Manuel d'utilisation / d'entretien du produit N1022A du fabricant Agilent Technologies
Aller à la page of 24
Agilent N1022A Active Probe Adapter User Guide.
ii Notices © Agilen t T echnolo gies, Inc. 2003 No pa rt of t his man ual ma y be r epro- duced in any form or by any means (including electronic stor age and retrieval or translation into a foreign lan- guage ) witho ut pri or agre emen t and wr it- ten consent from Ag ilent T echnologies, Inc.
Contents Conte nts 1 N1022A Acti ve Probe Adapt er 1 General Saf ety Consid erati ons 2 Instru ment Mark ings 3 Using the Prob e Adapter 4 Inspecti ng the N1022A 4 Conne cting t he N1022 A 5 Static- S.
Conte nts 2 Conte nts.
1 N1022A A ctive Prob e Adapt er N1022 A Active Pr obe Adapter The N1 022A Acti ve Prob e Adapte r provid es the i nterface to conne ct an Agilent Infiniimax Active Probe to an Agil ent 86100 -series In finiium DCA. T o use the N10 22A re quires th at firmware r evision A.
2 Gener al Sa fety Cons idera tio ns General Safety Considerat ions This produ ct has be en desi gned and tested in accord ance with the stand ards listed in the Ma nufactu rer’ s De clara tion of Co nform ity , and has been suppli ed in a safe c ondit ion.
3 Inst rume nt Ma rki ngs Instrument Markin gs The ins tructi on ma nual symb ol. The prod uct is ma rked with t his warni ng s ymbol when it is n ecess ary for th e us er t o re fer t o th e instr uctions in the ma nual. The las er radiati on symbol.
4 Using the Probe Adapter Using the Probe Adapter There are a few things to consid er when connecti ng the N 1022A Active Prob e Adapter to an active p robe an d the I nfiniium DCA. • The dynami c range of th e system wi ll be 3.2 V (6.4 Vp- p), which wit h probe o ff- set, c over s m ost d igit al te chn ologi es.
5 Conn ectin g the N1022A Tech nologies sales office . • If the ship ping cont ainer is damaged, o r the cus hioni ng materi als sho w signs of st ress, no tify the c arrier a s well as yo ur Agilen t Techno logies sal es offic e. Keep the shi pping material s for the c arrier' s inspe ction.
6 Connec ting t he N1022A panel. Connecti ng to a n Aux Pow er connect or prev ents the p robe fr om being automatically calibra ted. 3 Connect th e pro be ou tput to t he pr obe adap ter inpu t. T he probe tip should not be connect ed to a device under test at this ti me.
7 Conn ectin g the N1022A Figure 1. Probe C hannel Dial og Box 6 Click Close . You can no w change the pr obe tips or probe atte nuation. You can also perfo rm a probe calibrat ion. You can access inform ation a bout probe atte nuation a nd probe calibra tion in t he Infi niium DCA He lp syste m.
8 Stati c-Safe Work station Static-Safe Workst ation Elect rostati c discharge (ESD) can damage o r destroy el ectronic c omponents . All w ork on e lectroni c assemb lies sho uld be per formed at a stat ic-safe wo rk- statio n.
9 Reduci ng ESD Damage Both ty pes, when us ed tog ethe r , provi de a si gnific ant le vel of ESD pr otecti on. Of the two, on ly the ta ble-ma t and w rist -stra p combina tio n provide s ad equat e ESD p rotection w hen used alone.
10 Specifica tions and R egulatory Info rmation Specifications and Regulatory Information This sect ion l ists sp ecif icati ons an d ch aract eristi cs a nd regu lat ory in forma tion of t he prob e adapte r .
11 Specif ications Storage Humidity ( type t ested) 90% R.H . (non-con densing ) at +65 ° C (+149 ° F), 24 hr s Shock (typ e tested) 30 g (half sin e), 11 msec pulse Vibrati on (typ e test ed) Random, 0.21g rms, 5 to 50 0 Hz, 1 0 minu tes per axi s Sine, 0.
12 Specif ications *Other Infiniium instruments in com pliance with the Autoprobe interface can be added using t he Prob e dialo g box locate d in th e Infiniium grap hical u ser in terface . Figure 3. N1022A I/O Connections Connector I/O (see fi gure belo w) DCA Co nnector Probe Co nnector Power Connector Custom rugge dized 3.
13 Regula tory Information Regulator y Information Electro static Dis charge Immu nity - IEC 6100 0-4-2: 199 5 pass es criter ion B. According to IEC 61000-4-2: 1995, system measu rement may be affect.
14 Declara tion of Conform ity Declaration of Co nformity.
15 Service Service Service for the N1022A A ctive Prob e Adapte r is by replace ment only . T o order a new N 1022A, co ntact your local Agilent T e chnologies Sales Office. Re turn the f ailed pr obe adap ter to your lo cal Agilen t servic e cente r .
16 Returni ng the N1022A Returning the N102 2A When t he probe ad apter is returne d to an Agil ent ser vice offi ce for serv icing, i t must be adequat ely packag ed and ha ve a comp lete de scrip tion of the failur e sympto ms attached . Pleas e notify the ser vice offi ce before retur ning your pr obe adapt er for ser- vic e.
17 Agilen t T e chnologies Service Offic es CAUTIO N Cover e lectrica l connec tors to protect sens itive co mponents from elec trost atic damage. CAUTIO N Use origin al packagi ng or comparab le. Instru ment damage ca n result fro m using packaging materials oth er than the or iginal materials.
18 Agil ent Tech nol ogies Se rvi ce Offic es.
Index Inde x 1 A Accessories , 4 Agilent o ffices , 17 C CE mark , 3 cleaning , 2 , 4 , 17 CSA mark , 3 D Declaration of Conformity , 14 E EMC requirem ents , 13 environ menta l specifica tions , 10 E.
Index 2 Index.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Agilent Technologies N1022A c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Agilent Technologies N1022A - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Agilent Technologies N1022A, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Agilent Technologies N1022A va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Agilent Technologies N1022A, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Agilent Technologies N1022A.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Agilent Technologies N1022A. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Agilent Technologies N1022A ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.