Installation Manual 20002514, Rev BA December 2010 Micro Motion ® R-Series Coriolis Flow Sensors.
Safety and approval information This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EC declaration of conformity for directives that apply to this product.
Contents Chapter 1 Planning ...........................................................................................................................1 1.1 Installation checklist .......................................................................
Contents ii Micro Motion R-Series.
1 Planning Topics covered in this chapter: • Installation checklist • Best practices • Environmental limits 1.1 Installation checklist □ Make sure that the hazardous area specified on the sensor approval tag is suitable for the environment in which the sensor is installed.
Recommended sensor orientation Table 1-3: Liquids Gases Slurries □ Install the sensor so that the flow direction arrow on the sensor matches the actual forward flow of the process. 1.2 Best practices The following information can help you get the most from your sensor.
Environmental limits for R-Series sensors Figure 1-1: Maximum process temperature in °F (°C) 176 (80) 140 (60) 104 (40) 68 (20) 32 (0) –4 (–20) –40 (–40) –148 (–100) 158 (70) Ambient tem.
2 Mounting 2.1 Mount the sensor Use your common practices to minimize torque and bending load on process connections. Tip To reduce the risk of condensation or excessive moisture, the conduit opening should not point upward (if possible). The conduit opening of the core processor can be rotated freely to facilitate wiring.
3 Wiring Topics covered in this chapter: • Options for wiring • Connect 4-wire cable 3.1 Options for wiring The wiring procedure you follow depends on which electronics option you have. See Table 3-1 for the wiring options for each sensor electronics option.
Step 1: Cable preparation Cable layout Run conduit to sensor Metal conduit Wrap the drain wires twice around the shield and cut off the excess drain wires.
Step 2: Shield termination Assemble the Gland 1. Fold the shield or braid back over the clamping insert and 1/8 inch (3 mm) past the O-ring. 2. Install the gland body into the conduit opening on the core processor housing. 3. Insert the wires through gland body and tighten the gland nut onto the gland body.
Step 3: Terminal connections Connect the wires to the transmitter terminals (see the transmitter manual) Reinstall and tighten the core processor cover From Step 1 or 2 Connect the wires to the core p.
4 Grounding The sensor must be grounded according to the standards that are applicable at the site. The customer is responsible for knowing and complying with all applicable standards.
*20002514* 20002514 Rev BA 2010 Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459 www.micromotion.com Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 www.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Emerson 20002514 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Emerson 20002514 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Emerson 20002514, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Emerson 20002514 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Emerson 20002514, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Emerson 20002514.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Emerson 20002514. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Emerson 20002514 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.