Sierra Model 951 Digital Flo-Box Instruction Manual Part Number IM-951 Revision A 09-05 5 Harris Court, Building L Monterey , CA 93940 (831) 373-0200 (800) 866-0200 Fax (831) 373-4402 www .
Page 1 DESCRIPTION The Model 951 Digital Flow Box is a microprocessor-based digital indicator and power supply , capable of interfacing directly to a mass flow meter or a mass flow controller . The integrated +/-15 vdc @250ma power supply provides a well regulated, short circuit and thermal overload protected output.
INST ALLA TION AND WIRING INST ALLA TION The Model 951 enclosure is designated for panel mounting in a 1/8 DIN cutout. The cutout dimensions are shown below . T o panel mount the 951, perform the following steps: 1. Rotate the four pawl screws ( outside screws in each corner) several turns counter-clockwise to retract the pawls.
Page 3 CAL T ARE H1 H2 L1 L2 Pawl Screw (1 of 4 outside screws) These 4 screws are used to mount the instrument in a panel. Menu Switch S1 (Single Function Switch) Increment Switch S2 (Single Function.
Page 4 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | J1 P1 6 +15 vdc 8 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |.
Page 5 SWITCH DEFINITIONS Menu Switch (S1) The Menu switch (S1) is used to cycle through the flow controller setpoints, alarm setpoints, alarm hysteresis, calibration values and decimal point locations. The sequence is as follows. The designators, in parenthesis, identify text that will be displayed, momentarily , followed by the value.
Page 6 SWITCH DEFINITIONS (cont) Step and Min-Max Switch (S3) This switch has a dual function. When in the MENU sequence, it steps the digit to be incremented, one position to the right. Using S3 in conjunction with S2 allows rapid updating of all parameter values.
Page 7 FUNCTION DEFINITIONS The Mode switch (S1) cycles through the following functions. Flow Set Point (StPt) The Setpoint (0-5 vdc) command signal used to control the flow through a flow controller . The output voltage is 0-5 vdc. The setpoint output is defined as: Setpoint Output = StPt / CAL x 5.
Page 8 FUNCTION DEFINITIONS (cont) Alarm Set Point High 2 (SPH2) SPH2 is independent and functions identically to SPH1. SPH2 has its own open collector outputs and front panel LED (H2). Alarm Set Point Low 2 (SPL2) SPL2 is independent and functions identically to SPL1.
Page 9 FUNCTION DEFINITIONS (CONT) Decimal Point (dP) dP allows selection of decimal point locations for display . The locations selectable are x.xxxx, xx.
Page 10 OPERA TING INSTRUCTIONS TURN-ON INSTRUCTIONS Reference Figures 3 and 4 for proper input and output signal wiring. 1. Apply power to the instrument. 2. V erify the instrument readout blanks momentarily , then displays the value of the input signal.
Page 1 1 PROGRAM INSTRUCTIONS (cont) 2. Depress MODE (S1) switch once. The text followed by the current value of Alarm Set Point High 1 (SPH1) will be displayed. Use S2 and S3 to set Alarm Set Point High 1 to the desired value. Proceed to Step 3 to set Alarm Setpoint Low 1 (SPL1) or depress S4 to exit MODE sequence.
Page 12 PROGRAM INSTRUCTIONS (cont) 7. Depress MODE (S1) switch once. The text followed by the current value of Hysteresis Low will be displayed. Repeat Step 2 to set Alarm Hysteresis Low to the desired value. The maximum value for HL is 99 counts.
Page 13 Prior to connecting the 951 to the Flowmeter or Controller , verify that the Controller's +/-15 vdc requirement is no more than 40 0 ma. The 951 is capable of supplying up to 500ma for short periods of time ( in 1/2 or 1 hour increments).
Page 14 CALIBRA TION INSTRUCTIONS (cont) 6. Depress and hold CAL (S5) for approximately 3 seconds. The FS CAL LED should illuminate and the readout should display the FS Cal value within +/- 1 digit.
Page 15 RS232 SERIAL COMMUNICA TIONS (Special Order Only) The Parity (Even or None) and the Baud Rate (2400 or 9600) are selectable by setting DIP switches SW a and SWb, accessible from the right side of the instrument case.
Page 16 RS232 SERIAL COMMUNICA TIONS (Special Order Only) [Z] [CR] [LF] Zero the displayed reading [F] [CR] [LF] Calibrate Full Scale (Input signal is scaled to display Full Scale) [S] [counts] [CR] [LF] Set Full Scale display value.
MODEL SR5 PROGRAM MENU DIAGRAM Depress MODE ST ART Set Depress Depress Increments Flashing Digit Steps Flashing Digit to Right Set Depress MODE Depr.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Sierra 951 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Sierra 951 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Sierra 951, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Sierra 951 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Sierra 951, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Sierra 951.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Sierra 951. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Sierra 951 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.