Inst ruments Corp. 9 NI 9217 Operating Instructi ons and Specific ations Figure 3. Input Circuitry for On e Channel of the NI 9217 in 4-Wire Mode Each channel has one terminal, R TD+, to whic h yo u connect the positiv e lead of the R TD signal and one terminal, R TD–, to which yo u connect the negati ve lead of th e R TD signal.
NI 9217 Operating I nstruction s and Specificati ons 10 ni.com If yo u are u sing a 4-wire R TD, connect EX to the positi ve lead of the R TD. If yo u are u sing a 3-wire R TD, do not connect the R TD to the EX terminal.
© National Inst ruments Corp. 11 NI 9217 Operating Instructi ons and Specific ations Wiring for High-Vib ration Applications If an application is s u bject to high vibrat ion, National Instr u ments .
NI 9217 Operating I nstruction s and Specificati ons 12 ni.com Sleep Mode This mod u le s u pports a low-po wer sleep mode. S u pport for sleep mode at the system le vel depe nds on the chassis that the mod u le is pl u gged into. Refer to the chassis man u al for information abo u t s u pport for sleep mode.
© National Inst ruments Corp. 13 NI 9217 Operating Instructi ons and Specific ations Specifications The follo wing specifications a re typical for the range – 40 to 70 °C u nless otherwise noted. All v oltages are relat iv e to COM u nless otherwise noted.
NI 9217 Operating I nstruction s and Specificati ons 14 ni.com Con version time High-resol u tion mode....... .......... 200 ms per channel, 800 ms total for all channels High-speed mode .
© National Inst ruments Corp. 15 NI 9217 Operating Instructi ons and Specific ations Noise High-resol u tion mode......... ........ 0.003 °C High-speed mode .......... ........... ... 0 .02 °C Excitation c u rrent.................. ........... . 1 m A per channel Noise rejection Normal mode (50/60 Hz) High-resol u tion mode .
NI 9217 Operating I nstruction s and Specificati ons 16 ni.com High-resol u tion f ilter response 1,2 1 This im age is provid ed co u rtesy of Linear T echnology Corp. 2 High-speed filter response has the s ame characteristics as the high-resol u t ion filter response except that the first notch is at 14 kHz.
© National Inst ruments Corp. 17 NI 9217 Operating Instructi ons and Specific ations Overv oltage protection ....... ........... ... ±30 V between inp u ts MTBF .
NI 9217 Operating I nstruction s and Specificati ons 18 ni.com Physical Characteristics If yo u need to clean the mod u l e, wipe it with a dry to wel. Screw-terminal wiring ..... .............. ... 16 to 28 A WG copper cond u ctor wire with 7 mm (0.28 in.
© National Inst ruments Corp. 19 NI 9217 Operating Instructi ons and Specific ations Safety Maximum Voltage 1 Connect only v oltages that are within the follo wing limits. All terminals-t o-COM....... ........... .......±30 V Isolation Voltages Channel-to-channel.
NI 9217 Operating I nstruction s and Specificati ons 20 ni.com Caution Do not connect the NI 9217 to signals or u se for meas u rements within Meas u rement Categories III or IV .
© National Inst ruments Corp. 21 NI 9217 Operating Instructi ons and Specific ations Hazardous Locations U.S. (UL) .......... ......... ........... ........... . C lass I, Division 2, Gro u ps A, B, C, D, T4; Class I, Zone 2, AEx nC IIC T4 Canada (C-UL) .
NI 9217 Operating I nstruction s and Specificati ons 22 ni.com Operating h u midity (IEC 60068-2-56)....... ........... ............ 10 to 90% RH, noncondensing Storage h u midity (IEC 60068-2-56)....... ........... ............ 5 to 95% RH, noncondensing Maxim u m altit u de.
© National Inst ruments Corp. 23 NI 9217 Operating Instructi ons and Specific ations Electromagnetic Compatibility This prod u ct is designed to meet the req u ire ments of the follo wing standards o.
NI 9217 Operating I nstruction s and Specificati ons 24 ni.com ni.com/certification , search by mod u le n u mber or prod u ct line, and click the appropriate link in the Certification col u mn. Environmental Mana gement National Instr u ments is committe d to designing and man u fact u ring prod u cts in an en vironmentally res ponsible manner .
© National Inst ruments Corp. 25 NI 9217 Operating Instructi ons and Specific ations Calibration Yo u can obtain the calibration cert ificate and information abo u t calibration services for th e NI 9217 at ni.com/calibration . Calibration interval .
NI 9217 Operating I nstruction s and Specificati ons 26 ni.com Where to Go for Support The National Instr u ments W eb site is yo u r comple te reso u rce for technical s u pport.
© National Inst ruments Corp. 27 NI 9217 Operating Instructi ons and Specific ations K orea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 0 1 800 010 0793, Netherlands 31 .
© 2006–2008 Nati onal Instruments Corp. All rights reserved. 374187C-01 May08 National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instrument s Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments t rademarks.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté National Instruments NI 9217 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du National Instruments NI 9217 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation National Instruments NI 9217, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le National Instruments NI 9217 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le National Instruments NI 9217, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du National Instruments NI 9217.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le National Instruments NI 9217. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei National Instruments NI 9217 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.