Micro Motion ® F-Series Sensor Inst allation Manual Inst allation Manual P/N 20002298, Rev . C September 20 07.
.
Sensor In stallation: F-Series 1 Before Y ou Begin Before Y ou Begin This manual describes ho w to install a Micro Motion ® F-Series sensor . The follo wi ng information is provided in this manual: Customer service For technical assistance, phone the Micr o Motion Customer Service depa rtment: • In the U.
2 Sensor In stallation : F-Series Before Y ou Begin Specifications Full product specif ications can b e found in the F-Seri es Product Data Sheet, which is a v ailable from the Micro Motion web site at www .
Sensor In stallation: F-Series 3 Before Y ou Begin Figure 1 F-Series sensor with core p rocessor Approv al tag Flow direction arrow Core processor housing Calibration tag Process connection Approv al .
4 Sensor In stallation : F-Series Before Y ou Begin Figure 2 F-Series sensor with integrally mounted Model 1700/2700 transmitter Model 1700/2700 transmitter Approv al tag Flow direction arrow Core pro.
Sensor In stallation: F-Series 5 Before Y ou Begin Figure 3 F-Series sensor with junction box Junction bo x Approv al tag Flow direction arro w Approv al tag Process connection Calibration tag Purge f.
6 Sensor In stallation : F-Series Before Y ou Begin Figure 4 F-Series sensor with integra lly mounted IF T 9701 transmitter Calibration tag Flow direction arrow Model IFT9701 transmitter Approv al tag.
Sensor In stallation: F-Series 7 Before Y ou Begin Figure 5 High-temperature or extreme high-te mperature F-Series sensor with flexible conduit Flow direction arrow T ransmitter , core processo r , or.
8 Sensor In stallation : F-Series Determining a Location Step 1 Determining a Location Choose a location for the sensor based on the requirements descri bed in th is section. The foll o wing general guidelines can help you select an appropriate location fo r the sensor .
Sensor In stallation: F-Series 9 Determining a Locati on Figure 6 F-Series sensor ambie nt and process temperature limits (all models except high-temperature and extreme hig h-temperature models) Maxi.
10 Sensor In stallation : F-Series Determining a Location Figure 7 F-Series sensor ambient an d process temperature limits (high-tempera ture and extreme high-temperature mo dels) Hazardous area approv als may impose additional limits on ambient and process temperature.
Sensor In stallation: F-Series 11 Determining a Locati on Maximum wiring distances If the transmitter is mounted remote ly from the sensor , the maximum di stance between the sensor and transmitter depends on cable type.
12 Sensor In stallation : F-Series Orientin g the Se nsor Step 2 Orienting the Sensor The sensor will function prop erly in an y orientation if the sensor tubes remain f illed with process fluid. Micro Motion reco mmends or ienting F-Series sensors as sho wn in Figure 8.
Sensor In stallation: F-Series 13 Mounting the Se nsor Step 3 Mounting the Sensor Use your common practices to minimize torque and bendi ng load on process conn ections. Figure 9 illustrates ho w to mount the sensor . T o reduce the risk of co ndensation or excessi ve mois ture, the conduit opening should not point upward (if possib le).
14 Sensor In stallation : F-Series Mounting the Sensor Mounting the electronics of high-temperature and e xtreme high-temperature se nsors High-temperature and extreme high-tempera ture F-Series sensors come with a 32 ″ (812 mm) pre-installed flexible conduit.
Sensor In stallation: F-Series 15 Wiring Step 4 Wiring Improperly sealed housings can expose electron ics to moisture, which can cause measurement err or or flowmeter failure. Inspect and grease all ga skets and O-rings. Fully close and tighten all housing cover s and conduit openings.
16 Sensor In stallation : F-Series Wiring Figure 11 Micro Motion cable gland an d heat shrink 4. For conn ection at the core processor housing, pr epare shielded cable as follo ws (for armored cable, omit steps d, e, f, and g): a. Strip 4 1/2 inches ( 114 mm) of cab le jack et.
Sensor In stallation: F-Series 17 Wiring Figure 13 Applying the heat shrink g. Position gland clamping inse rt so the interior end is flush with the heat shrink. h. Fold the cl oth shield or braid and drain wires o ver the clampin g insert and approximately 1/8 inch (3 mm) past the O-ring.
18 Sensor In stallation : F-Series Wiring Figure 16 Connecting the wires at the core processor 7. Reinstall and tighten the co re processor housing co ver . 8. Additional wiring instructions for the tran smitter can be found in the transmitter manual.
Sensor In stallation: F-Series 19 Grounding Step 5 Grounding The sensor can be groun ded via the piping if the joints in the pipeline are ground- bonded. If the sen sor is not grounded via the piping, con nect a ground wire to the in ternal or exter nal grounding scre w , which is located on the core processo r housing or junction box.
20 Sensor In stallation : F-Series Purge Fitt ings Purge Fittings If the sensor has pur ge fittings, the y should remain seal ed at all times. After a purge plug has been remo v ed, the sensor case should b e purged w ith ar gon or nitrogen and resealed.
Sensor In stallation: F-Series 21 Purge Fittings 8. Make sure the pur ge f itting seals are tight so air cannot be dra wn b y suction into the sensor case.
22 Sensor In stallation : F-Series Return Policy Return Policy Micro Motion procedu res must be followed when retu rning equipment. These procedures ensure le gal compliance with go vernment transpor tation agencies and help pro vide a safe working en vironment for Micro Motion employees.
.
Micro Motion Inc. USA Worldwide Headqu arters 7070 Winchester Circle Boulder , Colorado 80301 T +1 303-527-5200 +1 800-522-6277 F +1 303-530-8459 Micro Motion Europe Emerson Process Management Neonstr.
Un point important après l'achat de l'appareil (ou même avant l'achat) est de lire le manuel d'utilisation. Nous devons le faire pour quelques raisons simples:
Si vous n'avez pas encore acheté Emerson F-SERIES SENSOR c'est un bon moment pour vous familiariser avec les données de base sur le produit. Consulter d'abord les pages initiales du manuel d'utilisation, que vous trouverez ci-dessus. Vous devriez y trouver les données techniques les plus importants du Emerson F-SERIES SENSOR - de cette manière, vous pouvez vérifier si l'équipement répond à vos besoins. Explorant les pages suivantes du manuel d'utilisation Emerson F-SERIES SENSOR, vous apprendrez toutes les caractéristiques du produit et des informations sur son fonctionnement. Les informations sur le Emerson F-SERIES SENSOR va certainement vous aider à prendre une décision concernant l'achat.
Dans une situation où vous avez déjà le Emerson F-SERIES SENSOR, mais vous avez pas encore lu le manuel d'utilisation, vous devez le faire pour les raisons décrites ci-dessus,. Vous saurez alors si vous avez correctement utilisé les fonctions disponibles, et si vous avez commis des erreurs qui peuvent réduire la durée de vie du Emerson F-SERIES SENSOR.
Cependant, l'un des rôles les plus importants pour l'utilisateur joués par les manuels d'utilisateur est d'aider à résoudre les problèmes concernant le Emerson F-SERIES SENSOR. Presque toujours, vous y trouverez Troubleshooting, soit les pannes et les défaillances les plus fréquentes de l'apparei Emerson F-SERIES SENSOR ainsi que les instructions sur la façon de les résoudre. Même si vous ne parvenez pas à résoudre le problème, le manuel d‘utilisation va vous montrer le chemin d'une nouvelle procédure – le contact avec le centre de service à la clientèle ou le service le plus proche.