Manuel d'utilisation / d'entretien du produit PN 51-389V du fabricant Emerson Process Management
Aller à la page of 34
Combination pH/ORP Sensor 389VP Instruction Man ual PN 51-389VP/rev ./F J anuary 2013.
This page left blank intentionally.
D ANGER: HAZ ARDOUS ARE A INST ALLA TION Eme rson de sig ns, m anu factu res , and tes ts its Ro se mou nt An alytical prod ucts t o mee t man y na tion al an d int e rna tio nal stand ards .
II About This Document This manual contains instruc tions for installation and opera tion of the Model 1066 Smar t T ransmitte r . The fo llowing list provides note s conc erning all revisions of this docume nt. Re v . Lev el Date Notes A 2/01 This is the initial re lease of the product manu al.
T able of Cont ents III 3 89VP Instruc tion Manual T able of Contents PN 51-389VP January 2013 Con te nt s Se cti on 1 : D e sc ri p ti on a nd Spe ci fi ca ti on s 1. 1 Fe at ur es an d A pp lic at io ns ..............................................
IV T able of Cont ents T able of Contents 389VP Instruc tion Manual January 2013 PN 51-389VP List o f T able s No. Title 3- 1 OR P o f S at ur at ed Qu inh ydr one S olu ti on (I n M il liv olt s) ......................................................
Description and Specific ations 1 3 89VP Instruc tion Manual Section 1: Description and specifications PN 51-389VP January 2013 F eatures and Applications 38 9V P tri ple -ju ncti on s ens ors are no w off er ed wi th SM AR T cap abil it ie s.
2 Descrip tion and Specific ations tanks, or ponds. The 389VP is suit able for virtually all applications and is comp atible with Ro semoun t Analytical and other manufacturers ’ instruments. Installation is ea sily achieved through the wide variet y of mounting confi gurations.
Ordering Information T he 389VP pH/ORP Sensor is off ered with a V ariopol (VP) connector and u ses a mating VP cable (purch ased se parate ly). A remote preampli fier (j-box or instrumen t) must be u sed with this sensor , unless (-70) preamplifier option is selected.
4 Descrip tion and Specific ations S ection 1: Description and specifications 389VP Instructio n Manual January 2013 PN 51-389VP This page left blank intentionally.
Installation 5 389VP Instruc tion Manual Section 2: Inst allation PN 51-389VP January 2013 Unpacking and inspec tion Inspect the outside o f the car ton f or an y dam age. If dam age is detected, contact the carrier imme- dia tely . Inspect the hardware .
6 Installation S ection 2: Inst allation 389VP Instruc tion Manual January 2013 PN 51-389VP Subme r sion M oun ting The 389 and 389VP Sensor has a 1 in. MNPT process connection at the back of the sensor . Utilizing a st andar d 1 in. union, the sensor may be mounted to a 1 in.
Installation 7 3 89VP Instruc tion Manual Section 2: Inst allation PN 51-389VP January 2013 FIGURE 2-1. Dim ensional Drawing WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH.
8 Installation S ection 2: Inst allation 389VP Instruc tion Manual January 2013 PN 51-389VP FIGURE 2-2. Subm ersion Inst allations DWG. NO. REV . 400389VP11 A DWG.
Installation 9 3 89VP Instruc tion Manual Section 2: Inst allation PN 51-389VP January 2013 FIGURE 2-3. Flow Through and Insertion Installations FIGURE 2-4.
10 Installation S ection 2: Inst allation 389VP Instruc tion Manual January 2013 PN 51-389VP F IGURE 2-6. VP6 81 Wiring F IGURE 2-7. VP6 1181 Wiring FIGURE 2-9.
Installation 11 3 89VP Instruc tion Manual Section 2: Inst allation PN 51-389VP January 2013 F IGURE 2-11. VP6 1181, 1050/1060, and 1003/1023 Wiring through Remote Junction Box F IGURE 2-12. VP6 2081 Wiring through Re mote Junction Box FIGURE 2-13. VP6 1055-22-32 Wiring FIGURE 2-14.
12 Installation S ection 2: Inst allation 389VP Instruc tion Manual January 2013 PN 51-389VP F IGURE 2-15. VP6 81 Wiring through Remot e Junction Box F IGURE 2-16. VP6 3081 & 4081 Wiring through Remot e Junction Box FIGURE 2-17. VP6 1054 Wiring FIGURE 2-18.
Installation 13 3 89VP Instruc tion Manual Section 2: Inst allation PN 51-389VP January 2013 F IGURE 2-19. C AS Installation, label information FIGURE 2-20.
14 Installation S ection 2: Inst allation 389VP Instruc tion Manual January 2013 PN 51-389VP F IGURE 2-22. SCL -(P/Q) Wiring F IGURE 2-23. 2700 Wiring FIGURE 2-23.
Installation 15 3 89VP Instruc tion Manual Section 2: Inst allation PN 51-389VP January 2013 F IGURE 2-25. Xmt VP8 Wiring. This is th e standard VP8 cab le wiring. VP8 ca ble assembly works both with VP6 and VP8 sensor connector FIGURE 2-26. 1056 VP8 Wiring.
16 Installation S ection 2: Inst allation 389VP Instruc tion Manual January 2013 PN 51-389VP F IGURE 2-27. 5081 VP8 Wiring. This is the standard VP8 cable wiring.
Start -up and Calibration 17 3 89VP Instruc tion Manual Section 3: Star t -up and Calibration PN 51-389VP January 2013 Sensor Prepar ation Shak e down the sensor to re move any air bubble s that may be prese nt at the tip of the pH gla ss bulb .
18 Start-up and Calibration S ection 3: Star t -up and Calibration 389VP Instruction Manual January 2013 PN 51-389VP 3.3 3.3.1 2 . Mea sure and record the pH of the proces s solution sample with a anothe r tempe rature com- pensa ted, calibrated pH instrument.
Mainte nance 19 3 89VP Instruc tion Manual Section 4: Maintenance PN 51-389VP January 2013 The 389VP Sensor requires minimum mainte nance . The sensor should be ke pt cle an and free of debris and sedime nt at all time s.
20 Mainte nance S ection 4: Maintenance 389VP Instruction Manual January 2013 PN 51-389VP 389VP ORP Pla tinum Elec trode Check The platinum electrode m ay be check ed a s f ollows.
Mainte nance 21 3 89VP Instruc tion Manual Section 4: Maintenance PN 51-389VP January 2013 476 +20 mV at 25°C when us ed with a sa turated KC l/AgCl ref ere nce e lec trode and platinum meas- u ring electrode.
22 Mainte nance S ection 4: Maintenance 389VP Instruction Manual January 2013 PN 51-389VP This page left blank intentionally.
T roubleshooting 23 3 89VP Instruc tion Manual Section 5: T roubleshooting PN 51-389VP January 2013 Sec tion 5: T rouble shoo ting y d e m e R e s u a C e l b a b o r P e l b u o r T Meter reads off scale (Display Defective preamplifier . Replace preamplifier (for code 02 reads overrange).
24 T roubleshooting S ection 5: T roubleshooting 389VP Instruction Manual January 2013 PN 51-389VP P ART DESCRIPTION 24281-00 15 ft. VP8 cable 24281-01 25 ft. VP8 cable 24281-02 2.5 ft. VP8 cable 24281-03 50 ft. VP8 cable 24281-04 100 ft.
Return of Materi al 25 3 89VP Instruc tion Manual Section 6: Return of Material PN 51-389VP January 2013 Sec tion 6: R eturn o f Ma te r i al 6.1 General T o expedite the repair and return of instruments, proper communica tion between the cu stomer and the factor y is impor tant.
26 R eturn of Mate rial S ection 6: Return of Material 389VP Instruc tion Manual January 2013 PN 51-389VP RETURNED BY: BILL TO: _____________________________ _____________________________ ____________.
War rant y 27 3 89VP Instruc tion Manual W arranty PN 51-389VP January 2013 W ARRANT Y Selle r warrants that the firmware will exe cute the programming instructions pro vided by Selle r , and that the.
Credit Cards for U.S. Purcha ses Only. 8 Emerson Process Management 2400 Barranca Parkway Ir vine , CA 92606 USA T el: (949) 757-8500 F ax: (949) 474-7250 rosemo untanaly tical.com © Rose mount Analy tical Inc. 2013 3 89VP Instruc tion Manual January 2013 PN 51-389VP ©2013 R ose mount An aly tica l, Inc.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Emerson Process Management PN 51-389V c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Emerson Process Management PN 51-389V - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Emerson Process Management PN 51-389V, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Emerson Process Management PN 51-389V va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Emerson Process Management PN 51-389V, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Emerson Process Management PN 51-389V.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Emerson Process Management PN 51-389V. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Emerson Process Management PN 51-389V ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.