4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 70-82-25-19 Rev. 1 7/99.
ii 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 1999 by Honeywell Inc.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual iii A bout This Document Abstract The purpose of this document is to support the installation operation and m aintenance of the 4909 CPVC Conductivity Cell Remov al/Insertion Assem bly.
iv 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Organization Telephone Address Honeywell TAC 1-800-243-9883 Voice 1100 Virginia Drive Fort Washington, PA 19034.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual v Contents 1. INTRODUCTION ....................................................................................................1 1.1 Overview ............................
vi 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4. MA INTENA NCE A ND REPLAC EMENT PA RTS .................................................17 4.1 To Clean The Cell ...................................................
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual vii Tables Table 4-1 Replacem ent Parts __________________________________________________________ 18 Table 4-2 Voltage.
.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 1 1. Introduction 1.1 Overview The 4909 Conduct ivit y C ell Ins ertion/R emoval A ssembly is design ed fo.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 2 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 1.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 3 1/ 4" NP T Plug Cabl e 1 - 1/ 2" NPT Nipple Conducti vity Cell P urg e Po rt C ompr es s ion H.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 4 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Cable C onductivity Cell Ad ap t er * * Au t oma t ic Tempera t ur e Com pensato r , w hen specifie d by cat a log nu mber, is m ounted insid e ad apter.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 5 A D C B Ter mina l Boar d As s e m b l y #6- 32 Bi ndi ng He ad Sc r e w & Wa s h e r Vi e w of B a .
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 6 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99.
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7 2. Specifications and Model Selection Guide 2.1 Specifications Cell Constants 0.01, 0.1,1.0, 5, 10, 20, 25 and 50 cm -1 . Electrode Material Nickel or Platinum , as specified.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 8 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Integral Automatic Temperature Compensation Refer to Section 2.2 – Model Selection Guid e Leadwire Tefzel covered, 18-g age cable (7 feet, 20 f eet, or Universal Head, as specified).
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 9 049 TABLE IV Selection 09 08 Leadwire Length ’ 7 ft.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 10 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 049 TABLE VII - OPTIONS Tagging None 0_ Linen L_ Stainless Steel S_ Certificate of Calibration No _ 0 Yes _ 1 Notes: 1.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 11 3. Installation 3.1 Requirements To insur e that a re presenta tive sample is being measured at all times, the solutio n must m o ve thr ough and completely purge out the cell channels or guard tube.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 12 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Allow at least 1/2 inch clearance bey ond the end of th e cell and 1/8 to 3/16 inch radius clearance surrounding the cell to p e rmit circulation of the solutio n.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 13 3. Using a strap wrench on the m ounting nipple, tighten the assembly an additional 1-1/2 to 2 turns. Do not use stillson or chain type wrenches as they m ay damage and weaken the CPVC plastic.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 14 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 3.5 Removal 1. REDUCE PROCESS PRESSURE TO 50 PSIG OR LESS. Sh ut off purge line, if used. Dis connect wi ring conn ections if X1 is selected for Table IV in the MSG.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 15 2.5" (64mm) App r ox 4.1" (105mm) App r ox 12.5" (317mm) X " See Bel ow App r ox 19.7" max . (502mm) Allow a minim um clearance o f 36" (914 m m) beyond this poin t for removal of electrode.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 16 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 2. 5" (64mm) Ap prox 4 .1" (105mm) Approx 12. 5" (317mm) X " See Below A pprox 19.
Maintenance and Replacement Parts 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 17 4. Maintenance and Replacement Parts If a series of below normal condu ctivity readings or abov e normal resistiv ity readings occu r, this m ay indicate that the cell is not filled with process solution.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 18 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4.3 Replacement Parts Table 4-1 Replacement Parts Description Part Number Complete Ass embly See Figs.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 19 5. Platinization and Platinum Black Only the electrodes havi ng cons tants from 5 t.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 20 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 A D C B View o f J unct ion Box Head w ith Cap Re moved WH I T E BLA CK RED GREEN Cell Temp. Com p.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 21 A A D C B View o f J unct i on Box He ad with Cap Removed WH I T E BLA CK RED GREEN Cell Temp.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 22 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 WH I T E BLACK RED GREE N Cell Temp. Comp. WK S H G R Five Point Termin al Board. Each Terminal W i ll Accept #16 G age Max .
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 23 WH I T E BLA CK RED GREEN Cell Temp.
.
.
Industrial Auto mation and Co ntrol Honey well, Inc. 1100 Virginia D rive Fort Wa shington, Pennsy lvania 19034.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Honeywell 4909 CPVC c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Honeywell 4909 CPVC - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Honeywell 4909 CPVC, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Honeywell 4909 CPVC va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Honeywell 4909 CPVC, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Honeywell 4909 CPVC.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Honeywell 4909 CPVC. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Honeywell 4909 CPVC ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.