TELEDYNE BROWN EGINEERING Analytical Instruments P/N M59608 ECO# 95-416 7/10/95 M ODEL 319R O XYGEN D ETECTOR I NSTRUCTION M ANUAL.
TELEDYNE BROWN EGINEERING Analytical Instruments Copyright © 1993 Teledyne Brown Engineering Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcr.
TELEDYNE BROWN EGINEERING Analytical Instruments Table of Contents 1.0 Introduction Main Features ............................................................................................... 1 Micro-Fuel Cell .......................................
TELEDYNE BROWN EGINEERING Analytical Instruments.
1 319R Introduction 1.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Introduction The Teledyne Brown Engineering Analytical Instruments (TBE) Model 319R Oxygen Detector measures the percent of oxygen in a wide range of gases and gas mixtures.
1.0 Introduction 319R 2 TELEDYNE BROWN ENGINEERING Analytical Instruments 4. Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell, warranted for 6 months, with a life expectancy of 6 months. 5. Analog Voltage Output. The Model 319R provides customer connections to external devices, such as computers, recorders, alarms, etc.
3 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Installation & Operation Precautions 1. Read this chapter in its entirety before installing the unit. 2. Upon receipt of the instrument, inspect the entire unit for damage.
4 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments The analysis unit should be installed in the process at a maximum distance of 6 feet away from the main unit, depending upon the level of electromagnetic noise in the operating environment.
5 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Alarm Terminal Connection SPDT relay contacts are provided for each of the 3 alarms. There are 9 terminals available on the back panel of the system for alarm circuit connection.
6 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Setting Alarms The alarms can be adjusted over the full scale range, and they are set as a percent of full scale.
7 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Calibration To calibrate the 319R: 1. Place the power switch in the ON position. 2. Allow the LCD reading to stabilize. 3. Expose the sensor to ambient air (20.9%) or instrument grade air.
8 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments.
9 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Maintenance Replacing the Fuse If the unit will not turn on when plugged in and switched ON (via the power switch), or you suspect a blown fuse, replace the fuse using the following procedure: 1.
3.0 Maintenance 319R 10 TELEDYNE BROWN ENGINEERING Analytical Instruments 5. Replace the fuse holder into its receptacle, pushing in firmly until it clicks. In units with DC power, the fuse is located on the rear panel above the ON/OFF switch. Open the fuse holder by unscrewing and removing the knob marked FUSE.
11 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Warranty Conditions The Class Z-2 cell used in the 319R is warranted for six (6) months of service from the date of shipment. Customers having warranty claims must return the cell in question to the factory for evaluation.
3.0 Maintenance 319R 12 TELEDYNE BROWN ENGINEERING Analytical Instruments.
13 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Appendix Specifications Ranges : 0–200 PPM, 0–1999 PPM O 2 , and 0– 25% calibration range Sensitivity: 1% of full scale Signal Output : Standard: 0–10 VDC, negative ground 4-20 mA, negative ground plus contact closures for range identification.
14 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Spare Parts List QTY. P/N DESCRIPTION 1 C57304 PC board, main 1 A33748 Thermistor assembly 1 C6689-Z2 Micro-Fuel Cell, Z-2 4 F1189 Fuse, 1A (AC) 4 F51 Fuse, ½A (DC) 1 A59264 Probe to analyzer cable, 6 feet A minimum charge of US $20.
15 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Drawing List C-57490 Outline diagram A-58154 Probe interconnection diagram D-57302 Electronic PCB schematic diagram C-57305 Power sup.
16 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Material Safety Data Sheet Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19. Mini-Micro-Fuel Cells, all classes.
17 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Flash Point: na Flammable Limits: na LEL: na UEL: na Extinguishing Media: Use extinguishing media appropriate to surrounding fire conditions.
18 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Contact of electrolyte with skin or eyes will cause a burning sensation and/or feel soapy or slippery to touch. Other symptoms of exposure to lead include loss of sleep, loss of appetite, metallic taste and fatigue.
19 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Section VIII – Control Measures Eye Protection: Chemical splash goggles Hand Protection: Rubber gloves Other Protective Clothing: A.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Teledyne 319R c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Teledyne 319R - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Teledyne 319R, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Teledyne 319R va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Teledyne 319R, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Teledyne 319R.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Teledyne 319R. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Teledyne 319R ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.