Manuel d'utilisation / d'entretien du produit T360U du fabricant Teledyne
Aller à la page of 21
Manual Addendum Model T360U CO 2 Analyzer (Addendum to T360 Manual, PN 07272) © TELEDYNE ADVANCED POLLUT ION INSTRUMENTATION (TAPI) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.
.
i ABOUT THIS ADDENDUM This addendum, part number 07273, is to be used in conjunction with the Model T360 operation manual, part number 07272. W here operation of th e Model T360U diverges from that of the Model T360, this addendum takes precedence.
ii This page intentionally left blank.
iii TABLE OF CONTENTS About This Addendum ................................................................................................................ i Revision History .............................................................................
iv This page intentionally left blank..
5 1. OVER VIEW The Model T360U is a close derivative of the Model T360 CO Analyzer. The main differences are as follows: The optical bench is longer (14 m), making the instrument more sensitive at low CO 2 levels. The instrument has a user-selectable full scale range of 0-100 PPB to 0-100 PPM of CO 2 .
6 PARAMETER SPECIFICATION AC Power 100 – 120V 50/60 Hz (120W) 220 – 240 V 50/60 Hz (144W) Environmental Conditions Installation Category (Over voltage Category) II Pollution Degree 2 Standard I/O .
7 4. PNEUMA TIC CONNECTIONS This section provides information on the pneumatic configur ations. Figure 4-1 shows the basic pneumatic flow of the T360U.
8 4.1. B ASIC AND Z ERO /S PAN V ALVE O PTION C ONFIGURATIONS CAUTION Sample and calibration gases should only come into contac t with PTFE (Teflon), FEP, glass, stainless steel or brass. Do NOT operate without first removing dust plugs from rear panel fixtures.
9 Ne edle val ve , s e t f l ow t o 8 00 cc /m i n (+/- 10 % ) C hassis Sa m ple Ze ro Ex hau st Ve n t S p a n Pre s sur e Sp an To P u rg e Sa mpl e gas Fr o m P u r g e Fl o w m e te r Pres sur e r.
10 4.2. M AKING THE P NEUMATIC C ONNECTIONS CAUTION Venting should be outsi de the shelter or immediate area s urrounding the instrument. 1. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long.
11 5. RELA Y ALARM OUTPUTS (ST ANDARD CONFIGURA TION) There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2 exceeded.
12 This page intentionally left blank..
T360U SPARE PARTS LIST Ref PN 07460, 05 Apr 2013, 14:24 PARTNUMBER DESCRIPTION 037600000 AKIT, EXP KIT, SODA LIME CH47 037860000 ORING, TEFLON, RETAINING RING, 47MM (KB) 039060100 ASSY, PRESSURE TRANS.
T360U SPARE PARTS LIST Ref PN 07460, 05 Apr 2013, 14:24 009450300 ASSY, ZERO/SPAN VALVES, CO 006110200 ASSY, MOTOR WHEEL HEATER 003291500 ASSY, THERMISTOR, BENCH/WHEEL 003290500 ASSY, THERMISTOR, BENCH 001761300 ASSY, FLOW CTRL, .010, 1/8", SS 001760400 ASSY, FLOW CTL, 800CC, 1/4" CONN-B 000941000 CD, ORIFICE, .
Interconnect List, T360U Reference: 073640100A Cable PN Signal Assembly PN J/P Pin Assembly PN J/P Pin 03995 CBL, MOTOR TO RELAY PCA GFC Drive - A Relay PCA 041350000 J6 1 GFC Motor 052380200 P1 1 GFC.
Interconnect List, T360U Reference: 073640100A Cable PN Signal Assembly PN J/P Pin Assembly PN J/P Pin FROM TO 06746 CBL, MB TO 06154 CPU GND Motherboard 058021100 P12 2 Shield RX0 Motherboard 0580211.
Interconnect List, T360U Reference: 073640100A Cable PN Signal Assembly PN J/P Pin Assembly PN J/P Pin FROM TO 06814 CBL ASSY, SYNC DEMOD , DC POWER DGND Relay PCA 041350000 J15 1 Sync Demod 032960000.
This page intentionally left blank. 18 07273A DCN6707.
07273A DCN6707 19.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Teledyne T360U c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Teledyne T360U - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Teledyne T360U, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Teledyne T360U va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Teledyne T360U, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Teledyne T360U.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Teledyne T360U. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Teledyne T360U ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.