USER GUIDE AND SPECIFICATIONS NI USB-6509 This document contains info rmation about using th e National Instruments USB-6509 data acquisition (D A Q) de vice with NI-D A Qmx 8.7 and later . The NI USB-6509 is a bus-po wered 96-channel, digital input/o utput (I/O) device.
NI 6509 User Guide an d Specifications 2 ni.com DIO Signal Connection ....... .............. ................. .............. .............. ..16 Protecting Inductive Loads ........ .............. ................. .............. ..........17 Sinking and Sourcing Examples.
© National Instruments Corporation 3 NI 6509 User Gu id e and Specifications Installing the Hardware The NI-DAQmx for USB Devices Getting Started Guide contains non-software-specific in formation about how to install USB devices.
NI 6509 User Guide an d Specifications 4 ni.com For secure desktop use, you can adhere the su pplied rubber no n-skid feet to the underside of the device, as sho wn in Figure 2. Note Do not apply t he rubber feet if you are panel mou nting the NI USB-6509 or stacking multiple NI USB-6509 devices.
© National Instruments Corporation 5 NI 6509 User Gu id e and Specifications Panel Mounting T o mount the NI USB-6509 to a board or panel, complete the following steps while referring to Figure 3. Figure 3. Mounting the NI USB-6509 on a Panel Notes Do not apply the rubber feet to the NI USB-6509 when panel m ounting the device.
NI 6509 User Guide an d Specifications 6 ni.com Hardware Over view Figure 4 shows the key functional components of the NI USB-6 509. Figure 4. NI USB-6509 Block Diagram Figure 5 shows the back panel of the NI USB-6509. Figure 5. NI USB-6509 Back Panel Refer to the Safety G uidelines section of this document for im portant safety information.
© National Instruments Corporation 7 NI 6509 User Gu id e and Specifications Programming Devices in Software National Instruments measurement de vices for USB are packaged with NI-DAQ driver software.
NI 6509 User Guide an d Specifications 8 ni.com Related Documentation The following documents contain information that you may fin d helpful as you use this u ser guide: • NI-DAQmx for USB Getting S.
© National Instruments Corporation 9 NI 6509 User Gu id e and Specifications Figure 6. Pin Assignments for the NI USB-6509 using the SH100 -100-F Refer to the Signal D escriptions section for information about the signals av ailable on this connector .
NI 6509 User Guide an d Specifications 10 ni.com R1005050 Pin Assignments Figure 6 shows the pin assignments for the NI USB-6509 usin g the R1005050 cable. The naming con vention for each pin is P X . Y , where X is the port (P) number , and Y i s the line number or name.
© National Instruments Corporation 11 N I 6509 User Guid e and Specifications Signal Descriptions Table 1 lists the signals and descriptions for all signals available on the NI USB-6509 device. Table 1. NI USB-6509 Signal Descriptions Pin Signal Name Description MSB LSB 1, 3, 5, 7, 9, 11, 13, 15 P2.
NI 6509 User Guide an d Specifications 12 ni.com Digital I/O Static DIO on NI USB-6509 Devices Y ou can use each of the NI USB-6509 DIO lines as a static digital input (DI) or digital output (DO ) line. Y ou can use static DIO lines to monitor or control digital signals.
© National Instruments Corporation 13 N I 6509 User Guid e and Specifications conditioning solution s for digital application s requiring high current dri ve. • If you conf igure a DIO line as an input, do not dri ve the line with voltages outside of its normal operat ing range.
NI 6509 User Guide an d Specifications 14 ni.com Figure 9. DIO Channel Configured for High DIO State with External Load Example: At power up, the device is configured for input and, by defau lt, all DIO lines are high. To pull one ch annel low, complete the following steps: 1.
© National Instruments Corporation 15 N I 6509 User Guid e and Specifications Also, make sure the pull-up resistor value is not so large that leakage current from the DIO line br ings the voltage at the re sistor below a TTL-high le vel of 2 V . Figure 10.
NI 6509 User Guide an d Specifications 16 ni.com The NI USB -6509 has a current-sense ci rcuit on VBUS (USB p ower rail). If the output channels source more than the specif ied current, the current is limited.
© National Instruments Corporation 17 N I 6509 User Guid e and Specifications Protecting Inductive Loads When inductive loads are co nnected to outputs, a large counter-electromotive force may occu r at switching time because of the energy stored in the inductive load.
NI 6509 User Guide an d Specifications 18 ni.com Figure 14. NI USB-6509 Sourcing Connection Example, <24 mA Driving a Relay >24 mA Figures 1 5 and 16 are examples of connecting the N I USB-6 509 to a relay that requires more than 24 mA of current.
© National Instruments Corporation 19 N I 6509 User Guid e and Specifications Driving SSRs Figure 17 shows an example of connecting the NI U SB-6509 to a solid-state relay (SSR). Figure 17. NI USB-6509 SSR Connection Example LED Indicator The LED indicator , located on the de vice back panel, indicates de vice status.
NI 6509 User Guide an d Specifications 20 ni.com +5 V Power A vailable at I/O Connector Pins 49 and 99 su pply +5 V po wer to the I/ O connector . The po wer dra wn from these two pins shares the po we r consumption with the transceive rs. The I/O connector power has a fuse for ov ercurrent protection.
© National Instruments Corporation 21 N I 6509 User Guid e and Specifications Internally , the filter uses two clocks: the sample clock and the filter clock.
NI 6509 User Guide an d Specifications 22 ni.com Programmable Power -Up States At power -up, the output dri ves on the NI USB-6509 are disabled. Al l lines are user-conf igurable for high-im pedance input, high ou tput, or low output. User-c onfigurable po wer-up states are useful for ensuring that the NI USB-6509 powers up in a kno wn state.
© National Instruments Corporation 23 N I 6509 User Guid e and Specifications Change Detection Example T able 4 sho ws a change detection example for six bits of one port. This example assumes the following line connections: • Bits 7, 6, 5, and 4 are connected to data lines from a four-bit TTL output device.
NI 6509 User Guide an d Specifications 24 ni.com The expiration signal that indicates an expired watchdog continues to assert until the watchdog is disarmed. After the watchdo g timer expires, the NI USB-6509 ignores any writes until the watchdog timer is disarmed.
© National Instruments Corporation 25 N I 6509 User Guid e and Specifications Power-on state .............. ................. ......... Configured as inputs, pull-up/pull-down (select able on the back panel switch ) Pull-up/pull-down resistor........
NI 6509 User Guide an d Specifications 26 ni.com The total current sinking/sourcing from one port cannot exceed 100 mA. T otal current sourced b y all DO line s simultaneously should not e xceed 215 mA for bus-po wered configuration.
© National Instruments Corporation 27 N I 6509 User Guid e and Specifications Safety This product is designed to meet the requirements of the follow ing standards of safety for electrical e quipment .
NI 6509 User Guide an d Specifications 28 ni.com CE Compliance This product meets the essential re quirements of applicable European Directives, as amended for CE markin g, as follows: • 2006/95/EC;.
© National Instruments Corporation 29 N I 6509 User Guid e and Specifications Safety Guidelines Caution Operate the hardware only as describe d in these operating instructions. The following section contains important safety infor mation that you must follow when installing and using the NI USB-6509.
NI 6509 User Guide an d Specifications 30 ni.com Operate the device at or belo w the Measurement Cate gory I 1 . Measurement circuits are subjected to working v oltages 2 and transient s tresses (overv oltage) from the circuit to which they are connected during measurement or test.
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for mor e informat ion about Nationa l Instruments trademarks. Other product and compan y names mentioned herein are t rademarks or trade names of their respective companie s.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté National Instruments NI USB-6509 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du National Instruments NI USB-6509 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation National Instruments NI USB-6509, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le National Instruments NI USB-6509 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le National Instruments NI USB-6509, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du National Instruments NI USB-6509.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le National Instruments NI USB-6509. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei National Instruments NI USB-6509 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.