Manuel d'utilisation / d'entretien du produit MMI-20014120 du fabricant Emerson Process Management
Aller à la page of 78
Inst allation and Configuration Manual P/N MMI-20014120, Rev . AB August 2012 Micro Motion ® 3098 Gas S pecific Gravity Meter.
©2012, Micro Motion, Inc . All rights reserved. Micro Motion is a re gistered trade name o f Micro Motion, I nc., Boulder , Colorado . The Micro Motion and Emerson log os are trademarks and service mar ks of Emerson Electric Co. All other trademarks are property of their respective owners.
Installation and Conf iguration Manual i Contents Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Safety guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ii Micro Motion 309 8 Gas Specific Gravity Met er Contents Chapter 4 Accuracy Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 4.1 Accuracy conside rations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation and Conf iguration Manual iii Contents C .2 Ne w and unused e quipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 C .3 Used equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
iv Micro Motion 309 8 Gas Specific Gravity Met er Contents.
Installation and Conf iguration Manual 1 Installation Procedure A ccuracy Considerations Electrical Connections Intro duction Chapter 1 Introduction This chapter contains an outline of ho w the 3098 sp ecif ic gra vity meter works, defines some of the terms commonly used in the manu al, and also gi ves some practical applications for the 3098.
2 Micro Motion 309 8 Gas Specific Gravity Met er Introduction 1.2 Specific gravity measurement Most major gas flo w metering systems require th e metered quantity to be presented in heat or standard volume units. T o achie ve this requirement, it is often necessary to make continuous and accurate measurements of spec if ic gravity .
Installation and Conf iguration Manual 3 Introduction Installation Procedure A ccuracy Considerations Electrical Connections Intro duction 1.3 Functional description Figure 1-2 Schematic diagram of a .
4 Micro Motion 309 8 Gas Specific Gravity Met er Introduction If the sample gas pressure rises abov e that of th e reference chamber pressure, the pressure control v alve opens to v ent the excess gas via an outlet ori f ice in the enclosure side, so that the sample gas pressure is reduced to equal the re ference gas pressure.
Installation and Conf iguration Manual 5 Introduction Installation Procedure A ccuracy Considerations Electrical Connections Intro duction Figure 1-3 T ypical 3098 specific gravity measuring system 1.
6 Micro Motion 309 8 Gas Specific Gravity Met er Introduction 1.4.2 Standard (base o r normal) density Standard (base or normal) density ( ρ s ) is the absolute density of a gas at standard (base or normal) conditions o f temperature an d pressure an d is commonly u sed for standa rd volu mne flo w determination from mass flo w measurement.
Installation and Conf iguration Manual 7 Introduction Installation Procedure A ccuracy Considerations Electrical Connections Intro duction 1.5 Physical prop erties of gas compounds 1.6 Applications The follo wing are typical applications where specif ic gravity measurement is an essential parameter .
8 Micro Motion 309 8 Gas Specific Gravity Met er Introduction 1.6.2 Wob be index measurement The burning characteristic of a ga s must be well establis hed for eff icient combustion and to ensure that no flame lift or flame light-back occurs on a par ticular b urner .
Installation and Conf iguration Manual 9 Installation Procedure Accuracy Considerations Electrical Connections Intro duction Chapter 2 Installation Procedure 2.1 Installation proce dure The procedure for installing the 30 98 in volv es the follo wing steps: 1.
10 Micro Motion 309 8 Gas Specific Gravity Met er Installation Procedure The 3098 specif ic gravity met er is contained inside an IP-rated enclo sure (which provides thermal insulation) and a mountin g system (consisting of a brack et and feet) to f ix the unit in place.
Installation and Conf iguration Manual 11 Installation Procedure Installation Procedure Accuracy Considerations Electrical Connections Intro duction 2.
12 Micro Motion 309 8 Gas Specific Gravity Met er Installation Procedure 2.6 Set-up pr ocedure – purge cycling and calibration Figure 2-2 Schematic diagram of a typical 3098 specific gravity measuring system The procedure for pur ging and calibrating the 3098 specif ic gravity met er is gi ven belo w (see Figure 2-2 for reference): 1.
Installation and Conf iguration Manual 13 Installation Procedure Installation Procedure Accuracy Considerations Electrical Connections Intro duction Purge cycling 10. Close v alve C. 11. When Control Pressure Indicator is at the desired v alue, shut v alve A and open v alve F .
14 Micro Motion 309 8 Gas Specific Gravity Met er Installation Procedure 27. Shut v alve B and disconnect the seco nd calibration gas bottle from pipe work. 28. Open the iso lation v alve D. 29. Open v alve A. 2.7 Outline dimensional drawings Figure 2-3 shows a 3098 specific gra vity meter witho ut an enclosure.
Installation and Conf iguration Manual 15 Installation Procedure Installation Procedure Accuracy Considerations Electrical Connections Intro duction Figure 2-3 3098 specific gravity me ter without an enclosure Dimensions in inches (mm) ∅ 12.4 (314) ∅ 8 (203.
16 Micro Motion 309 8 Gas Specific Gravity Met er Installation Procedure Figure 2-4 3098 specific gravity me ter with a small enclosure 20.3 (516) 19.7 (500) 12 (300) 1.
Installation and Conf iguration Manual 17 Installation Procedure Installation Procedure Accuracy Considerations Electrical Connections Intro duction Figure 2-5 3098 specific gravity mete r with a large enclosure 24.3 (616) 23.6 (600) 12 (300) 1.2 (30) 1 / 3 (8.
18 Micro Motion 309 8 Gas Specific Gravity Met er Installation Procedure.
Installation and Conf iguration Manual 19 Installation Procedure A ccuracy Considerations Electrical Connections Intro duction Chapter 3 Electrical Connections This chapter contains details and wiring.
20 Micro Motion 309 8 Gas Specific Gravity Met er Electrical Connectio ns Figure 3-2 Intercon nection diagram 3.2 EMC cabling and earthing T o meet the EC Directiv e for EMC (Electromagnetic Compatibility), it is r ecommended that the meter be connected using a su itable instrumentation cable an d earthed through the m eter body and pipew ork.
Installation and Conf iguration Manual 21 Electrical Connectio ns Installation Procedure A ccuracy Considerations Electrical Connections Intro duction Figure 3-3 Schema tic block diagram of meter circuit (2-wire system) Figure 3-4 Schema tic block diagram of meter circuit (3-wire system) 3.
22 Micro Motion 309 8 Gas Specific Gravity Met er Electrical Connectio ns When operat ing in a safe area with a 3-wire sy stem, the line resistance between meter and signal con verter must be greater than 40 ohms.
Installation and Conf iguration Manual 23 Electrical Connectio ns Installation Procedure A ccuracy Considerations Electrical Connections Intro duction Figure 3-7 7950 signal converter a nd gas specifi.
24 Micro Motion 309 8 Gas Specific Gravity Met er Electrical Connectio ns Figure 3-9 7950 signal converter and gas sp ecific gravity 3-wire system with shunt-diode safety barrier (hazardous area) Figu.
Installation and Conf iguration Manual 25 Electrical Connectio ns Installation Procedure A ccuracy Considerations Electrical Connections Intro duction Figure 3-12 7951 flow computer/7951 signal conver.
26 Micro Motion 309 8 Gas Specific Gravity Met er Electrical Connectio ns 3.5.4 7951 3-wire co nfiguration Figure 3-14 7951 flow computer/7951 signal converter gas specific gravity 3-wire system (safe.
Installation and Conf iguration Manual 27 Electrical Connectio ns Installation Procedure A ccuracy Considerations Electrical Connections Intro duction 3.6 System connections (customer’ s own equipment) 3.6.1 Non-hazardous areas Po wer supply to Density Meter: 15.
28 Micro Motion 309 8 Gas Specific Gravity Met er Electrical Connectio ns 3.6.4 Custo mer's equipment, 3-wire config uration Figure 3-19 Electrical connection s for meter 3-wire option used with .
Installation and Conf iguration Manual 29 Installation Procedure Accuracy Considerations Electrical Connections Intro duction Chapter 4 Accuracy Considerations This chapter provides a method for estimating th e accuracy of 3098 specif ic gravity meter measurements under v arious conditions.
30 Micro Motion 309 8 Gas Specific Gravity Met er Accuracy Considerations 4.1.3 Calculating parameters Density range at 20 °C Calculated using equation: Density meter temperature coefficient error In.
Installation and Conf iguration Manual 31 Accuracy Considerations Installation Procedure Accuracy Considerations Electrical Connections Intro duction Installation Procedure Accuracy Considerations Ele.
32 Micro Motion 309 8 Gas Specific Gravity Met er Accuracy Considerations W ith this decided, the 3098 specif ic gravity me ter can be calibrated by follo wing the calibration procedure described in Section 2.
Installation and Conf iguration Manual 33 Accuracy Considerations Installation Procedure Accuracy Considerations Electrical Connections Intro duction Installation Procedure Accuracy Considerations Ele.
34 Micro Motion 309 8 Gas Specific Gravity Met er Accuracy Considerations T E M P E R A T U R E C O E F F I C I E N T C A L I B R A T I O N C E R T I F I C A T E 3098 SPECIFIC GRAVITY METER Serial Numbers: Instrument 000001 Amplifier 000001 Cyclinder 000001 Pressure Test Units pressure tested to 300 p.
Installation and Conf iguration Manual 35 Specification Maintenance and F ault Finding Chapter 5 Maintenance and Fault Finding 5.1 Introduction This chapter deals with the reco mmended serv icing and maintenance that can be carried out under field condi tions, including calibration checks, fault- f inding procedures and simple maintenance.
36 Micro Motion 309 8 Gas Specific Gravity Met er Maintenance and Fault Finding 5.3.2 Instrument under -reads This is most probably due to a gas leak fro m the reference chamber .
Installation and Conf iguration Manual 37 Maintenance and Fault Finding Specification Maintenance and F ault Finding If the erratic signal is only present while there is a f l o w o f s a m p l e g a s t h rough the unit, then the fault is likely to be due to a malfunction of th e pressure contro l va lve, brought abou t by the presence of dirt.
38 Micro Motion 309 8 Gas Specific Gravity Met er Maintenance and Fault Finding 5.4.1 Main meter (3098 specific gr avity meter) removal (Figure 5-1) The instructions in this section ap ply only to 3098 specif ic gravity meters supplied with an enclosure (see Safety guidelines o n page 1).
Installation and Conf iguration Manual 39 Maintenance and Fault Finding Specification Maintenance and F ault Finding Figure 5-1 3098 specific gravity meter general a ssembly schematic (typical enclosure) 5.
40 Micro Motion 309 8 Gas Specific Gravity Met er Maintenance and Fault Finding 2. Use two of the remov ed M6 bolts to jack the meter from its housing using the tw o threaded holes found in the mounting housing (item 6). W ARNING! Do not try to lever the unit out – this will damage the sealing O-ri ng and the vibrating element.
Installation and Conf iguration Manual 41 Maintenance and Fault Finding Specification Maintenance and F ault Finding 4. Gently rest the 3098 specif ic gravity meter on its side and undo the six M6 bolts that lock the base plate to the reference cha mber .
42 Micro Motion 309 8 Gas Specific Gravity Met er Maintenance and Fault Finding 6. Re-assembly is the re verse of steps 2 and 1 abo ve. Note: Once the unit has been r eplac ed into its enclosure , a leak chec k must be performed be for e on-line operation.
Installation and Conf iguration Manual 43 Maintenance and Fault Finding Specification Maintenance and F ault Finding 5.4.7 Leak testing the 3098 specific gravity meter Leaks incurred during servicing can be cate gorized under two main headings: • Reference chamber leaks a.
44 Micro Motion 309 8 Gas Specific Gravity Met er Maintenance and Fault Finding.
Installation and Conf iguration Manual 45 Specification Maintenance and F ault Finding Chapter 6 Specifications 6.1 3098 speci fic gravity meter specifications Note: Some parts of this specif ication (marked with *) cannot b e guarant eed for 3098 specif ic gravity meters supplied without an IP-rated enclosur e (see Safety guidelines on pa ge 1).
46 Micro Motion 309 8 Gas Specific Gravity Met er Specifications 6.1.2 Electrical 6.1.3 Mechanical 6.1.4 Safety For A TEX/IECEx installations, see the A TEX/IECEx Safety Instructions booklet and PED Safety Instructions booklet (a v aila ble at www .micromotion.
Installation and Conf iguration Manual 47 Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization Appendix A Performance Optimization A.1 Introduction The 3098 sp ecific gra vity meter uses a vibrating el ement density sens or that is located within a pressure regulating system.
48 Micro Motion 309 8 Gas Specific Gravity Met er Performa nce Optimizat ion A.1.2 The non-ideal behaviour of gases This beha viour will af fect the operation of the me asurement system since the measurement of density at the operating condition is not on ly related to the relati ve dens ity of the gas b ut also to its compressibility factors.
Installation and Conf iguration Manual 49 Performa nce Optimizat ion Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization A.2 Recommended calibration methods From the previous descriptions it can be apprec iate d that there is a choice of calibration procedures.
50 Micro Motion 309 8 Gas Specific Gravity Met er Velocity of sound effect The velocity of sound effect on the sensor is such that: The velocity of sound in a gas can be determined as follo ws: For an.
Installation and Conf iguration Manual 51 Performa nce Optimizat ion Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization Compressibility factor The normal or b.
52 Micro Motion 309 8 Gas Specific Gravity Met er Performa nce Optimizat ion The Z factors of the calibration gases and the sample gases should be calcul ated at both base and operating conditions in order to es tablish the compressi bility factor V F and then tab ulated as shown by the ex ample in Ta b l e A - 2 .
Installation and Conf iguration Manual 53 Performa nce Optimizat ion Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization Legend for T able A1 Summary This appe.
54 Micro Motion 309 8 Gas Specific Gravity Met er Performa nce Optimizat ion T able A-1 At base conditions At sample gas conditions of ° C and bars 1 Gas type 2 Molecular mass M (1) (1) Column 2 data derived fr om p age x.
Installation and Conf iguration Manual 55 Performa nce Optimizat ion Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization Ta b l e A - 2 At base conditions At sample gas conditions of ° C and bar s 1 Gas type 2 Molecular mass M (1) (1) Column 2 data derived fr om page x.
56 Micro Motion 309 8 Gas Specific Gravity Met er Performa nce Optimizat ion Example Calcul ations Column 5 Metha ne Nitrogen Sample A Sample B Sample C Column 6 Metha ne Nitrogen Sample A Sample B Sample C Column 7 Methane = 1+7(0.00017+6E –5x16.04–1.
Installation and Conf iguration Manual 57 Performa nce Optimizat ion Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization Column 8 Methane = 0.9899 Nitrogen = 0.9988 Sample A = 0.9895 Sample B = 0.9896 Sample C = 0.
58 Micro Motion 309 8 Gas Specific Gravity Met er Performa nce Optimizat ion An alternati ve method for deri ving modified v alues of Y to produce nil error is gi ven on the ne xt page. Column 12 Methane 0 Nitrogen 0 Sample A 0.13 Sample B 0.16 Sample C 0.
Installation and Conf iguration Manual 59 Performa nce Optimizat ion Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization Alternati ve simplif ied method of der.
60 Micro Motion 309 8 Gas Specific Gravity Met er Performa nce Optimizat ion.
Installation and Conf iguration Manual 61 Principles of Operation Cer tified System Dra wings Return P olicy Perf ormance Optimization Appendix B Principles of Operation B.
62 Micro Motion 309 8 Gas Specific Gravity Met er Principl es of Operation Since conditions of constant volume and quantity e xis t for the reference gas, its density and molecular weight are constant an d from equation 5.
Installation and Conf iguration Manual 63 Principles of Operation C er tified System Drawings Return P olicy Perf ormance Optimization Appendix C Return Policy C.
64 Micro Motion 309 8 Gas Specific Gravity Met er Return Policy.
Installation and Conf iguration Manual 65 Principles of Operation Cer tified System Dra wings Return P olicy Perf ormance Optimization Appendix D Certified System Drawings D.1 General All certified dra wings in this manual are giv en here for planning purposes only .
66 Micro Motion 309 8 Gas Specific Gravity Met er Certified System Drawings Figure D-1 CSA system drawing, gas gr oups A, B, C, and D (2-wire o ption) shunt diode barrier.
Installation and Conf iguration Manual 67 Certified System Drawings Principles of Operation Cer tified System Dra wings Return P olicy Perf ormance Optimization Principles of Operation Cer tified Syst.
68 Micro Motion 309 8 Gas Specific Gravity Met er Figure D-3 CSA system drawing, gas gr oups A, B, C, and D (2-wire o ption) isolating interface unit s.
Installation and Conf iguration Manual 69 Certified System Drawings Principles of Operation Cer tified System Dra wings Return P olicy Perf ormance Optimization Principles of Operation Cer tified Syst.
70 Micro Motion 309 8 Gas Specific Gravity Met er Certified System Drawings.
.
Micro Motion Inc. USA Worldwide Headqu arters 7070 Winchester Circle Boulder , Colorado 80301 T +1 303-527-5200 +1 800-522-6277 F +1 303-530-8459 Micro Motion Europe Emerson Process Management Neonstr.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Emerson Process Management MMI-20014120 c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Emerson Process Management MMI-20014120 - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Emerson Process Management MMI-20014120, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Emerson Process Management MMI-20014120 va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Emerson Process Management MMI-20014120, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Emerson Process Management MMI-20014120.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Emerson Process Management MMI-20014120. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Emerson Process Management MMI-20014120 ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.