INSTRUCTION MANUAL FOR TRACE OXYGEN ANALYZER MODEL 311-D RANGES: 0-100 PPM; 0-1000 PPM; 1% AND 100% TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CALIF. 91749 TELEPHONE: (626) 934-1500 FAX: (626) 961-2538 TWX: (910) 584-1887 TDYANYL COID EASYLINK: 628-31862 Web: www.
TABLE OF CONTENTS MODEL 311D 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1 1.3.1 Micro-Fuel Cell 1 1.3.2 Reliable Calibration 1 1.3.3 Integral Power Supply 2 1.3.4 Accuracy and Response 2 1.3.5 Compact Packaging 3 1.
1. INTRODUCTION 1.1 Description : The Teledyne Analytical Instruments (TAI) Model 311-D is a portable trace oxygen analyzer which can be opera ted without an external power source, and reliably calibrated without the use of cumbersome, questionable, “certified” calibration gases.
After the electronics have been properly zeroed (a one-time factory operation), the instrum ent cannot produce an output indication in the absen ce of oxygen; therefore, the need for a “zero” standardization gas is obviated.
Further disassembly may be accom plished by rem oving the backplate assemb ly from its four mounting standoffs, and laying the two separated assem blies out as illustrated on the “Analyzer Wiring Diagram”. The diagram is included among the drawings at the rear of the manual.
necessary or desirable, it must have a metallic diaphragm. Regulators with organic or plastic diaphragms are permeable to oxygen, and if used in the sampling system, will lead to high oxygen readings.
dictated by the customer’s application. If the instrument were being used to certify the oxygen content of a product for delivery, then a calibrati on prior to certification would certainly be in order.
If, on the other hand, the instrument is not to be used imm ediately after calibration, and a low ppm oxygen gas is being employed as a purge, allo w the m anifold to be purged overnight, and then disconnect both male fittings. ALWAYS DISCONNECT THE SOURCE FITTING FIRST, AND IMMEDIATELY THEREAFTER, THE VENT FITTING .
every two (2) days of continuous use. To rech arge the batteries, place the range switch in the “OFF” position and connect the power cord to a convenience outlet.
No tools are required to replace the cell in the in strument. Sim ply unscrew (ccw) the plug at the bottom of the analyzer, and the cell will drop out of the m anifold cavity. Remove the new cell from its package, and carefu lly rem ove the shorting clip.
1. Disconnect cell. 2. Move range switch to “CAL” position. 3. Adjust P1 such that the output of A3, pin 6, measures between 0 and +0.5 mV, ideally +-.3 mV. 4. Adjust P2 for 0 ± 1 mV at output of A2, pin 6. 5. Verify that the offset is the same on all ranges.
OPERATING TEMPERATURE RANGE: 30 ° TO 125 ° F. RESPONSE AND RECOVERY: AT THE SPECIFIED FLOWRATE (0.25 SCFH), THE SENSOR OUTPUT SHOULD BE, IN TEN SECONDS, WITHIN ± 10% OF AN INTRODUCED GAS HAVING A C.
EFFECT OF CO2 ON B-2C CELL LIFE -11-.
RECOMMENDED SPARE PARTS LIST MODEL 311D QTY. P/N DESCRIPTION 5 F-39 MICRO FUSE, 0.25 AMP., FAST-BLOW (FOR 210-240 VAC) 5 F-51 MICRO FUSE, 0.50 AMP., FAST-BLOW (FOR 100-125 VAC) 1 C-6689-A2C MICRO-FUEL.
DRAWING LIST MODEL 311D B-26473 PICTORAL DIAGRAM C-28339 SCHEMATIC C-26606 W IRING DIAGRAM NOTE: The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinator.
.
Date: 6/20/2006 Time : 01:45:07 PM Title: Indsrsvr01drawingsMASTER26000C26606-14.dwg.
Date: 7/20/04 Time : 10:56:54 AM Title: G:MASTER28000C28339-8.DWG.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Teledyne 311-D c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Teledyne 311-D - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Teledyne 311-D, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Teledyne 311-D va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Teledyne 311-D, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Teledyne 311-D.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Teledyne 311-D. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Teledyne 311-D ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.