OPERATION MANUAL MODEL 701 ZERO AIR GENERATOR TELEDYNE ADVANCED POLLUTION INSTRUMENTATION DIVISION (TAPI) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledyne.com WE B S IT E: www.
.
01671H (DCN6051) i ABOUT THIS MANUAL This M701 Zero Air Generator operation manual is comprised of multiple documents as listed below. Part No. Rev Name/Description 01671 H Operation Manual 02052 M List, Spare Parts (in Appendix A) with Sales for most recent version.
ii 01671H (DCN6051) REVISION HISTORY M701 Manual 01671 2011 April 8, Rev H, per DCN 6051 Document P/N Rev DCN Change Summary Operation Manual 01671 H 6051 Correct name in title: from “…Module.
01671H (DCN6051) iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these mess ages carefully. Each safety message is associated with a safety alert symbol, and are placed throughout this manual and inside the instrument.
iv 01671H (DCN6051) CONSIGNES DE SÉCURITÉ Des consignes de sécurité import antes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instrumen ts. Veuillez lire attentivement ces consignes.
01671H (DCN6051) v TABLE OF CONTENTS TABLE OF CONTENTS .................................................................................. V LIST OF FIGURES ........................................................................................ VII LIST OF TABLES .
vi 01671H (DCN6051) 5.11 P RESSURE R EGULATOR ............................................................................................ 5-8 5.12 P RESSURE G AUGE ......................................................................................
01671H (DCN6051) vii LIST OF FIGURES F IGURE 3-1: D IMENSIONS ......................................................................................................................... 3-2 F IGURE 3-2: F RONT P ANEL ....................................
viii 01671H (DCN6051) This page intentionally left blank..
01671H (DCN6051) 1-1 1 INTRODUCTION 1.1 Preface Teledyne API is pleased that you have pur chased the Model 701 (or M701) Zero Air System. The M701 is an excellent source of clean, dry air for dilution calibrators. It also may be used as a source of purge air for permeation tube ovens or burner air for FID analyzers.
INTRODUCTION Teledyne API Model 701 Zero Air Generator 1-2 01671H (DCN6051) 1.2 Using This Manual This manual has the following data structures: TABLE OF CONTENTS Outlines the contents of the manual in the or der the information is presented. This is a good overview of the topics covered in the manual .
01671H (DCN6051) 2-1 2 SPECIFICATIONS AND WARRANTY 2.1 Specifications Table 2-1. Specifications PARAMETER SPECIFICATION Output STD 20 SLPM @ 30 psig Max delivery pressure 35 psig (Except in some specials that have restricted output) Dewpoint < 15 SLPM 15 SLPM < -20 C < -10 C SO 2 <0.
SPECIFICATIONS and WARRANTY Teledyne API Model 701 Zero Air Generator 2-2 01671H (DCN6051) 2.2 Warranty WARRANTY POLICY (02024D) ( DCN 4473) Prior to shipment, Teledyne API equipment is thor oughly inspected and tested. Should equipment failure occur, Teledyne API assures it s customers that prompt serv ice and support will be available.
01671H (DCN6051) 3-1 3 GETTING STARTED 3.1 Unpacking CAUTION To avoid personal injury, always use two persons to lift and carry the Model 701 1. Verify that there is no shipping damage. If there are signs of damage, immediately advise the shipper, then Teledyne API.
GETTING STARTED Teledyne API Model 701 Zero Air Generator 3-2 01671H (DCN6051) Figure 3-1: Dimensions.
Teledyne API Model M701 Zero Air Generator GETTING STARTED 01671H (DCN6051) 3-3 3.3 Front and Rear Panels Figure 3-2 and Figure 3-3 illustrate the M odel 701 front and rear panels.
GETTING STARTED Teledyne API Model 701 Zero Air Generator 3-4 01671H (DCN6051) 3.4 Electrical And Pneumatic Connections All connections, electric and pneumatic , are made at the rear panel (Figure 3-3).
Teledyne API Model M701 Zero Air Generator GETTING STARTED 01671H (DCN6051) 3-5 3.5 Power-Up 1. Turn on the front panel POWER SWITCH. See Figure 3-2. 2. The front panel power indicator should light. 3. The cooling fan should start immediately. 4. The compressor should start after a few seconds delay.
GETTING STARTED Teledyne API Model 701 Zero Air Generator 3-6 01671H (DCN6051) This page intentionally left blank..
01671H (DCN6051) 4-1 4 OPTIONAL HARDWARE AND SOFTWARE This section includes descriptions of the har dware and software options available for the M701 instrument.
OPTIONAL HARDWARE and SOFTWARE Teledyne API Model 701 Zero Air Generator 4-2 01671H (DCN6051) CAUTION THE AIR EXITING TH E SCRUBBER IS AT 300 C (575 F). THIS HOT AIR ENTERS THE COPPER TUBING WHERE IT IS COOLED BEFORE IT ENTERS THE NO SCRUBBER.
Teledyne API Model M701 Zero Air Generator OPTIONAL HARDWARE and SOFTWARE 01671H (DCN6051) 4-3 4.4 Rack Mount Kits (OPT 20A, OPT 20B & OPT 21) There are several options for mounting the analyzer in standard 19” racks.
OPTIONAL HARDWARE and SOFTWARE Teledyne API Model 701 Zero Air Generator 4-4 01671H (DCN6051) This page intentionally left blank..
01671H (DCN6051) 5-1 5 OPERATION 5.1 Overview Refer to Figure 5-1 , Figure 5-2 , and Figure 5-3 for the following: The M701 dries and scrubs ambient air to produce zero air. The compressor draws air in from the rear panel bulkhead union and inlet filter.
OPERATION Teledyne API Model 701 Zero Air Generator 5-2 01671H (DCN6051) When air usage is high (say 5-20 LPM), the compressor may run continuously. When air usage is low, the pressure switch tu rns the compressor off until the storage tank pressure drops to 35 psig, and then turns the compressor on again.
Teledyne API Model M701 Zero Air Generator OPERATION 01671H (DCN6051) 5-3 Figure 5-1: M701 Pneumatic Diagram with HC/ CO and Dew Point Sensor Options.
OPERATION Teledyne API Model 701 Zero Air Generator 5-4 01671H (DCN6051) Figure 5-2: M701 Pneumatic Diagram with CO and Dew Point Sensor Options.
Teledyne API Model M701 Zero Air Generator OPERATION 01671H (DCN6051) 5-5 Figure 5-3: Internal Chassis Layout, Plan View.
OPERATION Teledyne API Model 701 Zero Air Generator 5-6 01671H (DCN6051) 5.2 Compressor T he compressor is a single-cylinder oscillating piston type driven by a split capacitor AC motor. The compressor is dry; that is, ther e are no lubricants which can contaminate the compressed air.
Teledyne API Model M701 Zero Air Generator OPERATION 01671H (DCN6051) 5-7 5.6 Pressure Relief Valve The pressure relief valve is a safety device designed to limit the maximum pressure to which the M701 can be subjected. It is set to open at 90 psig. This can be quite loud when the valve opens and sounds like steam e scaping.
OPERATION Teledyne API Model 701 Zero Air Generator 5-8 01671H (DCN6051) 5.9 Storage Tank The storage tank serves two functions. As its name implies, it stores air so that when the demand is low, the compressor can be turned off to conserve energy, allowing the storage tank to supply the air requirements.
Teledyne API Model M701 Zero Air Generator OPERATION 01671H (DCN6051) 5-9 5.12 Pressure Gauge The front panel-mounted pressure gauge s hows approximately the regulated air pressure available to a calibrator. The gauge will usually indicate 30 psig. If the flow is very low, as in a standby condition, the gauge may read a little higher, say 33 psig.
OPERATION Teledyne API Model 701 Zero Air Generator 5-10 01671H (DCN6051) 5.14 CO Scrubber (option) The CO scrubber catalytically oxidizes CO to CO 2 . The catalyst is proprietary and operates at room temperature; howev er, it is heated to approximately 70 C to prevent water condensation.
Teledyne API Model M701 Zero Air Generator OPERATION 01671H (DCN6051) 5-11 The M701 Control PCB is microcontroller based, and all timing is derived from the AC line. It automatically senses the line fr equency (50/60 Hz) and provides the following functions: 1.
OPERATION Teledyne API Model 701 Zero Air Generator 5-12 01671H (DCN6051) Figure 5-4: Control Board Layout.
01671H (DCN6051) 6-1 6 MAINTENANCE 6.1 Schedule NOTE The operations outlined in this section are to be performed by qualified maintenance personnel only. Before performing any maintenance, cycle the power on/off and drain any water that may be present (refer to instructions for “ Shut Down, Storage and Transportation” in Section 3.
MAINTENANCE Teledyne API Model 701 Zero Air Generator 6-2 01671H (DCN6051) 3. Remove any dirt or dust with a vacuum cleaner. Do not use an air jet. This will only redistribute the dirt and will not remove it. WARNING There are high voltages present while the M701 is plugged in 6.
Teledyne API Model 701 Zero Air Generator MAINTENANCE 01671H (DCN6051) 6-3 6.4 Replacing The Charcoal Scrubber 1. Turn off the M701 and wait for the pressure to go to zero. 2. Open the front panel, or remove the top cover. 3. Remove the 1/4” tubing connected to the top of the scrubber canister.
MAINTENANCE Teledyne API Model 701 Zero Air Generator 6-4 01671H (DCN6051) 6.6 Replacing The CO - CO 2 Scrubber 1. Turn off power and unplug the M701. 2. Disconnect the CO-CO 2 scrubber at the Controller. 3. Lift out the scrubber by removing the two screws under the chassis and the two pneumatic fittings.
Teledyne API Model 701 Zero Air Generator MAINTENANCE 01671H (DCN6051) 6-5 6.7 Replacing The HC Scrubber 1. Turn off power and unplug the M701 power cord. 2. Disconnect the HC scrubber pow er at the Control PCA. 3. Remove the heat shield from the HC scrubber.
MAINTENANCE Teledyne API Model 701 Zero Air Generator 6-6 01671H (DCN6051) 6.8 Servicing The Regenerative Dryer It is unlikely that the Regenerative Drye r should ever need service. Should the molecular sieve become contaminated by oil or other external contaminants, the dryer will need to be cleaned and recharged.
Teledyne API Model 701 Zero Air Generator MAINTENANCE 01671H (DCN6051) 6-7 18. Place the upper manifold on top of the tw o columns. Be sure that the columns are really in the grooves. (Wiggle the columns sideways. You should feel that the columns are trapped in the grooves.
MAINTENANCE Teledyne API Model 701 Zero Air Generator 6-8 01671H (DCN6051) Figure 6-1: Regenerative Dryer.
Teledyne API Model 701 Zero Air Generator MAINTENANCE 01671H (DCN6051) 6-9 6.9 Changing The 4-Way Valve (in the Regenerative Dryer) 1. Remove the Regenerative Dryer from the chassis as described in 5.9 above. 2. Disconnect the two tubing connections and t he electrical connection to the 4-way valve.
MAINTENANCE Teledyne API Model 701 Zero Air Generator 6-10 01671H (DCN6051) NOTE This procedure checks the majority of the components and fittings. If you believe that there is still a leak, please contact Customer Service at (858) 657-9800.
01671H (DCN6051) 7-1 7 TROUBLESHOOTING This section presents possible problems and their causes and solutions. Problem Probable Cause Corrective Action Unplugged power cord. Plug in power cord at both ends. Incorrect line voltage. Refer to Table 3-1 for the correct line voltage.
TROUBLESHOOTING Teledyne API Model 701 Zero Air Generator 7-2 01671H (DCN6051) Problem Probable Cause Corrective Action Screws securing the compressor to the base plate are loose. Tighten the screws. Instrument vibrates excessively. (continued) The tubing attached to the compressor is vibrating against a component.
01671H (DCN6051) 8-1 8 A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne API considers the prevention of damage caused by the discharge of static electricity to be extremely important part of making sure that your analyzer continues to provide reliable service for a long time.
A Primer on Electro-Static Discharge Teledyne API Model 701 Zero Air Generator 8-2 01671H (DCN6051) Table 8-1. Static Generation Voltages for Typi cal Activities MEA NS OF GENERATION 65-90% RH 10-25% .
Teledyne API Model 701 Zero Air Generator A Primer on Electro-Static Discharge 01671H (DCN6051) 8-3 Potentially damaging electro-stat ic discharges can occur: Any time a charged surface (including the human body) discharges to a device.
A Primer on Electro-Static Discharge Teledyne API Model 701 Zero Air Generator 8-4 01671H (DCN6051) A charge can be induced onto the conductive surface and/or discharge triggered in the presence of a charged field such as a large static charge clinging to the surface of a nylon jacket of someone walking up to a workbench.
Teledyne API Model 701 Zero Air Generator A Primer on Electro-Static Discharge 01671H (DCN6051) 8-5 Simply touching a grounded piece of metal is insufficient . While this may temporarily bleed off static charges present at the time, once you stop touching the grounded metal new static charges will immediately begin to re- build.
A Primer on Electro-Static Discharge Teledyne API Model 701 Zero Air Generator 8-6 01671H (DCN6051) Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of the instrument chassis. This will safely c onnect you to the same gr ound level to which the instrument and all of it s components are connected.
Teledyne API Model 701 Zero Air Generator A Primer on Electro-Static Discharge 01671H (DCN6051) 8-7 2. Never carry the component or assembly wit hout placing it in an anti-ESD bag or bin.
A Primer on Electro-Static Discharge Teledyne API Model 701 Zero Air Generator 8-8 01671H (DCN6051) 8.5.5 Packing Components for Return to Teledy ne API’s Customer Service Always pack electronic components and assemb lies to be sent to Teledyne API’s Customer Service in anti-ESD bins, tubes or bags.
01671H (DCN6051) A-1 APPENDIX A - SPARE PARTS Note Use of replacement parts other than those supplied by TAPI may result in non- compliance w ith European standard EN 61010-1. Note Due to the dynamic nature of part numbers, please refer to the Website at ht tp :/ /www.
A-2 01671H (DCN6051) This page intentionally left blank..
SPARE PARTS LIST M701H Ref: 04825F (DCN5895) Part Numbe r Description 005960000 AKIT, EXPEND, 6LBS ACT CHARCOAL 005970000 AKIT, EXPENDABLE, 6LB PURAFIL 006900100 RETAINER PAD CHARCOAL, LARGE, 2-1/4&qu.
Reference 07346A (DCN5895) 1 2-5 6-10 11-20 21-30 017960000 ASSY, FAN M701/M701H 1 2 4 8 066180000 PCA, CONTROL BOARD, M701H 1 2 4 036260000 ASSY, 4-WAY VALVE, PARKER, M701 1 2 4 8 066780000 * ASSY, PUMP, W/BASE, 100-115V, GAST 1 2 FL0000044 FILTER, COALESCING, .
Reference 07344A (DCN5895) 005960000 AKIT, EXP, 6LBS ACT CHARCOAL 1 005970000 AKIT, EXP, 6LBS PURIFIL 1 006900100 RETAINER PAD CHARCOAL, LARGE, 2-1/4" 8 016920000 AKIT, EXP KIT, MOL SIEVE 1 05727.
LEVEL 1 SPARES KIT M701H Ref: 07345A (DCN5895) Part Numbe r Description 03626000A Valve, 4-way, regenerative scrubber 066180000 PCA, CONTROL BOARD, M701H CB0000004 Fuse, 3A, SB (HC temp control) FA000.
01671H (DCN6051) B-1 APPENDIX B - Schematics.
B-2 01671H (DCN6051) This page intentionally left blank..
1 1 2 2 3 3 4 4 D D C C B B A A Tit le Num ber Revi sion Si ze A Da te: 10 /1 1/2010 Sh eet of Fil e: N:P CBM GR.. 06619 -1B .Sch Doc Dr awn By: 701/701H Controller Boa rd (AC POWER) 06619 B 1 5 1 2 3.
1 1 2 2 3 3 4 4 D D C C B B A A Tit le Num ber Revi sion Si ze A Da te: 10 /1 1/2010 Sh eet of Fil e: N:P CBM GR. . 06619 -2B .Sch Doc Dr awn By: 701/701H Controller Boa rd (DC POWER) 06619 B 2 5 T P1.
1 1 2 2 3 3 4 4 D D C C B B A A Tit le Num ber Revi sion Si ze A Da te: 10 /1 1/2010 Sh eet of Fil e: N:P CBM GR.. 06619 -3B .Sch Doc Dr awn By: 701/701H Controller Boa rd (CPU) 06619 B 3 5 1 3 2 X1 4. 00 M Hz + C6 10 /35 1 2 3 4 5 6 Progr am Po rt J10 KBI NT S CL SDA 1 2 3 4 5 6 7 8 9 10 11 12 13 14 JP1 HEA DE R 7X2 S CL SDA TX RX RX TX R2 2.
1 1 2 2 3 3 4 4 D D C C B B A A Tit le Num ber Revi sion Si ze A Da te: 10 /1 1/2010 Sh eet of Fil e: N:P CBM GR. . 06619 -4B .Sch Doc Dr awn By: 701/701H Controller Boa rd (I/O) 06619 B 4 5 A0 1 A1 2.
1 1 2 2 3 3 4 4 D D C C B B A A Tit le Num ber Revi sion Si ze A Da te: 10 /1 1/2010 Sh eet of Fil e: N:P CBM GR.. 06619 -5B .Sch Doc Dr awn By: 701/701H Controller Boa rd (SENSORS) 06619 B 5 5 R32 6.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Teledyne 701 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Teledyne 701 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Teledyne 701, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Teledyne 701 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Teledyne 701, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Teledyne 701.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Teledyne 701. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Teledyne 701 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.