Manuel d'utilisation / d'entretien du produit DP97 du fabricant Omega Vehicle Security
Aller à la page of 24
.
CONTENTS PAGE 1. INTRODUCTION 1 Full description of the DP97 Front Panel information at a glance 2. INSTALLATION 3 Supply Connection Input connections – Sensors 3. OPERATION 4 i) To read sensor temperature directly ii) To read differential temperature iii) Matching DP97 to calibrated sensors iv) Error indications v) ¾ wire recognition 4.
1. INTRODUCTION A high precision portable thermometer for metrology and other exacting laboratory applications, the DP97 is a proven instrum ent used world wide as a laboratory and site standard in pharmaceuticals, medical, food, environmental testing, R & D, and general industrial applications.
Front Panel Information at a Glance: when you switch on To obtain a reading from a temperature sensor, select channel A or B (1) and select the required units (2).
2. INSTALLATION Back Panel General a) The standard version is for bench - top use. Retractable legs can be extended to provide a convenient viewing angle.
3. OPERATION Ensure that the instrument is connected to a suitable supply or has charged batteries and that the required input connections are made. Switch “on” using the rear mounted rocker swit ch. The display will show “Pr 2.1” or “Pr 1.4” the programme designation.
iii) Matching DP97 to Calibrated Sensors It is possible to automatically match the DP97 to a Pt100 sensor with up to 10 dedicated calibration values. Such values, published in an associated certificate indicate true Ω versus °C values corresponding t o accurately known reference temperatures.
If using a shorter leadout from the probe or thicker gauge cable, pleas e ensure that the DP97 is in the correct wire recog nition mode. This can be checked by pressing the SPAN s s button ONCE. Two numbers will be displayed e.g. 3 - 4. The first digit being the X parameter (e.
Page 6 The calibration kit option (two precision resistors) is required to perform re - calibration. No additional instruments are required since the DP97 self - calibrates as follows: Caution: Do not proceed unless specified resistors are available. Failure to do so may invalidate the warranty.
Page 7 Switch off the Instrument 1) Remove the jumper lead JP1. This is VITAL. Should the instrument be switched on with JP1 In position, the calibration is lost. 2) Replace the instrument cover. The instrument is now ready for normal use. The central processor is software driven and no other adjustments or modifications should be attempted.
Page 8 6. RS232 COMMUNICATIONS General Rem ote control will be seized by sending an RS232 character, and relinquished by sending the LOCAL command. While in remote control the front panel buttons will be locked out (inoperative) and the right hand decimal point will indicate remote operation (loca l lock - out).
Page 9 Command set: Display parameters: UNITS OHMS UNITS C UNITS K UNITS F INPUT A INPUT B INPUT A - B Analogue parameters: SPAN Value in Ohms ZERO Value in Ohms Measurement: MEASURE Return to Local: LOCAL Output Protocol Replies will depend on the command implemented, and wil l be numeric or – with commas and spaces as delimiters.
Page 10 For digital probe matching within the instrument (version 2.1 only) additional commands are used as follows: CORRECTIONS A (turns OFF B corrections) Turns A on B (turns OFF A corrections) Turn.
Note: Use of the DP97 “ CORRECTIONS PC SOFTWARE ” supplied with the instrument facilitate the correction procedure by means of a much simplified, user - friendly approach described below.
Page 13.
Page 14.
b) Using the software From MAIN MENU: - To input calibration values into the DP97 (up to 10 may be used per channel): º Click EDIT to obtain EDIT corrections menu. º Enter values agains t #01 through #10 but note that i) All 10 need not be entered ii) When returning to main menu the Ù values are sorted into ascending order.
Stored values can be viewed at any time as follows: º From Main Menu, click CHANNEL to select channels A OR B º Click READ Stored values can be changed at any time by repeating input procedure.
General Ra nge/Sensor type Pt100 to IEC 751 (ITS 90 refers). – 199.99 to 849.99°C.R o = 100 Ω . Two input channels, each 3 or 4 wire connection with automatic recognition (with manual overrride). Overall Instrument ±0.02°C ± 1 digit for range - 200°C to +500°C.
Charger requirement 10 – 11.5V d.c. 1A. Power Consumption 10W nominal. Max 20W when battery is charging. Series Mode Rejection 60dB @ 50Hz (50mV rms applied). Common Mode Rejection 30V rms applied between input and earth produces no measurable effect.
Communications RS232 (standard) Remote control and m easure, Isolated, 9600 Baud, 8 data, no parity, 1 stop bit. PC Software (standard) PC software running in MS DOS allows programming (standard) of custom calibration and preview of set values from a PC.
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) mo nth grace period to the normal one (1) year product warranty to cover handling and shipping time.
USA: One Omega Drive, P.O. Box 4047 ISO 9001 Certified Stamford CT 06907 - 0047 TEL: (203) 359 - 1660 FAX: (203) 359 - 7700 e - mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L5A1 TEL: (514) 856 - 6928 FAX: (514) 856 - 6886 e - mail: info@omega.
Process Measurement a nd Control? OMEGA…Of Course! Shop online at www.omega.com TEMPERATURE þ Thermocouple, RTD 7 Thermistor Probes, Connectors, Panels & Assemblies þ Wire: Thermocouple, RTD &.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Omega Vehicle Security DP97 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Omega Vehicle Security DP97 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Omega Vehicle Security DP97, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Omega Vehicle Security DP97 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Omega Vehicle Security DP97, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Omega Vehicle Security DP97.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Omega Vehicle Security DP97. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Omega Vehicle Security DP97 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.