MAXO 2 ® + A / AE USER’S GUIDE & OPERA TING INSTRUCTIONS BAT CAL MAXO 2 ® + A MAXO 2 ® + AE R217M40 Rev .D medicalManual.RevD.qxd 9/9/04 4:18 PM Page 1.
PREF ACE This manual describes the function, operation and maintenance of the MAXO 2 ® + hand-held oxygen analyzer . A member of Maxtec ® 's MAXO 2 ® analyzer line of oxygen analyzers and monitors, the MAXO 2 ® + utilizes the MAX-250+ oxygen sen- sor and is engineered for long life, maximum reliability and stable performance.
Classification Protection against electric shock: Internally powered equipment. Protection against water: IPX1 Mode of Operation: Continuous Sterilization: See section 7.
1.0 SYSTEM OVERVIEW 1.1 Base Unit Description The MAXO 2 ® + analyzer provides unparalleled performance and reliability due to an advanced design that includes the following features and operational benefits.
1.4 MAX-250+ Oxygen Sensor The MAX-250+ oxygen sensor offers stability and extra life. The MAX-250+ is a galvanic, partial pressure sensor that is specific to oxygen. It consists of two electrodes (a cathode and an anode), a teflon membrane and an electrolyte.
• If you are unsure about the displayed O 2 percentage. (see Factors influencing accurate readings.) A simple calibration may be made with the sensor open to static Ambient air . For optimum accuracy Maxtec ® recommends that the Sensor be placed in a closed loop circuit where gas flow is moving across the sensor in a controlled manner .
• Calibrate the MAXO 2 ® + at the same pressure as the sample gas. • If sample gases flow through tubing, use the same apparatus and flow rates when calibrating as when measuring. 3.4 Humidity Effects Humidity (non-condensing) has no effect on the performance of the MAXO 2 ® + other than diluting the gas, as long as there is no condensation.
FIGURE 4 Lock Lever Carefully , bring the two halves of the case together while positioning the wires so they are not pinched between the two case halves. The gasket separating the halves will be captured on the back case half. Reinsert the three screws and tighten until the screws are snug.
exhaled breath or secretions, unless you intend to dispose of the sensor , flow diverter and tee adapter after use. • Clean the sensor with a cloth moistened with iso propyl alcohol (65% alcohol/water solution).
9.2 Standard Replacement Parts Part Number Item R125P02-011 MAX-250+ Oxygen Sensor R125P03-002 MAX-250E Oxygen Sensor R115P85 MAX-250ESF Oxygen Sensor R217P08 Gasket RP06P25 #4-40 Pan Head Stainless Steel Screw R217P16-001 Front Assembly (Includes Board & LCD) R217P11-002 Back Assembly R217P24 Coiled Cable Assembly R217P09-001 Overlay 9.
14 MANUF ACTURED BY MAXTEC, INC. 10.0 WARRANTY The MAXO 2 ® + Analyzer is designed for medical oxygen delivery equipment and systems. Under normal operating conditions, Maxtec ® warrants the MAXO 2 .
17 WWW .MAXTECINC.COM 16 MANUF ACTURED BY MAXTEC, INC. medicalManual.RevD.qxd 9/9/04 4:18 PM Page 22.
medicalManual.RevD.qxd 9/9/04 4:18 PM Page 24.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté MaxTech MAXO2 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du MaxTech MAXO2 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation MaxTech MAXO2, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le MaxTech MAXO2 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le MaxTech MAXO2, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du MaxTech MAXO2.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le MaxTech MAXO2. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei MaxTech MAXO2 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.