Notice d'utilisation officielle

Invacare 10

68 pages·1.4 MB·Matériel d'oxygénothérapie·PDF·ENDEITNLJA
Invacare 10
Table des matières 31 sections accessories · ENGLISH · FRANÇAIS · special Notes · … AfficherMasquer
Page 1 of the Invacare 10 user manual
Page 1 · cliquer pour agrandir
Page 1
Invacare 10 diagram (page 1)
Fig. 1

Dealer: This manual MUST be given to the end user. User: Before using this product, read this manual and save for future reference.

For more information regarding Invacare products, parts, and services,

Page 2 of the Invacare 10 user manual — ACCESSORIES
Page 2 · cliquer pour agrandir
Page 2

WARNING

WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT.

ACCESSORIES

There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care provider for recommendations on which of these devices will be best for you. They should also give you advice on the proper usage, maintenance, and cleaning.

WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.

ENGLISH 2 Part No 1145728

Page 6 of the Invacare 10 user manual — NOTICE
Page 6 · cliquer pour agrandir
Page 6

SPECIAL NOTES

SPECIAL NOTES

Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words

SIGNAL WORDMEANING DANGERDanger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNINGWarning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTIONCaution indicates a potentially hazardous situation which, if not avoided, may result in property damage.

NOTICE

The information contained in this document is subject to change without notice.

DANGER Users MUST not smoke while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located. No smoking signs should be prominently displayed. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.

Page 7 of the Invacare 10 user manual —  DANGER
Page 7 · cliquer pour agrandir
Page 7

SECTION 1 —GENERAL GUIDELINES

SECTION 1 — GENERAL GUIDELINES

In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed.

CAUTION

“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”

WARNING

SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.

DANGER

Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel.

Page 8 of the Invacare 10 user manual — SECTION 1 —GENERAL GUIDELINES
Page 8 · cliquer pour agrandir
Page 8

SECTION 1 —GENERAL GUIDELINES

WARNING

The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air, and some that will not, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, it is necessary that all sources of ignition be kept away from the product and preferably out of the room in which it is being used. NO SMOKING signs should be prominently displayed.

A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless recommended by Invacare.

For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.

If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water, call qualified technician for examination and repair.

Keep the oxygen tubing, cord, and unit away from heated or hot surfaces, including space heaters, blankets, stoves and similar electrical appliances.

DO NOT move or relocate concentrator by pulling on the cord.

NEVER drop or insert any object into any opening.

Page 9 of the Invacare 10 user manual — Radio Frequency Interference
Page 9 · cliquer pour agrandir
Page 9

SECTION 1 —GENERAL GUIDELINES

WARNING

NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like. Fill humidifier with water to the level shown by the manufacturer. DO NOT overfill.

Keep unit at least three inches away from walls, draperies, furniture, and the like.

Invacare recommends that crush-proof oxygen tubing be used with this product and not exceed 50 ft. (15.2m) in length. Recommended use for Platinum 10 is up to 50 ft. in length of high flow tubing.

Maintenance

The Invacare Oxygen Concentrator was specifically designed to minimize routine preventive maintenance. Only professionals of the healthcare field or persons fully conversant with this process such as factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator.

Radio Frequency Interference

Most electronic equipment is influenced by Radio Frequency Interference (RFI). Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment.

Page 10 of the Invacare 10 user manual — SECTION 1 —GENERAL GUIDELINES
Page 10 · cliquer pour agrandir
Page 10

SECTION 1 —GENERAL GUIDELINES

WARNING

To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To Persons

Avoid using while bathing. If continuous usage is required by the physician’s prescription:

The concentrator MUST be located in another room at least seven feet from the bath.

DO NOT come in contact with the concentrator while wet.

DO NOT place or store product where it can drop into water or other liquid.

DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.

A product should NEVER be left unattended when plugged in.

This device is to be used only in accordance with the prescription of a physician and this owner’s manual. If at any time the patient or attendant conclude that the patient is receiving an insufficient amount of oxygen, the supplier and/or physician should be contacted IMMEDIATELY.

DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.

Close supervision is necessary when this product is used near children or physically-challenged individuals.

Use this product for only intended use as described in this manual.

DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.

Page 11 of the Invacare 10 user manual — Polarized Plug Instruction
Page 11 · cliquer pour agrandir
Page 11

SECTION 1 —GENERAL GUIDELINES

WARNING

Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance.

In certain circumstances oxygen therapy can be hazardous. Invacare recommends that you seek medical advice before using this machine.

This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining.

Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.

Polarized Plug Instruction

As a safety feature, this appliance has a polarized plug (one blade is wider than the other). This plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. DO NOT attempt to defeat this safety feature.

Page 12 of the Invacare 10 user manual — IOH200
Page 12 · cliquer pour agrandir
Page 12

SECTION 2 —FEATURES

SECTION 2 — FEATURES

Oxygen Purity Indicator Lights

(SensO2 model only)/ Fault

Invacare 10 — SECTION 2 —FEATURES diagram (page 12)
Fig. 2

and Power Indicator Lights

Cabinet Filter (one of two shown)

ACCESSORIES (NOT SHOWN)Remote Humidifier/Flowmeter Accessory -

IRCRF5 (Platinum 5 and XL Only)

PreciseRX™ Pediatric Humidifier/Flowmeter Accessory - IRCPF16 (Platinum 5 and XL Only)

Invacare 10 — SECTION 2 —FEATURES diagram (page 12)
Fig. 3

HomeFill II home oxygen compressor -

IOH200

NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the HomeFill II home oxygen compressor. The outlet fitting does not affect concentrator performance. Refer to the HomeFill II owner's manual, part number 1100873, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet fitting. For more information about the HomeFill II, contact your Invacare dealer.

Show 6 more labels / short text on this page
Oxygen Outlet · (LX model only) · Flowmeter · Power Switch · Circuit Breaker · Elapsed Time Meter
Page 13 of the Invacare 10 user manual — SECTION 3 —HANDLING
Page 13 · cliquer pour agrandir
Page 13

SECTION 3 —HANDLING

SECTION 3 — HANDLING

The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator.

If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare. Unpacking NOTE: For this procedure, refer to FIGURE 3.1.

1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator is to be used IMMEDIATELY, retain containers and packing materials for storage until use of the concentrator is required. Inspection
1. Inspect/examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components. Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place objects on top of repackaged concentrator.
Page 14 of the Invacare 10 user manual — SECTION 4 —SPECIFICATIONS
Page 14 · cliquer pour agrandir
Page 14

SECTION 4 —SPECIFICATIONS

SECTION 4 — SPECIFICATIONS

Electrical Requirements120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current InputPlatinum 5 and XL: 4.3 A
Platinum 104.7 A
Sound LevelPlatinum 5 and XL: 50 dB Average
Platinum 1058 dB Average

Altitude: Platinum 5 and XL - Up to 6,000 ft. (1828 meters) above sea level without degradation of concentration levels. From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below 90% efficiency

Platinum 10 - Up to 4,000 ft. (1230 meters) above sea level without degradation of concentration levels. From 4,000 ft. (1230 meters) to 13,129 ft. (4000 meters) below 90% efficiency

*Oxygen Output Concentration Levels:

All 5LXO2/5LX models 95.6% to 87% at 0.5 to 5 L/min. and all 10LXO2/10LX models 95.6% to 87% at 2 L to 10 L/min.

Maximum Outlet PressurePlatinum 5 and XL: 5 psi ± 0.5 psi. (34.5 kPa ± 3.45 kPa.)
Platinum 109 psi ± 0.5 psi. (62.0 kPa ± 3.45 kPa.)

*NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).

Page 15 of the Invacare 10 user manual — 5LX INDICATORS:
Page 15 · cliquer pour agrandir
Page 15

SECTION 4 —SPECIFICATIONS

5LXO2 INDICATORS:

! "

! " #

$% & '( %)!!* "

O 2

5LX INDICATORS:

Flow Range: Platinum 5 and XL - 0.5 to 5 L/min. For flowrates less than 1 L/min., we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory.

Platinum 10 - 2 L/min. to 10 L/min

Low Flow Alarm: Platinum 5 - 0 L/min. to 0.5 L/min., Platinum 10 - 0 L/min. to 1 L/min. Rapid Audible Alarm Beeping (No Accessories Connected)

Average Power Consumption:

Platinum 5 and XL400W Platinum 10: 585W

Pressure Relief Mechanism Operational at:

35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)

Change in maximum recommended flow when back pressure of 7kPa is applied:

0.7 L/min.

Show 14 more labels / short text on this page
· · · · · · · · · · · · ·
Page 16 of the Invacare 10 user manual — SECTION 4 —SPECIFICATIONS
Page 16 · cliquer pour agrandir
Page 16

SECTION 4 —SPECIFICATIONS

Filters: Cabinet (2), Outlet HEPA and Compressor Inlet.

Safety System: Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm. SensO2 Oxygen System (SensO2 Model) Low Flow Alarm.

Width18 3/8 in. ± 3/8 in. (46.7 cm ± 1 cm)
Height26 3/8 in. ± 3/8 in. (67.0 cm ± 1 cm)
Depth14 3/8 in. ± 3/8 in. (36.5 cm ± 1 cm)
WeightPlatinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ± 1 kg)
Platinum 1053 lbs ± 2 lbs (24.0 kg ± 1 kg)
Shipping WeightPlatinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ± 1 kg)
Platinum 1060 lbs ± 2 lbs (27.2 kg ± 1 kg)

Operating Ambient Temperature:

50°F - 95°F (10°C - 35°C) at 20-60% relative humidity

Exhaust: Platinum 5 and XL: Less than Ambient + 35°F (+19°C).

Platinum 10: Less than Ambient + 45°F (+25°C).

Oxygen Output:

Platinum 5 and XL: Less than Ambient + 4°F (+2°C).

Platinum 10: Less than Ambient + 8°F (+4.4°C).

Cabinet: Impact Resistant flame-retardant plastic cabinet that conforms to UL 94-V0.

Regulatory Listing: Double Insulated Product ETL certified complying with UL 1097 ETL certified to CSA C22.2 No. 68

Electrical: No extension cords.

Page 17 of the Invacare 10 user manual — SECTION 4 —SPECIFICATIONS
Page 17 · cliquer pour agrandir
Page 17

SECTION 4 —SPECIFICATIONS

Placement: No closer than three inches from any wall, furniture, draperies, or similar surfaces.

Tubing: 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). Platinum 10: Recommended use up to 50 ft high flow tubing with high flow cannula at all flow rates.

Relative Humidity20 to 60%

Time of Operation: Up to 24 hours per day.

Recommended Storage and Shipping Temperature:

-20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity

Page 18 of the Invacare 10 user manual — Select a Location
Page 18 · cliquer pour agrandir
Page 18

Introduction

Your oxygen concentrator is intended for individual use in the home. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.

Your provider will show you how to use your oxygen concentrator. He/She should be contacted with any questions or problems regarding your oxygen concentrator. This owner’s manual will tell you about your concentrator and will serve as a reference as you use your concentrator.

Select a Location

You may select a room in your house where using your oxygen concentrator would be most convenient. Your concentrator can be easily rolled from room to room on its casters.

Your oxygen concentrator will perform best when operated under the following conditions. Usage in environments other than those described may result in the need for increased equipment maintenance. The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and/or fumes.

Page 19 of the Invacare 10 user manual — Plug in Power Cord
Page 19 · cliquer pour agrandir
Page 19

Recommended Guidelines for Optimum Performance

Temperature50°F - 95°F (10°C - 35°C)

Electrical: No extension cords.

Placement: No closer than three inches from the wall, furniture, draperies, or similar surfaces.

Tubing and Cannula: 7 ft cannula with a maximum 50 ft of crush-proof Tubing (DO NOT pinch).

Platinum 10: Recommended use up to 50 ft high flow tubing with high flow cannula at all flow rates.

Environment: Smoke and soot-free. No confined spaces (Example: No closets).

Relative Humidity20 to 60%

Time of Operation: Up to 24 hours per day.

Flow Rate Platinum 5 and XL: From 0.5 L/min. to 5 L/min. For flow rates less than 1 L/min., use the Pediatric Flowmeter Accessory.

Platinum 10From 2 L/min. to 10 L/min
Minimum Operating Time30 Minutes

Plug in Power Cord

1. Plug in power cord to an electrical outlet.
Page 20 of the Invacare 10 user manual —  WARNING
Page 20 · cliquer pour agrandir
Page 20

Connect Humidifier (If So Prescribed)

WARNING

DO NOT overfill humidifier. Replace the humidifier cap and securely tighten. DO NOT reverse the oxygen input and output connections. Water from the humidifier bottle will travel through the cannula back to the patient.

Invacare 10 —  WARNING diagram (page 20)
Fig. 4

FIGURE 5.1 Filling the Humidifier

2. Place the humidifier bottle in the humidifier compartment.
Invacare 10 —  WARNING diagram (page 20)
Fig. 5

FIGURE 5.2 Humidifier Compartment

Humidifier Bottle · Humidifier Bottle

Page 21 of the Invacare 10 user manual — Connect Humidifier (If So Prescribed)
Page 21 · cliquer pour agrandir
Page 21
3. Remove the filter access door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door open. Refer to Detail “A” in FIGURE 5.3.
4. The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter. Refer to Detail “B” in FIGURE 5.3.
5. Attach it to the humidifier by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. Refer to Detail “C” in FIGURE 5.3.
Page 22 of the Invacare 10 user manual — DETAIL “B”
Page 22 · cliquer pour agrandir
Page 22

DETAIL “A”

Plate Groove (Insert Flat Head Screwdriver Here)

DETAIL “B”

DETAIL “C”

Invacare 10 — DETAIL “B” diagram (page 22)
Fig. 6

FIGURE 5.3 Attaching the Humidifier Bottle Adapter

6. Attach oxygen tubing from the humidifier bottle to oxygen outlet connector on the oxygen concentrator (FIGURE 5.4).
7. Attach the cannula/patient supply tubing to the humidifier bottle outlet (FIGURE 5.4).
8. After assembly, ensure that oxygen is flowing through the cannula.
Show 7 more labels / short text on this page
Filter Access Door · Humidifier Bottle · Adapter · Filter Access · Door · Humidifier Bottle · Adapter
Page 23 of the Invacare 10 user manual — Power Switch
Page 23 · cliquer pour agrandir
Page 23
Invacare 10 — Power Switch diagram (page 23)
Fig. 7 · Power Switch

FIGURE 5.4 Attaching the Oxygen Tubing

WARNING For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.

Power Switch

1. Press power switch to on position. All the panel lights and the audible alarm will come on for one second, indicating that the unit is functioning properly. After one second, only the GREEN system ok/power light will stay on.
Invacare 10 — Power Switch diagram (page 23)
Fig. 8 · Power Switch

On/Off (I/O) Power Switch

FIGURE 5.5 Power Switch

Show 5 more labels / short text on this page
Oxygen Outlet · Oxygen · Connector · Tubing · Humidifier Outlet
Page 24 of the Invacare 10 user manual — Flowrate
Page 24 · cliquer pour agrandir
Page 24

Flowrate

1. Turn the flowrate knob to the setting prescribed by your physician or therapist.

WARNING DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.

2. If the flowrate on the flowmeter ever falls below 0.5 L/min. for Platinum 5 (1 L/min. for Platinum 10) for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than 0.5 L/min. for Platinum 5 (1 L/min. for Platinum 10), the LOW FLOW audible alarm will go off.

Invacare 10 — Flowrate diagram (page 24)
Fig. 9 · Flowrate

FIGURE 5.6 Flowrate

Flow Knob · Ball

Page 25 of the Invacare 10 user manual — Explanation of Oxygen Purity Indicator Lights
Page 25 · cliquer pour agrandir
Page 25

SensO2 Oxygen Purity Indicator (If Your Unit Features the O2 Sensor Feature)

This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.

Initial Startup of the Concentrator

When the unit is turned on, the GREEN light will come on (System OK/O2 greater than 85%). After five minutes, the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values.

Explanation of Oxygen Purity Indicator Lights

GREEN light (O2) - Normal Operation.

YELLOW light ( ) - Call supplier IMMEDIATELY. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that backup oxygen is nearby.

RED light ( )- Total unit shutdown. Switch IMMEDIATELY to backup oxygen supply. Call supplier IMMEDIATELY.

GREEN light - with YELLOW light flashing - Call supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may continue to use the concentrator.

Page 26 of the Invacare 10 user manual — I/O
Page 26 · cliquer pour agrandir
Page 26

O 2

FIGURE 5.7 SensO2 Oxygen Purity Indicator (If Your Unit

Features the O2 Sensor Feature)

Indicators (If Your Unit Does Not Feature the O 2 Sensor)

LX models indicator Light Explanation

RED light ( ) - Total Unit Shut-Down. Switch IMMEDIATELY to a back-up oxygen supply. Call supplier IMMEDIATELY.

GREEN light ( )- On/Off. System okay.

I/O

FIGURE 5.8 Indicators (If Your Unit Does Not Feature the O2

Show 14 more labels / short text on this page
· · · · · · · · · · · · · Sensor)
Page 27 of the Invacare 10 user manual — Cleaning the Cabinet Filter
Page 27 · cliquer pour agrandir
Page 27

SECTION 6 —MAINTENANCE

SECTION 6 — MAINTENANCE

Routine Maintenance

Cleaning the Cabinet Filter

1. Remove each filter and clean at least once a week depending on environmental conditions.
2. Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
3. Dry the filters thoroughly before reinstallation.
Page 28 of the Invacare 10 user manual — Humidifier
Page 28 · cliquer pour agrandir
Page 28

SECTION 6 —MAINTENANCE

Cabinet Filter (located on both

Invacare 10 — SECTION 6 —MAINTENANCE diagram (page 28)
Fig. 10

FIGURE 6.1 Cleaning the Cabinet Filter

Cabinet

WARNING Unplug the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet.

1. Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.

Humidifier

1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten (10) parts water and one (1) part vinegar.
3. Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.

sides of concentrator)

Page 29 of the Invacare 10 user manual — AS REQUIRED
Page 29 · cliquer pour agrandir
Page 29

SECTION 6 —MAINTENANCE

WARNING DO NOT overfill humidifier bottle.

MODEL NO. IRC SERIAL NO.

ANNUALLY, DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS

INVACARE PLATINUM OXYGEN CONCENTRATOR

PREVENTIVE MAINTENANCE RECORD

EVERY 180 DAYS FOR IRC5LX (EVERY 90 DAYS FOR IRC10LX)

*Refer to Preventive Maintenance section of Service Manual.

Record Elapsed Hours On Hour Meter

Check Prescribed L/min. Flow Rate

ON EACH INSPECTION

(UNITS WITHOUT SensO2)

Check Compressor Inlet Filter*

Rebuild Top End of Compressor

Check Outlet HEPA Filter*

Check Power Loss Alarm

Record Date of Service

(SensO2 MODELS)

AS REQUIRED

FIGURE 6.2 Preventive Maintenance Record

Show 6 more labels / short text on this page
Check Oxygen Concentration · Clean/Replace Cabinet Filter(s) · Check Oxygen Concentration · Clean Cabinet Filter(s) · Replace Exhaust Muffler · Clean Heat Exchanger
Page 30 of the Invacare 10 user manual — SECTION 7 —TROUBLESHOOTING GUIDE
Page 30 · cliquer pour agrandir
Page 30

SECTION 7 —TROUBLESHOOTING GUIDE

SECTION 7 — TROUBLESHOOTING GUIDE

SYMPTOM: PROBABLE CAUSE: SOLUTION:

Alarm: Main Power Loss: 1. Insert plug into outlet.

Short beeps, long pause

1. Power cord NOT plugged in.
2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet.

Concentrator not operating, power switch On.

2. No power at outlet. 3. Call Service Provider.

Beep Beep 3. Tripped circuit breaker.

AlarmContinuous

System Failure: 1. Unit overheating due to blocked air intake.

1a. Remove and clean cabinet filters.

Concentrator not operating, Power Switch On.

1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture.

2. Insufficient power at outlet.

Beeeeeeeep

3. Internal repairs required
2. DO NOT use extension cords. Move to another electrical outlet or circuit.
3. Call Service Provider.
Page 31 of the Invacare 10 user manual — SECTION 7 —TROUBLESHOOTING GUIDE
Page 31 · cliquer pour agrandir
Page 31

SECTION 7 —TROUBLESHOOTING GUIDE

SYMPTOM: PROBABLE CAUSE: SOLUTION:

YELLOW or RED light Illuminated.

1. Low oxygen purity.*
1. Clean or Replace filters.
2. Kinked or blocked tubing, cannula or humidifier.*
2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON.

Alarm: Continuous On red light only.

3. Flowmeter set at 0.5 L/min. for Platinum 5 and 1 L/min. for Platinum 10 or less.*
3. Check flowmeter is set to 1.0 L/min. or more. (Refer to Recommended Guidelines for Optimum Performance on page 19.
4. Unit overheating due to blocked air intake.

* Only applies to 5LXO2 and 10LXO2 SensO2 models.

5. Insufficient power at outlet.

4.a. Remove and clean cabinet filters.

6. Internal repairs required.

4b. Move concentrator at least three inches from walls, draperies, and furniture.

5. DO NOT use extension cords. Move to another electrical outlet or circuit.
6. Call Service Provider.

GREEN light with YELLOW light flashing.

1. Internal repairs required.
1. Call Service Provider.
Page 32 of the Invacare 10 user manual — SECTION 7 —TROUBLESHOOTING GUIDE
Page 32 · cliquer pour agrandir
Page 32

SECTION 7 —TROUBLESHOOTING GUIDE

SYMPTOM: PROBABLE CAUSE: SOLUTION:

AlarmRAPID

LOW FLOW ALARM 1a. Kinked or blocked tubing, cannula or humidifier.

1a. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back On.

Beep Beep... Beep...Beep

1b. Flowmeter set at 0.5 L/min. for Platinum 5 and 1 L/min. for Platinum 10 or less.

1b. Reset flowmeter to prescribed flowrate. If prescribed flowrate is below 0.5 L/min., use the Pediatric Flowmeter IRCPF16 (Only with Platinum 5 and XL).

Page 33 of the Invacare 10 user manual — LIMITED WARRANTY
Page 33 · cliquer pour agrandir
Page 33

LIMITED WARRANTY

LIMITED WARRANTY

This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer. This warranty is not extended to any other person or entity and it is not transferable or assignable to any subsequent purchaser or owner. Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person. This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state.

Invacare Corporation warrants its Platinum XL Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period of five years from date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty.

Invacare Corporation warrants its Platinum 10 Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period of three years from date of purchase from Invacare or a dealer, with a copy of the seller’s invoice required for coverage under this warranty.

DO NOT OPEN OR ATTEMPT TO SERVICE THE 4-WAY VALVE BODY. This will void any and all warranty applicable to the valve body. If within such warranty periods any such product shall be proven to Invacare Corporation's satisfaction to be defective, such product shall be repaired or replaced, at Invacare Corporation's option. This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare Authorized Service Centers. It does not apply to the labor performed by the purchaser or user. This warranty does not include normal wear and tear or shipping charges. Invacare Corporation's sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement. Routine maintenance items, such as filters, are excluded from this warranty.

For warranty service, please contact Invacare Corporation's Service Department at the toll free number on the back cover. Upon receiving notice of an alleged defect in a product, Invacare Corporation will issue a serialized return authorization. It shall then be the responsibility of the purchaser to return the entire unit or remove, at purchaser's cost, the defective component part(s) identified, pack the component part(s) in a manner to avoid shipping

Page 34 of the Invacare 10 user manual — LIMITED WARRANTY
Page 34 · cliquer pour agrandir
Page 34

LIMITED WARRANTY

damage and to ship the component part(s) to either Invacare Corporation's plant or service center as specified by Invacare Corporation in advance. Defective component part(s) MUST be returned for warranty inspection using the serial number as identification within thirty (30) days of return authorization date. Do not return products to our factory without prior consent. C.O.D. shipments will be refused; please prepay shipping charges.

LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION'S CONTROL.

THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.

SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.

THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.

Dépannage

Dépannage

Problèmes fréquents, causes probables et solutions.

ProblèmeCause possibleSolution
Alarm:Main Power Loss:1. Insert plug into outlet. p.30
Short beeps, long pause1. Power cord NOT plugged in.2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet. p.30
Concentrator not operating, power switch On.2. No power at outlet.3. Call Service Provider. p.30
Beep......Beep......3. Tripped circuit breaker. p.30
Alarm: Continuous Concentrator not operating, Power Switch On. Beeeeeeeep....System Failure: 1. Unit overheating due to blocked air intake. 2. Insufficient power at outlet. 3. Internal repairs required1a. Remove and clean cabinet filters. 1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture. 2. DO NOT use extension cords. Move to another electrical outlet or circuit. 3. Call Service Provider. p.30
YELLOW or RED light Illuminated. Alarm: Continuous On red light only. * Only applies to 5LXO and 2 10LXO SensO 2 2 models.1. Low oxygen purity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min. for Platinum 5 and 1 L/min. for Platinum 10 or less.* 4. Unit overheating due to blocked air intake. 5. Insufficient power at outlet. 6. Internal repairs required.1. Clean or Replace filters. 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON. 3. Check flowmeter is set to 1.0 L/min. or more. (Refer to Recommended Guidelines for Optimum Performance on page 19. 4.a. Remove and clean cabinet filters. 4b. Move concentrator at least three inches from walls, draperies, and furniture. 5. DO NOT use extension cords. Move to another electrical outlet or circuit. 6. Call Service Provider. p.31
GREEN light with YELLOW light flashing.1. Internal repairs required.1. Call Service Provider. p.31
Alarm: RAPID Beep....Beep... Beep...BeepLOW FLOW ALARM 1a. Kinked or blocked tubing, cannula or humidifier. 1b. Flowmeter set at 0.5 L/min. for Platinum 5 and 1 L/min. for Platinum 10 or less.1a. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back On. 1b. Reset flowmeter to prescribed flowrate. If prescribed flowrate is below 0.5 L/min., use the Pediatric Flowmeter IRCPF16 (Only with Platinum 5 and XL). p.32

Poser une question sur : Invacare 10

Les utilisateurs et notre équipe répondent généralement sous 24 h — et votre question servira aussi aux suivants.

Add your email to get notified when it’s answered — no spam, ever.

0 questions